Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K253114

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-10-16

    (22 days)

    Product Code
    Regulation Number
    878.4350
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TargetCool™ (Cooling mode / Boosting mode) is indicated for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures, minor sprains or other minor sports injuries, and as an adjunct to rehabilitative treatment (e.g., intermittent cold with stretch).

    TargetCool™ (Freezing mode) is indicated for the surgical destruction of target tissue by applying cryogenic gases at extreme low temperatures

    • Molluscum Contagiosum
    • Skin Tags
    • Actinic Keratosis
    • Lentigo
    • Verruca Plana
    • Verruca Vulgaris
    • Verruca Lesions
    • Genital Lesions
    • Seborrheic Keratosis
    Device Description

    TargetCool™ is a handheld device that can deliver rapid, precise, and controlled cooling to the skin tissue. The TargetCool™ device consists of a main device, nozzle, a guard, a filter, and a cartridge.

    The main device produces controlled cooling based on thermoelectric cooling, which controls the temperature of the targeted area. TargetCool™ displays the skin temperature measured in real time, the set cooling temperature and time, and the device status through the LCD display. Also, if the measured temperature is below -1°C and lasts for more than 1 second, the status light blinks in blue with a beep sound.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K253348

    Validate with FDA (Live)

    Device Name
    TargetCool-e
    Date Cleared
    2025-10-15

    (15 days)

    Product Code
    Regulation Number
    878.4350
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TargetCool-e (Cooling mode / Boosting mode) is indicated for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures, minor sprains or other minor sports injuries, and as an adjunct to rehabilitative treatment (e.g., intermittent cold with stretch).

    TargetCool-e (Freezing mode) is indicated for the surgical destruction of target tissue by applying cryogenic gases at extreme low temperatures

    • Molluscum Contagiosum
    • Skin Tags
    • Actinic Keratosis
    • Lentigo
    • Verruca Plana
    • Verruca Vulgaris
    • Verruca Lesions
    • Genital Lesions
    • Seborrheic Keratosis
    Device Description

    TargetCool-e is a handheld device that can deliver rapid, precise, and controlled cooling to the skin tissue. The TargetCool-e device consists of a main device, nozzle, a guard and a cartridge.

    The main device produces controlled cooling based on thermoelectric cooling, which controls the temperature of the targeted area. TargetCool-e displays the skin temperature measured in real time, the set cooling temperature and time, and the device status through the LCD display. Also, if the measured temperature is below -1°C and lasts for more than 1 second, the status light blinks in blue with a beep sound.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K230599

    Validate with FDA (Live)

    Device Name
    TargetCool™
    Date Cleared
    2023-06-22

    (111 days)

    Product Code
    Regulation Number
    878.4350
    Age Range
    2 - 120
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TargetCool™ (Cooling mode / Boosting mode) is indicated for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures, minor sprains or other minor sports injuries, and as an adjunct to rehabilitative treatment (e.g., intermittent cold with stretch).

    TargetCool™ (Freezing mode) is indicated for the surgical destruction of target tissue by applying cryogenic gases at extreme low temperatures

    • Molluscum Contagiosum
    • Skin Tags
    • Actinic Keratosis
    • Lentigo
    • Verruca Plana
    • Verruca Vulgaris
    • Verruca Lesions
    • Genital Lesions
    • Seborrheic Keratosis
    Device Description

    TargetCool™ is a handheld device that can deliver rapid, precise, and controlled cooling to the skin tissue. The TargetCool™ device consists of a main device, nozzle, a guard, a filter, and a cartridge.

    The main device produces controlled cooling based on thermoelectric cooling, which controls the temperature of the targeted area. TargetCool™ displays the skin temperature measured in real time, the set cooling temperature and the device status through the LCD display. Also, if the measured temperature is below -1 ℃ and lasts for more than 1 second, the status light blinks in blue with a beep sound.

    The principle of pain relief through skin cooling is as the skin cools by spraying the cryogen onto the skin, the nerve conduction velocity (NCV) of the skin decreases, increasing the pain threshold (PTH) and pain tolerance (PT). Through this mechanism, pain in the skin is relieved. The advances represented by TargetCool™ lie in the fact that the device executes this process in a precise, controlled, and rapid way.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the TargetCool™ device. However, it does not include detailed acceptance criteria or a specific study proving the device meets those criteria with performance metrics, sample sizes, or ground truth establishment relevant to AI/ML device evaluations.

    The document focuses on demonstrating substantial equivalence to predicate devices based on intended use, technological characteristics, and safety. The performance data section refers to bench testing (compliance with standards, biocompatibility, hazard analysis) and mentions "Clinical publications and clinical data demonstrated the safety and effectiveness of the TargetCoolTM for treatment of pediatric patient," but it does not provide specifics about these clinical studies in the context of acceptance criteria.

    Therefore, many of the requested details cannot be extracted from the provided text.

    Here's what can be inferred and what is missing:

    1. A table of acceptance criteria and the reported device performance:

    Feature/TestAcceptance Criteria (Inferred)Reported Device Performance (Inferred)
    Cooling ModeTemperature of 2-4 °C within 5 secondsAchieves 2-4 °C within 5 seconds
    Boosting ModeTemperature of 2-4 °C, sustained for 3-4 minutes (for saline solution)Achieves 2-4 °C, sustained for 3-4 minutes (for saline solution)
    Freezing Mode (CO2 cryoablation)Temperature of -79°C (using CO2)Achieves -79°C (using CO2)
    Gas dispensing rate (Freezing Mode)0.578 g/sec0.578 g/sec
    Gas Volume65g cartridge capacityUses 65g cartridge
    Safety Feature (Temperature monitoring)Alarm and status light blinking if skin temperature is determined to be less than -1 °C for 1 second.Alarm and status light blinks if skin temp is < -1 °C for 1 sec.
    Electrical SafetyCompliance with IEC 60601-1Complies with IEC 60601-1
    Electromagnetic Compatibility (EMC)Compliance with IEC 60601-1-2Complies with IEC 60601-1-2
    BiocompatibilityConformance with ISO 10993 (for patient-contacting materials)Not applicable as material is identical to cleared predicate.
    Hazard AnalysisIdentified hazards mitigatedDevice hazard analysis completed and risk-control implemented to mitigate identified hazards.
    Overall Performance (Bench Testing)All specifications met acceptance criteria; supports substantial equivalence and safe operation.Testing results support all specifications met acceptance criteria; passed all testing.
    Clinical Safety and Effectiveness (Pediatric Use)Demonstrated safety and effectiveness (details not provided)Clinical publications and clinical data demonstrated safety and effectiveness for pediatric patients.

    Missing Information (Not found in the provided text):

    1. Sample size used for the test set and the data provenance: Not mentioned. The document refers to "Clinical publications and clinical data" but does not detail the nature, size, or origin of this data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not mentioned.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not mentioned.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a cryosurgical unit, not an AI/ML diagnostic or assistive device that would typically involve human readers.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a medical device, not an AI algorithm in the typical sense of standalone performance. Its performance relates to its physical parameters (temperature, flow rate) and clinical outcomes, not an algorithm's output.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): For clinical effectiveness, it would likely be based on clinical outcomes data (e.g., successful tissue destruction for freezing mode, pain/swelling reduction for cooling mode). However, the document doesn't specify how ground truth was established for the "clinical data" it refers to.
    7. The sample size for the training set: Not applicable and not mentioned, as this is not an AI/ML device requiring a training set in the conventional sense.
    8. How the ground truth for the training set was established: Not applicable and not mentioned.

    Summary of the Study and Evidence Presented:

    The core of the submission (K230599) is a substantial equivalence (SE) demonstration for the TargetCool™ device against two predicate devices (K221234 - an earlier version of TargetCool™ and K203481 - CryoVIVE). The study to prove acceptance criteria primarily seems to be a combination of:

    • Bench Testing: Demonstrating compliance with recognized standards (ISO 13485, IEC 60601-1, IEC 60601-1-2, ISO 10993 for biocompatibility-by-proxy). These tests verify the device's physical and electrical safety and performance parameters like temperature accuracy within specified ranges.
    • Safety Features Validation: Proving the alarm system for low skin temperature functions as designed.
    • Clinical Data/Publications Review: The document states that "Clinical publications and clinical data demonstrated the safety and effectiveness of the TargetCoolTM for treatment of pediatric patient." However, it does not elaborate on the specifics of these studies (e.g., study design, sample size, primary endpoints, results). This clinical evidence likely serves to support the overall safety and effectiveness claims rather than precise numerical acceptance criteria.

    The substantial equivalence argument is based on comparing the intended use, technological characteristics, and principles of operation to the predicate devices. Differences, such as the Boosting mode and modifications to the freezing nozzle, are addressed by arguing they "do not raise new questions of safety and effectiveness."

    Ask a Question

    Ask a specific question about this device

    K Number
    K221234

    Validate with FDA (Live)

    Device Name
    TargetCool
    Manufacturer
    Date Cleared
    2022-07-22

    (84 days)

    Product Code
    Regulation Number
    878.4350
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TargetCool™ is indicated for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures, minor sprains or other minor sports injuries, and as an adjunct to rehabilitative treatment (e.g., intermittent cold with stretch).

    Device Description

    TargetCool™ is a handheld device that can deliver rapid, precise, and controlled cooling to the skin tissue. The TargetCool™ device consists of a main device Cooling-Nozzle, a guard, a filter, and a cartridge. In Cooling mode, the main device produces controlled cooling based on thermoelectric cooling, which controls the temperature (2-4°C) of the targeted area. TargetCool™ displays the skin temperature measured in real time, the set cooling temperature and time, and the device status through the LCD display. Also, if the measured temperature is below -1 ℃ and lasts for more than 1 second, the status light blinks in blue with a beep sound.

    AI/ML Overview

    The provided document is a 510(k) Summary for the medical device TargetCool™ (K221234). It compares this device to a predicate device, also named TargetCool (K220674).

    Here's an analysis of the acceptance criteria and the study information, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document presents a comparison between the subject device (TargetCool™ K221234) and its predicate (TargetCool K220674) to establish substantial equivalence. While explicit acceptance criteria are not tabulated with numerical performance, the comparison acts as a de facto set of criteria since the subject device is stated to be "Same" as the predicate for all listed characteristics.

    Acceptance Criteria (from Predicate Device K220674)Reported Device Performance (TargetCool™ K221234)Satisfied?
    Indications for Use: Temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures, minor sprains or other minor sports injuries, and as an adjunct to rehabilitative treatment (e.g., intermittent cold with stretch).Temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures, minor sprains or other minor sports injuries, and as an adjunct to rehabilitative treatment (e.g., intermittent cold with stretch).Yes
    Component: Main system, Trigger, LCD Display, Cooling-Nozzle, Guard, Filter, CartridgeMain system, Trigger, LCD Display, Cooling-Nozzle, Guard, Filter, CartridgeYes
    Mechanism of Action: The unit blows very low-temperature gas at temperature and time settings, onto the desired treatment areaThe unit blows very low-temperature gas at temperature and time settings, onto the desired treatment areaYes
    Cryogen Type: CO2CO2Yes
    Temperature: Reaching 2-4 °C within 5 secReaching 2-4 °C within 5 secYes
    Treatment Duration: 0~60 sec0~60 secYes
    Gas Volume: 65g cartridge65g cartridgeYes
    Safety feature: Alarm and status light blinking if the temperature of the skin is less than -1 °C for 1 second.Alarm and status light blinking if the temperature of the skin is determined to be less than -1 °C for 1 second.Yes

    Study Proving Device Meets Acceptance Criteria:

    The document states: "TargetCool™ is the same as the device cleared in K220674. No changes have been made to the device to change the contraindication. TargetCool™, the subject device, is equivalent to the predicate device."
    This indicates that the "study" proving the device meets the acceptance criteria is primarily a direct comparison and declaration of sameness to a previously cleared device (K220674), rather than new, independent performance testing for the subject device against the listed criteria.

    2. Sample size used for the test set and the data provenance

    The document does not describe a separate "test set" in the context of a clinical study for the K221234 device. Instead, it relies on the predicate device's clearance. It mentions "Clinical publications and clinical data demonstrated the safety and effectiveness of the TargetCoolTM for treatment of pediatric patient" but does not provide details on the sample size or provenance of this data. Given the context of establishing substantial equivalence to an identical device, new clinical testing specific to K221234 for comparing against acceptance criteria is not detailed.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. No new test set for the K221234 device is described, as the substantial equivalence is based on the device being functionally identical to the predicate (K220674). The ground truth for the safety and effectiveness of the predicate device (and thus implicitly the subject device) would have been established during its regulatory review, likely based on scientific principles of cryotherapy and any historical clinical data available for similar devices, but this document does not provide details about that.

    4. Adjudication method for the test set

    Not applicable. There is no new "test set" with expert adjudication described for K221234 in this document.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a cryosurgical unit, not an AI-assisted diagnostic or imaging device. Therefore, MRMC studies and AI assistance are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical medical device, not an algorithm, so standalone algorithm performance testing is not relevant.

    7. The type of ground truth used

    The ground truth for the functionality and safety of the subject device (K221234) is established by its direct equivalence to the predicate device (K220674), which itself would have been cleared based on a combination of bench testing, adherence to standards (electrical safety, EMC, biocompatibility), and existing clinical literature/data for cryotherapy devices. The statement "Clinical publications and clinical data demonstrated the safety and effectiveness of the TargetCoolTM" refers to the scientific basis for the predicate device's (and thus the subject device's) application.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device that would require a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/machine learning device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220674

    Validate with FDA (Live)

    Device Name
    TargetCool
    Manufacturer
    Date Cleared
    2022-03-31

    (23 days)

    Product Code
    Regulation Number
    878.4350
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TargetCool™ is indicated for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures, minor sprains or other minor sports injuries, and as an adjunct to rehabilitative treatment (e.g., intermittent cold with stretch).

    Device Description

    TargetCool™ is a handheld device that can deliver rapid, precise, and controlled cooling to the skin tissue. The TargetCool™ device consists of a main device Cooling-Nozzle, a guard, a filter, and a cartridge,

    In Cooling mode, the main device produces controlled cooling based on thermoelectric cooling. which controls the temperature (2-4℃) of the targeted area. TargetCool™ displays the skin temperature measured in real time, the set cooling temperature and time, and the device status through the LCD display. Also, if the measured temperature is below -19C and lasts for more than 1 second, the status light blinks in blue with a beep sound.

    The principle of pain relief through skin cooling is as follows: As the skin cools by spraying the cryogen onto the skin, the nerve conduction velocity (NCV) of the skin decreases, increasing the pain threshold (PTH) and pain tolerance (PT). Through this mechanism, pain in the skin is relieved. The advances represented by TargetCool™ lie in the fact that the device executes this process in a precise, controlled, and rapid way.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the TargetCool™ device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Skin surface temperature reaches 2-4℃Skin surface temperature reached 2-4℃ within 5 seconds.
    Skin surface temperature is maintained at 2-4℃ for the maximum spraying time (60 seconds)Skin surface temperature was maintained at 2-4℃ for the maximum spraying time (60 seconds).
    Compliance with ISO 13485:2016 (Quality management systems for medical devices)TargetCool™ complies with all applicable standards, including ISO 13485:2016.
    Compliance with IEC 60601-1 (Electrical safety)TargetCool™ complies with all applicable standards, including IEC 60601-1.
    Compliance with IEC 60601-1-2 (Electromagnetic compatibility)TargetCool™ complies with all applicable standards, including IEC 60601-1-2.
    Biocompatibility (ISO 10993)Biocompatibility testing (ISO 10993) was deemed not applicable for TargetCool™ because the only patient-contacting material (polycarbonate on the guard) is the exact same material used in the predicate device, CryoVIVE (K203481), which was previously FDA-cleared and presumably met biocompatibility requirements.
    Device hazard analysis completed and risk-control implemented to mitigate identified hazardsA device hazard analysis was completed and risk-control implemented to mitigate identified hazards.
    All specifications met acceptance criteria of each module and interaction of processesTesting results support that all the specifications have met the acceptance criteria of each module and interaction of processes.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The text states, "TargetCool Temperature accuracy tests were conducted targeting three different treas (wrist, arm, back of neck) by three physicians at three different temperature settings (-10, -5, 5 degree C)." This implies the "test set" for temperature accuracy involved 3 anatomical areas and 3 temperature settings. The number of individual measurements or subjects is not explicitly stated beyond these categories.
    • Data Provenance: Not explicitly stated (e.g., country of origin). The study seems to be a prospective bench test/performance test conducted by the company to confirm device specifications rather than a clinical trial with patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: Three physicians were involved in conducting the temperature accuracy tests.
    • Qualifications of Experts: The specific qualifications (e.g., years of experience, specialty) of these three physicians are not specified in the provided text.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable/not mentioned. The study described is a performance test, not one that requires adjudication of expert interpretations. The physicians conducted the tests and observed the device's output.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, an MRMC comparative effectiveness study was not done. The document describes bench testing and a comparison to a predicate device based on technical specifications and indications for use, not a study evaluating human readers' improvement with or without AI.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • Yes, a standalone performance test was done. The "Performance Data" section details bench testing where the device's temperature accuracy and compliance with various standards were evaluated directly. This is a standalone performance assessment of the device's physical function.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: For the temperature accuracy tests, the ground truth was based on the measured skin surface temperature using presumably calibrated instruments in a controlled testing environment, compared against the device's set temperature. For the other performance criteria (electrical safety, EMC, etc.), the ground truth was adherence to established international standards (e.g., ISO, IEC).

    8. The Sample Size for the Training Set

    • Not applicable/not mentioned. The TargetCool™ device is a cryosurgical unit, a physical medical device that applies cold therapy. It is not an AI/ML algorithm that requires a training set in the conventional sense. The "training" or development of the device would involve engineering design and iterative testing, not data-driven machine learning.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable/not mentioned. As explained in point 8, this is not an AI/ML device requiring a training set with established ground truth from data.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1