510(k) Data Aggregation
(253 days)
The Tyto Stethoscope is an electronic stethoscope that enables transmission of auscultation sound data, whereby a clinician at one location on an IP network can listen to the auscultation sounds of a patient at a different location on the IP network with the signal carried on an IP connection between the two locations. The Tyto Stethoscope is intended to be used by lay users in non-clinical environments to capture auscultation sounds for transmission to a clinician for review. The Tyto Stethoscope is for medical diagnostics purposes only. The Tyto Stethoscope is not intended for self-diagnosis.
The Tyto Stethoscope (G3) system is designed for use by lay users in non-clinical environments to capture auscultation sounds for transmission to a clinician for review. It enables three types of stethoscope exams: Heart, Lungs and Heart Rate.
The Tyto Stethoscope (G3) system is a multiple functions system, the operation process of the system uses four (4) primary functional elements:
(1) The Tyto Stethoscope (composed of a TytoCare Device with Stethoscope adaptor supported with proprietary TytoCare Device App software).
The TytoCare Device App contains device software functions.
(2) A mobile device (e.g., a smartphone, not part of the Tyto Stethoscope device, not supplied by TytoCare, on which the proprietary TytoCare user App is running),
The TytoCare User App contains only other software functions.
(3) The TytoCare Server platform (composed of server hardware not part of the Tyto Stethoscope device, not supplied by TytoCare, on which the proprietary TytoCare server software is running).
The TytoCare server software is composed of two modules:
- TytoCare Server Software (Data Processing Software Module) – contains device software functions
- TytoCare Server Software (Server Management Software Module) – contains other functions
(4) A clinician receiving platform located in a clinical environment (e.g., a PC at the clinic, not part of the Tyto Stethoscope device, not supplied by TytoCare, on which the proprietary TytoCare Clinician Dashboard App is running).
The TytoCare Clinician Dashboard App is composed of two modules:
- TytoCare Clinician Dashboard (Exam Software Module) – contains device software functions
- TytoCare Clinician Dashboard (Clinician software module) – contains other functions
Two operational flows are optional for using the Tyto Stethoscope (G3): store-and-forward flow and on-line exam flow. Both flows are essentially similar and share the same fundamental steps: performing one or more measurements using the Tyto Stethoscope (G3), sending to a clinician, review of the measurements by the clinician. While in the store-and-forward flow the user can record the measurements and send the recorded data to the clinician whenever convenient for him/her, an online flow may be executed only when also the clinician is available on-line.
N/A
Ask a specific question about this device
(181 days)
The Tyto Stethoscope is an electronic stethoscope that enables transmission of auscultation sound data, whereby a clinician at one location on an IP network can listen to the auscultation sounds of a patient on site or at a different location on the IP network with the signal carried on an IP connection between the two locations. The Tyto Stethoscope is intended to be used by professional users in a clinical environment or by lay users in a nonclinical environment. The device is for medical diagnostics purposes only. The device is not intended for self-diagnosis.
The Tyto Stethoscope (OTC) is designed for use by professional as well as lay users in clinical or non-clinical environments. It enables four types of stethoscope exams: Heart, Lungs, Heart Rate and Audio (Audio is for clinician only). The operation process of the Tyto Stethoscope uses four (4) primary functional elements: (1) The Tyto Stethoscope (composed of a Stethoscope Tip and a Base Unit – Tyto Device and supported with proprietary software). (2) A mobile device (e.g., a smartphone, not part of the Tyto Stethoscope device, not supplied by TytoCare, on which the proprietary TytoCare App is running), (3) The Tyto Server platform (composed of server hardware not part of the Tyto Stethoscope device, not supplied by TytoCare, on which the proprietary server software is running). (4) A clinician receiving platform located in a clinical environment (e.g., a PC at the clinic, not part of the Tyto Stethoscope device, not supplied by TytoCare, on which the proprietary Clinician App is running).
The provided text is a 510(k) summary for the Tyto Stethoscope (OTC). It mostly focuses on establishing substantial equivalence to predicate devices and does not contain detailed information about specific acceptance criteria or a study that rigorously proves the device meets those criteria in a quantitative sense as typically seen for AI/ML devices.
However, based on the general information provided about performance testing, we can infer some "acceptance criteria" through the lens of device specifications and the types of tests mentioned.
Here's an attempt to answer the questions based on the available text, acknowledging that much of the detailed study information is not present in this summary.
1. A table of acceptance criteria and the reported device performance
Based on the provided text, the "acceptance criteria" are broad and relate to meeting specifications and demonstrating similar performance and safety to predicate devices. The text doesn't provide specific quantitative "reported device performance" metrics against numeric acceptance thresholds, as it relies on the device being technologically identical to a previously cleared device.
| Acceptance Criteria Category | Reported Device Performance (Summary from Text) |
|---|---|
| Electrical Safety | Meets IEC 60601-1, IEC 60601-1-11 standards |
| EMC (Electromagnetic Compatibility) | Meets IEC 60601-1-2 standard |
| Biocompatibility | Assessment performed (assumed to meet relevant standards) |
| Stethoscope Performance | Demonstrated "stethoscope performance" (e.g., sound quality, frequency response for heart/lung sounds); considered comparable to predicate devices. Specific metrics are not provided. |
| Pseudo Clinical Performance | Demonstrated "pseudo clinical" performance, implying functionality in simulated clinical scenarios. Specific metrics are not provided. |
| Heart Rate Measurement Accuracy | Demonstrated accuracy in heart rate measurement. Specific metrics are not provided. |
| Wireless Coexistence | Verified wireless coexistence performance. Specific metrics are not provided. |
| Software Validation | Software validation testing performed. |
| Usability | Usability study performed. |
2. Sample size used for the test set and the data provenance
The document states, "A testing plan was developed and performed in order to verify that the Tyto Stethoscope (K160401) meets its specifications and demonstrates, based on comparable characteristics, similar performance and safety as compared to its predicate devices". It then lists categories of performance testing.
- Sample Size for Test Set: Specific sample sizes for any of the listed tests (e.g., stethoscope performance, pseudo clinical, heart rate measurement) are not provided in this summary.
- Data Provenance: Not explicitly stated. Given the device is an electronic stethoscope and the context is regulatory, it's highly likely that the performance testing involved simulated scenarios, physical measurements, and potentially human subjects for usability or sound perception. The original clearance (K160401) would contain more details.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided in the 510(k) summary. The summary refers to "pseudo clinical" and "stethoscope performance" testing, which would typically involve expert assessment, but no details are given.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not provided in the 510(k) summary.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
The Tyto Stethoscope (OTC) is described as an electronic stethoscope for transmitting auscultation sounds to a clinician. It enables the transmission of raw sound data for human clinicians to interpret. There is no indication in this summary that it incorporates AI for interpretation or diagnosis, nor is there any mention of an MRMC comparative effectiveness study involving AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Since there is no mention of an AI algorithm performing diagnosis or interpretation, this question is not applicable. The device's function is to transmit data for a human clinician.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
Given the device's function (electronic stethoscope for sound transmission and heart rate), the "ground truth" for performance testing would likely involve:
- For Stethoscope Performance: Comparison to established acoustic standards, reference stethoscopes, or expert audiologist/physician assessments of sound quality.
- For Heart Rate Measurement: Comparison to ECG, manual palpation, or other validated heart rate measurement devices.
- For Pseudo Clinical: Likely comparison to expert clinical assessment of scenarios or validated simulated patient sounds.
However, the specific type of ground truth used for each test is not detailed in this summary.
8. The sample size for the training set
The provided text describes performance testing for an electronic stethoscope, not an AI/ML algorithm that requires training data in the traditional sense. Therefore, the concept of a "training set" as it relates to AI models is not applicable here. The device's functionality is based on hardware and software for sound acquisition and transmission, not data-driven learning.
9. How the ground truth for the training set was established
As explained in point 8, the concept of a "training set" for an AI/ML algorithm is not applicable to this device as described.
Ask a specific question about this device
(250 days)
The Tyto Stethoscope is an electronic stethoscope that enables transmission of auscultation sound data, whereby a clinician at one location on an IP network can listen to the auscultation sounds of a patient on site or at a different location on the IP network with the signal carried on an IP connection between the two locations. The Tyto stethoscope is intended for use by professional users in a clinical environment or in conjunction with patients or lay users in a non-clinical environment.
The Tyto Stethoscope is designed for use by professional as well as lay users in clinical or nonclinical environments. It enables four types of stethoscope exams: Heart, Lungs, Heart Rate and Audio (Audio is for clinician only). The operation process of the Tyto Stethoscope uses four (4) primary functional elements:
- (1) The Tyto Stethoscope (composed of a Stethoscope Tip and a Base Unit and supported with proprietary software)
- (2) A mobile device (e.g., a smartphone, not part of the Tyto Stethoscope device, not supplied by TytoCare, on which the proprietary TytoCare App is running),
- (3) The Tyto Server platform (composed of server hardware not part of the Tyto Stethoscope device, not supplied by TytoCare, on which the proprietary server software is running)
- (4) A clinician receiving platform located in a clinical environment (e.g., a PC at the clinic, not part of the Tyto Stethoscope device, not supplied by TytoCare, on which the proprietary Clinician App is running).
Two operational flows are optional for using the Tyto Stethoscope: store-and-forward flow and on-line exam flow. Both flows are essentially similar and share the same fundamental steps: performing one or more measurements using the Tyto Stethoscope, recording the data and sending to a clinician, review of the recorded measurements by the clinician, and user receiving a written summary from the clinician presenting his/her assessment and/or recommendations. While in the store-and-forward flow the user can perform the measurements and send the recorded data to the clinician whenever convenient for him/her, an on-line flow may be executed only when also the clinician is available on-line.
This document provides information about the Tyto Stethoscope, for which a 510(k) premarket notification was reviewed by the FDA. The document summarizes the device's intended use, description, and performance testing, but it does not explicitly state specific acceptance criteria or provide a detailed study that proves the device meets those criteria in the format requested.
Therefore, many of the requested fields cannot be directly extracted from the provided text. However, based on the information given, here's what can be inferred or stated:
Missing Information:
The provided document details the FDA's clearance for the Tyto Stethoscope but does not include:
- A table of specific acceptance criteria and detailed reported device performance.
- Sample size used for the test set or its provenance.
- Number of experts, their qualifications, or adjudication methods for establishing ground truth.
- Whether an MRMC comparative effectiveness study was done or effect size.
- Whether a standalone algorithm-only performance study was done.
- The sample size for the training set or how its ground truth was established.
Information Extracted from the Document:
-
A table of acceptance criteria and the reported device performance:
- Acceptance Criteria: Not explicitly stated in a tabular format. The document generally implies that the device should function as expected and be as safe and effective as predicate devices. The performance testing section lists categories of tests performed.
- Reported Device Performance: The document states, "Test results indicated that the Tyto Stethoscope functions as expected and is as safe and effective as its predicate devices for its intended use." No specific quantitative performance metrics are provided.
-
Sample size used for the test set and the data provenance: Not mentioned in the provided text.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not mentioned in the provided text.
-
Adjudication method for the test set: Not mentioned in the provided text.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not mentioned in the provided text. The device is a "Stethoscope," not an AI diagnostic tool in the sense of image interpretation assistance.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The Tyto Stethoscope is an electronic stethoscope that records and transmits audio. Its "performance testing" likely refers to the fidelity of audio capture, heart rate measurement, and wireless transmission, as opposed to an AI algorithm making a diagnosis. The document does not describe a standalone algorithm-only performance as might be applicable to an AI diagnostic device.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not mentioned in the provided text. The testing seems to focus on functional performance (e.g., stethoscope performance, heart rate measurement) and safety rather than diagnostic accuracy against a disease ground truth.
-
The sample size for the training set: Not mentioned in the provided text.
-
How the ground truth for the training set was established: Not mentioned in the provided text.