Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K192958

    Validate with FDA (Live)

    Date Cleared
    2020-04-02

    (163 days)

    Product Code
    Regulation Number
    868.2775
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TwitchView System is used for the quantitative monitoring of neuromuscular transmission by means of electromyography.

    Device Description

    Not Found

    AI/ML Overview

    I am sorry, but the provided text from the FDA document (K192958) for the TwitchView System does not contain the detailed information required to answer your request about acceptance criteria, device performance, study design, and ground truth establishment for an AI/ML device.

    The document is a 510(k) clearance letter for an Electrical Peripheral Nerve Stimulator (Product Code: KOI), which is a device for quantitative monitoring of neuromuscular transmission via electromyography. It is not an AI/ML powered device, and therefore, the kind of study details you are asking for (e.g., sample size for test set, data provenance, expert adjudication, MRMC studies, standalone performance, training set details) are not relevant or present in this specific clearance.

    The document primarily focuses on:

    • Confirming substantial equivalence to a predicate device.
    • Indications for Use.
    • General controls and regulatory requirements for a Class II device.

    The information you are requesting is typically found in:

    • Premarket Approval (PMA) applications for novel high-risk devices, which often involve extensive clinical trial data.
    • De Novo requests for novel low-to-moderate risk devices.
    • Specific sections within 510(k) summaries or detailed review memoranda for AI/ML devices that explicitly describe performance studies, often in response to special controls or FDA guidance for AI/ML.

    Since the provided text does not contain the requested information, I cannot fill out the table or describe the study details.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172843

    Validate with FDA (Live)

    Date Cleared
    2018-03-29

    (191 days)

    Product Code
    Regulation Number
    868.2775
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TwitchView System is used for the quantitative monitoring of neuromuscular transmission by means of electromyography.

    Device Description

    The TwitchView System has been designed to provide a quantitative neuromuscular transmission monitor with a modernized user interface and a single-use electrode array that is placed on the hand of the patient. The degree of neuromuscular block is measured with the TwitchView System by stimulating a peripheral nerve (the ulnar nerve at the wrist), and by evaluating the response of the hand via electromyography. The TwitchView System consists of three components: 1. TwitchView Monitor, 2. TwitchView Charging Station, 3. Single-Use Electrode Array.

    AI/ML Overview

    The provided text is related to a 510(k) premarket notification for the "TwitchView System," a medical device used for monitoring neuromuscular transmission. However, the document does not contain the specific information requested about acceptance criteria and a study proving the device meets those criteria, particularly in the context of an AI/ML-driven device.

    This document describes technical aspects, regulatory compliance (IEC standards, biocompatibility, electrical safety, etc.), and a comparison to predicate devices, but it does not mention any AI or machine learning components, nor does it describe a study involving a test set, expert ground truth, multi-reader multi-case studies, or standalone algorithm performance. The device primarily relies on electromyography (EMG) measurements and nerve stimulation, which are well-established non-AI technologies in medical monitoring.

    Therefore, I cannot fulfill the request to describe the acceptance criteria and the study that proves the device meets those criteria, specifically concerning AI/ML performance, because the provided text does not contain this information.

    The document focuses on demonstrating substantial equivalence to predicate devices based on:

    • Intended Use: Quantitative monitoring of neuromuscular transmission.
    • Technological Characteristics: Stimulating a peripheral nerve with constant current and measuring muscle response. The key difference from the predicate (TOF Watch) is the use of electromyography (TwitchView) versus accelerometry (TOF Watch) for muscle response detection, with the document stating these methods are clinically equivalent.
    • Performance Standards & Testing: Compliance with IEC 60601-2-40 (for electromyographs and evoked response equipment), PART 898 (for electrode lead wires and patient cables), biocompatibility, software verification/validation, electrical safety, EMC, and various bench performance tests (stimulation accuracy, EMG accuracy, battery life, etc.).

    There is no mention of an AI/ML component, and thus no related acceptance criteria, test sets, or validation studies for AI performance.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1