Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K152865

    Validate with FDA (Live)

    Date Cleared
    2016-09-27

    (363 days)

    Product Code
    Regulation Number
    868.5750
    Age Range
    18 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TRACOE® smart Cuff Manager is intended to maintain the HVLP (high volume low pressure) cuff pressure, of an endotracheal tube or tracheostomy tube, within the range of 20 to 30 cm H20 through passive control.

    It is for single patient use, and indically ventilated or spontaneously breathing patients. It is used under medical supervision in hospitals, pre-hospital (EMS), extended care facilities or outpatient clinics and is suitable for interor intra-facility transport.

    Device Description

    The TRACOE® smart Cuff Manager is an accessory device for high volume low pressure (HVLP) cuffed endotracheal/tracheostomy tubes to maintain the cuff pressure within a range of 20 cm H₂O to 30 cm H₂O. This device includes a spherical, hard shell (transparent) outer casing with an elastic inner balloon (blue), that is attached to a base housing with both a male and female luer connector. The male connector is inserted into the inflation line (pilot balloon valve of the endotracheal/tracheostomy tube HVLP cuff) and the female connector is attached to a cuff inflation device (syringe or cuff pressure monitor). When the TRACOE® smart Cuff Manager is connected to the cuff inflation line the inner balloon is inflated with a cuff inflation device until the balloon diameter is between 2/3 - 3/4 of the outer casing. The cuff inflation device is then removed from the TRACOE® smart Cuff Manager, and after a few minutes the cuff pressure is rechecked to confirm the cuff is properly inflated inner balloon of the TRACOE® smart Cuff Manager will act as a pressure reservoir, allowing the air to flow between the balloon and the endotracheal/tracheostomy cuff, through passive control, in order to maintain the pressure range.

    When the endotracheal/tracheostomy tube cuff pressure increases, due to compression through patient motion, coughing etc., the surplus air flows slowly into the inner balloon of the TRACOE® smart Cuff Manager through the integrated damping function. The damping function prevents sudden changes in cuff pressure which could result in a gap between the trachea. When the cuff pressure is reduced (following the patient motion) the inner balloon will allow the stored air to freely flow (without damping) back into the endotracheal/tracheostomy tube cuff in order to reinstate the cuff pressure to the prescribed range of 20 cm H₂O to 30 cm H₂O.

    The TRACOE® smart Cuff Manager is for single patient use, and provided sterile for ease of use when connecting in a sterile environment. The device is applicable for mechanically ventilated or spontaneously breathing patients under medical supervision in hospital (EMS), extended care facilities or outpatient clinics and is suitable for inter- or intra-facility transport.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the TRACOE® smart Cuff Manager. It details the device's intended use, its technical characteristics compared to predicate devices, and a summary of non-clinical testing performed to establish substantial equivalence.

    Here's the information regarding acceptance criteria and the study that proves the device meets those criteria:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly present a table of acceptance criteria and reported device performance in a numerical format. However, it states the key performance objective and describes how testing confirmed it.

    The primary performance criterion for the TRACOE® smart Cuff Manager is to maintain the High Volume Low Pressure (HVLP) cuff pressure of an endotracheal tube or tracheostomy tube within the range of 20 to 30 cm H2O through passive control.

    The reported device performance, demonstrated through non-clinical testing, is that the device successfully meets this criterion. The document states:

    • "Mechanical testing confirmed luer connections, pressure reservoir filling, damping function and maintaining the cuff pressure range (20 cm H2O to 30 cm H2O) under multiple situations (e.g., extended time periods, cuff compression)."
    • "Comparison testing with the predicate devices PressureEasy (K833327) and Lanz (K791045) included maintaining the cuff pressure during volume/pressure changes and over an extended period of time."
    • "The results of this testing demonstrated that the TRACOE® smart Cuff Manager is substantially equivalent to the legally cleared predicate devices."
    • "The TRACOE® smart Cuff Manager has passed all defined criteria."

    2. Sample size used for the test set and the data provenance:

    The document describes non-clinical performance and mechanical testing, as well as comparison testing with predicate devices. However, it does not specify the sample size for these tests (e.g., how many devices were tested, how many testing cycles). The tests are performed on the device itself (TRACOE® smart Cuff Manager) and in comparison with predicate devices.

    The data provenance is from non-clinical testing performed by the manufacturer, TRACOE medical GmbH, in Germany. The tests are inherently prospective as they are conducted specifically to validate the device's performance against predefined criteria.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable as the described study is a non-clinical performance and mechanical testing, not a study evaluating expert interpretation or clinical decision-making. The "ground truth" here is the physical measurement of cuff pressure and the device's ability to maintain it within a specified range, as measured by instruments.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This information is not applicable for non-clinical performance and mechanical testing. Adjudication methods are typically used in clinical studies involving expert reviews or subjective assessments, which is not the nature of the described tests.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No, an MRMC comparative effectiveness study was not done. The device (TRACOE® smart Cuff Manager) is a passive mechanical device for maintaining cuff pressure, not an AI-assisted diagnostic or interpretive tool that would involve human readers. Clinical testing was explicitly stated as "not required to demonstrate substantial equivalence."

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    The described testing is essentially standalone performance testing of the mechanical device. The device operates through "passive control" without active human intervention once set up, making the non-clinical tests akin to standalone performance evaluations in a controlled environment.

    7. The type of ground truth used:

    The ground truth used for this device is physical measurement of cuff pressure. This is a quantitative, objective measurement taken by instruments to determine if the device successfully maintains the pressure within the specified range (20-30 cm H2O).

    8. The sample size for the training set:

    There is no training set mentioned or implied because this is a mechanical device, not a machine learning or AI algorithm that requires training data.

    9. How the ground truth for the training set was established:

    As there is no training set, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1