Found 2 results

510(k) Data Aggregation

    K Number
    K232802

    Validate with FDA (Live)

    Date Cleared
    2023-10-12

    (30 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TMINI™ Miniature Robotic System is indicated as a stereotaxic instrumentation system for total knee replacement (TKA) surgery. It is to assist the surgeon by providing software-defined spatial boundaries for orientation and reference information to identifiable anatomical structures for the accurate placement of knee implant components.

    The robotic device placement is performed relative to anatomical landmarks as recorded using the system intraoperatively and based on a surgical plan determined preoperatively using CT based surgical planning tools.

    It includes a handheld robotic device, an optical sensor navigation system and accessories, software system, surgical instruments and accessories.

    The targeted population has the same characteristics as the population that is suitable for the implant(s) compatible with the TMINI™ Miniature Robotic System. The TMINIM Miniature Robotic System is to be used with the following knee replacement systems in accordance with the indications and contraindications:

    • · Enovis™ EMPOWR Knee System®
    • · Ortho Development® BKS® and BKS TriMax® Knee System
    • · Total Joint Orthopedics Klassic® Knee System
    • · United® U2™ Knee Total Knee System
    Device Description

    Like its predicate, the TMINI™ Miniature Robotic System consists of three primary components: a three-dimensional, graphical, Preoperative Planning Workstation (TPLAN Planning Station), an Optical Tracking Navigation Console (TNav) and a Robotically Controlled Hand-held Tool (TMINI Robot) that assists the surgeon in preparing the bone for implantation of TKA components.

    The TPLAN Planning Station uses preoperative CT scans of the operative leg to create 3D surface models for case templating and intraoperative registration purposes. The Planning Workstation contains a library of 510(k) cleared knee replacement implant(s). The surgeon can select an implant model from this library and manipulate the 3D representation of the implant in relation to the bone model to place the implant. Once the surgeon is satisfied with the implant location and orientation, the data is written to a file that is used to guide the robotically controlled hand-held tool.

    The handheld robotic tool is optically tracked relative to optical markers placed in both the femur and tibia and articulates in two degrees-of-freedom, allowing the user to place bone pins in a planar manner in both bones. Mechanical guides are clamped to the bone pins, resulting in subsequent placement of cut slots and drill quide holes such that the distal femoral and proximal tibial cuts can be made in the pre-planned positions and orientations, and such that the implant manufacturer's multi-planer cutting block can be placed relative to drilled distal femoral pilot holes.

    AI/ML Overview

    The provided text describes the TMINI™ Miniature Robotic System (Additional Knee Systems) and its substantial equivalence to a predicate device (K230202) for total knee replacement (TKA) surgery. The primary modification in this submission is the addition of compatibility with three new knee implant systems.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly list quantitative acceptance criteria in a table format with specific thresholds (e.g., "Accuracy must be within +/- X mm"). Instead, it states that performance testing was conducted and passed. The performance criteria are implicitly linked to the "cutting accuracy" and "cadaver lab validation testing."

    Acceptance Criteria CategorySpecific Acceptance Criteria (Inferred from text)Reported Device Performance
    Cutting Accuracy VerificationDevice must meet all test criteria and specifications for cutting accuracy, similar to the predicate.Passed (All test criteria and specifications met)
    Cadaver Lab Validation TestingDevice must meet all test criteria in simulated surgical testing in a cadaver model, similar to the predicate.Passed (All test criteria met)
    BiocompatibilityMaterials must pass standard biocompatibility tests (Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Pyrogenicity).Passed (All 5 tests passed the acceptance criteria of the test protocol)
    Risk AssessmentThe addition of new implant systems must not introduce new clinical hazards or increase the likelihood/severity of existing hazards.Passed (No new clinical hazards identified; existing risks unchanged)

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document does not explicitly state the sample size for the "cutting accuracy verification" or the "cadaver lab validation testing." It mentions "simulated surgical testing in a cadaver model," implying a limited number of cadavers were used, but the exact count is not given.
    • Data Provenance: The document does not specify the country of origin for the data. The "cadaver model" likely refers to ex vivo testing in a lab setting, not human clinical data. The review is for a 510(k) submission to the US FDA, so the data would have been submitted to support this application. It is implicitly a pre-market study, and given the nature of a 510(k) (seeking substantial equivalence), it would be considered prospective for the purpose of demonstrating the device's performance for this specific submission, even if the methods were established previously.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not provide any information regarding the number of experts, their qualifications, or how they established ground truth for the performance testing.

    4. Adjudication Method for the Test Set

    The document does not provide any information regarding an adjudication method for the test set. Given the "cutting accuracy" and "cadaver lab validation" descriptions, the "ground truth" likely involves precise physical measurements rather than subjective expert interpretations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically used for diagnostic or screening devices where human interpretation of images is a key component, often comparing human readers with and without AI assistance. The TMINI™ Miniature Robotic System is a surgical assistance device, and the testing described focuses on its mechanical accuracy and safety, not diagnostic interpretation. Therefore, there is no mention of human readers or an effect size of AI assistance on their performance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    The performance testing described ("cutting accuracy verification" and "cadaver lab validation testing") can be interpreted as standalone performance of the robotic system in achieving its intended surgical goal. While the system "assists the surgeon," the testing focuses on the robot's precision in placing bone pins and enabling cuts, which is an intrinsic performance characteristic of the algorithm and hardware. It demonstrates the direct output of the robotic system.

    7. The Type of Ground Truth Used

    The ground truth for "cutting accuracy verification" would likely be:

    • Precise Physical Measurements: Using highly accurate measurement tools (e.g., CMM, optical tracking systems, or specialized gauges) to quantify the deviation of the cuts or pin placements from the pre-planned surgical model.

    The ground truth for "cadaver lab validation testing" would similarly involve:

    • Physical Measurements Post-Procedure: Assessing the accuracy of implant component placement or bone cuts on the cadaveric specimens against the pre-operative plan. This might involve post-procedure imaging (e.g., CT scans) and subsequent measurement analysis.

    The biocompatibility testing ground truth is based on standard chemical and biological assay results.

    8. The Sample Size for the Training Set

    The document does not directly address or mention a "training set" in the context of an AI/ML algorithm. The TMINI™ system uses "software-defined spatial boundaries" and "preoperative CT scans" for planning. While there's software involved, the description doesn't suggest a machine learning model that undergoes a training phase with a large dataset. It appears to be a deterministic robotic system guided by a surgical plan created from CT data.

    9. How the Ground Truth for the Training Set Was Established

    As no "training set" for an AI/ML algorithm is explicitly mentioned or seems applicable to this device as described, the question of how its ground truth was established is not relevant based on the provided text. The "TPLAN Planning Workstation" has a "library of 510(k) cleared knee replacement implant(s)," and surgeons use this to manipulate 3D representations for planning, which then guides the robot. This implies a rules-based or model-based system rather than a machine learning one requiring a distinct training phase.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230202

    Validate with FDA (Live)

    Date Cleared
    2023-04-25

    (90 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TMINI™ Miniature Robotic System is indicated as a stereotaxic instrumentation system for total knee replacement (TKA) surgery. It is to assist the surgeon by providing software-defined spatial boundaries for orientation and reference information to identifiable anatomical structures for the accurate placement of knee implant components.

    The robotic device placement is performed relative to anatomical landmarks as recorded using the system intraoperatively and based on a surgical plan determined preoperatively using CT based surgical planning tools.

    It includes a handheld robotic device, an optical sensor navigation system and accessories, software system, surgical instruments and accessories.

    The targeted population has the same characteristics as the population that is suitable for the implant(s) compatible with the TMINI™ Miniature Robotic System. The TMINI™ Miniature Robotic System is to be used with the following knee replacement system(s) in accordance with the indications and contraindications: Enovis™ EMPOWR Knee System®.

    Device Description

    The TMINI™ Miniature Robotic System consists of three primary components: a three-dimensional, graphical, Preoperative Planning Workstation (TPLAN Planning Station), an Optical Tracking Navigation Console (TNav) and a Robotically Controlled Hand-held Tool (TMINI Robot) that assists the surgeon in preparing the bone for implantation of TKA components.

    The TPLAN Planning Station uses preoperative CT scans of the operative leg to create 3D surface models for case templating and intraoperative registration purposes. The Planning Workstation contains a library of 510(k) cleared knee replacement implant(s). The surgeon can select an implant model from this library and manipulate the 3D representation of the implant in relation to the bone model to optimally place the implant. Once the surgeon is satisfied with the implant location and orientation, the data is written to a file that is used to guide the robotically controlled hand-held tool.

    The handheld robotic tool is optically tracked relative to optical markers placed in both the femur and tibia and articulates in two degrees-of-freedom, allowing the user to place bone pins in a planar manner in both bones. Mechanical guides are clamped to the bone pins, resulting in subsequent placement of cut slots and drill guide holes such that the distal femoral and proximal tibial cuts can be made in the pre-planned positions and orientations, and such that the implant manufacturer's multi-planer cutting block can be placed relative to drilled distal femoral pilot holes.

    AI/ML Overview

    The provided text does not contain detailed information about specific acceptance criteria and a study proving the device meets them in the way described in the prompt. Instead, it focuses on demonstrating substantial equivalence to a predicate device (Rosa Knee System) for FDA clearance.

    However, I can extract the information related to performance testing that is mentioned as part of the substantial equivalence claim.

    Here's an attempt to answer based on the available information, highlighting where details are missing:

    The TMINI™ Miniature Robotic System (K230202) underwent performance testing to support its substantial equivalence to the legally marketed predicate, Rosa Knee System (K182964).

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryAcceptance Criteria (from predicate)Reported Device Performance (TMINI™)
    BiocompatibilityPassedPassed
    Cutting Accuracy< 0.75mm RMSEEquivalent or better than manual instrumentation (< 0.5mm RMSE)
    Cadaver Lab Validation TestingPassedPassed
    Software TestingPassedPassed
    Electromagnetic Compatibility and Electrical SafetyMeets IEC RequirementsMeets IEC Requirements

    2. Sample size used for the test set and the data provenance:

    • Cutting Accuracy: The text states "< 0.5mm RMSE" as the reported device performance. It also mentions "See Section 19 for Details" regarding Cutting Accuracy and Cadaver Lab Validation. Without access to Section 19, the specific sample size, data provenance (e.g., country of origin, retrospective/prospective), and whether this refers to a test set in the context of an AI study are not available in this document. This typically refers to a performance study on cadavers or simulated bone, not an AI test set with "data provenance" in the usual imaging sense.
    • For Biocompatibility, Software, EMC/Electrical Safety, and Cadaver Lab Validation, specific sample sizes and data provenance are not detailed in this summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the document. The performance tests mentioned (Biocompatibility, Cutting Accuracy, Cadaver Lab Validation, Software, EMC/Electrical Safety) are typically engineering and clinical performance evaluations, not diagnostic accuracy studies involving expert interpretation of data.

    4. Adjudication method for the test set:

    This information is not provided because the performance tests described do not involve adjudication by experts in the context of a diagnostic AI study.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is irrelevant for this submission. The TMINI™ Miniature Robotic System is described as a stereotaxic instrumentation system to assist surgeons in total knee replacement surgery by providing software-defined spatial boundaries and orientation/reference information. It is not an AI-powered diagnostic tool that assists human readers (e.g., radiologists) in interpreting images. Therefore, an MRMC study is not applicable to this device type.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    The device description indicates human-in-the-loop operation, as it "assists the surgeon" and involves a "handheld robotic device" and a "surgical plan determined preoperatively using CT based surgical planning tools." The performance claim for cutting accuracy is for the robotic system, implying its standalone mechanical accuracy, but it's always in the context of assisting a surgeon.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    For Cutting Accuracy, the ground truth would likely be a precise physical measurement of the achieved cut planes/pin placements compared to the pre-planned positions and orientations. This would be established through metrology or other highly accurate measurement techniques in a laboratory or cadaveric setting. For other tests like Biocompatibility, Software, EMC/Electrical Safety, and Cadaver Lab Validation, the ground truth would be adherence to established standards or clinical protocols.

    8. The sample size for the training set:

    This information is not provided. The TMINI™ system uses CT-based surgical planning tools. While software is involved, the primary focus of the performance testing presented is on the accuracy and safety of the robotic system's physical actions rather than a machine learning model's training data.

    9. How the ground truth for the training set was established:

    This information is not provided. If there is a machine learning component, details of its training data and ground truth establishment are not included in this summary. However, given the nature of the device (robotic surgical assistance), "training set" might refer to data used to develop the planning algorithms or robotic control, which would have different ground truth establishment methods than, for example, an image classification AI.