Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K043482

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-12-21

    (4 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reliance PI Process Indicator is a single use chemical indicator designed to change from the "START" dark pink color to the "ENDPOINT" orange color or lighter upon exposure to an effective dose of peracetic acid generated from Reliance™ DG Dry Germicide during the Reliance Endoscope Processing Cycle.

    The Reliance PI Process Indicator is intended to be used exclusively for the independent monitoring of the Reliance Endoscope Processing Cycle. The Reliance PI Process Indicator is not an indicator of load disinfection or cleaning efficacy.

    Device Description

    The Reliance PI Process Indicator is a single use chemical indicator strip that is designed to confirm an effective dose of peracetic acid (PAA), the active ingredient generated from Reliance DG Dry Germicide in the Reliance Endoscope Processor's HLD cycle.

    This test strip consists of a 0.2-square inch reagent-containing indicator pad attached to one end of a 0.2 x 3.25-inch polystyrene handle. The indicator pad contains an indicator dye with a background dye that together form a dark pink color.

    Upon exposure to the effective dose of PAA generated during the Reliance Endoscope Processing Cycle, the indicator dye only is oxidized by the Reliance DG Dry Germicide use dilution during the Reliance Endoscope Processing Cycle. During the processing cycle, the indicator pad color transitions through a lighter pink color until an orange coloration, based on the background dye, is visible.

    AI/ML Overview

    The Reliance™ PI Process Indicator is a chemical indicator designed to confirm an effective dose of peracetic acid (PAA) in the Reliance Endoscope Processor's High-Level Disinfection (HLD) cycle.

    Here's an analysis of its acceptance criteria and the supporting study:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Endpoint Specifications)Reported Device Performance
    Complete Color Change (Pass): When exposed to PAA at a dose ≥ 11,500 mg/L PAA min generated from Reliance DG Dry Germicide during the Reliance Endoscope Processing Cycle, there will be a complete change to endpoint color (orange) or lighter in 100% of the indicator strips.Sensitivity Studies: The Reliance PI Process Indicator develops the complete ENDPOINT color (pass) when the dose of PAA is ≥ 11,500 mg/L PAA min. Simulated-Use Study (27 cycles): Demonstrated "complete" PI readings in all 27 cycles, indicating detection of intended PAA levels.
    Incomplete Color Change (Fail): When exposed to the minimum recommended dose of 9000 mg/L PAA min or less, there will be an incomplete color change in 100% of the indicator strips.Sensitivity Studies: Fails to reach the complete ENDPOINT color (incomplete endpoint reading, not acceptable) when the PAA dose is < 9000 mg/L PAA min. Simulated-Use Study (9 cycles): Demonstrated an "incomplete" PI reading in each of the 9 cycles when processed with a PAA dose of 9000 mg/L PAA min (minimum recommended dose).
    Specificity: Components of the Dry Germicide's container cannot independently cause a color change to ENDPOINT.Specificity Testing: Confirmed that the two components of the Dry Germicide's container cannot independently cause a color change to ENDPOINT.
    Repeatability and Precision: Demonstrated over numerous repeated assays using varying parameters of use.Repeatability and Precision: Confirmed over numerous repeated assays using varying parameters of use.
    Stability of PI Color: Confirmed when reading delays occur in the closed processor or after the PI is removed from the processor.Stability of PI Color: Confirmed when reading delays occur in the closed processor or after the PI is removed from the processor.
    Stability of Strips (Unopened): 2 years of storage under labeled storage conditions.Stability of Strips in unopened bottles: Established through 24 months of storage under labeled storage conditions.
    Shelf-life (Opened): 6 months after opening.Use-life of the opened bottle: Validated through 6 months after opening.
    Performance in Reliance Processing Cycle: Performance validated with and without worst-case washing phases, with maximum air purge time, and with conditions of delayed PI reading.Performance in Reliance Processing Cycle: Validated under these varying conditions.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sensitivity Studies: The exact sample size for the "Sensitivity Studies" is not explicitly stated as a number of strips but implies multiple tests to determine the PAA dose thresholds for pass/fail.
    • Simulated-Use Studies:
      • One study involved 27 cycles using eight flexible endoscopes to detect intended PAA levels.
      • Another study involved 9 cycles using three worst-case flexible endoscopes to test the minimum recommended PAA dose (9000 mg/L PAA min).
    • Data Provenance: Not explicitly stated, but the submission is from STERIS Corporation in Mentor, Ohio, USA, suggesting the studies were conducted within the US or under their oversight. The studies appear to be prospective as they were specifically designed and executed for device validation.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    The document does not explicitly state the number of experts or their qualifications used to establish the ground truth for interpreting the process indicator's color change or the PAA concentration measurements. However, it does mention "correctly interpreted by the clinical users" in an in-use study, implying clinical personnel were involved in observing the indicator results in a real-world setting, but their expert role in establishing the ground truth for validation is unclear.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method (e.g., 2+1, 3+1, none) for the test set. The evaluation of the indicator's color change (pass/fail) seems to be based on predefined objective criteria related to specific PAA concentrations, rather than subjective expert consensus requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A multi-reader multi-case (MRMC) comparative effectiveness study was not conducted in the context of comparing human readers' performance with and without AI assistance, as this device (a chemical indicator) does not involve AI or human interpretation in the same way an imaging diagnostic device would. It's a binary "pass/fail" chemical reaction.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    This is not applicable as the device is a physical chemical indicator, not an algorithm. Its performance is inherent to its chemical reaction to PAA, and while human observation is required to read the indicator, the "performance" described refers to the indicator's objective chemical response.

    7. Type of Ground Truth Used

    The ground truth used for the device's performance relies on objective chemical measurements of peracetic acid (PAA) concentrations.

    • "Kinetic Studies validated the Reliance Endoscope Processor's delivery of peracetic acid in concentrations appropriate for PI performance testing."
    • "The complete ENDPOINT color (pass) when the dose of PAA is >= 11,500 mg/L PAA min."
    • "Fails to reach that color (incomplete endpoint reading, not acceptable) when the PAA dose is < 9000 mg/L PAA min."
      This indicates that the ground truth for "effective dose" and "minimum recommended dose" was established by precise measurements of PAA concentration (mg/L PAA min).

    8. Sample Size for the Training Set

    The document does not specify a separate "training set" as this is a chemical indicator, not a machine learning model. The validation tests described serve to establish its performance characteristics against defined chemical criteria.

    9. How the Ground Truth for the Training Set Was Established

    As there's no "training set" in the machine learning sense, this question is not applicable. The underlying "ground truth" for the device's function (i.e., known PAA concentrations and their effect on the indicator) would have been established through controlled laboratory experiments and chemical analysis during the development and kinetic studies of the indicator itself.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1