Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K083180

    Validate with FDA (Live)

    Date Cleared
    2009-02-12

    (107 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Avantis Medical Systems, Inc. Third Eye is indicated for use in the instrument channel of a conventional colonoscope to provide retrograde illumination and visualization of the colon for diagnostic purposes.

    Device Description

    The Third Eye System consists of both disposable portions and facility equipment. The disposable portions include the Third Eye Retroscope and Caps. The facility equipment portions of the device include the Third Eye Video Processor and accessories. The Third Eye is designed as an auxiliary device for use during a colonoscopy procedure. After a colonoscope has been advanced to the cecurn, the Third Eye Retroscope is inserted through the instrument channel of the colonoscope. As the Third Eye Retroscope emerges from the distal tip of the instrument channel of the colonoscope, it automatically bends 180 degrees to form a "J" shape. The Third Eye then provides a continuous retrograde image of the colon throughout the process of withdrawal of the colonoscope. The Third Eve Video Processor is a piece of electrical equipment that acts as an interface for the Third Eye Retroscope, the video output signals, and user input controls. The output video signals from the Video Processor are displayed on a monitor for viewing by the physician. Application software that runs on the Video Processor allows the user to adjust the light intensity and various picture settings of the Third Eve Retroscope image.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided text for the Avantis Medical Systems' Third Eye Retroscope Auxiliary Endoscopy System:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implicit)Reported Device Performance
    Safety: Device modifications do not raise new or unresolved safety issues. Adherence to general controls provisions (registration, labeling, GMP)."Modifications made to the device and accessories did not change the fundamental scientific technology of the device. The testing described demonstrates that the differences in the device components and accessories do not raise any new or unresolved issues for safety and efficacy." "Risks to patients in terms of adverse events are comparable to those reported in the literature for traditional screening colonoscopies. Use of the Third Eye did not increase the risks to the patients..."
    Effectiveness/Performance: Device performs as intended."Bench verification testing was conducted to verify that the modified device meets the design inputs and intended performance characteristics. Results demonstrate that modifications made to the Third Eye design performs as intended." "Results from a prospective, multi-center post-market study demonstrate that retrograde viewing of the colon with the Third Eye improves the detection of polyps and adenomas compared to traditional colonoscopy alone. Results were statistically significant."
    Clinical Benefit: Improves detection of polyps and adenomas."Avantis Medical Systems, Inc. has determined, based on clinical data, that the Third Eye improves the detection of polyps and adenomas in patients for whom colonoscopies are indicated." "Use of the Third Eye did not increase the risks to the patients but offered a statistically significant benefit of improved detection."
    Substantial Equivalence: To the predicate device (K070330) despite modifications."The Avantis Third Eye is substantially equivalent in operation and fundamental scientific technology to its previous version, cleared under K070330." "The modified device is substantially equivalent to the cleared predicate device in fundamental design materials, processes, sterile barrier packaging, and sterilization. All changes have been subjected to risk analysis and have been verified to meet current design specifications."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated in terms of a specific number of patients or cases. The document refers to a "multi-center post-market study."
    • Data Provenance:
      • Country of Origin: Not specified.
      • Retrospective or Prospective: Prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • This information is not provided in the document. The document states that the Third Eye "improves detection of polyps and adenomas," implying a comparison, but it does not detail how the ground truth for "polyps and adenomas" was established in the clinical study (e.g., standard colonoscopy findings, pathology, or a panel of experts).

    4. Adjudication Method for the Test Set

    • This information is not provided in the document.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    • It states that a clinical study "demonstrate(s) that retrograde viewing of the colon with the Third Eye improves the detection of polyps and adenomas compared to traditional colonoscopy alone." This indicates a comparative effectiveness study, however, it's not explicitly labeled as an MRMC study and the specific effect size quantifying "how much human readers improve with AI vs without AI assistance" is not provided. The document only states that the results were "statistically significant." The device itself is an auxiliary endoscope, not an AI, so the framing of "human readers improve with AI vs without AI assistance" isn't directly applicable in this context. It's about improved detection with the device vs. without.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • No, this was not done. The device is an auxiliary endoscopic system designed for human use during a colonoscopy to provide additional visualization. It is not an algorithm that operates independently.

    7. The Type of Ground Truth Used

    • The document states the study measured "detection of polyps and adenomas." While not explicitly defined as pathology, in the context of colonoscopy clinical trials, pathology from biopsied or resected lesions is the generally accepted gold standard for confirming polyps and adenomas. However, the document does not definitively state that pathology was the sole or primary ground truth method.

    8. The Sample Size for the Training Set

    • This information is not applicable/not provided. The Third Eye Retroscope is a medical device, not an AI algorithm that requires a "training set" in the machine learning sense. The clinical study described is for validation of the device's performance.

    9. How the Ground Truth for the Training Set Was Established

    • This information is not applicable/not provided for the same reason as point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K070330

    Validate with FDA (Live)

    Date Cleared
    2007-04-24

    (78 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Avantis Medical Systems, Inc. Third Eye Retroscope Auxiliary Endoscopy System is intended for use in the instrument channel of a standard colonoscope to provide retrograde illumination and visualization of the colon for diagnostic purposes.

    Device Description

    The Third Eye Retroscope Auxiliary Endoscopy System is designed as an auxiliary device for use during a colonoscopy procedure. After a colonoscope has been advanced to the cecum, the Third Eye Retroscope auxiliary Endoscopy device is inserted through the instrument channel of the colonoscope. As the auxiliary device emerges from the distal tip of the instrument channel of the colonoscope, the auxiliary device automatically bends 180 degrees to form a "J" shape. The device then provides a continuous retrograde image of the colon throughout the process of withdrawal of the colonoscope. The Third Eye Retroscope is deployed through the instrument channel of a colonoscope in what is commonly referred to as a mother-daughter configuration. As a result, only approximately 4 cm of the distal tip of the device protrudes from the distal end of the colonoscope's instrument channel except for a few seconds during insertion and removal, when up to 9 cm of the distal tip extends beyond the end of the colonoscope. Any contact of the device with the colonic mucosa is transitory. The system consists of both disposable portions, and facility equipment. The disposable portions include the Third Eye Retroscope and Cap. The Third Eye Retroscope is an endoscope that consists of a distal section, flexible insertion tube and proximal section. The distal section houses an imaging sensor, light source and related circuitry. The flexible insertion tube consists of a standard braided catheter with electrical wires located inside the lumen. These wires carry signals from the distal region to the proximal region. The proximal section includes a torquing knob, flexible wire extension region, and electrical connector. The Third Eye Cap with polarizing filter is placed on the distal end of the colonoscope. The facility equipment portions of the device include the Third Eye Retroscope Video Processor and accessories. The video processor is a piece of electrical equipment that acts as an interface box for the Third Eve Retroscope, the video output signals, and user input controls. The output video signals from the video processor are displayed on a monitor for viewing by the physician. Application software that runs on the video processor allows the user to adjust various picture settings of the Third Eye Retroscope image and adjust the light intensity of the Retroscope. Additionally, the software facilitates the capture of still images, the export of these still images, and allows a user to save system settings under a profile name.

    AI/ML Overview

    This document is a 510(k) summary for the Avantis Medical Systems, Inc. Third Eye™ Retroscope™ Auxiliary Endoscopy System. It primarily focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed acceptance criteria and a study design with performance metrics typically associated with AI/ML device evaluations.

    Therefore, many of the requested data points (like sample size for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, training set sample size, and ground truth for training set) are not applicable or not provided in this type of regulatory submission. This document predates the widespread use of sophisticated AI/ML in medical devices and the associated regulatory expectations for performance studies.

    However, I can extract the available information related to performance and create a table for what is provided.

    Here's the breakdown of the information based on the document:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative acceptance criteria or detailed performance metrics. Instead, it relies on demonstrating substantial equivalence to predicate devices in terms of intended use, operation, materials, and safety through in vitro bench testing.

    Feature/ParameterAcceptance Criteria (Implied by Substantial Equivalence)Reported Device Performance (as per submission)
    Intended Use"Indication for use" similar to predicate devices."Intended for use in the instrument channel of a standard colonoscope to provide retrograde illumination and visualization of the colon for diagnostic purposes." This is deemed similar to the combined intended uses of the predicate devices for colonoscopy, with the difference being a retrograde view.
    OperationSimilar operation to predicate devices (e.g., insertion method, image generation, control via video processor).The Retroscope is inserted into the colonoscope's instrument channel similar to biopsy forceps. It provides images of the colon similar to a colonoscope, but with a retrograde view. Controlled by the Third Eye Video Processor similar to how a colonoscope is controlled by a Video System Center.
    MaterialsSimilar materials to predicate devices.(Not explicitly detailed in this summary, but presumed to be comparable to materials used in similar endoscopes/accessories to ensure biocompatibility and function).
    SafetySimilar safety profile to predicate devices.The device is evaluated through in vitro bench testing to demonstrate safety. Contact with colonic mucosa is described as transitory.
    Imaging PerformanceEquivalent "Imaging Performance" to predicate colonoscopes.Provides retrograde illumination and visualization. The software allows adjustment of picture settings and light intensity.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not specified. The document only mentions "in vitro bench testing" without providing details on participants or specific datasets.
    • Data Provenance: Not specified. Given it's "in vitro bench testing," it likely refers to laboratory testing and not human or animal study data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Number of Experts/Qualifications: Not applicable/not specified. The testing described is "in vitro bench testing," which typically does not involve expert readers establishing ground truth for diagnostic accuracy in the way AI/ML systems are evaluated.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Adjudication Method: Not applicable/none specified. This type of adjudication is relevant for studies involving human interpretation and ground truth establishment, which isn't described here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No. This document is for a traditional medical device (an auxiliary endoscope system), not an AI/ML diagnostic system. Therefore, an MRMC comparative effectiveness study regarding "human readers improve with AI vs without AI assistance" was not performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: Not applicable. The Third Eye Retroscope is an auxiliary endoscopy system used by a physician (human-in-the-loop). It's a visual aid, not an autonomous diagnostic algorithm. While it has "application software that runs on the video processor," this software adjusts picture settings and light intensity, and facilitates image capture, rather than providing autonomous diagnostic interpretations. The document refers to "in vitro bench testing" for performance evaluation, which is likely about image quality, functional integrity, and safety attributes of the device itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Type of Ground Truth: Not explicitly stated or required for this type of 510(k) submission which relies on substantial equivalence based on physical and functional characteristics of the device tested via "in vitro bench testing." For imaging performance, the "ground truth" would likely be objective measurements of image resolution, clarity, field of view, illumination, and mechanical reliability, compared against expected performance or established standards.

    8. The sample size for the training set

    • Training Set Sample Size: Not applicable. This device is not an AI/ML product developed using training datasets in the conventional sense. Its "application software" serves control and display functions, not machine learning.

    9. How the ground truth for the training set was established

    • Ground Truth for Training Set: Not applicable for the reasons mentioned above.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1