Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K210436

    Validate with FDA (Live)

    Date Cleared
    2021-04-12

    (59 days)

    Product Code
    Regulation Number
    886.5925
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TOTAL30™ (lehfilcon A) spherical soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes with up to approximately 1.50 diopters (D) of astigmatism that does not interfere with visual acuity.

    TOTAL30™ Asphere (lehfilcon A) soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes with up to approximately 1.50 diopters of astigmatism that does not interfere with visual acuity.

    TOTAL30™ for Astigmatism (lehfilcon A) toric soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes with up to 6.00 diopters (D) of astigmatism.

    TOTAL30™ Multifocal (lehfilcon A) soft contact lenses are indicated for the optical correction of presbyopia with or without refractive ametropia (myperopia) in phakic or aphakic persons with non-diseased eyes who may require a reading addition of +3.00 diopters (D) or less and who may have up to approximately 1.50 diopters (D) of astigmatism that does not interfere with visual acuity.

    TOTAL30™ Multifocal Toric (lehfilcon A) soft contact lenses are indicated for the optical correction of presbyopia with or without refractive ametropia (myopia and byperopia) in phakic persons with non-diseased eves who may require a reading addition of +3.00 diopters (D) or less and who may have up to 6.00 diopters (D) of astigmatism.

    The lenses are to be prescribed for daily wear, with removal for cleaning and disinfection (chemical, not heat) prior to reinsertion, or disposal, as recommended by the eve care professional. Lenses should be discarded with a new pair each month, or more often, if recommended by the eye care professional.

    Device Description

    The subject device, TOTAL30™ (lehfilcon A) soft contact lenses, are made from a surface treated, silicone containing hydrogel lens material that is approximately 55% water and 45% lehfilcon A. The color additive Reactive Blue 247 is added to the lens material to create a light blue-green edgeto-edge color to make it easier to see when handling. The lenses contain two benzotriazole monomers to block UVA and UVB radiation and, additionally, reduce transmittance in the range from 380 nm to 450 nm.

    Lehfilcon A represents a Group 5B silicone hydrogel contact lens material ("enhanced oxygen permeable materials') according to ISO 18369-1:2017, Ophthalmic optics - Contact lenses - Part 1: Vocabulary, classification system and recommendations for labelling specifications, as follows:

    Group Suffix: 5B
    Hydrogel Material: High water subgroup
    Description: A subgroup of Group 5 which contains more than 50 % water and no ionic monomer or oligomer at pH 6 to pH 8

    Lehfilcon A lens designs include spherical, asphere, toric, multifocal toric contact lenses in the following parameter ranges:
    Diameter Range: 13.0 to 15.0 mm
    Base Curve Range: 8.0 to 9.2 mm
    Power Range: -20.00 D to +20.00 D
    Center Thickness: varies with design and power (Example: 0.08 mm for -3.00 D spherical)
    Cylinder Power (Toric) -0.25 D to -10.00 D
    ADD Power (Multifocal) LO, MED, HI

    Lenses have the following properties:
    Refractive index: 1.40 (hydrated)
    Water content: 55% by weight in normal saline
    Surface water content: ≥ 90%
    Oxygen permeability: 123 x 10 -11 [(cm2/sec)(ml O2 /ml.mmHg)] measured at 35 °C (normalized Dk-Polarographic method)
    Light transmittance: >85%
    UV Transmittance: The transmittance characteristics are less than 1% in the UVB range of 280 nm to 315 nm and less than 10% in the UVA range of 315 to 380 nm for the entire power range.

    Lehfilcon A contact lenses are supplied sterile in sealed blister packs containing isotonic phosphate buffered saline solution (PBS) with polymeric wetting agents.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for Alcon's TOTAL30 contact lenses. It details the steps taken to demonstrate substantial equivalence to a predicate device, CooperVision Biofinity.

    Here, we will synthesize the information to describe the acceptance criteria and the study that proves the device meets them, focusing on aspects relevant to device performance and ground truth establishment.

    Based on the provided document, the acceptance criteria and proof of adherence are primarily demonstrated through a comparison against a predicate device and adherence to established ISO standards and FDA guidance for contact lenses, rather than a typical AI/software device evaluation with specific performance metrics like sensitivity, specificity, etc.

    However, we can infer "acceptance criteria" as meeting the safety and performance standards demonstrated by the predicate device and relevant industry standards. The "study that proves the device meets the acceptance criteria" refers to the nonclinical and clinical performance testing.

    1. A table of acceptance criteria and the reported device performance:

    Since this is a contact lens submission (medical device), the "acceptance criteria" are not reported as specific numerical thresholds for metrics like accuracy, sensitivity, or specificity as would be common for an AI/software device. Instead, they relate to demonstrating comparable safety, effectiveness, and performance to a legally marketed predicate device and compliance with applicable standards. The "reported device performance" is largely qualitative and comparative.

    Here's an interpretation based on the document:

    Acceptance Criterion (Inferred from Regulatory Requirements/Predicate Comparison)Reported Device Performance (TOTAL30 Lehfilcon A)
    Safety:
    Biocompatibility (non-toxic, non-irritating, non-sensitizing)Confirmed through cytotoxicity, ocular irritation/toxicity (rabbits), sensitization, acute systemic toxicity (mice), and genotoxicity testing. Concluded to be non-toxic, non-irritating, and non-sensitizing. Existing biocompatibility data for packaging materials also applies.
    Sterility and StabilitySupplied sterile in sealed blister packs. Ongoing stability study demonstrates lenses remain sterile and stable for the labeled expiration date.
    Compatibility with Lens Care ProductsConfirmed compatibility with commonly available cleaning/disinfection solutions (peroxide, MPDS), preserved/unpreserved saline, and rewetting drops per ISO 11981:2017. Uptake/release of preservatives (PQ, Aldox, PHMB, alexidine) comparable to control lenses per ISO 11986:2017.
    Effectiveness/Performance:
    Comparable Clinical Performance (Vision, Comfort, Fit, Handling, Health)Clinical study showed similar clinical performance between TOTAL30 (lehfilcon A) and the predicate Biofinity (comfilcon A) in clinically relevant areas of vision, comfort, fit, handling, and health when worn on a daily wear basis. Primary effectiveness endpoint (Snellen VA at distance) showed similar performance.
    Physical-Chemical PropertiesEstablished per GxP conditions and ISO 18369-2, -3, -4 for: Refractive Index (1.40), Oxygen Permeability (123 x 10^-11), Ion Permeability, Mechanical Properties, Wetting Contact Angle, Transmittance Properties (>85% light, <1% UVB, <10% UVA), Percent Water Content (55%), Residuals and Extractables, Lens Parameters, Package Saline Properties (pH, osmolality). Concluded to be consistent with industry-marketed lenses and equivalent to predicate.
    Meeting Intended UseIndicated for optical correction of refractive ametropia/presbyopia for specified conditions (myopia, hyperopia, astigmatism, reading addition). Demonstrated as safe and effective for daily wear given the aforementioned nonclinical and clinical results.
    Substantial Equivalence to Predicate DeviceConcluded that nonclinical and clinical data demonstrate the subject device is "as safe, as effective, and performs as well as or better than the legally marketed predicate device." Any differences do not adversely affect safety and effectiveness.
    Compliance with Regulatory Standards/GuidanceAll testing conducted in accordance with May 1994 FDA 510(k) Guidance Document for Class II Contact Lenses and ISO standards (ISO 18369-1:2017, ISO 9394, ISO 10993 series, ISO 11981:2017, ISO 11986:2017, ISO 11980:2012, 21 CFR Part 58 GLP). Biocompatibility testing conducted per GLP, and physical-chemical testing per GxP.

    2. Sample size used for the test set and the data provenance:

    • Sample Size (Clinical Study): 119 subjects enrolled, 118 randomized and exposed to study lenses (236 eyes).
      • 78 subjects (test group: lehfilcon A)
      • 40 subjects (control group: comfilcon A)
    • Data Provenance:
      • Country of Origin: US (8 study sites in the US).
      • Retrospective or Prospective: Prospective, randomized, stratified, controlled, double-masked, parallel group study.
    • Nonclinical/Bench Testing: The document does not specify exact sample sizes for each nonclinical test (e.g., number of lenses tested for oxygen permeability). It refers to "a series of" tests performed under GxP conditions and according to ISO standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This is not applicable in the context of this 510(k) submission. "Ground truth" in this context is established by the accepted methods of evaluating contact lens safety and performance, which include:

    • Clinical observation and objective measurements: Performed by eye care professionals at the study sites (though their qualifications and numbers are not detailed, it's implied they are qualified for conducting such clinical trials). The primary effectiveness endpoint was Snellen VA (visual acuity) at distance, which is an objective measurement. Other endpoints included manifest refraction, keratometry, and biomicroscopy findings.
    • Subjective ratings: From subjects regarding vision, comfort, and handling.
    • Nonclinical/Bench testing: Results from laboratory measurements according to established ISO standards.
    • Adverse event reporting: Monitored throughout the study.

    There isn't a "ground truth" derived from a panel of independent expert adjudicators as one might see in an imaging AI study.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable. This is a clinical study for a medical device (contact lens), not a test set for an AI algorithm requiring adjudication of interpretations. The study involved direct clinical observation, objective measurements, and subject reporting.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is a submission for a contact lens, not an AI-powered diagnostic device that assists human readers/interpreters. The clinical study was a direct comparison between the new contact lens and a predicate contact lens.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. TOTAL30 is a physical medical device (contact lens), not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for evaluating the contact lens's performance and safety is multifaceted and comprises:

    • Objective Clinical Measurements: Snellen Visual Acuity (VA) at distance, manifest refraction, keratometry, biomicroscopy findings (e.g., assessing ocular health, tissue responses).
    • Subjective Clinical Data: Patient-reported outcomes on vision, comfort, and handling.
    • Nonclinical/Bench Test Results: Quantitative measurements of physical and chemical properties (e.g., oxygen permeability, water content, refractive index, transparency, mechanical properties) obtained through standardized laboratory methods.
    • Biocompatibility Test Results: Laboratory and animal study data demonstrating absence of toxicity, irritation, or sensitization.
    • Comparison to Predicate Device: The performance and safety profile of the new lens are benchmarked against an already legally marketed, established predicate device.

    Essentially, the "ground truth" is established by comprehensive conformance to recognized standards and demonstrated equivalence to a safe and effective predicate device across multiple domains (biocompatibility, physical properties, and clinical performance).

    8. The sample size for the training set:

    Not applicable. This is a contact lens, not an AI algorithm. There is no concept of a "training set" for this type of device.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for a contact lens.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1