510(k) Data Aggregation

    K Number
    K172500

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-03-01

    (195 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Technoweb Sugical Mask is intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.

    Device Description

    Technoweb Surgical Mask is a type of surgical mask covers the user's nose and mouth. And provides a physical barrier to fluids and particulate materials. Technoweb Surgical Mask is a flat, pleated tie-on or elastic ear-loop mask consisting of three nonwoven layers; inner and outer cover web with middle different filter web sandwiched in between.

    AI/ML Overview

    This document describes the FDA's substantial equivalence determination for the Technoweb Surgical Mask (K172500) to the predicate device, 3M VFlexTM Health Care Particulate Respirator and Surgical Mask, Model 1805/1805S (K121069).

    Here's an analysis of the provided information concerning acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Technoweb Surgical Mask are based on performance testing to demonstrate substantial equivalence to the predicate device. These criteria are primarily focused on the physical characteristics and protective capabilities of surgical masks.

    Acceptance Criteria CategorySpecific Criteria/Tests (Technoweb Surgical Mask)Reported Device Performance (Technoweb Surgical Mask)Predicate Device (3M VFlexTM Model 1805/1805S)Comparison
    Fluid ResistanceASTM F1862Fluid resistance at 12 mmHg.YES (implies meeting a similar standard)Similar
    MaterialOuter cover web: Polypropylene spunbond Stiffener web: Polypropylene spunbond Filter web: Polypropylene Inner web: Polypropylene Nose-Clip: Aluminum Staple: Steel Headband: NylonAll listed materials are used.Outer cover web: Felt Stiffener web: N/A Filter web: Felt Inner web: Felt Nose-Clip: Aluminum Staple: N/A Headband: Polyisoprene Nose foam: Plastic coatingSimilar (some material differences noted but deemed acceptable for equivalence)
    DesignFlat fold with ear loopsFlat fold with ear loopsFlat fold with Advanced Electrostatic Media (implied design feature aiding filtration)Similar
    Indications for UseIntended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.Meets this indication.Same indication.Same
    Single UseYesYesYesSame
    BiocompatibilityBiocompatibility assessment performed on patient contact materials.Passed (implied by conclusion of safety and equivalence).(Not explicitly detailed for predicate, but assumed to meet safety standards)Implied as similar by equivalence conclusion.
    Safety and Performance (General)Laboratory testing regarding characteristics was performed to verify safety and performance.Passed (implied by conclusion of safety and equivalence).(Not explicitly detailed for predicate, but assumed to meet safety standards)Implied as similar by equivalence conclusion.
    DimensionsTM-R is 212mm(W), 74mm(H) and model TM-S is 177.8mm(W), 138 mm(H).Reported dimensions.(Not explicitly detailed for predicate in the table, but assumed to be comparable for "similar" rating)Similar

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not provide specific details on the sample sizes used for the various performance tests (e.g., fluid resistance, material testing, biocompatibility). It generally states "Laboratory testing regarding characteristics was performed on Technoweb surgical mask." It also doesn't specify the country of origin of the data or whether the studies were retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Since this is a submission for a medical device (surgical mask), the "ground truth" is typically established by recognized national or international standards and test methods (e.g., ASTM F1862 for fluid resistance). The document does not mention the use of human experts to establish "ground truth" for the performance characteristics of the mask in the same way one might for an AI diagnostic imaging device. Instead, the performance is measured against established scientific criteria.

    4. Adjudication Method

    Not applicable for this type of device and testing. Performance is measured objectively against predefined standards, not through expert adjudication of subjective assessments.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is a physical medical device (surgical mask), not an AI algorithm intended for diagnostic interpretation or human-in-the-loop assistance.

    6. Standalone Performance Study (Algorithm Only)

    Not applicable. This is not an algorithm. The "standalone performance" refers to the intrinsic physical and protective properties of the mask as tested against standards.

    7. Type of Ground Truth Used

    The ground truth used for evaluating the Technoweb Surgical Mask's performance is based on established industry standards and test methods, such as ASTM F1862 for fluid resistance, and general biocompatibility testing methodologies. These are objective, empirically defined criteria for surgical mask performance.

    8. Sample Size for the Training Set

    Not applicable. This is a physical device, not an AI model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for a physical surgical mask.