510(k) Data Aggregation

    K Number
    K203219

    Validate with FDA (Live)

    Date Cleared
    2020-12-29

    (57 days)

    Product Code
    Regulation Number
    882.5600
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. The Trevo Retriever is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should start within 6 hours of symptom onset.
    2. The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
    3. The Trevo Retriever is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion of the internal carotid artery (ICA) or middle cerebral artery (MCA)-MI segments with smaller core infarcts (0-50 cc for age < 80 years, 0-20 cc for age ≥ 80 years). Endovascular therapy with the device should start within 6-24 hours of time last seen well in patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy.
    Device Description

    The Trevo Retriever consists of a flexible, tapered core wire with a shaped section at the distal end. Platinum markers at the distal end allow fluoroscopic visualization. In addition, the shaped section is also radiopaque. Retriever dimensions are indicated on product label. The Retriever delivery wire has a hydrophilic coating on the distal 101cm length to reduce friction during use. The Retriever has a shaft marker to indicate proximity of Retriever tip relative to Microcatheter tip. A Torque Device is provided with the Retriever to facilitate manipulation and retrieval. The Retriever comes preloaded in an insertion tool to introduce the Retriever into a Microcatheter.

    AI/ML Overview

    This document describes the Stryker Neurovascular Trevo XP ProVue Retriever and Trevo NXT ProVue Retriever, neurovascular mechanical thrombectomy devices reviewed under K203219. The submission confirms substantial equivalence to previous devices (K190779 and K200117) made by the same manufacturer. The core argument for substantial equivalence relies on the fact that the indications for use, materials, manufacturing processes, design, technology, biocompatibility, packaging, and sterilization methods remain the same, with the only change being a modification to the core wire dimension.

    Here's an analysis of the provided information concerning acceptance criteria and the study proving adherence:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance values for each test. Instead, it provides a summary of various bench tests and their "Conclusions," stating that "Device integrity met acceptance criteria," "Deliverability force met acceptance criteria," etc. The specific quantitative acceptance values are not detailed.

    TestTest Method SummaryConclusions (Reported Device Performance)
    Insertion Tool Use, DimensionsEvaluate device integrity after use of insertion tool. Device prepared per IFU, inspected, subjected to 3 runs of simulated use in a clinically relevant vasculature model, and inspected after each run.Device integrity met acceptance criteria.
    Retriever / Microcatheter Deliverability (Track Test- First Push)Measure the force required to track a retriever through a microcatheter placed in a clinically relevant vasculature model.Deliverability force met acceptance criteria.
    Retriever In-Vivo ResheathabilityEvaluate device integrity after re-sheathing into microcatheter. Device prepared per IFU, inspected, subjected to 3 runs of simulated use in a clinically relevant vasculature model, and inspected after each run.Device integrity met acceptance criteria.
    Retractability of Retriever into Balloon Guide CatheterEvaluate device integrity after retracting retriever into Balloon Guide Catheter. Device prepared per IFU, inspected, subjected to 3 runs of simulated use in a clinically relevant vasculature model, and inspected after each run.Device integrity met acceptance criteria.
    Retriever Shaft Kink ResistanceEvaluate device integrity after tracking through a tortuous anatomy. Device prepared per IFU, inspected, subjected to 3 runs of simulated use in a clinically relevant vasculature model, and inspected after each run.Device integrity met acceptance criteria.
    Reloadability into Insertion ToolEvaluate device integrity after reloading retriever into insertion tool. Device prepared per IFU, inspected, subjected to 3 runs of simulated use in a clinically relevant vasculature model, and inspected after each run.Device integrity met acceptance criteria.
    Retriever Shaped Section Torque/Tensile DurabilityDetermine durability of shaped section after multiple torque and tensile loading cycles. Apply two clockwise (CW) or counter clockwise (CCW) revolutions and load proximal 50% of exposed shaped section to 1.7 lbf five times, then pull to failure. Acceptance: Device shall withstand six cycles at 1.7 lbf before fracture.Torque/tensile durability met acceptance criteria.
    Retriever Proximal Curl ResistanceDetermine if curl formation is present after the required revolutions. Apply 5 clockwise (CW) or 5 counter clockwise (CCW) revolutions. Acceptance: Core wire shall not form a curl or fracture.Curl resistance met acceptance criteria.
    Retriever Mid Joint Tensile StrengthDetermine mid joint tensile strength. Identify mid solder joint(s) and subject to tensile testing. Acceptance: Device shall withstand 1.7 lbf minimum before wire or solder joint failure.Mid Joint Tensile met acceptance criteria.
    Retriever platinum wire and joint durabilityDetermine durability of platinum wire weaves and solder joints. Identify Mid-Joint, position tubing with Retriever inside, wrap tubing containing shaped section around Dowel Pin (one layer) until entire shaped section is wrapped around pin, unwrap and repeat for 6 times. Unsheath and examine.Platinum wire and joint met acceptance criteria.
    Retriever platinum wire joint tensile strengthDetermine platinum wire solder joint tensile strength. Free Platinum wires, unweave from Nitinol struts. Subject joint to tensile testing. Acceptance: Joint shall withstand 0.1 lbf minimum before wire or solder joint failure.Platinum wire solder joint met acceptance criteria.
    ISO FractureTesting for fracture based on EN ISO 11070:2014, Annex F. Retriever wire wound around a cylindrical former (<=10x max OD) for at least eight complete turns, then unwound and examined for fractures.Fracture resistance met acceptance criteria.
    ISO FlexureTesting for resistance to damage by flexing based on EN ISO 11070:2014, Annex G. Retriever delivery wire subjected to 20 cycles of repeated reverse bending and straightening, then examined for damage and flaking.Flexure resistance met acceptance criteria.
    Retrievability of Retriever with Intermediate CatheterEvaluate device integrity after retrieval of Retriever with Intermediate Catheter. Device prepared per IFU, inspected, subjected to 3 runs of simulated use in a clinically relevant vasculature model, and inspected after each run.Device integrity met acceptance criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states that for some bench tests (e.g., Insertion Tool Use, Resheathability, Retractability, Kink Resistance, Reloadability, Retrievability), the device was subjected to "3 runs of simulated use." For other tests like torque/tensile durability, curl resistance, and various tensile strength tests, specific run counts or minimum forces are mentioned, but not the total number of devices or components tested to achieve those results.

    The data provenance is bench testing only. No patient data (clinical or retrospective/prospective) was used for evaluation of the modified device in this submission, as per their statement: "No animal or clinical studies were conducted as there is no change to the indications for use or the fundamental scientific technology."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. The tests are bench tests comparing the physical properties of the modified device to the predicate. There is no "ground truth" established by human experts in the context of diagnostic interpretation or clinical outcomes for this submission.

    4. Adjudication Method for the Test Set

    Not applicable. This was predominantly bench testing of device components, not a clinical study requiring adjudication of expert interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device (mechanical thrombectomy retriever), not an AI-powered diagnostic or assistive technology. No MRMC study was conducted.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a mechanical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    For the bench tests, the "ground truth" or reference standard would be the established engineering specifications and performance characteristics of the predicate device, or relevant industry standards (e.g., ISO standards mentioned). The assessment is against these predefined engineering and safety specifications, not against clinical outcomes or pathology reports.

    8. The Sample Size for the Training Set

    Not applicable. This is a mechanical device, not a machine learning algorithm requiring a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As above, this is a mechanical device, not an algorithm based on a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190779

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-06-25

    (90 days)

    Product Code
    Regulation Number
    882.5600
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. The Trevo Retriever is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should start within 6 hours of symptom onset.
    2. The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
    3. The Trevo Retriever is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion of the internal carotid artery (ICA) or middle cerebral artery (MCA)-M1 segments with smaller core infarcts (0-50 cc for age < 80 years, 0-20 cc for age ≥ 80 years). Endovascular therapy with the device should start within 6-24 hours of time last seen well in patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy.
    Device Description

    The Trevo Retriever consists of a flexible, tapered core wire with a shaped section at the distal end. Platinum markers at the distal end allow fluoroscopic visualization. In addition, the shaped section is also radiopaque. Retriever dimensions are indicated on product label. The Retriever has a hydrophilic coating to reduce friction during use. The Retriever has a shaft marker to indicate proximity of Retriever tip relative to Microcatheter tip. A torque device is provided with the Retriever to facilitate manipulation and retrieval. An insertion tool is provided to introduce the Retriever into a Microcatheter. The Insertion Tool is a sheath in which the Retriever comes preloaded. Once half the retriever's length is inserted into the microcatheter, the insertion tool is removed. Retrievers have a modified proximal end that permits attachment of the Abbott Vascular DOC Guide Wire Extension (REF 22260). Joining the Guide Wire Extension to the Retriever facilitates removal or exchange of a catheter while maintaining the Retriever position in anatomy. After the exchange has been completed, the extension can be detached.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Stryker Neurovascular Trevo XP ProVue Retriever. This document primarily focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a detailed clinical study with acceptance criteria for an AI device. The information requested in your prompt (e.g., acceptance criteria for device performance, sample sizes for test/training sets, expert qualifications, ground truth establishment methods, MRMC studies) is typically found in regulatory submissions for AI/ML-based medical devices, which assess algorithm performance.

    The provided document does not contain the specific type of information needed to fully answer your request regarding acceptance criteria and performance data for an AI-powered device.

    However, I can extract the information that is present concerning the Trevo Retriever, which is a mechanical thrombectomy device, and interpret it in the context of device performance, even if it's not an AI device.

    Let's break down what is available and what is not:

    What is Available (relevant to a physical medical device, not AI/ML):

    • Device Type: Neurovascular Mechanical Thrombectomy Device (a physical medical device, not an AI/ML algorithm).
    • Purpose: To restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke.
    • Performance Data - Non-Clinical:
      • Test: Simulated Use
      • Test Method Summary: "Simulated Use testing utilized a neurovascular model with a re-circulating water bath at 37℃ to simulate the human arterial circulation and to assess the device's ability to retrieve the clot and achieve recanalization."
      • Conclusion/Performance: "Simulated Use met acceptance criteria."
    • Performance Data - Clinical:
      • Data Source: Analysis of Real World Data (RWD) from the Trevo Retriever Registry.
      • Purpose: To assess safety or effectiveness on the combination use of Trevo Retriever and intermediate catheter aspiration approach.
      • Additional Support: Comprehensive review of relevant clinical literature.

    What is NOT available (and would be expected for an AI/ML device as per your prompt):

    • A table of acceptance criteria and reported device performance for an AI/ML algorithm. The document only states that "Simulated Use met acceptance criteria" without specifying the precise quantitative criteria or results.
    • Sample sizes used for the test set and data provenance (e.g., country of origin, retrospective/prospective). The "Simulated Use" refers to a bench test, not a patient test set, and the "Trevo Retriever Registry" is mentioned for clinical data, but no specifics about its use as a "test set" for performance evaluation against acceptance criteria are given.
    • Number of experts used to establish ground truth for the test set and their qualifications. This is relevant for AI/ML evaluation.
    • Adjudication method (e.g., 2+1, 3+1, none) for the test set.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and its effect size. This is specific to AI clinical utility.
    • If a standalone (algorithm only without human-in-the-loop performance) was done.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc.).
    • The sample size for the training set.
    • How the ground truth for the training set was established.

    Summary in relation to your prompt, given the provided document:

    Since the provided text describes a physical medical device (a thrombectomy device) and not an AI/ML algorithm, most of the requested points related to AI/ML software validation (like training set, ground truth experts, MRMC studies) are not applicable and thus not present in the document.

    The "acceptance criteria" mentioned in the non-clinical performance section only states that the "Simulated Use met acceptance criteria," but it does not specify what those criteria were (e.g., success rate, time to recanalization, etc.) nor does it provide the exact performance metrics achieved.

    Therefore, I cannot construct the table or provide the detailed answers to most of your questions based on the provided text, as it does not pertain to an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K150616

    Validate with FDA (Live)

    Date Cleared
    2015-06-12

    (94 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Trevo XP ProVue Retriever (4x30mm) is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

    Device Description

    The Trevo XP ProVue Retriever (4x30mm) consists of a flexible, tapered core wire with a shaped section at the distal end. It is designed to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke. Radiopaque platinum wires in the shaped section and radiopaque markers on the distal end allow fluoroscopic visualization. Retriever dimensions are indicated on the product label. The Retriever has a hydrophilic coating to reduce friction during use. A torque device is provided with the Retriever to facilitate manipulation. The torque device is used to lock the core wire to the microcatheter during the procedure. Locking of the torque device to the wire allows the microcatheter and Retriever to be retracted as a system during clot retrieval. An insertion tool is provided to introduce the Retriever into a microcatheter. The proximal end of the device is compatible with the Abbott guide wire extension to facilitate removal or exchange of a catheter while maintaining the Retriever position in the vessel.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Trevo XP ProVue Retriever (4x30mm). This document focuses on demonstrating substantial equivalence to a predicate device, rather than proving performance against specific acceptance criteria for a new clinical outcome. Therefore, many of the requested items (e.g., sample size for test set, number of experts for ground truth, MRMC study, training set details) are not explicitly present in the provided text as they are typically associated with de novo approvals or significant device modifications requiring new clinical evidence.

    However, I can extract information related to the acceptance criteria for bench and animal testing, and the study methodologies employed to demonstrate substantial equivalence.

    Here's the information derived from the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for each test were "meets acceptance criteria." The device performance is generally stated as having met these unspecified criteria. The document emphasizes substantial equivalence to the predicate device through bench and animal testing.

    TestAcceptance CriteriaReported Device Performance
    Dimensional VerificationMeets acceptance criteriaDimensional verification meets acceptance criteria.
    Retriever Mid Joint Tensile StrengthMeets acceptance criteriaRetriever Mid Joint Tensile Strength meets acceptance criteria.
    Retriever Tip Tensile StrengthMeets acceptance criteriaRetriever Tip Tensile Strength meets acceptance criteria.
    Retriever Shaped Section Radial ForceMeets acceptance criteriaRetriever Shaped Section Radial Force meets acceptance criteria.
    Retriever / Vessel Interaction (Tip Flexibility)Meets acceptance criteriaRetriever/Vessel Interaction (Tip Flexibility) meets acceptance criteria.
    Retriever Torque Tensile DurabilityMeets acceptance criteriaRetriever Torque Tensile Durability meets acceptance criteria.
    Retriever Platinum Wire Joint StrengthMeets acceptance criteriaRetriever Platinum Wire Joint Strength meets acceptance criteria.
    Retriever Platinum Wire and Joint DurabilityMeets acceptance criteriaRetriever Platinum Wire and Joint Durability meet acceptance criteria.
    RadiopacityMeets acceptance criteriaRadiopacity meets acceptance criteria. (Due to same platinum weave wires and configuration as predicate).
    Retriever / Microcatheter DeliverabilityMeets acceptance criteriaRetriever/Microcatheter Deliverability meets acceptance criteria.
    Particulate EvaluationAll samples meet acceptance criteriaAll samples meet acceptance criteria.
    Coating Integrity EvaluationAll samples meet acceptance criteriaAll samples meet acceptance criteria.
    Simulated UseMeets acceptance criteriaSimulated Use meets acceptance criteria.
    Animal Studies (Safety and Usability)Met acceptance criteriaThe safety and usability results from the animal studies met acceptance criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    • Bench Testing: The sample sizes for individual bench tests are not explicitly stated in the summary document. The phrasing "All samples meet acceptance criteria" suggests that a sufficient number of devices were tested per manufacturing and quality control standards, but the exact N is not provided.
    • Animal Testing: The document mentions that a "larger diameter device (6x30mm) than the subject device (4x30mm)" was used for animal studies as a "worst case." The number of animals used is not specified.
    • Data Provenance: The bench and animal studies were conducted by Concentric Medical, Inc. (the submitter). These are prospective studies performed specifically for this 510(k) submission. No country of origin for the data is explicitly mentioned beyond the submitter's location in Mountain View, California, USA.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Bench Testing: No external experts are mentioned for establishing ground truth for the bench tests. These tests are industrial performance evaluations against internal specifications.
    • Animal Testing: The animal studies were conducted in compliance with GLP regulation (21 CFR Part 58), which implies qualified personnel (e.g., veterinarians, pathologists, study directors) were involved in the conduct and evaluation, but their specific number or qualifications are not detailed.
    • Clinical Data: No clinical study was performed for this 510(k) submission (stated in the document as "No clinical study was performed as there is no change to the indications for use or the fundamental scientific technology for the subject device."). Therefore, no experts were used to establish ground truth from a clinical test set.

    4. Adjudication Method for the Test Set

    Not applicable, as no human-read test set or clinical study requiring adjudication is described. The evaluations were based on quantifiable bench tests and animal observations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. The document explicitly states: "No clinical study was performed as there is no change to the indications for use or the fundamental scientific technology for the subject device."

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to a medical device in the context of hardware, specifically a "thrombus retriever" catheter. The concept of "standalone algorithm performance" is not applicable here as it is not an AI/software device.

    7. The Type of Ground Truth Used

    • Bench Testing: Ground truth for bench tests is based on engineering specifications and physical measurements, which are inherently objective and quantitative (e.g., tensile strength, radial force, dimensions).
    • Animal Testing: Ground truth in animal studies would typically involve pathological examination and observations for adverse events (e.g., "evidence of arterial transmural dissection or perforation"). This is based on direct biological and histological assessment.

    8. The Sample Size for the Training Set

    Not applicable. This device is a physical medical instrument (thrombus retriever), not a machine learning algorithm requiring a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. This device is a physical medical instrument (thrombus retriever), not a machine learning algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K143077

    Validate with FDA (Live)

    Date Cleared
    2015-04-06

    (161 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Trevo XP ProVue Retriever (6x25 mm) is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

    Device Description

    The Trevo XP ProVue Retriever (6x25mm) consists of a flexible, tapered core with a shaped section at the distal end. It is designed to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke. Radiopaque platinum wires in the shaped section and radiopaque markers on the distal end allow fluoroscopic visualization. Retriever dimensions are indicated on the product label. The Retriever has a hydrophilic coating to reduce friction during use. A torque device and an insertion tool are provided with the Retriever. The proximal end of the device is compatible with the Abbott guide wire extension to facilitate removal or exchange of a catheter while maintaining the Retriever position in the vessel.

    AI/ML Overview

    The provided text describes the Trevo XP ProVue Retriever (6x25 mm) and summarizes the testing conducted to support its substantial equivalence to a predicate device (Trevo XP ProVue Retriever (4x20mm)). However, it does not contain specific acceptance criteria for each test or detailed performance metrics in the format of a table comparing criteria to reported performance. Instead, it generally states that tests "meet acceptance criteria" or provides qualitative conclusions.

    Therefore, for aspects requiring quantitative acceptance criteria and reported device performance (like a comparative table), a direct answer cannot be fully generated as that level of detail is not explicitly present. Similarly, information regarding sample size for test sets, data provenance, expert involvement for ground truth, adjudication methods, MRMC studies, standalone performance with specific metrics, or training set details are not provided in the given document.

    Based on the available information, here's what can be inferred and stated:

    1. A table of acceptance criteria and the reported device performance

    The document lists various tests and provides a "Conclusions" column for each, generally stating that the device "meets acceptance criteria." It does not provide the numerical acceptance criteria themselves or quantitative reported device performance in most cases.

    TestAcceptance Criteria (Not explicitly stated quantitatively in the document for most tests)Reported Device Performance (Qualitative conclusions)
    Dimensional VerificationNot specified quantitativelyDimensional verification meets acceptance criteria.
    Retriever Mid Joint Tensile StrengthNot specified quantitativelyRetriever Mid Joint Tensile Strength meets acceptance criteria.
    Retriever Tip Tensile StrengthNot specified quantitativelyRetriever Tip Tensile Strength meets acceptance criteria.
    Retriever Shaped Section Radial ForceNot specified quantitativelyRetriever Shaped Section Radial Force meets acceptance criteria.
    Retriever/Vessel Interaction (Tip Flexibility)Not specified quantitativelyRetriever/Vessel Interaction (Tip Flexibility) meets acceptance criteria.
    Retriever Torque Tensile DurabilityNot specified quantitativelyRetriever Torque Tensile Durability meets acceptance criteria.
    Retriever Platinum Wire Joint StrengthNot specified quantitativelyRetriever Platinum Wire Joint Strength meets acceptance criteria.
    Retriever Platinum Wire and Joint DurabilityNot specified quantitativelyRetriever Platinum Wire and Joint Durability meet acceptance criteria.
    RadiopacityVisual assessment of device under fluoroscopy.Radiopacity meets acceptance criteria.
    Retriever / Microcatheter DeliverabilityNot specified quantitativelyRetriever/Microcatheter Deliverability meets acceptance criteria.
    Particulate EvaluationMaximum allowable number/size of particulates generated during simulated use.All samples meet acceptance criteria (i.e., total number and size of particulates generated were within limits for ≥10µm, ≥25µm, and ≥50µm size ranges).
    Coating Integrity EvaluationNo delamination at coating edges, minimal defects along coating length after tracking.Neither Baseline nor Post-tracking test articles exhibited delamination at coating edges. Baseline showed no defects; Post-tracking showed minimal defects. Coating functions as intended and exhibited appropriate integrity.
    Simulated UseSuccessful removal of simulated thrombus, no adverse events.Simulated Use meets acceptance criteria (i.e., device successfully retrieved simulated thrombus in a neurovascular model).
    Animal Studies (Acute)No arterial transmural dissection or perforation due to Stentriever use.Histopathology consistent with arterial healing after routine catheterization. Angiography showed no evidence of vessel dissection, perforation, or thrombosis.
    Animal Studies (Chronic)No arterial transmural dissection or perforation; no stenosis, vessel irregularity, or pseudoaneurysm at treatment sites.Histopathology consistent with arterial healing. No evidence of vessel dissection or perforation. Angiography showed no evidence of dissection, perforation, stenosis, vessel irregularity, intimal flap, or pseudoaneurysm at Day 0 and Day 30. Animal studies meet acceptance criteria and are comparable to the predicate device.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not explicitly stated for most tests.
      • For Animal Studies: 12 test articles (devices) were deployed (4 per animal) across 6 treatment runs in both internal maxillary arteries (IMAs) of animals, resulting in the assessment of 6 arteries (2 per animal) for both acute and chronic studies.
    • Data Provenance:
      • Simulated Use: Conducted using a silicone neurovascular model cast from actual human neurovascular arteries.
      • Animal Studies: Conducted in swine, in compliance with GLP regulation (21 CFR Part 58). This is prospective animal data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The tests primarily involve engineering performance metrics, simulated use outcomes, and animal study observations (angiography, histopathology) conducted by unspecified personnel. There is no mention of external human experts establishing ground truth for the test set in the context of diagnostic interpretation or similar tasks.

    4. Adjudication method for the test set

    This information is not provided in the document. The tests described are primarily physical/mechanical performance tests, simulated use, and animal studies. Adjudication methods like "2+1" or "3+1" are typically used for establishing ground truth in clinical or image-based diagnostic studies involving human interpretation.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no mention of an MRMC comparative effectiveness study, AI assistance, or human reader performance improvement in this document. The device is a mechanical thrombectomy retriever, not an AI-powered diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable as the device is a medical instrument (thrombectomy retriever) and not an algorithm or AI system. Its performance is inherent to the device itself (mechanical, physical, and biocompatibility characteristics), not an algorithm that operates standalone. The animal and simulated use studies demonstrate its "standalone" performance as a device.

    7. The type of ground truth used

    • For physical/mechanical tests (e.g., tensile strength, radial force, deliverability, particulate evaluation): The "ground truth" is defined by established engineering specifications and physical measurements, and compliance with these objective criteria.
    • For Radiopacity: Visual assessment under fluoroscopy.
    • For Coating Integrity: Visual assessment, comparison of test articles, and consideration of other correlated tests (lubricity, durability, particulate generation).
    • For Simulated Use: The ability to retrieve simulated thrombus in a physical model, guided by procedural instructions derived from the Instructions for Use.
    • For Animal Studies: Histopathology and angiography results are used to assess vessel response, dissection, perforation, thrombosis, stenosis, etc., against predefined safety criteria.

    8. The sample size for the training set

    This information is not provided and is not applicable to the type of device (a medical instrument). Training sets are typically associated with machine learning or AI models, which this device is not.

    9. How the ground truth for the training set was established

    This information is not provided and is not applicable for the same reasons as #8.