Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K223231

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-02-24

    (129 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Ti-Largo™ Cervical Interbody Cages are indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. Ti-Largo™ Cervical Interbody Cages are used to facilitate intervertebral body fusion in the cervical spine and are placed via an anterior approach at the C2 to T 1 disc levels using autograft bone. Ti-Largo™ Cervical Interbody Cages are to be used with supplemental fixation. Patients should have at least six (6) weeks of non-operative treatment with an intervertebral cage.

    Device Description

    FloSpine Ti-Largo™ Cervical Interbody Cages are keystone shaped devices that have a central lumen and smaller graft windows throughout the structure of the implant. The implants are additively manufactured (ADM) using grade 23 titanium alloy powder per ASTM F3001. The implants are provided sterile (gamma radiation, 25 kGy minimum dose), are intended for single use only, and should not reused under any circumstances. Components from this system should not be used in conjunction with components from other company products.

    Ti-Largo™ Cervical Interbody Cages are intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the cervical spine. The Ti-Largo™ Cervical Interbody System was developed for anterior cervical fusion with implants available in keystone shaped footprints ranging in size between 11 x 14mm and 16 x 18mm, thickness ranging from 5 to 14mm in 1 mm increments, 7degree lordosis for all sizes, and 0-degree lordosis for the smallest and largest sizes. The surface texture of the 3D structure of the implants resists expulsion forces.

    AI/ML Overview

    The provided text does not contain information about acceptance criteria and a study proving a device meets acceptance criteria in the context of an AI/ML medical device. Instead, it is a 510(k) summary for the FloSpine Ti-Largo™ Cervical Interbody System, which is a physical medical device (an intervertebral body fusion device made of titanium alloy).

    Therefore, I cannot provide the requested information regarding AI/ML device performance, ground truth, expert adjudication, or training/test set sample sizes, as these concepts are not applicable to the presented document.

    The document discusses performance data for the physical interbody system, which includes mechanical and sterilization testing to demonstrate substantial equivalence to predicate devices.

    Here's what I can extract regarding the physical device's performance data:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document lists performance tests conducted but does not explicitly state the numerical acceptance criteria for each test or the specific reported device performance values (e.g., actual compression strength in Newtons or number of cycles to failure). It simply states that the outcomes met "all acceptance criteria."

    Test CategoryTest Standard/TypeReported Performance
    Mechanical TestingStatic CompressionMet acceptance criteria
    Dynamic Compression FatigueMet acceptance criteria
    Static TorsionMet acceptance criteria
    Dynamic Torsion FatigueMet acceptance criteria
    Static ExpulsionMet acceptance criteria
    SubsidenceMet acceptance criteria
    Wear Debris AnalysisMet acceptance criteria
    SterilizationGamma Radiation (SAL 10⁻⁶)Validated based on VDmax method (ANSI/AAMI/ISO 11137-1 and 11137-2)

    2. Sample size used for the test set and the data provenance: Not applicable. These are mechanical and sterilization tests, not studies on a test set of data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth and expert adjudication are not relevant for mechanical and sterilization testing of a physical implant.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is not an AI/ML product.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable to mechanical and sterilization testing.

    8. The sample size for the training set: Not applicable.

    9. How the ground truth for the training set was established: Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1