510(k) Data Aggregation

    K Number
    K121620

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-12-11

    (193 days)

    Product Code
    Regulation Number
    872.4120
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for use in surgical procedures including osteotomy, osteoplasty, periodontal surgery and implantation.

    Device Description

    This product is a device using by purpose of bone cutting, polishing when operate implant for dentist's office using ultrasonic vibration because it is supplied electric and water in outside

    AI/ML Overview

    The provided text is a 510(k) summary for the DMETEC SmarThor Ultrasonic Surgery device. It primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a standalone study with acceptance criteria, performance metrics, and detailed study design as might be done for a novel device or AI/software.

    Therefore, the following information is extracted or inferred based on the nature of a 510(k) submission for this type of device (an ultrasonic bone-cutting instrument).

    Here's a breakdown of the information based on your request, highlighting what is present and what is not:


    1. Table of Acceptance Criteria and Reported Device Performance

    For this 510(k) submission, the "acceptance criteria" are implied by demonstrating substantial equivalence to the predicate device (EMS Piezon Master Surgery). The performance is reported by comparing the specifications of the SmarThor to the predicate device.

    Characteristic / Acceptance Criteria (Implied)DMETEC (SmarThor) Reported PerformanceEMS (Piezon Master Surgery) Predicate Performance
    Indication for UseThe device is intended for use in surgical procedures including osteotomy, osteoplasty, periodontal surgery and implantation.The EMS Piezon Master Surgery is an ultrasonic bone-cutting instrument for use in surgical dentistry.
    Supply Voltage24V DC100-240 V AC
    Power Consumption (max)50W max105VA
    Maximum Output Power with Load50W25W
    Frequency Range Available24-32 kHz24-32 kHz
    Transformer Input100-240 VAC, 50-60Hz100/220 VAC/50-60Hz
    Transformer Output24V DC/ 2.7A24V AC/ 1.25A
    Main Components1 Unit Body, 1 Hand-piece, 1 Tip1 Unit Body, 1 Hand-piece, 5 Tips
    Tip Material (contact with patient)Stainless steelStainless steel
    BiocompatibilityMeets ISO 10993-5:2009 (Grade 0 cytotoxicity, 0% sensitization, no oral mucosa irritation/mortality)(Implied predicate meets standards)
    Electrical SafetyMeets IEC 60601-1:1988 + A1:1991 + A2:1995(Implied predicate meets standards)
    Electromagnetic Compatibility (EMC)Meets IEC 60601-1-2: 2007(Implied predicate meets standards)
    Software ValidationSoftware validation report shows substantial equivalence and proper performance.(Implied predicate has validated software)
    Device Life5 years(Not explicitly stated for predicate in summary)

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not applicable in the context of this 510(k) submission. No specific "test set" of cases or patients was used for evaluating performance in the way an AI/software study would. The evaluation is based on non-clinical bench testing and comparison of specifications to a predicate device, as well as biocompatibility testing on biological samples.
    • Data Provenance:
      • Non-clinical tests (biocompatibility, electrical safety, EMC, software validation): These tests would have been performed by DMETEC CO., LTD. in Korea (country of origin). These are "prospective" tests conducted specifically for this submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Not applicable. This 510(k) does not involve expert-established ground truths for performance evaluation in the way a diagnostic AI would. The "ground truth" for compliance with standards (e.g., biocompatibility) is defined by the standard itself, and the comparison to the predicate relies on published specifications and post-market experience.

    4. Adjudication Method for the Test Set

    • Not applicable. There was no clinical "test set" requiring adjudication by experts.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    • No, an MRMC study was not done. This device is a physical instrument, and its 510(k) submission relies on demonstrating substantial equivalence through technical specifications, non-clinical bench testing, and comparison with a predicate device, not through a comparative effectiveness study involving human readers and interpretations.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a medical device (ultrasonic surgery instrument), not an AI algorithm. Its performance is inherent in its physical and electrical characteristics.

    7. The Type of Ground Truth Used

    • For non-clinical tests (biocompatibility, electrical safety, EMC): The ground truth is established by the requirements and methodologies defined in the cited international standards (e.g., ISO 10993-5:2009, IEC 60601-1, IEC 60601-1-2).
    • For comparison to the predicate: The ground truth is the published technical specifications and intended use of the legally marketed predicate device (EMS Piezon Master Surgery), as well as its established safety and effectiveness through "post-market experience."

    8. The Sample Size for the Training Set

    • Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K113152

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-06-14

    (233 days)

    Product Code
    Regulation Number
    872.4120
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A device intended for use surgical procedure Including osteotomy, osteoplasty, Periodontal surgery and implantation

    Device Description

    This product is a device using by purpose of bone cutting, polishing when operate implant for dentist's office using ultrasonic vibration because it is supplied electric and water in outside

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the DMETEC Ultrasonic Surgery Machine:

    Acceptance Criteria and Device Performance

    The provided document primarily focuses on demonstrating substantial equivalence to a predicate device (EMS Piezon Master Surgery) rather than defining explicit, quantitative acceptance criteria for de novo performance. The acceptance criteria are implicitly met by showing that the DMETEC device is "similar in functions and efficiency" and has "similar specification and performance" to the predicate.

    Here's a table comparing the DMETEC device's reported performance against the predicate, which serves as the de facto "acceptance criteria" through substantial equivalence:

    CharacteristicPredicate Device (EMS Piezon Master Surgery) Acceptance Criteria (Implicit)DMETEC Device (Surgy Star, AIC Surgery) Reported Performance
    Indications for UseSurgical dentistry procedures including osteotomy, osteoplasty, periodontal surgery, and implantation.Surgical procedure including osteotomy, osteoplasty, periodontal surgery, and implantation.
    Supply Voltage100-240 V AC24V DC
    Power Consumption (max)105 VA50W max
    Maximum Output Power (load)25W50W
    Frequency Range Available24-32 kHz24-32 kHz
    Transformer Input100/220 VAC/50-60Hz100-240 VAC, 50-60Hz
    Transformer Output24V AC/ 1.25A24V DC/ 3.75A
    Main Components1 Unit Body, 1 Hand-piece, 5 Tips1 Unit Body, 1 Hand-piece, 10 Tips
    Tip MaterialStainless steelStainless steel

    Note: The differences, particularly in supply voltage and output power, are acknowledged, but the submission argues for substantial equivalence based on the overall functional similarity and safety.

    Study Information

    The document does not describe a classic clinical study with an explicit test set and detailed performance metrics as one might expect for novel technology. Instead, it relies on demonstrating substantial equivalence through non-clinical testing and comparison to post-market experience of the predicate device.

    1. Sample size used for the test set and the data provenance:

      • No specific sample size for a clinical "test set" is mentioned in the context of device performance evaluation.
      • Data provenance: The document mentions "post market experience" of the predicate device. This suggests retrospective data from the market performance of the EMS Piezon Master Surgery. No specific country of origin is mentioned for this "post market experience."
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. There is no mention of a defined "test set" requiring expert ground truth establishment for clinical performance.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable, as no clinical test set requiring adjudication is described.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This device is a surgical instrument, not an imaging or diagnostic AI device. Therefore, MRMC studies and "human readers" improving with/without AI are not relevant.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is a physical surgical instrument, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for substantial equivalence appears to be derived from documented performance and safety records of the predicate device (EMS Piezon Master Surgery), as well as compliance with recognized standards (IEC 60601-1, ISO 10993). For biocompatibility, lab tests (in vitro cytotoxicity, skin sensitization, oral mucosa irritation) served as the basis for assessing safety.
    7. The sample size for the training set:

      • Not applicable. As this isn't an AI/machine learning device, there is no "training set."
    8. How the ground truth for the training set was established:

      • Not applicable, for the same reason as above.

    Summary of Non-Clinical Tests Conducted:

    • Biocompatibility Tests:
      • In vitro cytotoxicity test (ISO 10993-5 criteria: evaluated as Grade 0, no cell lysis or toxicity).
      • Skin sensitization test (scores were zero, 0% sensitization rates at 24 hours).
      • Oral mucosa irritation test (no mortality, no body weight changes, no differences between saline and test group).
    • Electric Safety and Electromagnetic Compatibility:
      • Tested according to EN 60601-1 and IEC 60601-1-2. Device met the requirements.
    • Software Validation:
      • A software validation report indicated the device "performs as it should" and is substantially equivalent.

    Overall Conclusion of Studies:

    The manufacturer concludes that the DMETEC Ultrasonic Surgery Machine (Surgy Star, AIC Surgery) is substantially equivalent to the predicate device based on:

    • Similar intended use.
    • Similar specifications and performance, despite some numerical differences (e.g., output power).
    • Successful completion of non-clinical tests (biocompatibility, electrical safety, electromagnetic compatibility, software validation) demonstrating safety.
    • Comparison to the post-market experience of the predicate device to infer similar functions and efficiency.