Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K210012

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-07-09

    (186 days)

    Product Code
    Regulation Number
    870.5100
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tamarin Blue PTCA RX Dilatation Catheters are indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving myocardial perfusion.

    Device Description

    The Tamarin Blue PTCA RX Dilatation Catheter is a standard Rapid Exchange (RX) PTCA catheter with a single lumen proximally, a dual lumen distally, a semi-compliant inflatable balloon and a soft, tapered distal tip to aid in crossing tight stenoses. One lumen of the catheter's dual lumen is used for inflation and deflation of the balloon, and the other lumen allows guide wire access through the distal part of the catheter. The maximum recommended guide wire diameter is 0.014". A luer lock fitting (hub) allows connection with an inflation device. Two radiopaque markers on the guide wire lumen tubing provide visual reference points for balloon positioning across the stenosis. A hydrophilic coating has been applied on the distal portion of catheter i.e., up to the RX port, to improve catheter pushability. The balloon material expands to a known diameter at a specific pressure as defined in the compliance table supplied with the catheter.

    The device is supplied sterile and intended for one-time use.

    AI/ML Overview

    This document is a 510(k) summary for the Tamarin Blue PTCA RX Dilatation Catheter (K210012). It details the device's comparison to predicate and reference devices and non-clinical testing performed to establish substantial equivalence.

    Here's an analysis to extract the requested information, although it's important to note that this document pertains to a medical device (catheter) and not an AI/ML-driven diagnostic or image analysis device, which are typically the subject of the detailed acceptance criteria and study design questions you've posed. Therefore, many of the requested fields related to AI/ML performance, ground truth establishment by experts, MRMC studies, training sets, etc., will not be applicable or present in this type of submission.

    The information consistently aligns with a submission for a physical medical device, not a software/AI device.

    Here's the closest interpretation of your request based on the provided document:


    Device Name: Tamarin Blue PTCA RX Dilatation Catheter

    Device Type: Medical Device (PTCA Dilatation Catheter)

    Acceptance Criteria and Device Performance:

    The primary "acceptance criteria" for this device, as detailed in a 510(k) submission, revolve around demonstrating substantial equivalence to a previously cleared predicate device. This is typically achieved through non-clinical bench testing to ensure the device performs as intended and meets established safety and performance standards for its class. The document explicitly states that the technological characteristics (components, design, sterilization method, shelf life, operating principle) are identical to the predicate device, with the only notable difference being the addition of a hydrophilic coating, which was tested and shown not to raise new questions of safety or effectiveness.

    1. A table of acceptance criteria and the reported device performance:

    Since this is a physical medical device, "acceptance criteria" are not reported as classical sensitivity/specificity/AUC metrics for a diagnostic algorithm. Instead, they are engineering and performance benchmarks validated through bench testing. The document states that the device "met all acceptance criteria."

    Acceptance Criteria CategoryReported Device Performance
    Coating PerformanceHydrophilic coating performed as intended and did not impact functionality.
    Coating Integrity & Durability(Passed, implied by "performed as intended")
    Balloon Preparation, Deployment, & Retraction(Passed, implied by "met all acceptance criteria")
    Balloon Rated Burst PressurePassed (Rated Burst Pressure: 16 atm - identical to predicate)
    Balloon Fatigue(Passed, implied by "met all acceptance criteria")
    Balloon ComplianceSemi-compliant (identical to predicate)
    Balloon Inflation/Deflation(Passed, implied by "met all acceptance criteria")
    Particulate(Passed, implied by "met all acceptance criteria")
    BiocompatibilityVerified to be the same as predicate/reference devices based on material similarities; subject device is biocompatible.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not explicitly stated as a "sample size" in the context of clinical studies for diagnostic accuracy. The non-clinical bench testing would involve a certain number of manufactured devices for each test (e.g., a specific batch size for burst pressure testing, a number of devices for fatigue testing). These numbers are not detailed in this summary.
    • Data Provenance: The testing is non-clinical bench testing, meaning it's performed in a laboratory, not from patient data. Therefore, "country of origin of the data" and "retrospective or prospective" are not applicable.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This concept is not applicable here as the "ground truth" for a physical device is its adherence to engineering specifications and performance standards, validated through objective bench testing. There are no "experts" establishing ground truth in the sense of medical image interpretation for diagnosis.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable. This is for objective bench testing, not subjective human interpretations that require adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is for an AI/ML-driven diagnostic/image analysis device. This product is a physical medical catheter.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is a physical medical catheter, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for this device is based on engineering specifications, established performance standards (e.g., burst pressure limits), and material properties. Compliance is demonstrated through objective bench testing.

    8. The sample size for the training set:

    Not applicable. There is no AI/ML "training set" for this physical device.

    9. How the ground truth for the training set was established:

    Not applicable. There is no AI/ML "training set" for this physical device.


    Summary of the Study:

    The study proving the device meets acceptance criteria is a non-clinical bench testing program. The purpose of this testing was to demonstrate that the Tamarin Blue PTCA RX Dilatation Catheter (K210012), with the addition of a hydrophilic coating, is substantially equivalent to its predicate device (Tamarin Blue PTCA RX Dilatation Catheter, K112735) and equivalent to a reference device (Filao NC RX PTCA Dilatation Catheter, K141933) for the coating aspects.

    The testing evaluated various performance characteristics, including:

    • Coating Inspection
    • Coating Integrity (Performance and Durability)
    • Balloon preparation, deployment and retraction
    • Balloon Rated Burst
    • Balloon Fatigue
    • Balloon compliance
    • Balloon inflation/deflation
    • Particulate

    The conclusion from these tests was that the hydrophilic coating performed as intended, did not impact the functionality of the device, and that the subject device "met all acceptance criteria" and is biocompatible. The technological characteristics, intended use, and indications for use were found to be identical or substantially similar, with any differences not raising new questions of safety or effectiveness.

    Ask a Question

    Ask a specific question about this device

    K Number
    K112735

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-11-02

    (409 days)

    Product Code
    Regulation Number
    870.5100
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tamarin Blue® RX PTCA Dilatation Catheters are indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving myocardial perfusion.

    Device Description

    The Tamarin Blue® PTCA RX Dilatation Catheter is a standard Rapid Exchange (RX) PTCA catheter with a single lumen proximally, a dual lumen distally, a semi-compliant inflatable balloon and a soft, tapered distal tip to aid in crossing tight stenoses. One lumen of the catheter's dual lumen is used for inflation and deflation of the balloon, and the other lumen allows guide wire access through the distal part of the catheter. The maximum recommended guide wire diameter is 0.014". A luer lock fitting (Hub) allows connection with an inflation device. Two radiopaque markers on the guide wire lumen tubing provide visual reference points for balloon positioning across the stenosis. The balloon material expands to a known diameter at specific pressure as defined in the compliance table supplied with the catheter.

    AI/ML Overview

    Here is an analysis of the provided 510(k) summary regarding the Tamarin Blue® PTCA RX Dilatation Catheter, focusing on acceptance criteria and the supporting study information:

    Acceptance Criteria and Device Performance

    The provided document describes a medical device, the Tamarin Blue® PTCA RX Dilatation Catheter, and its non-clinical testing. It asserts that the device met all the predetermined acceptance criteria of design verification and validation. However, the document does not explicitly list the acceptance criteria values or the specific reported device performance values for each test. It only states that the device successfully met these unlisted criteria.

    Therefore, a table with specific acceptance criteria and reported performance cannot be created from the provided text. The document broadly states the types of tests performed.

    Test CategoryAcceptance Criteria (Not explicitly stated)Reported Device Performance (Not explicitly stated)
    BiocompatibilityWould involve meeting standards like ISO 10993 (e.g., cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility, material-mediated pyrogenicity, implantation). Specific limits for extractables or biological responses would be predefined.The device materials were confirmed biocompatible based on testing results, indicating they met the requirements of ISO 10993 Part 1, 2, 4, 5, 10, 11, 12, ASTM F756-00, and 21 CFR 58 (GLP regulations).
    Balloon ComplianceSpecific diameter at specific pressure (e.g., nominal pressure, rated burst pressure). Defined compliance curve within a specified tolerance from the predicate or design specification.The balloon material expands to a known diameter at specific pressure as defined in the compliance table supplied with the catheter (table not provided in summary, but implies data exists and was accepted).
    Balloon Burst PressureMinimum burst pressure values for different balloon diameters and lengths, along with a specified safety factor.Device passed testing for balloon burst pressure. This implies the observed burst pressures met or exceeded the predefined minimums.
    Balloon Fatigue EnduranceAbility to withstand multiple inflation/deflation cycles (e.g., number of cycles at specific pressure without failure, typically 3 or more cycles).Device passed testing for balloon fatigue (repeated balloon inflation) endurance, implying it met the required number of cycles without failure.
    Balloon Inflation/Deflation PerformanceTime to inflate and deflate within specified limits; complete deflation allowing re-insertion/repositioning.Device passed testing for balloon inflation/deflation performance.
    Bond StrengthsMinimum tensile or shear strength values for critical bonds (e.g., balloon-to-catheter shaft, hub-to-catheter shaft) to prevent dislodgement during use.Device passed testing for bond strengths.
    Catheter Dimensions/ProfileVerification of outer diameter, balloon length, shaft diameter, guidewire lumen diameter, distal tip taper against design specifications.Device passed testing for catheter dimensions and balloon profile.
    Catheter Body Min. Burst StrengthMinimum pressure at which the catheter shaft maintains integrity.Device passed testing for catheter body minimum burst strength.
    Device PreparationCompatibility with specified guidewires (e.g., 0.014" diameter) and introducer sheaths; smooth passage without damage.Device passed testing for device preparation (guidewire and introducer compatibility).
    Flexibility and KinkAbility to navigate tortuous anatomy without kinking or permanent deformation. Measured by kink resistance at a certain bending radius or force.Device passed testing for flexibility and kink, indicating it meets the required performance for navigation.
    TorqueAbility to transmit rotational forces from the proximal end to the distal end efficiently (e.g., torque response, 1:1 torqueability).Device passed testing for torque.
    RadiopacityVisibility of markers and potentially the balloon under fluoroscopy, typically compared to a reference material or having a minimum visibility score.Two radiopaque markers provide visual reference points, and the device passed testing for radiopacity, confirming their visibility.
    Coating IntegrityAdherence and durability of any coating, typically assessed after simulated use or handling. No delamination or flaking.Device passed testing for coating integrity.
    Particulate EvaluationNumber and size of particulates released from the device, ensuring it's below acceptable limits (e.g., ISO 10993-12, or specific limits for medical devices).Device passed testing for particulate evaluation, indicating any released particulates are within acceptable limits.

    Study Information

    The provided document describes a non-clinical study involving various in vitro tests to demonstrate the safety and effectiveness of the device. It is not a clinical study involving human subjects or AI algorithms.

    1. Sample size used for the test set and the data provenance:

      • Sample Size: Not specified for any of the individual non-clinical tests. The document only lists the types of tests performed.
      • Data Provenance: The tests are non-clinical (laboratory-based) and conducted according to international standards (ISO, ASTM) and US federal regulations (21 CFR 58 GLP). The country of origin of the data is not explicitly stated, but the sponsor is Natec Medical Ltd. from Mauritius.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. This document describes non-clinical engineering and materials testing, not a study requiring expert consensus for diagnostic ground truth.

    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable. Non-clinical physical and chemical tests have predefined objective pass/fail criteria, not subjective interpretation requiring adjudication.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. No MRMC or AI-related study was performed or mentioned. This device is a physical medical device (catheter), not an AI diagnostic tool.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI algorithm.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For non-clinical tests, the "ground truth" is typically defined by pre-established engineering specifications, material standards, and regulatory requirements. For example, a bond strength test would have a minimum force requirement (ground truth) that the device must meet. Biocompatibility has specific toxicological thresholds.

    7. The sample size for the training set: Not applicable. There is no training set mentioned as this is not an AI/machine learning study.

    8. How the ground truth for the training set was established: Not applicable. There is no training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1