Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K151945

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-02-18

    (219 days)

    Product Code
    Regulation Number
    872.6300
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Triumph Resin Dam is a resin-based, light-cured protective barrier material recommended to protect dental materials or tissues. For Use only by a Licensed Dentist. Rx Use Only.

    Device Description

    Triumph Resin Dam is a light-cured, resin dam material recommended to block out or protect dental materials or tissues. Triumph contains polymerizable monomers that when light-cured form a protective barrier. When set the resin has a low strength allowing its removal from the protected surfaces.

    AI/ML Overview

    This document is a 510(k) summary for a medical device called "Triumph Resin Dam," which is a resin-based, light-cured protective barrier material for dental use. The summary focuses on demonstrating substantial equivalence to a predicate device ("OpalDam") rather than presenting a study proving performance against acceptance criteria in the typical sense of a clinical trial for diagnostic AI.

    However, I can extract the relevant information from the document to construct an answer aligned with your request, interpreting "acceptance criteria" as the criteria for demonstrating substantial equivalence and "study" as the comparison tests performed.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    For this type of device (a Class I dental resin dam), the "acceptance criteria" are essentially the characteristics and performance deemed equivalent to a legally marketed predicate device. The document primarily focuses on demonstrating substantial equivalence to the Opal Dam (K971284) and referencing BriteSmile (K010935) for hardness comparison.

    Property / Acceptance Criteria (as per predicate equivalence)Reported Device Performance (Triumph Resin Dam)
    Intended UseResin barrier used to isolate and protect dental materials and tissues. (Matches Opal Dam)
    CompositionUrethane acrylates, silica, titanium dioxide, light curing chemistry. (Different from Opal Dam's Di-urethane di-methacrylate, filler. However, the document states "Formula differences...exist, however Triumph Resin Dam was found to be biocompatible...and has similar technological and performance characteristics.")
    Light Cure CapabilityYes (Matches Opal Dam)
    Light Cure Setting TimeYes, < 20 seconds (Matches Opal Dam)
    Biocompatibility (Cytotoxicity)Grade "0" - "Reactivity. None" (Found to be biocompatible per ISO 10993)
    Biocompatibility (Human Skin Irritation)"No reaction" (Found to be biocompatible per ISO 10993)
    Biocompatibility (Skin Sensitization)Grade "0" - "No visible change" (Found to be biocompatible per ISO 10993)
    Flexibility/RemovabilityFormulated to be flexible enough (not too brittle), allowing removal in one piece. (Similar to Opal Dam, despite hardness differences. Also similar to BriteSmile in this aspect). The document notes "This difference does not effect the ability of either material to perform as a resin barrier."
    Mechanical Properties (Hardness)Hardness was compared to BriteSmile (K010935), indicating Triumph is less hard than Opal Dam but "similiar" in properties to BriteSmile, which is also formulated for one-piece removal.
    Mechanical Properties (Depth of Cure)Greater depth of cure than Opal Dam and BriteSmile. (Although different, this is not stated as a negative and is considered acceptable for substantial equivalence).

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This document does not describe a "test set" in the context of diagnostic performance typically associated with AI. The evaluations are for biocompatibility and physical/mechanical properties.

    • Biocompatibility Tests (Cytotoxicity, Human Skin Irritation, Skin Sensitization): Standardized tests (ISO 10993) were performed. The sample sizes for these specific tests are not detailed in this summary but would be defined by the ISO standard. The provenance of the data is not specified beyond being generated for this submission.

    • Physical/Mechanical Properties (Hardness, Depth of Cure, Setting Time): These are laboratory measurements. Sample sizes are not specified for each measurement, but they would be conducted on samples of the device material. Data provenance is not further detailed.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable to the evaluation of this device. The "ground truth" for biocompatibility is established by the standardized methods of ISO 10993, and for physical properties by laboratory testing procedures. There are no "experts" establishing conventional ground truth in the sense of clinical diagnoses.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. There is no adjudication method described as this is not a diagnostic study involving human interpretation.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a physical dental material, not an AI or diagnostic tool, and therefore no MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical dental material, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's evaluation relies on:

    • Standardized Biological Evaluation Protocols: For biocompatibility, ISO 10993 serves as the "ground truth" framework.
    • Physical/Chemical Measurement Standards: For properties like setting time, depth of cure, and hardness, these are determined through established laboratory measurement techniques, often against internal specifications or industry standards.
    • Comparison to Predicate Device: The primary "ground truth" for regulatory clearance is demonstrating that the device is substantially equivalent to a legally marketed predicate, meaning it performs similarly and raises no new questions of safety or effectiveness.

    8. The sample size for the training set

    Not applicable. This device is a physical dental material, not an AI model, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1