Found 2 results

510(k) Data Aggregation

    K Number
    K040470

    Validate with FDA (Live)

    Date Cleared
    2004-04-28

    (64 days)

    Product Code
    Regulation Number
    876.5990
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the Tripter-X1 Compact Extracorporeal Shock Wave Lithotripter is the fragmentation of urinary tract stones (i.e. renal calyceal, renal pelvic, and upper ureteral stones).

    Device Description

    The Compact is a transportable Electrohydraulic Extracorporeal Shock Wave Lithotripter, which consists of a Shock Wave Generator (SWAG), a Motorized Floating Treatment Table (MFT), and control means.

    AI/ML Overview

    The provided document is a 510(k) summary for the Tripter X-1 Compact Extracorporeal Shock Wave Lithotripter with a Sonographic Localization Option. It explicitly states that no clinical tests were performed (Section 7, "Clinical Tests").

    Therefore, the following information cannot be extracted from the provided text:

    1. Table of acceptance criteria and the reported device performance: No acceptance criteria or performance data from clinical or standalone studies are mentioned.
    2. Sample size used for the test set and the data provenance: No test set data is mentioned.
    3. Number of experts used to establish the ground truth for the test set and their qualifications: No ground truth establishment for a test set is mentioned.
    4. Adjudication method for the test set: No adjudication method is mentioned.
    5. Multi Reader Multi Case (MRMC) comparative effectiveness study, including effect size: No MRMC study was performed.
    6. Standalone (algorithm only) performance: This device is a medical device (lithotripter), not an AI algorithm, so standalone performance in that context is not applicable. Even if it were, no performance studies were conducted.
    7. Type of ground truth used: Not applicable as no studies were performed.
    8. Sample size for the training set: Not applicable as no AI/algorithm training is mentioned for this device.
    9. How the ground truth for the training set was established: Not applicable as no AI/algorithm training is mentioned.

    The submission claims that "The Compact with the Sonographic Localization Option meets the requirements for a special 510(k) by the virtue of being a minor modification, which does not change the fundamental technology or reduce safety and effectiveness, of the Company's predicate device, the Tripter X-1 Compact." This implies that its acceptance criteria were met by demonstrating substantial equivalence to predicate devices without the need for new clinical performance studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K023535

    Validate with FDA (Live)

    Date Cleared
    2003-01-17

    (88 days)

    Product Code
    Regulation Number
    876.5990
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fragmentation of urinary tract stones (i.e. renal calyceal, renal pelvic, and upper ureteral stones).

    Device Description

    The Duet is a transportable Electrohydraulic Extracorporeal Shock Wave Lithotripter, which consists of a Shock Wave Generator (SWAG), a Motorized Floating Treatment Table (MFT), and control means. The Shock Wave Generator can be operated in 4 modes: a) Bottom Reflector only. b) Top reflector only. c) Alternate mode (asynchronous). d) Simultaneous mode (synchronous).

    AI/ML Overview

    Here's an analysis of the provided text regarding the device's acceptance criteria and study, structured to answer your specific questions:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text does not explicitly state specific quantifiable acceptance criteria such as "fragmentation rate of X% for stones of Y size" or "success rate of Z%". Instead, it relies on regulatory compliance and the device being a "minor modification" with "same fundamental scientific technology and intended use" as predicate devices.

    Acceptance CriterionReported Device Performance
    SafetyMeets IEC and UL standards (e.g., IEC 60601-1, UL 2601-1, etc.)
    Effectiveness"does not reduce safety and effectiveness" compared to predicate device
    Technological Equivalence"same fundamental scientific technology and intended use as predicate devices"
    Intended UseIndicated for "Fragmentation of urinary tract stones (i.e. renal calyceal, renal pelvic, and upper ureteral stones)."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document explicitly states: "No clinical tests were performed." Therefore, there is no test set in the traditional sense of a clinical study involving human subjects or patient data.
    • Data Provenance: Not applicable, as no clinical tests were performed.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Not applicable, as no clinical tests were performed and thus no ground truth derived from expert review of patient data.

    4. Adjudication Method for the Test Set

    • Not applicable, as no clinical tests were performed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    • No, an MRMC comparative effectiveness study was not done. The document explicitly states "No clinical tests were performed."

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Study was Done

    • This question is not applicable to this device. The "Duet" is an Extracorporeal Shock Wave Lithotripter (ESWL), which is a physical medical device, not an AI algorithm or software. Its performance is related to mechanical and acoustic properties, not software-based analysis.

    7. The Type of Ground Truth Used

    • Not applicable in the context of clinical or AI-based performance. The "ground truth" for this submission appears to be compliance with engineering and electrical safety standards (as listed in Section 7 - Performance Testing) and the established safety and effectiveness profile of its predicate devices.

    8. The Sample Size for the Training Set

    • Not applicable. This device is a physical medical device and not an AI algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. This device is a physical medical device and not an AI algorithm.

    Summary of the Device Submission's Approach:

    This 510(k) submission for the Direx Systems Corporation Tripter X-1 Compact Duet is a "special 510(k)" submission. This type of submission is used for modifications to a manufacturer's own legally marketed device (the predicate device) where the modification does not affect the intended use or fundamental scientific technology, and does not raise new questions of safety or effectiveness.

    Because it was deemed a "minor modification," the submission relied heavily on:

    • Substantial equivalence to predicate devices: The core argument is that the Duet shares the "same fundamental scientific technology and intended use" as its own prior device (Tripter X-1 Compact) and other legally marketed ESWL devices.
    • Compliance with recognized electrical and safety standards: Extensive testing against IEC and UL standards confirmed its safety from an engineering perspective.
    • Lack of requirement for new clinical data: The FDA agreed that "No clinical tests were performed" were necessary because the modification ("enables the use of 2 reflectors instead of one") was not considered to introduce new risks or alter the fundamental mechanism of action that would necessitate new human trials.