    K Number
    K152339

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-06-24

    (310 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Unifine Pentips range of pen needles are intended for use with pen injector devices for the subcutaneous injection of drugs, including insulin.

    Device Description

    Unifine® Pentips® and Unifine® Pentips® Plus are sterile, non-toxic, non-pyrogenic, single use pen needles for the subcutaneous injection of drugs from pen injectors. The pen needles are used by consumers, caregivers, and healthcare professionals. They are available in a variety of lengths (4mm, 5mm, 6mm, 8mm and12mm) and gauges (29G, 31G and 32G).

    The pen needle assembly consists of a double-ended cannula, a needle hub, a needle shield, primary container and sterility seal that covers the assembly inside the primary container.

    The hub has internal threads, which allows the pen needle to be screwed onto a pen injector device. The Non Patient (NP) end of the cannula (inside the hub) penetrates the rubber septum of the cartridge in the pen injector device. The Patient end and NP end of the cannula are lubricated using a silicone based lubricant to aid in injection and penetration into the rubber septum. The inner needle shield is injection molded and assembled over the Patient end of the needle to protect the needle from damage and accidental needle-sticks. The pen needle assembly is then inserted into an iniection molded outer container and sealed with a protective peel-away label to ensure a sterility barrier and tamper evidence. The outer container can be used to remove the hub with cannula from the pen injector device after the injection is completed. The peel-away label is pre-printed with information, which includes the lot number, expiry date, and size (length) and gauge of the needle. The individual needle assemblies are then packaged in bags and/or cartons, and placed into shippers with appropriate labeling. The shipper cases are then placed on pallets for sterilization.

    Unifine Pentips Plus pen needles differ from Unifine Pentips pen needles because the primary container consists of two 'chamber contains a new, unused pen needle and the second chamber is empty. The empty chamber serves as a built-in pen needle remover, and may prevent accidental needlesticks. Users can insert their used needle into the chamber, then unscrew from the injection pen. The primary container is designed to hold the used pen needle until it can be disposed of in a suitable sharps container. The actual pen needle is identical between Unifine Pentips and Unifine Pentips Plus in both the physical appearance and functionality.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for medical devices, specifically pen needles. It is not a study proving an AI/ML medical device meets acceptance criteria. Therefore, most of the requested information regarding AI/ML device performance, ground truth establishment, sample sizes for training/test sets, expert adjudication, and MRMC studies, cannot be extracted from this document.

    This document focuses on demonstrating substantial equivalence of the Unifine® Pentips® and Unifine® Pentips® Plus pen needles to legally marketed predicate devices, primarily K973899 (Unifine® Pentips®). This is a regulatory pathway for Class II medical devices in the US, where the manufacturer needs to show that their new device is as safe and effective as an already cleared device.

    Here's what can be extracted based on the provided text, and where the requested information is not applicable or not present:


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for these pen needles are primarily defined by adherence to ISO standards and internal performance tests. The document states that the devices meet these standards/criteria.

    Test / CharacteristicAcceptance Criteria (Standard / Description)Reported Device Performance
    Dimensions for needlesISO 11608-2:2012 Section 4.2.2Meets standard
    Needle pointsISO 11608-2:2012 Section 4.5Meets standard
    Freedom from defectsISO 11608-2:2012 Section 4.6Meets standard
    LubricationISO 11608-2:2012 Section 4.7Meets standard
    Dislocation of measuring point at patient endISO 11608-2:2012 Section 4.8Meets standard
    Bond between hub and needle tubeISO 11608-2:2012 Section 4.4Meets standard
    Ease of assembly/disassemblyISO 11608-2:2012 Section 4.10Meets standard
    SterilizationISO 11608-2:2012 Section 4.11, ISO 11137-2:2007 (UK), ISO 11135-1:2006 (Italy)Meets standard
    Flow rateISO 11608-2:2012 Section 4.3Meets standard
    Needle dose accuracyISO 11608-2:2012 Section 4.9, 11.4.2Meets standard
    Needle hub torque removalISO 11608-2:2012 Section 4.9, 11.4.3Meets standard
    Force required to remove pen needle from primary containerEstablished internal acceptance criteria based on device specificationsMeets acceptance criteria
    Force required to fit pen needle into primary container (UP)Established internal acceptance criteria based on device specificationsMeets acceptance criteria
    Force to remove needle shield from pen needleEstablished internal acceptance criteria based on device specificationsMeets acceptance criteria

    Note: The document states "The devices comply with the acceptance criteria established based on the specifications of the devices" for internal tests, but does not provide the specific numerical criteria for these tests.


    Regarding the study that proves the device meets acceptance criteria (for an AI/ML context):

    This document describes a regulatory submission for a physical medical device (pen needles), not an AI/ML software device. Therefore, the concepts of "test set," "training set," "ground truth," "experts for ground truth," "adjudication," and "MRMC studies" as they apply to AI/ML performance evaluation are not relevant here. The "study" described is a series of non-clinical performance (bench) tests and biocompatibility tests to demonstrate substantial equivalence to existing devices.

    Here's why the specific questions are not applicable or cannot be answered from this document:

    2. Sample sized used for the test set and the data provenance:

    • Not Applicable in AI/ML context: This is a physical device. "Test set" refers to samples tested during bench testing.
    • Sample Size: The document does not specify the number of pen needles tested for each performance test. It only states that the tests were performed and results "meet standard" or "meet acceptance criteria."
    • Data Provenance: Not applicable in the AI/ML sense. Data is from laboratory bench testing of manufactured pen needles. The manufacturing locations mentioned are the UK (Owen Mumford) and Italy (Artsana S.p.A.).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable: There is no "ground truth" in the AI/ML sense (e.g., expert-annotated images) for a physical device like a pen needle. The "ground truth" is defined by the objective physical and mechanical properties measured against established ISO standards and internal specifications. Bench testing is typically performed by trained technicians.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable: No expert adjudication process as would be used for clinical image interpretation or AI model validation. Bench tests have objective, measurable outcomes.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable: This type of study (MRMC, human readers, AI assistance) is for clinical decision support or diagnostic AI devices. This document is for a mechanical device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable: This is not an algorithm or software device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Not Applicable in AI/ML context: The "ground truth" for these pen needles is defined by objective physical and mechanical measurements against established ISO standards (e.g., dimensions, flow rate, bond strength) and internal engineering specifications. Biocompatibility is assessed through standardized biological tests.

    8. The sample size for the training set:

    • Not Applicable: This is not an AI/ML device, so there is no "training set."

    9. How the ground truth for the training set was established:

    • Not Applicable: As above, no training set or AI/ML ground truth. The "ground truth" for manufacturing and quality control of pen needles is based on engineering specifications and adherence to international standards for medical devices.
    Ask a Question

    Ask a specific question about this device

    K Number
    K973899

    Validate with FDA (Live)

    Device Name
    UNIFINE PENTIPS
    Date Cleared
    1997-12-04

    (51 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Unifine Pentip are intended for the subcutaneous injection of drugs from prefilled cartridges used in conjunction with variable dose delivery systems such as Pen Injectors. Its intended use is substantially equivalent to that of the BD Microfine.

    The Unifine Pentip is single use, disposable hypodermic single lumen needle designed for use with multidose injection devices that use prefilled cartridges.

    The protective paper is removed from the Pentip which is then screwed onto the device cartridge housing and then used in the prescribed manner.

    One of the main uses of the Pentip is for use with pen Injectors for the delivery of insulin which is supplied in the form of prefilled cartridges. The patient places a fresh Pentip needle onto the Pen Injector, dials a dose, inserts the Pentip into the skin and delivers the dose.

    Device Description

    The Owen Mumford Unifine Pentip is visually and mechanically extremely similar to the BD Microfine Pentip to which substantial equivalence is claimed.

    AI/ML Overview

    I am sorry, but the provided text does not contain the detailed study information required to fill out the table and answer all the questions about acceptance criteria and device performance.

    The document is a 510(k) submission for the Unifine Pentip, claiming substantial equivalence to the Becton Dickinson Microfine Pentip. While it states that the performance is "substantially equivalent," it does not provide specific acceptance criteria, quantitative performance metrics, or details about a study conducted to demonstrate this equivalence.

    Here's a breakdown of what cannot be extracted from the provided text:

    • Acceptance Criteria Table: No specific criteria (e.g., success rate, accuracy, failure rate) are mentioned.
    • Reported Device Performance: No quantitative performance data is provided for the Unifine Pentip.
    • Sample size for the test set: Not mentioned.
    • Data provenance (country, retrospective/prospective): Not mentioned.
    • Number of experts used for ground truth: Not applicable, as no such study is described.
    • Qualifications of experts: Not applicable.
    • Adjudication method: Not applicable.
    • MRMC comparative effectiveness study: Not mentioned.
    • Effect size of human reader improvement: Not applicable.
    • Standalone (algorithm only) performance study: Not applicable, as this is a physical medical device, not an AI algorithm.
    • Type of ground truth used: Not applicable, as no study establishing ground truth is described.
    • Sample size for the training set: Not applicable, as this is a physical medical device, not an AI algorithm.
    • How ground truth for training set was established: Not applicable.

    The document focuses on regulatory approval based on substantial equivalence to an existing marketed device, citing similarity in design, intended use, and operational principles, and noting that the product has been marketed in Europe without adverse reports. It does not include a detailed clinical or performance study with quantified acceptance criteria as might be expected for an AI/algorithm-based device.