Found 6 results

510(k) Data Aggregation

    K Number
    K252948

    Validate with FDA (Live)

    Date Cleared
    2025-10-07

    (21 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TrueBeam™, TrueBeam STx and Edge™ Systems are intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediatric patients. The TrueBeam radiotherapy system can produce CBCT images that can be used in image guided radiation therapy, and the simulation and planning for adaptive radiation therapy.

    The TrueBeam, TrueBeam STx and Edge Systems may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngioma, pituitary tumors, spinal cord tumors, hemangioblastoma, orbital tumors, ocular tumors, optic nerve tumors, and skull based tumors), head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, larynx, oropharynx, nasopharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, esophageal cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal tumors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carcinoma), genitourinary tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as squamous cell, basal cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, cavernous malformation, trigeminal neuralgia, chordoma, glomus tumors, hemangiomas, and medically refractory essential tremor (indicated for adults only)), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and skin), pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, craniopharyngioma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, cavernous malformation, and skull base tumors), and low-dose radiotherapy for adults with medically refractory osteoarthritis.

    VitalBeam® is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediatric patients.

    VitalBeam may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngioma, pituitary tumors, spinal cord tumors, hemangioblastoma, orbital tumors, ocular tumors, optic nerve tumors, and skull based tumors), head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, larynx, oropharynx, nasopharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, esophageal cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal tumors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carcinoma), genitourinary tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as squamous cell, basal cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, cavernous malformation, trigeminal neuralgia, chordoma, glomus tumors, and hemangiomas), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and skin), pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, craniopharyngioma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, cavernous malformation, and skull base tumors), and low-dose radiotherapy for adults with medically refractory osteoarthritis.

    Device Description

    The TrueBeam and VitalBeam Radiotherapy System is a medical linear accelerator that delivered therapeutic radiation to patient in accordance with the physician's prescription.

    The system consists of two major components – a photon, electron and diagnostic kV X-ray radiation beam producing component that is installed in a radiation-shielded vault and a control console area located outside the treatment room.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K232870

    Validate with FDA (Live)

    Date Cleared
    2023-12-21

    (97 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TrueBeam-TrueBeam STx-Edge: The TrueBeam ™ radiotherapy delivery system is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.

    VitalBeam: The VitalBeam system is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.

    Device Description

    The TrueBeam and VitalBeam Radiotherapy System is a medical linear accelerator that delivered therapeutic radiation to patient in accordance with the physician's prescription.

    The system consists of two major components – a photon, electron and diagnostic kV X-ray radiation beam producing component that is installed in a radiation-shielded vault and a control console area located outside the treatment room.

    AI/ML Overview

    The provided text does not contain information about an AI/ML device or its acceptance criteria. Instead, it is an FDA 510(k) clearance letter for the Varian Medical Systems TrueBeam, TrueBeam STx, EDGE, and VitalBeam (4.1) radiation therapy systems.

    The document discusses:

    • Device Name & Regulation: TrueBeam, TrueBeam STx, EDGE, and VitalBeam (4.1), regulated as a Medical Charged-Particle Radiation Therapy System (21 CFR 892.5050, Class II, Product Code: IYE).
    • Intended Use & Indications for Use: The devices are intended for stereotactic radiosurgery and precision radiotherapy for various lesions, tumors, and conditions throughout the body in adult and pediatric patients. A key update in version 4.1 is the incorporation of changes to the CBCT image reconstruction algorithm to allow for the use of CBCT images in the simulation and planning for adaptive radiation therapy. However, it explicitly states that TrueBeam v4.1 cannot deliver adaptive dosimetry on a real-time, per-fraction basis, but supports off-line adaptive treatment planning or delivery.
    • Predicate Device: TrueBeam / TrueBeam STx / Edge / VitalBeam (K231317), which is version 4.0.
    • Technological Characteristics: The subject device (v4.1) and predicate device (v4.0) share the same medical linear accelerator, couch, integrated treatment and imaging consoles, multi-leaf collimators, and supported treatment techniques. The significant difference is the software change to the CBCT image reconstruction algorithm for adaptive radiation therapy planning.
    • Performance Data: Verification and validation of the modified software and hardware were conducted to assess image quality and dose calculation accuracy against planning CT images for suitability in adaptive planning. Software was considered a "major" level of concern. Human factors validation study was conducted per IEC 62366. The system complies with various IEC and ISO standards related to safety, EMC, quality management, risk management, and biocompatibility.
    • Clinical Testing: No animal or clinical tests were submitted to establish substantial equivalence.

    Since the device described is a medical linear accelerator for radiation therapy and not an AI/ML device, the specific questions regarding AI/ML acceptance criteria, test set details (sample size, provenance, expert adjudication, MRMC studies, standalone performance), ground truth, and training set details are not applicable to the provided text. The document focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance testing and compliance with regulatory standards for a radiation therapy system.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231317

    Validate with FDA (Live)

    Date Cleared
    2023-09-12

    (127 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TrueBeam-TrueBeam STx-Edge: The TrueBeam™, TrueBeam STx and Edge™ Systems are intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere radiation therapy is indicated for adults and pediatric patients. The TrueBeam, TrueBeam STx, and Edge Systems may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, pitutary tumors, spinal cord tumors, hemangioblastoma, orbital tumors, ocular tumors, and skull based tumors), head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, nasopharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal tumors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carcinoma), genitourinary tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as squamous cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, cavernous malformation, trigeminal neuralgia, chordoma, gand hemangiomas), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and skin) and pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, meningioma, metastasis, meduloblastoma, nasopharyngeal tumors, arteriovenous malformation, and skull base tumors), and medically refractory essential tremor (indicated for adults only).

    VitalBeam: VitalBeam® is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediative patients. VitalBeam may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngiona, pituitary tumors, spinal cord tumors, hemangioblastoma, orbital tumors, optic nerve tumors, and skull based tumors), head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, larynx, oropharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, esophageal cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal tumors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carcinoma), genitourinary tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as squamous cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, cavernous malformation, trigeminal neuralgia, chordoma, glomus tumors, and hemangiomas), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and skin) and pediatric tumors (such as glioma, pituitary tumors, hemangioblastoma, craniopharyngioma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, cavernous malformation, and skull base tumors).

    Device Description

    The TrueBeam and VitalBeam Radiotherapy System is a medical linear accelerator that delivered therapeutic radiation to patient in accordance with the physician's prescription.

    The system consists of two major components – a photon, electron and diagnostic kV X-ray radiation beam producing component that is installed in a radiation-shielded vault and a control console area located outside the treatment room.

    AI/ML Overview

    The provided text does not contain information about acceptance criteria or a study that proves a device meets such criteria for an AI/ML component. The document is a 510(k) summary for Varian Medical Systems' TrueBeam, TrueBeam STx, Edge, and VitalBeam radiotherapy systems (K231317).

    It focuses on demonstrating substantial equivalence to a predicate device (K213977) by outlining the device's intended use, indications for use, technological characteristics, and compliance with various safety and performance standards. It explicitly states: "No animal or clinical tests are being submitted to establish substantial equivalence with the predicate device."

    Therefore, I cannot fulfill your request for:

    1. A table of acceptance criteria and reported device performance.
    2. Sample sizes or data provenance for a test set.
    3. Number and qualifications of experts for ground truth establishment.
    4. Adjudication method for a test set.
    5. MRMC comparative effectiveness study details.
    6. Standalone performance data.
    7. Type of ground truth used.
    8. Sample size for the training set.
    9. How ground truth for the training set was established.

    The document discusses hardware and software verification and validation, human factors validation, and electrical safety/EMC testing, all against established standards, but not a specific study evaluating an AI/ML component against performance acceptance criteria. The "Templated-based fiducial detection algorithm for Auto Beam Hold" is mentioned as a new feature, suggesting an algorithmic component, but no specific performance study for this algorithm is detailed in the provided summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223839

    Validate with FDA (Live)

    Date Cleared
    2023-05-22

    (151 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TrueBeam-TrueBeam STx-Edge: The TrueBeam™, TrueBeam STx and Edge™ Systems are intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere radiation therapy is indicated for adults and pediatric patients. The TrueBeam, TrueBeam STx, and Edge Systems may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngioma, pituitary tumors, hemangioblastoma, orbital tumors, ocular tumors, optic nerve tumors, and skull based tumors), head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, nasopharynx, nasopharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, esophageal cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal tumors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carvinoma), gentourinary tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as squamous cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, cavernous malformation, trigeminal neuralgia, chordoma, glomus tumors, and hemangiomas), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, avernous malformation, and skull base tumors), and medically refractory essential tremor (indicated for adults only). VitalBeam: VitalBeam® is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediatric patients. VitalBeam may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngioma, pitutary tumors, spinal cord tumors, hemangioblastoma, orbital tumors, optic nerve tumors, and skull based tumors), head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, larynx, oropharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, esophageal cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal turnors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carcinoma), genitourinary tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as squamous cell, basal cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, trigeminal neuralgia, chordoma, glomus tumors, and hemangiomas), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and skin) and pediatric tumors (such as glioma, pituitary tumors, hemangioblastoma, craniopharyngioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, cavernous malformation, and skull base tumors).

    Device Description

    The TrueBeam and VitalBeam Radiotherapy System is a medical linear accelerator that delivered therapeutic radiation to patient in accordance with the physician's prescription. The system consists of two major components – a photon, electron and diagnostic kV X-ray radiation beam producing component that is installed in a radiation-shielded vault and a control console area located outside the treatment room.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification summary for the Varian Medical Systems TrueBeam™, TrueBeam STx™, Edge™, and VitalBeam radiotherapy delivery systems. This submission primarily focuses on expanding the indications for use to include "medically refractory essential tremor" for the TrueBeam™, TrueBeam STx™, and Edge™ Systems.

    The document states that the subject device has the same technological characteristics as its previously cleared predicate device (K213977). Therefore, the provided text does not describe a new study to prove the device meets acceptance criteria for a newly designed or significantly modified product. Instead, it relies on the previously established performance, safety, and effectiveness of the predicate device and a literature review to support the expanded indication.

    Here's an analysis based on the information provided, recognizing that it's a 510(k) for an expanded indication rather than a de novo submission for a new device:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not provide a specific table of acceptance criteria with numerical performance targets (e.g., accuracy, precision) for the device itself and its reported performance as these were established during the clearance of the predicate device (K213977).

    Instead, the "Summary of Performance Testing" for the current submission K223839 states:

    CriterionReported Performance
    Performance SpecificationsEquivalent to predicate device (K213977)
    Principle of OperationSame as predicate device (K213977)
    Safety ProfileNo changes demonstrated by verification, validation, and safety standards testing (for predicate device K213977)
    Expanded Indication (medically refractory essential tremor) for TrueBeam™, TrueBeam STx™, Edge™Literature review shows similar outcomes and complication rates for SRS treatment of tremor conditions, comparable to traditional surgical options and Gamma Knife. Linear accelerator-based SRS is reported to be safe and efficacious for this indication when other options are unavailable/contraindicated.

    The "acceptance criteria" here implicitly refer to the continued safety and effectiveness of the device for its core functions, as established by the predicate device, and the literature-supported safety and efficacy of Stereotactic Radiosurgery (SRS) for medically refractory essential tremor when delivered by a linear accelerator.

    2. Sample Size Used for the Test Set and Data Provenance:

    • For the device's core performance: Not applicable to this specific submission, as it relies on previous testing of the predicate device (K213977).
    • For the expanded indication (medically refractory essential tremor): The "Summary of Clinical Data" mentions a literature review of over 1300 patients treated with SRS for tremor conditions.
      • Data Provenance: The data comes from published literature, implying a retrospective compilation of studies that likely originated from various countries and institutions. The document does not specify particular countries or whether the studies were prospective or retrospective, but typical clinical literature reviews will encompass both.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • For the device's core performance: Not applicable to this specific submission as it relies on the predicate.
    • For the expanded indication: The "ground truth" for the efficacy and safety of SRS for tremor conditions is established through published clinical outcomes. The decision to include "medically refractory essential tremor" as an indication is based on a review of these published studies. The document does not explicitly state how many experts conducted this literature review for Varian, nor their specific qualifications. However, such reviews are typically conducted by regulatory affairs professionals, sometimes with input from clinical experts (e.g., radiation oncologists, neurosurgeons) who would understand the clinical context and outcomes.

    4. Adjudication Method for the Test Set:

    Not applicable. This submission relies on a literature review of outcomes for an existing therapy, not on a new test set requiring multi-reader adjudication of images or data.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study was not done as described in this document. The submission is not about human readers improving with AI assistance; it's about the safety and efficacy of a medical linear accelerator system for an expanded treatment indication.

    6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance):

    Not applicable. This device is a radiotherapy delivery system, a hardware and software system that always involves human operators (radiation oncologists, medical physicists, radiation therapists) in its use for treatment planning and delivery. It is not an "algorithm-only" diagnostic AI device.

    7. Type of Ground Truth Used:

    • For the device's core performance: Not explicitly stated in this document but would typically involve physical measurements, phantom studies, and possibly clinical outcome data from previous device versions/studies.
    • For the expanded indication (medically refractory essential tremor): Clinical outcomes data from a literature review. This includes reported tremor improvement and complication rates from published studies using both linear accelerators and Gamma Knife for SRS treatment of tremor.

    8. Sample Size for the Training Set:

    Not applicable. This is not an AI/ML device that requires a training set in the typical sense. The "training" for the device's functionality would have been part of its original development and testing (for the predicate device), but this submission does not describe an AI training process.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no mention of a training set for an AI/ML model in this submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213977

    Validate with FDA (Live)

    Date Cleared
    2022-06-03

    (165 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TrueBeam-TrueBeam STx-Edge:
    The TrueBeam, TrueBeam STx, and Edge Systems are intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere radiation therapy is indicated for adults and pediatric patients.
    The TrueBeam, TrueBeam STx, and Edge Systems may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngioma, pituitary tumors, spinal cord tumors, hemangioblastoma, orbital tumors, ocular tumors, optic nerve tumors, and skull based tumors), head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, nasopharynx, sinonasal, salivary gland, and thyroid cancer), thoracic turnors (such as lung cancer, esophageal cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal tumors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carcinoma), genitourinary tumors (such as prostate, bladder, testicular, and kidney), breast turnors, sarcomas, lymphoid tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as squamous cell, basal cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, cavernous malformation, trigeminal neuralgia, chordoma, glomus tumors, and hemangiomas), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and skin) and pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, and skull base tumors).

    VitalBeam:
    VitalBeam® is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediatric patients.
    VitalBeam may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma. meningioma, craniopharyngioma, pituitary tumors, spinal cord tumors, hemangioblastoma, orbital tumors, optic nerve tumors, and skull based tumors), head and neck tumors primary of the head and neck, oral cavity, hypopharynx, larynx, oropharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, esophageal cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal tumors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carcinoma), genitourinary tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as squamous cell, basal cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, trigeminal neuralgia, chordoma, glomus tumors, and hemangiomas), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and skin) and pediatric tumors (such as elioma, evendymoma, pituitary tumors. hemangioblastoma, craniopharyngioma, metastasis, medulloblastoma, nasopharyageal tumors, arteriovenous malformation, cavernous malformation, and skull base tumors).

    Device Description

    The TrueBeam and VitalBeam Radiotherapy System is a medical linear accelerator that delivered therapeutic radiation to patient in accordance with the physician's prescription.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study information based on the provided text, formatted to your requests. However, please note that the document is a 510(k) summary for a radiotherapy system, which focuses on demonstrating substantial equivalence to a predicate device for regulatory approval. It is not a clinical study of an AI/ML device, and therefore many of the requested fields related to AI performance, ground truth establishment, and expert involvement are not applicable or not provided in this type of document.


    Acceptance Criteria and Device Performance

    This document describes a medical device (TrueBeam, TrueBeam STx, Edge, and VitalBeam radiotherapy systems) that has undergone updates. The "acceptance criteria" here refer to conformance to regulatory standards and a demonstration of substantial equivalence to a predicate device, rather than performance metrics for an AI algorithm. The performance is assessed through verification, validation, and safety standards testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is not an AI/ML device with specific performance metrics like sensitivity, specificity, or AUC, the "performance" here refers to compliance with safety and functional standards.

    Acceptance Criterion (Implicit)Reported Device Performance
    Conformance to FDA Quality System Regulation (21 CFR §820)Hardware and software verification and validation testing conducted, and confirmed conformance.
    Conformance to ISO 13485 Quality Management System standardDesign and development, including verification and validation testing, conducted within an established Quality System compliant to EN ISO 13485:2016.
    Conformance to ISO 14971 Risk Management StandardDesign and development, including verification and validation testing, conducted within an established Quality System compliant to EN ISO 14971:2012. Outcome: No Discrepancy Reports (DRs) remaining with a priority of Safety Intolerable.
    Software documentation as per FDA GuidanceProvided documentation as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
    Software Safety LevelSoftware considered a "major" level of concern.
    Biocompatibility (patient-contact materials)No change to patient-contact materials biocompatibility in this medical device. Conformance to ANSI/AAMI/ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1" confirmed.
    Human factors validation (IEC 62366)Validation study conducted according to IEC 62366 to verify that the device "performs well as intended for the intended users, uses, and use environments."
    Electrical safety (IEC 60601-1)Testing conducted. System complies with IEC 60601-1 standards for safety.
    Electromagnetic compatibility (EMC) (IEC 60601-1-2)Testing conducted. System complies with IEC 60601-1-2 standard for EMC.
    Conformance to other recognized standards (e.g., IEC 60601-2-1)Conforms to various listed standards (e.g., IEC 60601-1-6, IEC 62304, IEC 60825-1, IEC 60601-1-3, IEC 61217, IEC 62274, ISO 10993-1, IEC 60601-2-68, ISO 15223-1). Full details refer to Summary of Use of Voluntary Consensus Standards document in Section 09.
    Intended Use and Indications for UseUnchanged from predicate device.
    Principle of OperationNo changes from predicate device.
    Safety ProfileResults of verification, validation, and safety standards testing demonstrate no changes to the safety profile of the device. Outcome: The product conformed to the defined user needs and intended uses, and there were no discrepancy reports (DRs) remaining which had a priority of Safety Intolerable. Therefore, the device is substantially equivalent to the predicate device.

    2. Sample Size Used for the Test Set and Data Provenance

    This document describes a regulatory submission for a radiotherapy system, not an AI/ML diagnostic or prognostic device that typically uses "test sets" of patient data for performance evaluation in the typical sense. The testing described is primarily hardware and software verification and validation, along with compliance to international standards. Therefore, information about a "sample size for a test set" or "data provenance" (country, retrospective/prospective) related to patient data for AI evaluation is not applicable and not provided in this document.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Given that this is not an AI/ML device being evaluated for diagnostic or prognostic accuracy on patient data, the concept of "ground truth" and "experts establishing ground truth" in the context of a "test set" is not applicable here. The "experts" involved would be engineers, physicists, and regulatory specialists confirming compliance with technical and safety standards.

    4. Adjudication Method for the Test Set

    As this is not an AI/ML study involving human interpretation of data where adjudication is needed, an "adjudication method" is not applicable and not described in the document.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study comparing human readers with and without AI assistance was not done and is not relevant for this type of medical device submission. This document pertains to a radiotherapy treatment delivery system, not an AI-assisted diagnostic or interpretation tool.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was done

    This is not applicable as the TrueBeam is a medical linear accelerator for radiation therapy delivery, not an AI algorithm intended for standalone performance evaluation in the way this question implies. Any "algorithms" (e.g., iCBCT reconstruction) are components of the system, and their performance is evaluated as part of the overall system's functional and safety requirements, not as standalone AI products.

    7. The Type of Ground Truth Used

    The "ground truth" in this context is the adherence to established engineering, safety, and regulatory standards, along with the device's functional specifications. It's established through:

    • Engineering specifications and design documents: Dictating how the device should perform.
    • International Consensus Standards: Like IEC 60601 series, IEC 62304, ISO 13485, ISO 14971, which define safety and performance requirements.
    • Verification and Validation Testing: Empirical testing to confirm the device meets these specifications and standards.
    • Clinical use (implied/previous predicate data): The safety and efficacy of radiation therapy itself, and the predicate device, form the basis. The current submission demonstrates that the updated device maintains that safety and efficacy.

    It is not expert consensus on patient pathology, or outcomes data in the sense of a clinical trial for a novel AI diagnostic.

    8. The Sample Size for the Training Set

    This is not applicable as the device is not an AI/ML system that undergoes "training" on a dataset in the typical machine learning sense. The development of the device follows traditional engineering design, verification, and validation processes. While some components like the iCBCT algorithm might involve data for development, the concept of a "training set" for the whole system, as understood for AI, isn't present in this document.

    9. How the Ground Truth for the Training Set was Established

    Since there is no "training set" in the AI/ML context for the device as a whole, this question is not applicable. In the case of component algorithms (like iCBCT), any "ground truth" for their development would be based on known physics, simulation data, and expert validation of imaging quality, but this detail is not provided in a 510(k) summary focused on substantial equivalence.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171733

    Validate with FDA (Live)

    Date Cleared
    2017-07-12

    (30 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TrueBeam and Edge Systems are intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediatric patients.

    The TrueBeam and Edge Systems may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngioma, pituitary tumors, spinal cord tumors, hemangioblastoma, orbital tumors, ocular tumors, optic nerve tumors, and skull based tumors), head and neck turnors (such as unknown primary of the head and neck, oral cavity, hypopharynx, nasopharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal tumors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carcinomaly tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid tumors (such as Hodgkin's lymphoma), skin cancers (such as squamous cell, basal cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, cavernous malformation, trigeminal neuralgia, chordoma, glomus tumors, and hemangiomas), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and skin) and pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, craniopharyngioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, cavernous malformation and skull base tumors).

    Device Description

    The TrueBeamTM Radiotherapy Delivery System is a medical linear accelerator that integrates the previously cleared Trilogy Radiotherapy system and associated accessories into a single device.

    The system consists of two major components, a photon, electron, and diagnostic kV X-ray radiation beam-producing component that is installed in a radiation-shielded vault and a control console area located outside the treatment room.

    AI/ML Overview

    The provided text describes the regulatory clearance for the Varian TrueBeam, TrueBeam STx, and Edge Radiotherapy Delivery Systems but does not contain information about acceptance criteria or a study proving device performance against such criteria.

    The document is a 510(k) premarket notification decision letter from the FDA, along with the device's Indications for Use and a 510(k) summary. It primarily addresses the substantial equivalence of the new devices to a previously cleared predicate device (K162472).

    Here's what can be extracted and what is missing based on your request:

    1. A table of acceptance criteria and the reported device performance:

    • Not provided. The document focuses on regulatory compliance and equivalence to a predicate device, not on specific performance metrics or acceptance criteria for a validation study.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Not provided. There is no mention of a test set with patient data. The "testing" referred to is primarily hardware and software verification and validation, as well as adherence to safety standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    • Not applicable. This information is for studies involving interpretation or diagnosis, which is not the focus of this clearance document.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. As above, this is for studies involving interpretation or diagnosis.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a radiotherapy delivery system, not an AI-assisted diagnostic or interpretive tool. Therefore, an MRMC study related to human reader performance with AI assistance is irrelevant to this submission.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a hardware and software system for delivering radiation therapy, not a standalone algorithm in the sense of AI for diagnosis or interpretation. The document mentions software verification and validation, but this is for the control of the device, not for an autonomous diagnostic function.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not applicable. Ground truth, in the context of diagnostic or interpretive AI, is not relevant to this device's clearance. The "ground truth" for a radiotherapy delivery system would relate to its physical performance (e.g., dose delivery accuracy, beam shaping accuracy) which is assessed through engineering tests and phantom measurements, not human-read data.

    8. The sample size for the training set:

    • Not provided. This device is not an AI/ML model that would have a "training set" in the typical sense. Its software is developed and validated against requirements and specifications, not trained on data to learn a task.

    9. How the ground truth for the training set was established:

    • Not applicable. No training set as per the typical understanding of AI/ML.

    Summary of available information related to validation/testing from the provided document:

    • Type of Testing: Hardware and software verification and validation testing, as well as conformance to FDA Quality System Regulation (21 CFR §820), ISO 13485 (Quality Management System), ISO 14971 (Risk Management), and other FDA recognized consensus standards (listed in the table on page 4). This includes electrical safety (IEC 60601-1) and electromagnetic compatibility (IEC 60601-1-2).
    • Software Level of Concern: The software for this device was considered "major" level of concern, meaning "A failure or latent flaw in the software could directly result in serious injury or death to the patient or operator."
    • Conclusion: The tests demonstrated "conformance to applicable requirements specifications and assured hazard safeguards functioned properly." The conclusion states that the devices are substantially equivalent to their predicate device based on these tests.

    In essence, the provided document is a regulatory approval notice based on demonstrating safety, effectiveness, and substantial equivalence to a predicate device through engineering validation and adherence to recognized standards, rather than a clinical performance study with specific acceptance criteria for diagnostic accuracy or similar metrics.