Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K133539

    Validate with FDA (Live)

    Device Name
    TOTAL ACROSS
    Manufacturer
    Date Cleared
    2014-03-26

    (128 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TOTAL across™ crossing support catheter is intended to guide and support a guidewire during the access of peripheral arteries, allow for wire exchanges, and provide a conduit for the infusion of saline solutions or diagnostic contrast agents.

    Device Description

    The TOTAL across is an over the wire (OTW) single lumen catheter composed of a spiral hypotube which provides a smooth transition from proximal to distal part, a proximal female hub and a distal flexible part with an atraumatic tapered tip.

    The distal end of the catheter features a lubricious, hydrophilic coating and 2 radiopaque markers. The TOTAL across is 0.014" (0.36mm) guidewire compatible, and features working lengths of 100cm and 150cm.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study proving device performance, based on the provided text:

    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text details various tests performed on the TOTAL across™ crossing support catheter. For each test, the general statement is made that "All of the pre-determined acceptance criteria were met and the testing passed." Specific numerical acceptance criteria are not explicitly stated in this summary, but the successful completion of each test implies adherence to predefined standards.

    Test CategoryReported Device Performance
    Surface Inspection and Device DimensionsMet pre-determined acceptance criteria; testing passed.
    Device PreparationMet pre-determined acceptance criteria; testing passed.
    TrackabilityMet pre-determined acceptance criteria; testing passed.
    FlexibilityMet pre-determined acceptance criteria; testing passed.
    Kink testMet pre-determined acceptance criteria; testing passed.
    Introducer Sheath CompatibilityMet pre-determined acceptance criteria; testing passed.
    Guidewire CompatibilityMet pre-determined acceptance criteria; testing passed.
    Catheter Bond StrengthMet pre-determined acceptance criteria; testing passed.
    Tip PullMet pre-determined acceptance criteria; testing passed.
    Torque StrengthMet pre-determined acceptance criteria; testing passed.
    RadiopacitySatisfactory and comparable to predicate device's visibility.
    Coating EvaluationMet pre-determined acceptance criteria; testing passed.
    Particulate evaluationMet pre-determined acceptance criteria; testing passed.
    Catheter Body Burst PressureMet pre-determined acceptance criteria; testing passed.
    Contrast Media and Saline Flow RateMet pre-determined acceptance criteria; testing passed.
    Hub Test (ISO 594-1:1986 & 594-2:1998)Met pre-determined acceptance criteria; testing passed.
    CrossabilityMet pre-determined acceptance criteria; testing passed.
    Biocompatibility Studies:
    Cytotoxicity Study (ISO MEM Elution)Met pre-determined acceptance criteria; testing passed.
    ISO Guinea Pig Maximization SensitizationMet pre-determined acceptance criteria; testing passed.
    ISO Intracutaneous Study in RabbitsMet pre-determined acceptance criteria; testing passed.
    ISO Systemic Toxicity Study in MiceMet pre-determined acceptance criteria; testing passed.
    USP Pyrogen studyMet pre-determined acceptance criteria; testing passed.
    Bacteria Reverse Mutation StudyMet pre-determined acceptance criteria; testing passed.
    ASTM Hemolysis StudyMet pre-determined acceptance criteria; testing passed.
    C3a Complement Activation AssayMet pre-determined acceptance criteria; testing passed.
    SC5b-9 Complement Activation AssayMet pre-determined acceptance criteria; testing passed.
    In-vivo Thromboresistance StudyMet pre-determined acceptance criteria; testing passed.
    ASTM Partial Thromboplastin timeMet pre-determined acceptance criteria; testing passed.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not provide specific numerical sample sizes for the "test set" (bench testing or biocompatibility testing). It refers to the tests being conducted on "the TOTAL across" and "the finished TOTAL across," implying samples of the device were used.

    The data provenance is described as "bench testing" and "biocompatibility testing." This typically implies laboratory-based, controlled studies, not human clinical trials. The country of origin of the data is not specified, but the applicant (Medtronic Vascular) is based in Santa Rosa, CA, USA, and an alternate contact is based in Roncadelle, ITALY. The testing was conducted in accordance with FDA guidance documents and ISO standards, which are internationally recognized. The studies are prospective in nature, as they involve testing new iterations of the device to ensure conformity to standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable to this submission. The "ground truth" for the performance and biocompatibility studies of a medical device like this catheter is based on established engineering principles, international standards (ISO, ASTM), and FDA guidance documents, not on human expert consensus derived from image interpretations or clinical assessments. The "experts" involved would be the engineers, scientists, and technicians conducting the testing and verifying the results against the predefined specifications. Their qualifications are inherent in their ability to perform and interpret these specialized tests.

    4. Adjudication Method for the Test Set

    This is not applicable to this type of device and testing. Adjudication methods like 2+1 or 3+1 are used in clinical trials or studies involving human assessment (e.g., image interpretation) where there might be disagreement among experts. For bench and biocompatibility testing, the results are typically quantitative and objective, measured against predefined acceptance criteria. Any discrepancies would be resolved through re-testing or thorough investigation of the testing methodology, not through expert adjudication in the traditional sense.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices that assist human readers in interpreting medical images or data. The TOTAL across™ is a physical medical device (a crossing support catheter), and its performance is evaluated through bench testing and biocompatibility studies, not through human-in-the-loop diagnostic tasks. Therefore, there is no AI component or human reader improvement effect size to report.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    A standalone performance evaluation (algorithm only) was not done as this device does not contain an AI algorithm. Its performance is assessed through its physical and material properties in a standalone manner (i.e., the device itself is tested without human intervention in its function, unlike, for example, a surgical tool where human skill is key to its performance).

    7. The Type of Ground Truth Used

    The "ground truth" for the performance tests relies on:

    • Established engineering specifications and design requirements: These are the inherent standards the device is designed to meet, such as specific dimensions, strength, flexibility, etc.
    • Applicable FDA guidance documents and ISO/ASTM standards: These provide the methodologies and acceptance criteria for evaluating various characteristics (e.g., ISO 594 for hub tests, ANSI/AAMI/ISO 10993 for biocompatibility).
    • Comparison to predicate devices: For some characteristics, the ground truth for "acceptable" performance is established by demonstrating comparability to legally marketed predicate devices. For example, radiopacity was considered "satisfactory and comparable to the predicate device's visibility."

    8. The Sample Size for the Training Set

    This is not applicable. The TOTAL across™ is a physical medical device and not an AI/machine learning model, so there is no "training set" in the context of data for an algorithm. The "training" in the manufacturing context refers to the manufacturing process that creates the device, but not a data set.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable for the reasons stated in point 8. There is no training set for an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1