Search Results
Found 1 results
510(k) Data Aggregation
(163 days)
Total Bile Acids Assay is intended for the in vitro quantitative determination of total bile acids (TBA) in human serum samples. Total bile acids are metabolized in the liver and serve as a marker for normal and abnormal liver function. Serum total bile acids are increased in patients with liver disease.
Total Bile Acids Assay contains a bile acid calibrator. The calibrator is design to be used with the assay for the quantitative determination of TBA in serum.
Total Bile Acids Assay has control design to be used with the assay for the quantitative determination of TBA in serum.
The Diazyme's Total Bile Acids assay is a clinical test kit, intended for quantitative determination of total bile acids in serum by an enzymatic method.
The Diazyme's Total Bile Acids assay is comprised of Reagent 2, and a calibrator.
The reagents of the assay kit are stable liquid formulation that allows ease of use coupled with enhanced performance characteristics. In the presence of Thio-NAD, the enzyme 3ox-hydroxysteroid dehydrogenase (3-a-HSD) converts bile acids to 3-keto steroids and Thio-NADH. The reaction is reversible and 3-0-HSD can convert 3-keto steroids and Thio-NADH to bile acids and Thio-NAD. In the presence of excess NADH, the enzyme cycling occurs efficiently and the rate of formation of Thio-NADH is determined by measuring specific change of absorbance at 405nm.
Here's a summary of the acceptance criteria and the study details for the "Total Bile Acids Assay" device, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state formal "acceptance criteria" in a tabular format with specific thresholds before presenting the results. Instead, it presents performance characteristics that were likely used to demonstrate substantial equivalence to the predicate device.
| Performance Characteristic | Acceptance Criteria (Implicit/Inferred) | Reported Device Performance |
|---|---|---|
| Precision | Good precision (low CV%) for both intra-assay and inter-assay measurements, across different concentration levels. | Intra-Assay: 8 μM = 3.9% CV%, 23 μM = 1.3% CV% |
| Inter-Assay: 8 μM = 2.9% CV%, 23 μM = 2.6% CV% | ||
| Correlation with Predicate | "Good correlation" with the legally marketed predicate device (Trinity's method). | Correlation coefficient = 0.99 |
| Interference | "Less than 1% interference" at specified concentrations for common interfering substances. | Triglyceride: 750 mg/dl = <1% interference |
| Bilirubin: 50 mg/dl = <1% interference | ||
| Ascorbic Acid: 50 mg/dl = <1% interference | ||
| Hemoglobin: 500 mg/dl = <1% interference | ||
| Accuracy | (Inferred from combination of correlation, precision, and interference) Device should demonstrate "excellent accuracy" and "good agreement" with the predicate. | Demonstrated by correlation, precision, and interference studies. "Excellent accuration accuration" (typo in original) and "good agreement" claimed. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: The document does not specify the exact sample size for the test set used in the correlation study with the predicate device. It mentions testing "clinical patient samples" in the conclusion but doesn't provide a number. For the precision studies, two concentrations (8 μM and 23 μM) were tested, but the number of replicates is not stated. For interference, it states "pooled human sera" were used, but not the number of samples.
- Data Provenance: The document does not explicitly state the country of origin for the data. The studies appear to be retrospective in the sense that they are laboratory performance evaluations rather than new clinical trials with patient recruitment, but this is an inference based on the type of data presented (precision, correlation, interference). There's no indication of prospective clinical study design.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable to this type of device (in vitro diagnostic assay). For a diagnostic assay, "ground truth" is typically established by reference methods or clinical diagnosis, not by experts reviewing images or other subjective data. The performance is assessed against known values (for precision, linearity) or by comparison with an accepted predicate device (for correlation).
4. Adjudication Method for the Test Set
This information is not applicable as there is no mention of expert review or adjudication for the test set. Performance is based on quantitative measurements.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
This information is not applicable. This device is an in vitro diagnostic assay for measuring total bile acids, not an imaging-based AI device that would involve human readers or image interpretation.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done
This device is an in vitro diagnostic assay, which inherently operates "standalone" in the sense that it quantifies a specific analyte (total bile acids) without human intervention in the measurement process itself, beyond sample preparation and loading. The results are generated by the assay's chemical/enzymatic reaction and detection system. This isn't an AI algorithm in the typical sense that would have a "human-in-the-loop" component for interpretation of its output.
7. The Type of Ground Truth Used
The ground truth used for this device's evaluation includes:
- Known concentrations: For evaluating precision and linearity (though linearity data is not detailed here, it's a standard part of IVD validation).
- Comparison to a legally marketed predicate device: "Trinity's method" was used as the comparator for assessing correlation, implying its results are considered a valid reference for comparison.
- Spiked samples: For interference studies, known interfering substances were added to "pooled human sera" in controlled concentrations.
8. The Sample Size for the Training Set
This information is not applicable. This device is a traditional enzymatic assay, not a machine learning or AI model that requires a "training set" in the context of algorithm development. Its "training" is inherent in the chemical and enzymatic reactions designed by the manufacturer.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable for the same reasons as point 8.
Ask a specific question about this device
Page 1 of 1