510(k) Data Aggregation

    K Number
    K252390

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-10-29

    (90 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Telescope™ Guide Extension Catheter is intended to be used in conjunction with guide catheters to access discrete regions of the coronary and/or peripheral vasculature, and to facilitate placement of interventional devices.

    Device Description

    The Telescope™ Guide Extension Catheter is a single-lumen rapid exchange catheter designed to act as an extension to a traditional guide catheter. The Telescope™ Guide Extension Catheter is intended to be used with guide catheters to access discrete regions of the coronary and/or peripheral vasculature, and to facilitate placement of interventional devices.

    The Telescope™ Guide Extension Catheter device is 150cm in length and consists of a hydrophilic-coated single-lumen distal guide segment connected to a stainless-steel polytetrafluoroethylene (PTFE) coated proximal pushwire.

    The device is offered in two sizes 6F and 7F, with a hub at the proximal end of the pushwire that is used for device identification.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K183353

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-03-22

    (109 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Telescope™ Guide Extension Catheter is intended to be used in conjunction with guide catheters to access discrete regions of the coronary and/or peripheral vasculature, and to facilitate placement of interventional devices.

    Device Description

    The Telescope guide extension catheter is a single-lumen rapid exchange catheter. The guide extension catheter is designed to act as an extension to a traditional guide catheter and to facilitate the delivery of interventional devices into the vasculature. The guide extension catheter is intended to be used within the coronary and/or peripheral vasculature to provide support.

    The Telescope™ Guide Extension Catheter is offered in sizes compatible with 6F and 7F guide catheters as indicated by the product model numbers below.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the Medtronic Telescope™ Guide Extension Catheter. It outlines the device's indications for use, its substantial equivalence to a predicate device, and the non-clinical testing performed to establish this equivalence.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly list specific quantitative "acceptance criteria" in the format of a table with numerical thresholds. Instead, it describes various "Design Verification/In-Vitro Bench Testing" and "Biocompatibility Testing" categories that were performed to demonstrate substantial equivalence to the predicate device. The implied acceptance criterion for each test is that the device's performance is acceptable and does not raise new questions of safety and effectiveness, making it comparable to the predicate.

    Here's an interpreted table based on the testing categories mentioned:

    Acceptance Criteria (Implied)Reported Device Performance
    Design Verification/In-Vitro Bench Testing:
    Simulated UseNo new safety or effectiveness issues raised.
    DeliverabilityNo new safety or effectiveness issues raised.
    Tip compressionNo new safety or effectiveness issues raised.
    Pushwire IntegrityNo new safety or effectiveness issues raised.
    TorqueNo new safety or effectiveness issues raised.
    Lumen PatencyNo new safety or effectiveness issues raised.
    Maximum outer diameterNo new safety or effectiveness issues raised.
    Catheter effective lengthNo new safety or effectiveness issues raised.
    Kink resistanceNo new safety or effectiveness issues raised.
    Tensile StrengthNo new safety or effectiveness issues raised.
    Coating IntegrityNo new safety or effectiveness issues raised.
    Coating ParticulateNo new safety or effectiveness issues raised.
    Biocompatibility Testing (in compliance with ISO 10993-1):
    CytotoxicityNo new safety or effectiveness issues raised.
    Acute Systemic ToxicityNo new safety or effectiveness issues raised.
    Material Mediated PyrogenicityNo new safety or effectiveness issues raised.
    ASTM HemolysisNo new safety or effectiveness issues raised.
    Complement ActivationNo new safety or effectiveness issues raised.
    Partial thromboplastin timeNo new safety or effectiveness issues raised.
    In-vivo ThromboresistanceNo new safety or effectiveness issues raised.
    Irritation / Intracutaneous ReactivityNo new safety or effectiveness issues raised.
    SensitizationNo new safety or effectiveness issues raised.

    2. Sample Size Used for the Test Set and the Data Provenance:

    The document does not specify the exact sample sizes used for each individual test within the "Design Verification/In-Vitro Bench Testing" and "Biocompatibility Testing." It broadly states that "non-clinical data" was generated.

    The data provenance is from Medtronic Ireland, the manufacturer, through in-vitro bench testing and biocompatibility testing. This is typically considered prospective for the purposes of generating data to support the 510(k) submission, as the tests were specifically conducted for the purpose of demonstrating substantial equivalence of the new device. The country of origin for the data generation would be Ireland, where Medtronic Ireland is located.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    This information is not provided in the document. The document describes bench testing and biocompatibility testing, which typically involve laboratory personnel and engineers, rather than clinical experts establishing ground truth in the context of diagnostic interpretation.

    4. Adjudication Method for the Test Set:

    This information is not applicable and not provided. The testing described is objective, technical testing (e.g., tensile strength, lumen patency) and biocompatibility testing against defined standards. It does not involve human interpretation or adjudication of results in the way a clinical study would.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not done. The document explicitly states: "No clinical investigation has been performed for this device."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    This concept is not applicable to this device. The Telescope™ Guide Extension Catheter is a physical medical device, not a software algorithm or AI model. Therefore, "standalone algorithm performance" is not a relevant metric.

    7. The Type of Ground Truth Used:

    For the "Design Verification/In-Vitro Bench Testing," the ground truth is established by engineering specifications, industry standards, and internal Medtronic test protocols that ensure the device meets its functional and performance requirements. For "Biocompatibility Testing," the ground truth is established by international standards, specifically ISO 10993-1: 2009/ Cor 1: 2010, which outlines the requirements for biological evaluation of medical devices.

    8. The Sample Size for the Training Set:

    This information is not applicable and not provided. As this is a physical medical device submission and no AI/machine learning component is mentioned, there is no "training set" in the context of an algorithm.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable as there is no training set for an AI/machine learning algorithm.