Found 4 results

510(k) Data Aggregation

    K Number
    K222837

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-12-08

    (79 days)

    Product Code
    Regulation Number
    870.4875
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Turbo-Power is indicated for laser atherectomy of de novo or restenotic lesions in native infrainguinal atteries and for the treatment of femoropopliteal artery in-stent restenosis (ISR) in bare nitinol stents, with adjunctive Percutaneous Transluminal Angioplasty (PTA).

    Device Description

    The Turbo-Power (2.0mm, 2.3mm) Laser Atherectomy Catheter are laser atherectomy devices designed for use with the CVX-300™ Excimer Laser System and Philips Laser System. The Turbo-Power (2.0mm) and (2.3mm) Laser Atherectomy Catheters are sterile, single use, prescription only devices used for peripheral atherectomy. The Turbo-Power (2.0mm) and (2.3mm) are used to ablate lesions with reference vessel diameters of ≥3.0mm. Turbo-Power (2.0mm) and (2.3mm) Laser Atherectomy Catheters are comprised of 2 subassemblies:

      1. Catheter Subassembly
      1. Motor Drive Unit (MDU) Subassembly
        The working length of the Turbo Power Laser Atherectomy Catheter is constructed of multiple optical fibers arranged eccentrically around a 0.018" (0.46 mm) quidewirecompatible lumen. The PTFE quidewire lumen tip is attached to a stainless steel torque wire which is connected to the MDU at the proximal end of the working length. The multifiber laser catheter transmits ultraviolet energy from the Spectranetics CVX-300 Excimer Laser System or Philips Laser System through the tip of the laser to an obstruction in the patient's artery. The outer surface of the laser catheter working length is hydrophilic-coated, and the distal tip of the catheter contains a radiopaque marker band for in situ visibility. The ultraviolet energy transmitted from the CVX-300 laser system or Philips Laser System is used to photoablate multiple morphology lesions which may be comprised of atheroma, fibrosis, calcium, and thrombus, thus recanalizing diseased vessels. Photoablation is the process by which energy photons cause molecular bond disruption at the cellular level without thermal damage to surrounding tissue.
    AI/ML Overview

    This document is a 510(k) premarket notification for the Turbo-Power (2.0mm and 2.3mm) Laser Atherectomy Catheters. It describes a modification to an existing device (a new Printed Circuit Board Assembly and firmware) and aims to demonstrate substantial equivalence to a predicate device.

    Therefore, the study design described is a design verification and validation testing for the modified device, rather than a clinical trial or large-scale comparative effectiveness study typical for new medical device approvals. The acceptance criteria and the study performed are focused on confirming that the modified device functions as intended and meets the safety and performance requirements established for the original device.

    Here's an analysis based on the provided text:

    1. A table of acceptance criteria and the reported device performance:

    The document broadly states that "Functional testing was performed to confirm the devices with the new PCBA meet the relevant product requirements." and "The following testing was conducted to verify that the subject device met all acceptance criteria as required by the risk analysis that was performed." However, a specific table detailing individual acceptance criteria and corresponding reported device performance values (e.g., pass/fail for specific tests, quantitative measurements within certain tolerances) is not provided in this summary. The categories of testing performed are listed:

    Acceptance Criteria (General Categories)Reported Device Performance
    Functional Performance (relevant product requirements)Confirmed to meet relevant product requirements (Pass)
    Software PerformanceSoftware testing was performed (Pass)
    Medical Electrical SafetyMedical Electrical Safety Testing was performed (Pass)
    SterilizationAssessment concluded additional sterilization testing not required (Deemed Acceptable)
    Risk Analysis RequirementsMet all acceptance criteria as required by risk analysis (Pass)

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document does not specify the exact sample size (number of modified devices) used for the functional, software, and medical electrical safety testing. It mentions "devices with the new PCBA" in plural, implying more than one unit, but no specific number.
    • Data Provenance: The testing was conducted by The Spectranetics Corporation, based in Colorado Springs, Colorado, USA. The data is prospective as it involves new testing on the modified device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This type of information (number and qualifications of experts for ground truth) is typically relevant for studies involving subjective human interpretation, such as image reading in AI/diagnostic device studies. For this submission, which focuses on engineering verification of a physical device modification, the "ground truth" is established by the device's technical specifications and engineering requirements. Therefore, no external experts (like radiologists) were used to establish a "ground truth" in the diagnostic sense. The "experts" would be the internal engineers and quality personnel responsible for defining and executing the test protocols and assessing compliance. Their qualifications are implicitly assumed to be appropriate for device testing and validation.

    4. Adjudication method for the test set:

    Not applicable in the context of this device modification verification. Adjudication methods (like 2+1 or 3+1) are used to resolve disagreements among multiple human readers when establishing ground truth for diagnostic studies. For engineering and safety testing, passing or failing a test is typically based on predefined objective criteria, not subjective interpretation requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No, an MRMC comparative effectiveness study was not done. This document is for a medical device modification (a laser atherectomy catheter) and not an AI/CADe device where human reader improvement with AI assistance would be relevant. The submission aimed to demonstrate substantial equivalence of the modified device to its predicate, focusing on maintaining existing performance and safety.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is not an algorithmic device. The "performance" here refers to the physical and electrical function of the catheter and its associated components, not the output of a diagnostic algorithm.

    7. The type of ground truth used:

    The "ground truth" for this device modification is based on engineering specifications, functional requirements, and safety standards. The tests performed verify that the modified device continues to meet these objective, predefined criteria. It does not involve "expert consensus, pathology, or outcomes data" in the traditional sense of validating a diagnostic or AI-based device.

    8. The sample size for the training set:

    Not applicable. This is a medical device modification, not a machine learning or AI algorithm development. Therefore, there is no "training set" in the context of AI.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180694

    Validate with FDA (Live)

    Date Cleared
    2018-04-05

    (20 days)

    Product Code
    Regulation Number
    870.4875
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Turbo-Power is indicated for laser atherectomy of de novo or restenotic lesions in native infrainguinal arteries and for the treatment of femoropopliteal artery in-stent restenosis (ISR) in bare nitinol stents, with adjunctive Percutaneous Transluminal Angioplasty (PTA).

    Device Description

    The Turbo-Power (2.0mm, 2.3mm) Laser Atherectomy Catheter are laser atherectomy devices designed for use with the CVX-300™ Excimer Laser System. The Turbo-Power (2.0mm) and (2.3mm) Laser Atherectomy Catheters are sterile, single use, prescription only devices used for peripheral atherectomy. The Turbo-Power (2.0mm) and (2.3mm) are used to ablate lesions with reference vessel diameters of ≥3.0mm. Turbo-Power (2.0mm) and (2.3mm) Laser Atherectomy Catheters are comprised of 2 subassemblies: 1. Catheter Subassembly 2. Motor Drive Unit (MDU) Subassembly The working length of the Turbo Power Laser Atherectomy Catheter is constructed of multiple optical fibers arranged eccentrically around a 0.018" (0.46 mm) guidewirecompatible lumen. The PTFE guidewire lumen tip is attached to a stainless steel torque wire which is connected to the MDU at the proximal end of the working length. The multifiber laser catheter transmits ultraviolet energy from the Spectranetics CVX-300 Excimer Laser System through the tip of the laser to an obstruction in the patient's artery. The outer surface of the laser catheter working length is hydrophilic-coated, and the distal tip of the catheter contains a radiopaque marker band for in situ visibility. The ultraviolet energy transmitted from the CVX-300 laser system is used to photoablate multiple morphology lesions which may be comprised of atheroma, fibrosis, calcium, and thrombus, thus recanalizing diseased vessels. Photoablation is the process by which energy photons cause molecular bond disruption at the cellular level without thermal damage to surrounding tissue.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for Spectranetics, Inc.'s Turbo-Power (2.0mm, 2.3mm) Laser Atherectomy Catheters, seeking substantial equivalence to a predicate device (K172687).

    This document does not contain information related to an AI/ML powered medical device and therefore does not have the answers to questions 1-9. The device described is a physical medical device (laser atherectomy catheter) with minor design modifications (improved battery clip and battery pull tab), not an AI/ML algorithm.

    Therefore, I cannot fulfill the request to describe the acceptance criteria and study proving an AI/ML device meets the acceptance criteria based on the provided text. The traditional medical device testing mentioned (Motor Drive Unit Lifetime Test, Motor Drive Unit Functionality Test, Pull Tab Isolation Test) are specific to the physical catheter's mechanical performance, not the performance of an AI/ML model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172687

    Validate with FDA (Live)

    Date Cleared
    2017-11-27

    (82 days)

    Product Code
    Regulation Number
    870.4875
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Turbo-Power Laser Atherectomy Catheter is indicated for laser atherectomy of de novo restenotic lesions in native infrainguinal arteries and for the treatment of femoropopliteal artery in-stent restenosis (ISR) in bare nitinol stents, with adjunctive Percutaneous Transluminal Angioplasty (PTA).

    Device Description

    The 2.0mm and 2.3mm Turbo-Power Laser Atherectomy Catheters are laser atherectomy devices designed for use with the CVX-300™ Excimer Laser System. The 2.0mm and 2.3mmTurbo-Power Laser Atherectomy Catheters are sterile, single use, prescription only devices used for peripheral atherectomy. The 2.0mm and 2.3mm Turbo-Power Laser Atherectomy Catheters are used to ablate lesions with reference vessel diameters of ≥3.0mm. 2.0mm and 2.3mm Turbo-Power Laser Atherectomy Catheters are comprised of 2 subassemblies: 1. Catheter Subassembly 2. Motor Drive Unit (MDU) Subassembly The working length of the Turbo Power Laser Atherectomy Catheter is constructed of multiple optical fibers arranged eccentrically around a 0.018" (0.46 mm) guidewirecompatible lumen. The PTFE quidewire lumen tip is attached to a stainless steel torque wire which is connected to the MDU at the proximal end of the working length. The multifiber laser catheter transmits ultraviolet energy from the Spectranetics CVX-300 Excimer Laser System through the tip of the laser to an obstruction in the patient's artery. The outer surface of the laser catheter working length is hydrophilic-coated, and the distal tip of the catheter contains a radiopaque marker band for in situ visibility. The ultraviolet energy transmitted from the CVX-300 laser system is used to photoablate multiple morphology lesions which mav be comprised of atheroma, fibrosis, calcium, and thrombus, thus recanalizing diseased vessels. Photoablation is the process by which energy photons cause molecular bond disruption at the cellular level without thermal damage to surrounding tissue.

    AI/ML Overview

    The provided text is a 510(k) summary for the Spectranetics Turbo-Power Laser Atherectomy Catheters. The document states that "New clinical data was not required to demonstrate substantial equivalence," and much of the testing was "leveraged from the predicate Turbo-Power (2.3mm)" and summarized within this submission (which is not fully provided). Therefore, many of the requested details about acceptance criteria, specific performance metrics, sample sizes, and ground truth establishment for a new study are not available in the provided text because a new extensive clinical study for this specific 510(k) was not conducted.

    However, based on the information provided, here's what can be inferred or directly stated:

    1. A table of acceptance criteria and the reported device performance

    The document lists types of performance data collected but does not provide specific acceptance criteria or quantitative performance results in a table format.

    Acceptance Criteria CategoryReported Device Performance
    Design Verification and Validation TestingMet all acceptance criteria as required by the risk analysis.
    Simulated Use TestingPerformed. Outcomes not detailed.
    Functional TestingPerformed. Outcomes not detailed.
    Physical TestingPerformed. Outcomes not detailed.
    Laser TestingPerformed. Outcomes not detailed.
    SterilizationProduct adoption equivalency per AAMI TIR:28-2009.
    Pre-clinical GLP StudiesCompared usability and procedural safety to predicate devices. Supported premarket notification.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set Sample Size: Not specified for any new testing. Much of the data was leveraged from predicate devices. For pre-clinical GLP studies, the sample size is not disclosed.
    • Data Provenance: Not specified for any new testing. The phrase "pre-clinical GLP studies" suggests controlled laboratory or animal studies, not human clinical trials.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as new clinical data requiring expert-established ground truth was not explicitly detailed or required. The testing involved "Design Verification and Validation Testing" and "Pre-clinical GLP studies," which typically don't involve expert ground truth in the same way clinical imaging studies do.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as new clinical data requiring adjudication was not explicitly mentioned or required.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was mentioned. This device is a physical medical device (laser atherectomy catheter), not an AI-powered diagnostic tool. Therefore, the concept of "human readers improve with AI vs without AI assistance" does not apply.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable to a physical medical device like a laser atherectomy catheter. The device itself performs the intended function, often with human operation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the pre-clinical GLP studies, the ground truth would likely be based on direct observations, measurements of tissue effects, and safety endpoints within the experimental setup, rather than expert consensus on diagnostic images or pathology. The document does not provide details.

    8. The sample size for the training set

    Not applicable. This device is not an AI algorithm that requires a "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no "training set" for this physical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162561

    Validate with FDA (Live)

    Date Cleared
    2017-01-05

    (113 days)

    Product Code
    Regulation Number
    870.4875
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Turbo-Power is indicated for laser atherectomy of de novo and restenotic lesions in native infrainguinal arteries and for the treatment of femoropopliteal artery in-stent restenosis (ISR) in bare nitinol stents, with adjunctive Pecutaneous Transluminal Angioplasty (PTA).

    Device Description

    The Turbo-Power (2.0mm) Laser Atherectomy Catheter is a laser atherectomy device designed for use with the CVX-300™ Excimer Laser System. The Turbo-Power (2.0mm) Laser Atherectomy Catheter is a sterile, single use, prescription only device used for peripheral atherectomy.

    The Turbo-Power (2.0mm) is used to ablate lesions with reference vessel diameters of ≥3.0mm. Turbo-Power (2.0mm) Laser Atherectomy Catheter is comprised of 2 subassemblies:

      1. Catheter Subassembly
      1. Motor Drive Unit (MDU) Subassembly

    The working length of the Turbo Power Laser Atherectomy Catheter is constructed of multiple optical fibers arranged eccentrically around a 0.018" (0.46 mm) guidewirecompatible lumen. The PTFE quidewire lumen tip is attached to a stainless steel torque wire which is connected to the MDU at the proximal end of the working length. The multifiber laser catheter transmits ultraviolet energy from the Spectranetics CVX-300 Excimer Laser System through the tip of the laser to an obstruction in the patient's artery. The outer surface of the laser catheter working length is hydrophilic-coated, and the distal tip of the catheter contains a radiopaque marker band for in situ visibility. The ultraviolet energy transmitted from the CVX-300 laser system is used to photoablate multiple morphology lesions which mav be comprised of atheroma, fibrosis, calcium, and thrombus, thus recanalizing diseased vessels. Photoablation is the process by which energy photons cause molecular bond disruption at the cellular level without thermal damage to surrounding tissue.

    AI/ML Overview

    The provided document is a 510(k) summary for the Spectranetics Turbo-Power (2.0mm) Laser Atherectomy Catheter. It describes the device, its intended use, and substantial equivalence to a predicate device. However, it does not contain specific acceptance criteria or detailed results of a study proving the device meets those criteria, nor does it include information about sample sizes for test sets, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone performance, or training set details.

    The document mentions that various tests were conducted to validate and verify that the subject device met all acceptance criteria, as required by the risk analysis. These tests included:

    • Design Verification and Validation Testing: Simulated Use Testing, Functional Testing, Physical Testing, Laser Testing.
    • Sterilization: Product adoption equivalency per AAMI TIR:28-2009.
    • Biocompatibility: Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Direct Hemolysis, Indirect Hemolysis, In Vivo Thrombogenicity-Ovine Model, Genotoxicity - Ames Test, Material Mediated Pyrogenicity.
    • Pre-clinical Data: A pre-clinical GLP study was conducted to compare usability and procedural safety of the Turbo-Power (2.0mm) and Turbo-Power (2.3) laser catheters.

    The document does not provide the specific acceptance criteria for these tests or the reported performance values against those criteria in a table format. It only states that the device "met all acceptance criteria." It also explicitly states that new clinical data was not required to demonstrate substantial equivalence, which means there isn't a human-in-the-loop or standalone study with clinical outcomes described here.

    Therefore, many of the requested details cannot be extracted from this document.

    Here's a breakdown of what can and cannot be answered based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    • Cannot be provided. The document states that testing was performed and the device "met all acceptance criteria," but it does not specify what those criteria were or the actual performance results in a quantifiable manner.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Cannot be provided. The document mentions "Simulated Use Testing," "Functional Testing," "Physical Testing," "Laser Testing," and a "pre-clinical GLP study." However, it does not specify sample sizes for any of these tests or the data provenance (e.g., retrospective/prospective, country of origin). The pre-clinical study involved an "Ovine Model," indicating animal data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable/Cannot be provided. This question typically applies to studies involving human interpretation or clinical endpoints. The document describes engineering, biocompatibility, and animal (ovine model) studies. It does not mention any studies where human experts established ground truth for a test set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable/Cannot be provided. Similar to point 3, this is relevant for studies involving human assessment or interpretation, which are not detailed here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. The document explicitly states "New clinical data was not required to demonstrate substantial equivalence." An MRMC study would fall under clinical data. This device is a physical catheter, not an AI system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is a physical catheter, not an algorithm or AI system. Its performance is evaluated through engineering, biocompatibility, and pre-clinical animal studies, not standalone algorithmic performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Cannot be fully specified. For the engineering tests, ground truth would likely refer to engineering specifications and performance benchmarks. For the biocompatibility tests, it would be established standards (e.g., ISO standards for cytotoxicity, sensitization). For the pre-clinical GLP study using an ovine model, ground truth would have been based on observations and measurements from the animal subjects related to usability and procedural safety as defined by the study protocol.

    8. The sample size for the training set

    • Not applicable. The document describes a medical device (catheter), not an AI system that requires a "training set."

    9. How the ground truth for the training set was established

    • Not applicable. Same as point 8.