510(k) Data Aggregation

    K Number
    K240597

    Validate with FDA (Live)

    Device Name
    TERA HARZ CLEAR
    Manufacturer
    Date Cleared
    2024-03-05

    (1 days)

    Product Code
    Regulation Number
    872.5470
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TERA HARZ CLEAR is intended for the treatment of tooth malocclusions in patients with permanent dentition (i.e, all second molars). Utilizing a series of increments, it sequentially positions teeth by way of continuous gentle force. TERA HARZ CLEAR is intended exclusively for professional dental work. Fabrication of aligner with TERA HARZ CLEAR requires an additive manufacturing system (AMS) that includes compatible with the following :
    Design:
    Scanner 3Shape A/S TRIOS 3 Basic
    Design software 3Shape A/S 3Shape Ortho System™
    Printing:
    3D Printer UNIZ SLASH 2
    UNIZ Nbee
    SprintRay Inc. SprintRay Pro 95
    Post-Curing:
    Post-cure unit CureM U102H
    Tera Harz Cure THC

    Device Description

    TERA HARZ CLEAR is a series of clear aligners that are used to replace traditional orthodontic wires and brackets for the alignment of maloccluded or misaligned teeth. TERA HARZ CLEAR have 3 models : TC-85DAC, TR-07, TA-28. This series of aligner is intended for the treatment of tooth malocclusions in patients with permanent dentition (i.e., all second molars) by moves the teeth gently, and in small increments, from their original to their final treated position for improved dental alignment. This series of aligner is worn for approximately 1 week of 20 to 22 hours per day after (However, there are to be removed for eating and for cleaning) which it is replaced by the next stage aligners and are designed to be used in a sequence. This is repeated for duration as prescribed by the dental clinician. TERA HARZ CLEAR is a light-cured, methacrylate-based resin commonly can used in additive manufacturing for the production of dental structures such as sequential aligners. TERA HARZ CLEAR has stored in a black 1,000 g of HDPE bottle. It contains materials with colors of yellowish. This device is a liquid photo-curable material that is polymerized by UV laser at 405412 nm. It can be used to make a tooth model with a photo-curable polymer that is cured by ultraviolet light. The liquid UV curing resin is cured at a specific wavelength (395405 nm) by the photo-initiator contained in the resin. It is typically 100 um in thickness and is output at a resolution of 40 to 90 um on the x, y axis, and it is possible to produce threedimensional printed matter by curing lamination step by step a thickness of 100 um. However, scanner, design software, 3D printer and post-cure unit are not included with the device. These fabrications of TERA HARZ CLEAR are beginning with the dental clinician prescribing aligners to treat a patient's malocclusion, and decision to use methacrylate- based resins is made by the dental clinician. TERA HARZ CLEAR, an orthodontic appliance such as sequential aligner, is manufactured in a 3D printer that is compatible. The dental clinician can generate a digital file by scanning the patient's mouth directly using listed with Intraoral scanner under FDA Classification Product Code NOF, regulation872.3661. This digital file is a series of CAD files (.stl) for building models that can be used to fabricate aligners. Commonly used standard dental software is used by dental professionals to virtually design a sequential aligner and generate an industrystandard "STL" 3D dataset which reflects the intended shape and contour. The design software used is 3Shape Ortho System™ by 3Shape A/S (K180941). The specialized orthodontic treatment planning software has a 510k clearance for the intended use under FDA Classification Product Code PNN, regulation 872.5470. This software is used for management of 3D scanned orthodontic models, orthodontic diagnosis by measuring, analyzing, inspecting and visualizing 3D scanned orthodontic models, virtual planning of orthodontic treatments by simulating tooth movements, and design of sequential aligners based on 3D scanned orthodontic models. Once dental clinic manufacturing unit receive the data that *.stl CAD files of the treatment plan the 3D printer begins additive manufacturing. The dental clinician generates sequential 3D printed models replicating the approved treatment plan. The sequential aligner is 3D printed and cured in a post-cure unit. The fabricated aligners are cut to fit dentition, the cleaned and polished to remove rough edges by the dental clinician. The prescribing physician review and approves the sequential aligners areprovides them to the patient the confirming fit and design.

    AI/ML Overview

    The provided text describes the regulatory clearance of a dental device, "TERA HARZ CLEAR," an orthodontic plastic bracket (sequential aligner), by the FDA. The document focuses on demonstrating substantial equivalence to a predicate device, K180107 (Blue Sky Bio Aligner).

    The information provided is primarily related to bench testing of material properties, rather than performance of an AI/ML algorithm or a multi-reader, multi-case study. Therefore, many of the requested points regarding AI/ML performance, ground truth establishment for AI models, and human reader studies (MRMC) are not applicable or cannot be extracted from this document.

    Here's an analysis based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria are primarily derived from the ISO 20795-2:2013 standard, "Dentistry - Base Polymers - Part 2: Orthodontic base polymers." The device performance is compared to these criteria and the predicate device.

    Performance MetricAcceptance Criteria (ISO 20795-2:2013)Predicate Device (K180107) PerformanceTERA HARZ CLEAR (Subject Device) PerformanceDevice Meets Criteria?
    Ultimate Flexural Strength≥ 50 MPaAvg. 51.7 MPaTC-85DAC: Avg. 58.04 MPaTA-28: Avg. 70.26 MPaTR-07: Avg. 75.74 MPaYes
    Flexural Modulus≥ 1500 MPaAvg. 1560 MPaTC-85DAC: Avg. 1575 MPaTA-28: Avg. 1849 MPaTR-07: Avg. 1764 MPaYes
    Solubility≤ 5 µg/mm³Avg. 1.67 µg/mm³TC-85DAC: Avg. 1.56 µg/mm³TA-28: Avg. 1.30 µg/mm³TR-07: Avg. 1.39 µg/mm³Yes
    Water Sorption≤ 32 µg/mm³Avg. 17.51 µg/mm³TC-85DAC: Avg. 17.03 µg/mm³TA-28: Avg. 16.41 µg/mm³TR-07: Avg. 16.22 µg/mm³Yes
    PrecisionStandard deviation < 0.150 mmAvg. 0.122 mmTC-85DAC: Avg. 0.086 mmTA-28: Avg. 0.079 mmTR-07: Avg. 0.084 mmYes
    Freedom from porosityNo porosityNo porosityNo porosityYes
    Surface characteristicsSmooth, hard and glossySmooth, hard and glossySmooth, hard and glossyYes
    Shape capabilitySample edges reproducedSample edges reproducedSample edges reproducedYes
    ColourTransparent without changingTransparent without changingTransparent without changingYes
    AppearanceNo foreign materials, contaminations, or other hazardous defectsNo foreign materials, contaminations, or other hazardous defectsNo foreign materials, contaminations, or other hazardous defectsYes
    BiocompatibilityCompliant with ISO 10993-1510k clearance for aligner material, no biocompatibility testing performedBiocompatible according to ISO 10993-1Yes

    2. Sample size used for the test set and the data provenance

    The document does not specify exact sample sizes for each mechanical test. It mentions:

    • "All test results of Flexural Strength, Flexural Modulus, Water Solubility, and Water Sorption was similar to that of the predicate device that meet the requirements of ISO 20795-2:2013."
    • "The TERA HARZ CLEAR were outputted by each different output condition and each flexural strength were measured and the evaluation criteria of the all the specimens were more than 50 MPa."
    • "All translational measurements were within 0.150 mm of the target input value...There were no statistical differences in the difference in the intended and measured values observed from any of the groups."

    The testing was "bench testing" performed on the subject device and the predicate device. The document does not explicitly state the country of origin for the test data, nor whether it was retrospective or prospective. Given it's a 510(k) submission for a new device, the testing would generally be prospective for the subject device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This section is not applicable. The device is a physical orthodontic aligner, and the "ground truth" for the performance tests (e.g., flexural strength, solubility) is established by empirical measurements against established engineering and material science standards (ISO 20795-2:2013, ISO 10993). There were no human expert evaluations (e.g., radiologists interpreting images) involved in establishing the ground truth for these physical properties.

    4. Adjudication method for the test set

    Not applicable. This is not a study involving human interpretation or subjective assessment of a dataset that would require an adjudication method. The tests are objective measurements of physical properties.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document is for a physical medical device (orthodontic aligner), not an AI/ML powered device that would involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not a software algorithm; it is a physical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the performance characteristics of TERA HARZ CLEAR is based on international standards for dental materials (ISO 20795-2:2013) and biocompatibility standards (ISO 10993 series). This represents a type of objective, measurable ground truth based on established scientific and engineering principles. For manufacturing precision, the ground truth was the "target input value."

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/ML device that requires a training set or ground truth for training.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223355

    Validate with FDA (Live)

    Device Name
    Tera Harz Clear
    Manufacturer
    Date Cleared
    2022-11-03

    (1 days)

    Product Code
    Regulation Number
    872.5470
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TERA HARZ CLEAR is intended for the treatment of tooth malocclusions in patients with permanent dentition (i.e., all second molars). Utilizing a series of increments, it sequentially positions teeth by way of continuous gentle force. TERA HARZ CLEAR is intended exclusively for professional dental work. Fabrication of aligner with TERA HARZ CLEAR requires an additive manufacturing system (AMS) that includes compatible with the following:

    Design:
    Scanner: 3Shape A/S, TRIOS 3 Basic
    Design software: 3Shape A/S, 3Shape Ortho System™

    Printing:
    3D Printer: UNIZ, SLASH 2; SprintRay Inc., SprintRay Pro 95

    Post-Curing:
    Post-cure unit: CureM, U102H

    Device Description

    TERA HARZ CLEAR is a series of clear aligners that are used to replace traditional orthodontic wires and brackets for the alignment of maloccluded or misaligned teeth.

    This series of aligner is intended for the treatment of tooth malocclusions in patients with permanent dentition (i.e., all second molars) by moves the teeth gently, and in small increments, from their original to their final treated position for improved dental alignment. This series of aligner is worn for approximately 1 week of 20 to 22 hours per day after (However, there are to be removed for eating and for cleaning) which it is replaced by the next stage aligners and are designed to be used in a sequence. This is repeated for duration as prescribed by the dental clinician.

    TERA HARZ CLEAR is a light-cured, methacrylate-based resin commonly can used in additive manufacturing for the production of dental structures such as sequential aligners.

    TERA HARZ CLEAR has stored in a black 1,000 g of HDPE bottle. It contains materials with colors of yellowish. This device is a liquid photo-curable material that is polymerized by UV laser at 405412 nm. It can be used to make a tooth model with a photo-curable polymer that is cured by ultraviolet light. The liquid UV curing resin is cured at a specific wavelength (395405 nm) by the photo-initiator contained in the resin. It is typically 100 um in thickness and is output at a resolution of 40 to 90 µm on the x, y axis, and it is possible to produce three-dimensional printed matter by curing lamination step by step a thickness of 100 um.

    However, scanner, design software, 3D printer and post-cure unit are not included with the device.

    These fabrications of TERA HARZ CLEAR are beginning with the dental clinician prescribing aligners to treat a patient's malocclusion, and decision to use methacrylatebased resins is made by the dental clinician. TERA HARZ CLEAR, an orthodontic appliance such as sequential aligner, is manufactured in a 3D printer that is compatible.

    The dental clinician can generate a digital file by scanning the patient's mouth directly using listed with Intraoral scanner under FDA Classification Product Code NOF, regulation 872.3661.

    This digital file is a series of CAD files (.stl) for building models that can be used to fabricate aligners. Commonly used standard dental software is used by dental professionals to virtually design a sequential aligner and generate an industry-standard "STL" 3D dataset which reflects the intended shape and contour. The design software used is 3Shape Ortho System™ by 3Shape A/S (K180941). The specialized orthodontic treatment planning software has a 510k clearance for the intended use under FDA Classification Product Code PNN, regulation 872.5470. This software is used for management of 3D scanned orthodontic models, orthodontic diagnosis by measuring, analyzing, inspecting and visualizing 3D scanned orthodontic models, virtual planning of orthodontic treatments by simulating tooth movements, and design of sequential aligners based on 3D scanned orthodontic models.

    Once dental clinic manufacturing unit receive the data that *.stl CAD files of the treatment plan the 3D printer begins additive manufacturing. The dental clinician generates sequential 3D printed models replicating the approved treatment plan. The sequential aligner is 3D printed and cured in a post-cure unit. The fabricated aligners are cut to fit dentition, the cleaned and polished to remove rough edges by the dental clinician. The prescribing physician review and approves the sequential aligners are provides them to the patient the confirming fit and design.

    AI/ML Overview

    This document describes the premarket notification for the "TERA HARZ CLEAR" sequential aligner and its substantial equivalence determination by the FDA. The information provided focuses on the device's technical characteristics and performance testing against a predicate device and relevant standards.

    Here's a breakdown of the requested information based on the provided text:

    Acceptance Criteria and Device Performance Study

    The acceptance criteria are generally based on the ISO 20795-2:2013 standard for "Dentistry - Base Polymers - Part 2: Orthodontic base polymers," and comparisons to a predicate device (Blue Sky Bio Aligner, K180107).

    1. Table of Acceptance Criteria and Reported Device Performance

    Test ItemAcceptance CriteriaTERA HARZ CLEAR (UNIZ Slash 2) PerformanceTERA HARZ CLEAR (SprintRay PRO 95) PerformancePredicate Device (Blue Sky Bio Aligner) Performance
    Physical Properties
    Surface characteristicssmooth, hard, and glossysmooth, hard and glossysmooth, hard and glossysmooth, hard and glossy
    Shape capabilityedges are reproducededges are reproducededges are reproducededges are reproduced
    Colourtransparently without changingtransparently without changingtransparently without changingtransparently without changing
    Freedom from porosityno porosityno porosityno porosityno porosity
    Ultimate Flexural Strength≥5 MPaAvg. 5.92 MPaAvg. 6.02 MPaAvg. 5.17 MPa
    Flexural modulus≥50 MPaAvg. 69.10 MPaAvg. 72.38 MPaAvg. 56.26 MPa
    Solubility≤5 µg/mm³Avg. 1.54 µg/mm³Avg. 1.56 µg/mm³Avg. 1.67 µg/mm³
    Water sorption≤32 µg/mm³Avg. 17.45 µg/mm³Avg. 17.03 µg/mm³Avg. 17.51 µg/mm³
    Precisionstandard deviation less than 0.150 mm (implicitly)Uniz: Avg.0.072/mm (specific values provided)SprintRay: Avg.0.065/mm (specific values provided)Avg. 0.122/mm (specific values provided)
    AppearanceNo foreign materials, contaminations, or defectsNo evidence of foreign materials, contaminations,No evidence of foreign materials, contaminations,No evidence of foreign materials, contaminations,
    or any other defectsor any other defectsor any other defects
    Biocompatibility
    BiocompatibilityBiocompatible according to ISO 10993-1Biocompatible according to ISO 10993-1Biocompatible according to ISO 10993-1Predicate did not perform testing (510k clearance)

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: For physical properties testing (such as Ultimate Flexural Strength, Flexural Modulus, Solubility, Water Sorption, Surface characteristics, Shape capability, Colour, Freedom from porosity, and Precision), 5 samples of the "TERA HARZ CLEAR" device were used for each test. These samples were from the same lot and tested using two different 3D printers (UNIZ Slash 2 and SprintRay Pro 95).
    • Data Provenance: The document does not explicitly state the country of origin for the test data but indicates that the submitting company, Graphy Inc., is located in Seoul, Republic of Korea. The testing itself involved comparisons to a legally marketed predicate device (Blue Sky Bio Aligner) and adherence to international standards (ISO 20795-2:2013). The study appears to be prospective bench testing performed specifically for this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not mention the use of experts to establish a "ground truth" for the test set in the conventional sense of clinical evaluation. The ground truth for the physical and mechanical properties was established by adherence to the ISO 20795-2:2013 standard and direct comparison against a legally marketed predicate device's performance data.

    4. Adjudication Method for the Test Set

    No adjudication method is mentioned, as the study primarily involved bench testing of physical and mechanical properties against a standard and a predicate device.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was conducted or reported in this document. The device is a physical product (sequential aligner), and the evaluation focuses on its material properties and manufacturing validation rather than diagnostic performance or human-in-the-loop effectiveness.

    6. Standalone Performance (Algorithm Only)

    This section is not applicable as the "TERA HARZ CLEAR" is a physical medical device (sequential aligner), not a software algorithm or AI-powered diagnostic tool. The document describes the design software and 3D printing systems used in its fabrication, but the device itself is the physical aligner.

    7. Type of Ground Truth Used

    The ground truth for the device's performance was primarily based on:

    • International Standard Requirements: ISO 20795-2:2013.
    • Predicate Device Performance: Direct comparison of physical and mechanical properties with a legally marketed predicate device (Blue Sky Bio Aligner, K180107).
    • Manufacturing Validation: Verifying that manufactured devices meet predefined dimensional tolerances (0.150 mm of the target input value) using independent software and digital calipers.

    8. The Sample Size for the Training Set

    The document does not mention a "training set" in the context of an algorithm or AI. Since this is a physical medical device and not an AI/software product, the concept of a training set as typically understood in AI/ML studies is not applicable.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable for the reasons stated in point 8.