Found 1 results

510(k) Data Aggregation

    K Number
    K192645

    Validate with FDA (Live)

    Date Cleared
    2019-12-23

    (90 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Trigon™ Ti Stand-Alone Wedge Fixation System is intended to be used for internal bone fixation for bone fractures or osteotomies in the foot, such as:

    • Cotton (opening wedge) osteotomies of the medial cuneiform
    • Evans lengthening osteotomies

    The Trigon Ti wedges are intended for use with ancillary fixation.

    The Trigon Ti Stand-Alone Wedge Fixation System is not intended for use in the spine.

    Device Description

    The Trigon Ti Stand-Alone Wedge Fixation System is a family of additively manufactured titanium wedges used for angular correction of small bones of the foot. The wedges incorporate an organized lattice structure and two screw-receiving holes. The system rectangular and kidney shaped wedges may or may not include a window for containing grafting material. The wedges are designed in rectangular and kidney shaped footprints in a range of sizes (16x16mm to 20x22mm) and in multiple thicknesses (5 to 12mm). The associated 2.5mm diameter screws are designed in lengths of 10 to 20mm.

    When used with the provided screw fixation the Trigon wedges may be used with or without ancillary plating. When used without the provided screws Trigon wedges are intended for use with ancillary fixation. Windowless wedge designs that do not allow for graft placement should only be used with ancillary plating.

    AI/ML Overview

    The provided text describes a medical device, the Trigon™ Ti Stand-Alone Wedge Fixation System, and its 510(k) premarket notification. However, it does not contain any information about acceptance criteria or a study proving the device meets those criteria in the context of an AI/ML algorithm or software.

    The document primarily focuses on:

    • Regulatory review: The FDA's substantial equivalence determination for the device.
    • Device description: Physical characteristics, materials, and intended use of the bone wedge system.
    • Predicate devices: Identification of similar devices already on the market.
    • Non-clinical testing: Mechanical testing (compression, compression shear, expulsion, screw strength/pullout) to demonstrate substantial equivalence to predicates.

    Therefore, I cannot fulfill your request for information regarding acceptance criteria and a study proving the device meets those criteria, as it relates to AI/ML performance, because that information is not present in the provided text.

    Specifically, the document does not mention or provide data for any of the following:

    • AI/ML algorithm performance.
    • Test sets, training sets, or data provenance.
    • Experts establishing ground truth.
    • Adjudication methods.
    • Multi-reader multi-case (MRMC) studies.
    • Standalone algorithm performance.
    • Types of ground truth (beyond mechanical test results for a physical device).