510(k) Data Aggregation

    K Number
    K162960

    Validate with FDA (Live)

    Device Name
    TotalC-Ram
    Manufacturer
    Date Cleared
    2017-03-14

    (141 days)

    Product Code
    Regulation Number
    872.3275
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    It is intended for the permanent cementation of ceramic (including zirconia), composite and metal-based posts, crowns, bridges, veneers, inlays and onlays.

    Device Description

    TotalC-Ram is permanent, radio-opaque dual cured (auto/photopolymerizable) self-etch and self-adhesive resin cement.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a dental cement called "TotalC-RAM." It outlines the device's characteristics and compares it to predicate devices to demonstrate substantial equivalence to legally marketed devices.

    This document does not describe a study involving AI/ML medical devices based on imaging data or other complex data types that would require acceptance criteria related to accuracy, sensitivity, specificity, or human reader performance. Therefore, most of the requested information (points 1-9) about acceptance criteria, test sets, expert ground truth, MRMC studies, standalone performance, and training sets is not applicable to this document.

    Here's why and what information can be extracted:

    • This is a review for a dental cement (TotalC-RAM), which is a physical material used in dental procedures.
    • The comparison is based on physical and mechanical properties (e.g., shear bond strength, flexural strength, solubility, working time, setting time, film thickness, shelf life) and chemical composition.
    • Section 11, "Brief Description of Performance and Biocompatibility Testing," explicitly states that the submission is an Abbreviated 510(k) and relies on conformity with guidance documents and recognized consensus standards (like ISO standards for dental materials).
    • Section 12, "Clinical Performance Data," explicitly states: "Not applicable. Clinical performance testing has not been performed for the subject device." This confirms that there was no human reader study, no ground truth established by experts interpreting imaging, and no AI performance metrics.

    Therefore, it is impossible to populate the requested table and answer the specific questions (1-9) as they pertain to the evaluation of an AI-based system's performance. The information requested is relevant for software-based medical devices, especially those that analyze medical images or other clinical data to provide diagnostic or prognostic insights.

    To directly address your request, here's what can be inferred or stated based on the document's content, acknowledging the mismatch in context:

    1. A table of acceptance criteria and the reported device performance

    The "acceptance criteria" here are not in terms of AI performance metrics (like sensitivity/specificity) but rather the physical and mechanical properties being comparable to predicate devices or meeting ISO standards. The document doesn't explicitly state quantitative acceptance criteria for each property for the purpose of a performance study, but rather lists the measured performance of the device and compares it to predicates. The "acceptance" is that these values are within a range considered safe and effective, and contribute to substantial equivalence.

    PropertySubject Device Performance (TotalC-RAM)Predicate TotalCem PerformancePredicate DentoCem PerformanceImplicit "Acceptance Criteria" (Substantial Equivalence/ISO Standards)
    Chemical CompositionUDMA, Bis-GMA, TEGDMA (Matrix); Barium Glass, Fumed Silica (Fillers)Same as Subject DeviceSame as Subject DeviceCompatible with predicate, safe and effective materials as per standards.
    DeliveryAutomixAutomixAutomixFunctionally similar to predicates.
    Curing MethodLight/Self-CureLight/Self-CureLight/Self-CureFunctionally similar to predicates.
    RadiopacityYesYesYesVisible on X-rays, similar to predicates.
    Shear Bond Strength to Zirconia (MPa)11.22.83.8Demonstrated higher values than predicates, which supports its specific indication for zirconia cementation. While higher here, it's still considered substantially equivalent for the broader dental cement category.
    Flexural Strength (MPa)125.5170151Within acceptable range for dental cements, meeting ISO 4049 requirements.
    Solubility (µg/mm3)LOWLOWLOWLow enough to meet ISO standards, indicating stability.
    Working time at 22°C (min)4.542.5Within a practical range for dental procedures, similar to predicates.
    Setting time at 37 °C (min)2.534Within a practical range for dental procedures, similar to predicates.
    Film thickness (µg/mm3)101010Thin enough for effective seating of restorations, meeting ISO standards.
    Shelf life (Years)222Standard shelf life for such materials.

    Note: The "acceptance criteria" for a physical dental cement are primarily whether its properties fall within established ranges defined by international standards (e.g., ISO 4049, ISO 9917-1) and whether it is "substantially equivalent" to predicate devices, meaning it performs similarly and does not raise new questions of safety or effectiveness. The increased shear bond strength for zirconia is highlighted as a performance advantage, but not as a strict "acceptance criterion" that it must surpass predicates.

    2. Sample size used for the test set and the data provenance: Not Applicable. This is a material properties comparison, not a data-driven AI study. Laboratory tests are performed on material specimens.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not Applicable. Ground truth for material properties is established through standardized laboratory testing, not expert consensus on interpretations.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not Applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not Applicable. No human readers or AI assistance involved.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not Applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Ground truth for material properties is established through standardized laboratory testing methods as outlined by ISO standards (e.g., measuring flexural strength, solubility, bond strength under controlled laboratory conditions).

    8. The sample size for the training set: Not Applicable. This product is not an AI/ML device.

    9. How the ground truth for the training set was established: Not Applicable. This product is not an AI/ML device.