Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K083526

    Validate with FDA (Live)

    Date Cleared
    2009-03-13

    (105 days)

    Product Code
    Regulation Number
    868.5895
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Respironics Trilogy100 system provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. Trilogy100 is intended for pediatric through adult patients weighing at least 5 kg (11 lbs.).

    The device is intended to be used in home, institution/hospital, and portable applications such as wheelchairs and gurneys, and may be used for both invasive and non-invasive ventilation. It is not intended to be used as a transport ventilator.

    Device Description

    The Respironics Trilogy 100 Ventilatory Support System is a microprocessor controlled blower based pressure support, pressure control or volume controlled ventilator intended for the care of individuals who require mechanical ventilation. The ventilator is suitable for use in the institutional, home and transport settings and is applicable for adults and pediatric patients weighing at least 5 kg (11 lbs) who require the following types of ventilatory support.

    CPAP – Continuous Positive Airway Pressure
    S Spontaneous Ventilation
    S/T Spontaneous and Timed Ventilation
    T Timed Ventilation
    PC Pressure Control Ventilation
    PC-SIMV Pressure Controlled Synchronized Intermittent Mandatory Ventilation
    AC Assist Control Ventilation
    CV Control Ventilation
    SIMV Synchronized Intermittent Mandatory Ventilation

    A user interface displays clinical data and enables the operator to set and adjust certain clinical parameters.

    The Trilogy 100 Ventilator is fitted with alarms to alert the user to changes that will affect the treatment. Some of the alarms are pre-set (fixed), others are user adjustable.

    The Trilogy 100 Ventilator can deliver invasive (via ET tube) or non-invasively (via a mask)

    AI/ML Overview

    The provided text is a 510(k) premarket notification for the Respironics Trilogy 100 Ventilator. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than undergoing extensive clinical trials to establish new acceptance criteria and prove performance against those criteria. Therefore, the document does not contain the detailed information typically found in a study designed to establish and meet novel acceptance criteria for a new device.

    However, based on the information provided, I can infer the "acceptance criteria" as demonstrating substantial equivalence to predicate devices, and the "study" proving this as the compilation of technical characteristics and bench testing results comparing the Trilogy 100 to its predicates.

    Here's a breakdown of the available information, addressing as many of your points as possible:

    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Inferred from Substantial Equivalence Claim)Reported Device Performance (as described in the submission)
    Same intended use as predicate devices"The Respironics Trilogy100 system provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. Trilogy100 is intended for pediatric through adult patients weighing at least 5 kg (11 lbs.). The device is intended to be used in home, institution/hospital, and portable applications such as wheelchairs and gurneys, and may be used for both invasive and non-invasive ventilation." (Matches predicate's intended use).
    Same operating principle as predicate devices"Same operating principle."
    Same technology as predicate devices"Same technology."
    Same manufacturing process as predicate devices"Same manufacturing process."
    Similar modes of ventilation as predicate devices"The modes of ventilation on Trilogy 100 are similar to those on other currently marketed continuous ventilator." Specific modes listed include CPAP, S, S/T, T, PC, PC-SIMV, AC, CV, SIMV.
    Similar performance characteristics to predicate devices (e.g., Rise Time, Ramp, Triggering, Pressure Regulation, Alarms)"Trilogy 100 has similar performance characteristics to the predicate devices, as listed below..." The document details numerous features that are "either unchanged or similar" to the specific predicate devices (Respironics BiPAP Synchrony with AVAPS, Respironics BiPAP PLV Continuum, Pulmonetic LTV-1000, Newport Medical HT-50). These include: - Environment of Use - Physical Characteristics of Design, Manufacturing, Energy Delivery, Materials, Anatomical Sites, Energy Source - Rise Time, Ramp, Triggering, Pressure Regulation method - Alarm/Power Control Indicators, Patient Alarms, System Error Alarms - Device Settings (Tidal Volume, Pressure, Rate), Inspiratory Time Setting, Rise Time Setting - Protection against electric shock and patient applied part - Therapy features (Bi-Flex and AVAPS, where applicable to specific predicates) - Data Storage, Serial Interface, User Interface, Real-Time Pressure Display, Humidifier, Oxygen Safety Valve, Modem
    Compliance with specific standards (ASTM F1100-90, F1246-91, ISO 10651-2, ISO 10651-6, IEC 60601-1-2) and FDA guidance documents."Bench testing has confirmed that the Trilogy 100 Ventilator performs equivalently to the cited device predicates. Performance testing was conducted per the applicable sections of ASTM F1100-90, F1246-91, ISO 10651-2 and ISO 10651-6. EMC testing was performed per IEC 60601-1-2. Electrical, mechanical and environmental testing was performed in accordance with the FDA Draft Reviewers Guidance for Pre-market Notification Submission (1993). Software validation testing was performed per FDA's Guidance for the Content of Premarket Submissions for Software contained in Medical Devices (2005). The results of all testing demonstrate that all design and system requirements for the Trilogy 100 have been met."

    Detailed Study Information (as much as can be inferred from the document):

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size: Not explicitly stated as a "sample size" in the context of patient data. The "test set" here refers to the extensive bench testing and engineering verification and validation activities. The document implies a comprehensive set of tests covering various performance parameters and safety aspects according to industry standards.
      • Data Provenance: The nature of the testing (bench testing, EMC testing, electrical, mechanical, environmental testing, software validation) suggests the data was generated in a controlled laboratory environment. No country of origin for specific data is mentioned beyond Respironics, Inc. being in Monroeville, PA, USA. The testing is prospective in the sense that it was performed specifically for this submission to verify the device's design and performance.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable in the sense of medical experts establishing "ground truth" for patient data. The "ground truth" for this type of submission is defined by the technical specifications and performance requirements derived from predicate devices and applicable standards. The "experts" involved would be the engineers, quality assurance personnel, and regulatory affairs specialists at Respironics responsible for designing, testing, and documenting the device's compliance. Their qualifications are not detailed.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable as this is not a study involving human interpretation of medical data. Performance was measured against predefined technical specifications and standards, likely through automated or manual measurements by testing engineers.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study was done. This device is a ventilator, not an AI-assisted diagnostic tool for human readers.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • This is not an AI algorithm. The device itself (the ventilator) underwent standalone performance testing. The detailed bench testing and compliance with standards are the standalone performance evaluation of the device's functions.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • For this 510(k) submission, the "ground truth" is primarily based on:
        • Standard Specifications: Requirements outlined in ASTM F1100-90, F1246-91, ISO 10651-2, ISO 10651-6, IEC 60601-1-2.
        • Predicate Device Performance: Documented performance characteristics and safety profiles of the legally marketed predicate devices.
        • FDA Guidance Documents: Benchmarking against relevant FDA guidance for ventilators and software.
    7. The sample size for the training set:

      • Not applicable. This is not a machine learning or AI algorithm where a "training set" (implying data used to teach an algorithm) is relevant. The "design phase" and prior knowledge from predicate devices inform the design of the Trilogy 100.
    8. How the ground truth for the training set was established:

      • Not applicable for the reason stated above.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1