Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K111441

    Validate with FDA (Live)

    Date Cleared
    2011-08-19

    (87 days)

    Product Code
    Regulation Number
    884.1730
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Unimax Veress Needle is intended for percutaneous insertion into the peritoneal cavity for the purpose of insufflation with carbon dioxide to establish pneumoperitoneum prior to the placement of trocars during laparoscopic procedures.

    Device Description

    The Unimax Veress Needle is a sterile and single-use product. It incorporated a spring loaded blunt stylet mechanism similar to the needle. It is used to establish peritoneum prior to trocar and cannula insertion in laparoscopic procedures. The Veress Needle, available in 120mm and 150mm lengths, has applications in gynecological laparoscopy and other laparoscopic procedures.

    AI/ML Overview

    The Unimax Veress Needle is not an AI/ML device, but rather a traditional medical device (a surgical instrument). Therefore, the provided document does not contain the information typically associated with AI/ML device evaluations, such as studies on algorithm performance, human-in-the-loop performance, ground truth establishment for training/test sets, or sample sizes related to AI model development.

    The document describes the non-clinical testing performed to demonstrate the safety and effectiveness of the Unimax Veress Needle and its substantial equivalence to a predicate device.

    Here's the information extracted from the document based on the provided categories, with "N/A" for criteria not applicable to a non-AI/ML device:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (What was tested)Reported Device Performance (Test Results)
    Tip Pull TestMeets pre-defined acceptance criteria
    Switch OperationMeets pre-defined acceptance criteria
    Spring Obturator OperationMeets pre-defined acceptance criteria
    Needle Puncture Force TestMeets pre-defined acceptance criteria
    Cytotoxicity TestComplies with ISO 10993-5
    Intracutaneous Reactivity TestComplies with ISO 10993-10
    Maximization Sensitization TestComplies with ISO 10993-10
    EO Sterilization Validation StudyComplies with ISO 11135-1
    Ethylene Oxide Sterilization Residuals StudyComplies with ISO 10993-7

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document lists "A series of in vitro and in vivo preclinical physical, mechanical and biocompatibility tests" but does not specify the sample sizes (e.g., number of needles tested for each physical/mechanical test) or the exact provenance of the data (e.g., where the in vivo tests were conducted). It is indicated that the manufacturer is based in Taipei, Taiwan. The tests are non-clinical (preclinical), not clinical data, and appear to be prospective testing of the device itself rather than data collection from a population.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    N/A. This is a non-AI/ML device. Ground truth, in the context of AI, refers to expert-labeled data. For a physical medical device, performance is evaluated against engineering specifications and international standards, not against expert interpretation of data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    N/A. Adjudication methods are relevant for resolving discrepancies in expert labeling for AI ground truth. For this device, performance is measured against technical specifications and standard compliance.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    N/A. This is a non-AI/ML device. MRMC studies are used to evaluate diagnostic performance of AI-assisted systems.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    N/A. This is a non-AI/ML device. "Standalone performance" refers to the performance of an AI algorithm without human involvement.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For this medical device, the "ground truth" or standard for acceptance is adherence to recognized international standards (e.g., ISO 10993 series, ISO 11135-1) and demonstrating performance at least equivalent to the predicate device through physical and mechanical testing (e.g., Tip Pull Test, Switch Operation, Spring Obturator Operation, Needle Puncture Force Test). There is no "ground truth" in the AI sense for this type of product.

    8. The sample size for the training set

    N/A. This is a non-AI/ML device, so there is no training set for an algorithm.

    9. How the ground truth for the training set was established

    N/A. This is a non-AI/ML device, so there is no training set for an algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1