Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K222216

    Validate with FDA (Live)

    Date Cleared
    2022-12-22

    (150 days)

    Product Code
    Regulation Number
    870.2870
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pressure Monitoring Kit with TruWave Disposable Pressure Transducer is for use on patients requiring intravascular, intracranial, or intrauterine pressure monitoring.

    Device Description

    The TruWave Disposable Pressure Transducer is a sterile, single-use device that is used to monitor intravascular, intracranial, and intrauterine pressures. The pressure transducer has a straight, flow-through design across the pressure sensor, and is available with or without the integral flush device. The pressure sensor is a pressure sensitive silicon chip with two electrodes for excitation voltage and two electrodes for signal output. A transparent fluid path with an integral stopcock at one end and an integral flush device (either 3mL or 30mL) at the other end encloses the sensor. The enclosure of the TruWave Disposable Pressure Transducer has a pathway for air to enter the housing and acts as a vent. A disposable cable (available in 10-inch/25 cm and 48-inch/120 cm lengths) attached to the pressure transducer housing interfaces with an Edwards Lifesciences reusable cable that is specifically wired for the monitor being used. The TruWave Disposable Pressure Transducer can be mounted on the patient's arm using an arm strap or it may be mounted on an IV pole in a holder. The TruWave Disposable Pressure Transducer may be a component in various pressure monitoring kits or systems.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the TruWave Disposable Pressure Transducer. The information primarily focuses on demonstrating substantial equivalence to a predicate device through various non-clinical tests.

    However, the question asks for details related to acceptance criteria, device performance, and study information that would typically be found in a clinical study report or a more detailed performance study. The provided document does not contain a table of acceptance criteria with reported device performance metrics in a quantitative sense, nor does it detail a clinical study with human patients, ground truth establishment by experts, or MRMC studies.

    Instead, the document states: "All device acceptance criteria were met. Results of non-clinical testing show that the subject device TruWave Disposable Pressure Transducer meets its intended use and demonstrate that the device is as safe, as effective, and performs as well as the predicate device." It then lists types of non-clinical testing performed.

    Therefore,Based on the provided text, the following information can be extracted regarding acceptance criteria and the study:

    1. A table of acceptance criteria and the reported device performance:

    The document states "All device acceptance criteria were met" and that the non-clinical testing demonstrated the device "meets its intended use and demonstrate that the device is as safe, as effective, and performs as well as the predicate device." However, a specific table outlining quantitative acceptance criteria and their corresponding reported device performance values is not provided in the excerpt. The text mentions several types of testing performed to demonstrate equivalence and safety/effectiveness.

    Acceptance Criteria CategoryReported Device Performance
    Biocompatibility (ISO 10993-1:2018)Compliance demonstrated.
    Electromagnetic Compatibility (EMC)Complies with IEC 60601-1, IEC 60601-2-37, and IEC 60601-1-2 standards.
    Electromagnetic Immunity (EMI)Complies with IEC 60601-1, IEC 60601-2-37, and IEC 60601-1-2 standards.
    Electrical Safety (ANSI/AAMI BP22:1994, IEC 60601-2-34:2011)Defibrillator challenge, dielectric strength, liquid ingress, and leakage current tests performed and met.
    Device Performance (ANSI/AAMI BP22:1994, IEC 60601-2-34:2011)Accuracy, excitation/signal impedance, light sensitivity, symmetry, and overpressure tests conducted and met.
    Magnetic Resonance (MR) Safety (ASTM F2503-20)Performed and met standards.
    Mechanical Testing (ANSI/AAMI BP22:1994, Edwards' design requirements)Performed and met requirements.
    Usability/Human FactorsEvaluated by healthcare provider users. Output implied to be satisfactory as acceptance criteria were met.

    2. Sample size(s) used for the test set and the data provenance:

    The document refers to "non-clinical testing" and does not specify sample sizes in terms of number of devices or data points for each test category. No information on data provenance (e.g., country of origin, retrospective or prospective) is provided, as these are non-clinical (laboratory/bench) tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable and not provided. The testing described is non-clinical performance and safety testing against established standards, not a clinical study requiring expert consensus for ground truth. The "usability/human factors" evaluation was performed by "healthcare provider users," but without details on the number or specific qualifications.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This information is not applicable and not provided, as it pertains to clinical studies with expert reviewers and not the non-clinical testing described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable and not provided. The device is a physical pressure transducer, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This information is not applicable and not provided, as the device is a physical pressure transducer, not an algorithm.

    7. The type of ground truth used:

    For the non-clinical tests, the "ground truth" relates to compliance with established international and national standards (e.g., ISO 10993-1, IEC 60601 series, ANSI/AAMI BP22, ASTM F2503) and the manufacturer's internal design requirements. This is based on standardized measurement and testing protocols, not expert consensus, pathology, or outcomes data in a clinical sense.

    8. The sample size for the training set:

    This information is not applicable and not provided. The described tests are for device verification and validation, not for training a model or algorithm.

    9. How the ground truth for the training set was established:

    This information is not applicable and not provided, as there is no "training set" in the context of the reported non-clinical device testing.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183413

    Validate with FDA (Live)

    Date Cleared
    2019-05-01

    (142 days)

    Product Code
    Regulation Number
    870.2870
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pressure Monitoring Kit with TruWave Disposable Pressure Transducer is for use on patients requiring intravascular, intracranial, or intrauterine pressure monitoring.

    Device Description

    The Edwards Lifesciences Pressure Monitoring kit with TruWave disposable pressure transducer is a sterile, single-use kit that monitors intravascular blood pressure, intracranial pressure, and intrauterine pressure. The disposable sterile cable (available in 12-inch/30 cm and 48-inch/120 cm lengths) interfaces exclusively with an Edwards Lifesciences cable that is specifically wired for the patient monitor used to display the pressure data.

    The TruWave disposable pressure transducer has a straight, flow through design, where the fluid is passed across the pressure sensor. The DPT is available either with or without an integral flush device.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device, the TruWave Disposable Pressure Transducer. It describes the device, its intended use, and the regulatory determination of substantial equivalence.

    However, the provided text does not contain any information about acceptance criteria or a study proving the device meets acceptance criteria in the context of an AI/ML medical device, which is what the prompt is asking for. The device described (a pressure transducer) is a physical medical device, not an AI/ML algorithm.

    Therefore, I cannot extract the requested information (acceptance criteria table, sample sizes, ground truth establishment, expert qualifications, adjudication methods, MRMC studies, standalone performance, training set details) from the provided document as it is not relevant to an AI/ML device study.

    The document discusses:

    • Device Name: TruWave Disposable Pressure Transducer
    • Regulation Number: 21 CFR 870.2870 (Catheter Tip Pressure Transducer)
    • Indications for Use: For patients requiring intravascular, intracranial, or intrauterine pressure monitoring.
    • Substantial Equivalence: To predicate devices K171996 and K142749.
    • Testing: Mentions "successfully passed biocompatibility and functional testing," but no details on the specific performance metrics or acceptance criteria for these tests are provided, nor is there any mention of AI/ML performance metrics. The "functional testing" broadly refers to the device's physical operation, not an AI algorithm's performance.

    To answer your prompt, I would need a document detailing the clinical validation or performance study of an AI/ML medical device, including its acceptance criteria and how it was proven to meet them.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171996

    Validate with FDA (Live)

    Date Cleared
    2017-10-23

    (112 days)

    Product Code
    Regulation Number
    870.2870
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pressure Monitoring Kit with TruWave Disposable Pressure Transducer is for use on patients requiring intravascular, intracranial, or intrauterine pressure monitoring.

    Device Description

    The Edwards Lifesciences Pressure Monitoring Kit with TruWave disposable pressure transducer is a sterile, single-use kit that monitors intravascular blood pressure, intracranial pressure, and intrauterine pressure. The disposable sterile cable (available in 12-inch/30 cm and 48-inch/120 cm lengths) interfaces exclusively with an Edwards Lifesciences cable that is specifically wired for the patient monitor used to display the pressure data. The TruWave Disposable Pressure Transducer has a straight, flow through design, where the fluid is passed across the pressure sensor. The DPT is available either with or without an integral flush device.

    AI/ML Overview

    This document (K171996) is a 510(k) premarket notification for a medical device called the "TruWave™ Disposable Pressure Transducer." The key information is outlined in Section 5 – 510(k) Summary.

    Here's the breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Important Note: This document describes a device modification submission, not a submission for a completely novel device. Therefore, the "studies" focus on demonstrating that the changes made to the device (sterilization method and plasticizer) do not negatively impact its performance or safety compared to the predicate device. It is not an AI-powered device, so typical AI/ML study components (like expert consensus for image labeling, MRMC studies, or training set details) are not applicable.


    1. A table of acceptance criteria and the reported device performance:

    The acceptance criteria are essentially that the modified device performs "substantially equivalently" to the predicate device. The performance is reported as successful completion of various tests.

    Acceptance Criteria CategorySpecific Criteria/TestReported Device Performance
    Material Change ImpactThe change in plasticizer from DEHP to TOTM/DINCH® for different components should not alter the performance or safety."Testing was conducted to ensure that the change in sterilization method and change in materials did not alter the performance of the TruWave DPT kits." (Implies successful testing). The device "passed all biocompatibility testing." The device "successfully passed biocompatibility, functional testing, electrical performance and safety testing, and usability testing."
    Sterilization Change ImpactThe change in sterilization method from 100% Ethylene Oxide to E-beam radiation should not alter the performance or safety."Testing was conducted to ensure that the change in sterilization method and change in materials did not alter the performance of the TruWave DPT kits." (Implies successful testing). The device "passed all biocompatibility testing." The device "successfully passed biocompatibility, functional testing, electrical performance and safety testing, and usability testing."
    BiocompatibilityThe device must remain biocompatible after the material and sterilization changes.The products "passed all biocompatibility testing."
    Functional PerformanceThe device must continue to function as intended for intravascular, intracranial, or intrauterine pressure monitoring.The TruWave disposable pressure transducer kits "successfully passed... functional testing." The design verification (electrical) test results were updated in the IFU to comply with AAMI/ANSI BP22:1994/(R)2016, implying successful compliance.
    Electrical PerformanceThe electrical characteristics of the device must remain within specified limits, compliant with relevant standards (e.g., AAMI/ANSI BP22:1994/(R)2016).The TruWave disposable pressure transducer kits "successfully passed... electrical performance and safety testing." Clarifying the design verification (electrical) test results in the Instructions for Use 'Performance Specifications' section to comply with AAMI/ANSI BP22: 1994/(R) 2016.
    Safety TestingThe device must meet safety standards.The TruWave disposable pressure transducer kits "successfully passed... safety testing."
    UsabilityThe device should remain easy and safe to use.The TruWave disposable pressure transducer kits "successfully passed... usability testing."
    Substantial EquivalenceThe modified device must be substantially equivalent to the predicate device in terms of intended use/indications for use, and technology, with no new questions of safety or effectiveness."The subject TruWave Disposable Pressure Transducer kits are substantially equivalent to the predicate TruWave Disposable Pressure Transducer Kits (K141495 and K142749)." "Testing was conducted to ensure that the change in sterilization method and change in materials did not alter the performance of the TruWave DPT kits."

    Regarding the details of the study:

    Since this is a 510(k) for a modification to a physical medical device (a pressure transducer), the "study" is primarily a series of engineering, material, and biocompatibility tests rather than clinical trials or AI model validation studies. Therefore, many of the requested points are not applicable or explicitly stated in the provided summary.

    2. Sample sized used for the test set and the data provenance:

    • Sample Size: Not explicitly stated in the provided text for the specific tests (e.g., how many units were tested for functional or electrical performance). This is typical for a 510(k) summary, as the detailed test protocols and sample sizes would be in the full submission, not the public summary.
    • Data Provenance: Not specified (e.g., country of origin, retrospective/prospective). The tests were likely performed in a lab setting by the manufacturer (Edwards Lifesciences).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This question is not applicable to this type of device submission. "Ground truth" in this context would refer to the established performance characteristics and safety profiles of the predicate device, against which the modified device is compared via defined test methods (e.g., accuracy of pressure readings, material integrity). It doesn't involve expert labeling or interpretation of complex data like medical images.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • This is not applicable as there's no human interpretation or labeling of data requiring adjudication. Performance is measured against engineering specifications and industry standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done:

    • No, this was not done. MRMC studies are specific to evaluating the impact of AI/ML systems on human reader performance, typically in diagnostic imaging. This device is a physical transducer, not an AI/ML product.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No, this was not done. This question pertains to AI/ML algorithm performance. The device is a physical pressure transducer.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):

    • The "ground truth" implicitly used for this device comparison is the performance and safety data of the legally marketed predicate devices, as well as established engineering and medical device standards (e.g., AAMI/ANSI BP22:1994/(R)2016 for electrical performance, and general biocompatibility standards). The modified device must demonstrate that its performance remains within acceptable limits as defined by these standards and the predicate performance.

    8. The sample size for the training set:

    • This is not applicable. There is no "training set" as this is not an AI/ML device.

    9. How the ground truth for the training set was established:

    • This is not applicable. There is no "training set" for ground truth establishment.
    Ask a Question

    Ask a specific question about this device

    K Number
    K142749

    Validate with FDA (Live)

    Date Cleared
    2015-01-18

    (116 days)

    Product Code
    Regulation Number
    870.2870
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pressure Monitoring Kit with TruWave Disposable Pressure Transducer is for use on patients requiring intravascular, intracranial, or intrauterine pressure monitoring.

    Device Description

    The Edwards Lifesciences Pressure Monitoring Kit with TruWave disposable pressure transducer is a sterile, single-use kit that monitors intravascular blood pressure, intracranial pressure, and intrauterine pressure. The disposable sterile cable (available in 12-inch/30 cm and 48-inch/120 cm lengths) interfaces exclusively with an Edwards Lifesciences cable that is specifically wired for the patient monitor used to display the pressure data. The TruWave Disposable Pressure Transducer has a straight, flow-through design, where the fluid is passed across the pressure sensor. The DPT is available either with or without an integral flush device.

    AI/ML Overview

    The provided document describes a 510(k) premarket notification for the TruWave™ Disposable Pressure Transducer. The submission aims to demonstrate substantial equivalence to a legally marketed predicate device. This type of regulatory submission focuses on comparing a new device to an existing one, rather than proving novel efficacy or conducting extensive clinical trials as would be required for a PMA.

    Therefore, the information available pertains to the device's performance against a predicate and functional/safety testing, rather than a clinical study establishing diagnostic accuracy or effectiveness in a "predictive" sense, as would be typical for AI/ML devices. Many of the requested fields (such as multi-reader multi-case studies, expert adjudication for ground truth, or effect size of AI assistance) are not applicable to this type of device and submission.

    Here's an analysis based on the available information:

    1. Table of Acceptance Criteria and the Reported Device Performance

    Acceptance Criteria (Key Performance Specifications)Reported Device Performance
    Integral flush device flow rate3 mL/hr or 30 mL/hr (depending on the variant)
    Operating Pressure Range-50 to +300 mmHg
    Nonlinearity and Hysteresis±1.5% of reading or ± 1 mmHg, whichever is greater

    2. Sample size used for the test set and the data provenance

    The document does not specify a distinct "test set" in the context of clinical data for diagnostic accuracy. The testing performed was primarily performance testing to compare the device to a predicate. The document states:

    • "Performance testing was conducted to compare the proposed device to the predicate device."
    • "The TruWave disposable pressure transducer has successfully passed functional- performance post MRI exposure, the testing included pressure accuracy (nonlinearity and hysteresis)."

    The details of the sample size (e.g., number of units tested, number of measurements) for these performance tests are not provided. Similarly, the "data provenance" in terms of subject demographics or clinical setting is not applicable as it's not a clinical study involving human patients to establish diagnostic performance. The tests are likely bench or laboratory-based.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not Applicable. As this is not a diagnostic AI/ML device, there is no "ground truth" established by experts in the context of interpreting medical images or clinical data. The performance criteria (e.g., pressure accuracy) are determined by quantitative laboratory measurements against set standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not Applicable. There is no "adjudication method" described, as there are no expert interpretations or clinical diagnoses being made that would require adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This is not an AI-assisted device for human interpretation, therefore an MRMC study was not performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not Applicable. This is a hardware device (a disposable pressure transducer) and does not involve AI algorithms operating in a standalone capacity.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for evaluating the performance of this device is based on technical specifications and established measurement standards, rather than clinical outcomes or expert consensus on a diagnosis. For example, pressure accuracy (nonlinearity and hysteresis) would be measured against a highly accurate reference pressure measurement system.

    8. The sample size for the training set

    Not Applicable. This is a hardware medical device; it does not involve machine learning or a "training set" in the computational sense.

    9. How the ground truth for the training set was established

    Not Applicable. There is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141495

    Validate with FDA (Live)

    Date Cleared
    2014-09-03

    (89 days)

    Product Code
    Regulation Number
    870.2870
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pressure Monitoring Kit with TruWave Disposable Pressure Transducer is for use on patients requiring intravascular, intracranial, or intrauterine pressure monitoring.

    Device Description

    The Edwards Lifesciences Pressure Monitoring Kit with TruWave disposable pressure transducer is a sterile, single-use kit that monitors intravascular blood pressure, intracranial pressure, and intrauterine pressure. The disposable sterile cable (available in 12-inch/30 cm and 48-inch/120 cm lengths) interfaces exclusively with an Edwards Lifesciences cable that is specifically wired for the patient monitor used to display the pressure data. The TruWave Disposable Pressure Transducer has a straight, flow-through design, where the fluid is passed across the pressure sensor. The DPT is available either with or without an integral flush device.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the "Pressure Monitoring Kit with TruWave Disposable Pressure Transducers" (K141495). This is a submission to the FDA to demonstrate that the new device is substantially equivalent to a legally marketed predicate device, not a study proving the device meets acceptance criteria in the context of a new diagnostic algorithm or AI system.

    Therefore, the information typically requested in your query regarding acceptance criteria and performance studies of AI/diagnostic algorithms (e.g., sample size, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone performance, training sets, ground truth establishment) is not present in this regulatory document. This document focuses on demonstrating substantial equivalence to a predicate device through material biocompatibility and performance testing, without the need for the types of clinical performance studies you're asking about for AI/diagnostic algorithms.

    However, I can extract the information relevant to what is present, which is comparative analysis and functional/safety testing, rather than an AI algorithm's performance.

    Here's an interpretation based on the given information:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategoryReported Device Performance
    Material BiocompatibilitySuccessfully passed in compliance with applicable ISO 10993 requirements.
    Performance TestingSuccessfully passed performance testing.
    Substantial EquivalenceDemonstrated to be safe, effective, and substantially equivalent to the predicate device (Phoenix [TruWave] disposable pressure transducer) for its intended use.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The document states that "performance testing was conducted", but no details about the sample size (number of devices, test conditions, etc.) or data provenance are given.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. The evaluation of this medical device is based on "performance testing" and "biocompatibility testing," not on establishing ground truth for a diagnostic algorithm with expert review.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided in the document. Adjudication methods are typically relevant for diagnostic studies involving human interpretation or subjective assessments, which is not the nature of the testing described for this pressure transducer.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable as the device is a disposable pressure transducer, not an AI or diagnostic tool that assists human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the device is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The "ground truth" for this device would be its physical and functional specifications. The performance testing would have compared the device's output (e.g., pressure readings) against a known standard or established measurement. The document states that testing was done to "compare the proposed device to the predicate device", implying the predicate device's established performance serves as a comparative benchmark.

    8. The sample size for the training set

    This is not applicable as the device is a physical medical device, not an AI algorithm requiring a training set.

    9. How the ground truth for the training set was established

    This is not applicable as the device is a physical medical device, not an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1