Found 2 results

510(k) Data Aggregation

    K Number
    K130636

    Validate with FDA (Live)

    Date Cleared
    2013-06-07

    (88 days)

    Product Code
    Regulation Number
    892.2030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TeleradPRO Edge and TeleradPRO Edge HD-CCD digitizers are used for making digital copies of medical x-ray film, including printed and radiographic film. The target users of these devices are medical professionals or trained staff for use as secondary capture images for consultation, review and final interpretation.

    These devices are not intended to be used for primary diagnosis in mammography applications.

    Device Description

    The device consists of a film digitizer with single-film or multi-film feeder, external power adapter, and Windows driver software. The film digitizer will convert the X-ray film into digital data representing the X-ray film, and the windows driver software is used to import the digital data into a compatible software application. The digitizer is used with DICOM standard software that meets or exceeds ACR and DICOM standards for use of secondary capture images for consultation, review and final interpretation.

    The digitizer is connected to a PC through a USB 2.0 interface. The digitizer utilizes rollers driven by a stepper motor to feed the X-ray film past the scan optics. The scan optics consists of a white or blue LED illuminator, a lens, mirrors, and a CCD linear array detector.

    This device has no patient contact and does not supply a diagnostic result. The film digitizer only provides digital data representing the film.

    AI/ML Overview

    This document describes a 510(k) submission for the TeleradPRO Edge and TeleradPRO Edge HD-CCD film digitizers. The submission asserts substantial equivalence to a predicate device, the VIDAR TeleradPRO film digitizer (K102476).

    1. Table of Acceptance Criteria and Reported Device Performance

    The device is a medical image digitizer, and its performance is primarily characterized by technical specifications rather than diagnostic performance metrics like sensitivity or specificity. The "acceptance criteria" can be inferred from the stated technological characteristics and the claim of substantial equivalence to the predicate device.

    Acceptance Criteria (Inferred from Technological Characteristics)Reported Device Performance (TeleradPRO Edge)Reported Device Performance (TeleradPRO Edge HD-CCD)
    High Spatial Resolution600 dpi600 dpi
    Grayscale Bit Depth16-bit grayscale16-bit grayscale
    Optical Density Sensitivity (DMAX)5.1 OD5.5 OD
    Medical OD Range (incorporates noise & linearity)0.2 - 3.20.1 - 3.6
    Compliance with DICOM standards for secondary capture imagesUsed with DICOM standard softwareUsed with DICOM standard software
    USB 2.0 InterfaceConnected to PC through a USB 2.0 interfaceConnected to PC through a USB 2.0 interface

    2. Sample size used for the test set and the data provenance

    The provided document does not specify a separate "test set" in the context of a clinical study with patient data. The performance testing mentioned refers to technical evaluations of the digitizer's hardware and software capabilities. No patient data or clinical images are described as being used for a "test set" in the sense of evaluating diagnostic accuracy. The device "has no patient contact and does not supply a diagnostic result."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. As described in point 2, there is no mention of a test set requiring expert-established ground truth for diagnostic accuracy, as the device doesn't provide a diagnostic result.

    4. Adjudication method for the test set

    Not applicable. There is no mention of a test set requiring adjudication for diagnostic accuracy.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an X-ray film digitizer, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The device is a film digitizer, not an algorithm providing diagnostic output. Its function is to convert physical film to digital data.

    7. The type of ground truth used

    For the technical performance characteristics (spatial resolution, OD range, etc.), the ground truth would be established through calibrated measurement instruments and industry standards for film digitizer performance. For instance, DMAX and OD range would be measured using calibrated optical density strips.

    8. The sample size for the training set

    Not applicable. This device is a film digitizer; it does not utilize a "training set" in the context of machine learning.

    9. How the ground truth for the training set was established

    Not applicable. This device is a film digitizer; it does not utilize a "training set" in the context of machine learning.

    Study Proving Device Meets Acceptance Criteria:

    The document states: "VIDAR conducts extensive performance testing and the test results demonstrate the device meets the requirements for its intended use."

    However, the specific details of this "extensive performance testing" are not provided in this 510(k) summary. The submission relies on claiming substantial equivalence to the predicate device (VIDAR TeleradPRO film digitizer, K102476) rather than providing a detailed report of a new clinical or technical performance study for the TeleradPRO Edge and HD-CCD. The conclusion states: "In terms of intended use, function, safety, operating environmental conditions and effectiveness of the TeleradPRO Edge and TeleradPRO Edge HD-CCD it is determined to be substantially equivalent to the predicate device used for this application." This implies that the 'study' demonstrating acceptance criteria is primarily a comparison against the predicate device's established performance and specifications, along with internal technical verification tests.

    Ask a Question

    Ask a specific question about this device

    K Number
    K102476

    Validate with FDA (Live)

    Device Name
    TELERADPRO
    Date Cleared
    2010-10-07

    (38 days)

    Product Code
    Regulation Number
    892.2030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VIDAR TeleradPRO film digitizer is used for making digital copies of medical x-ray films.
    The TeleradPRO film digitizer is intended for making digital copies of medical xray films.

    Device Description

    The device consists of a film digitizer with single-film or multi-film feeder, external power adapter, and Windows driver software. The film digitizer will convert the X-ray film into digital data representing the X-ray film, and the windows driver software is used to import the digital data into a compatible software application. The Digitizer is connected to a PC through a USB 2.0 interface. The digitizer utilizes rollers driven by a stepper motor to feed the X-ray film past the scan optics. The scan optics consists of a white LED illuminator, a lens, mirrors, and a CCD linear array detector.

    AI/ML Overview

    The provided text describes the TeleradPRO X-Ray Film Digitizer and its substantial equivalence to a predicate device (TeleRADPro Film Digitizers, K993597). However, it does not contain detailed information about specific acceptance criteria or an in-depth study that proves the device meets those criteria in a quantitative manner as typically expected for medical device performance studies.

    The document primarily focuses on establishing substantial equivalence to a predicate device, which often relies on demonstrating similar technological characteristics and intended use, rather than presenting a novel clinical study with detailed performance metrics against predefined acceptance criteria.

    Based on the provided text, here's the information that can be extracted, and where information is missing:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Optical Resolution300 dpi
    Grayscale Depth16-bit
    Optical Density Sensitivity (DMAX)4.8 OD
    Medical OD Range0.2 - 3.2 (incorporates noise and linearity measurements)
    Intended UseMaking digital copies of medical x-ray films (met by design)
    Function, Safety, Operating Environmental Conditions, EffectivenessDetermined to be substantially equivalent to predicate device

    Missing Information: Specific quantitative acceptance thresholds for DMAX, medical OD range linearity, and noise are not provided, nor are detailed test results demonstrating compliance against these thresholds. The document states "VIDAR conducts extensive performance testing and the test results demonstrate the device meets the requirements for its intended use" but does not elaborate on the specific tests or their outcomes beyond the summary characteristics.


    2. Sample Size Used for the Test Set and Data Provenance

    This information is not provided in the document. The document refers to "extensive performance testing" and "Bench Testing" (Section 22, which is not included in the provided text), but it does not specify any test set sample sizes or the provenance of any data (e.g., country of origin, retrospective/prospective). It suggests that the testing was primarily to demonstrate the device's technical specifications and substantial equivalence, not necessarily a clinical or reader study with a specific test set of images.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. As no clinical or reader study with a "test set" and "ground truth" established by experts is described, this information is not applicable. The device is a digitizer, not an AI diagnostic tool, so "ground truth" would relate to the fidelity of the digitization process itself, not an interpretation of medical images.


    4. Adjudication Method for the Test Set

    This information is not provided in the document. See point 3 above.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    No MRMC study was done, nor is it relevant given the device is an X-ray film digitizer, not an AI diagnostic tool designed to assist human readers. The document does not mention any AI components.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

    The document describes "Performance Testing" and "Bench Testing" which would likely be standalone evaluations of the digitizer's technical specifications (e.g., resolution, DMAX, linearity). However, details of these studies are not provided beyond the summary characteristics listed in Section 1.


    7. The Type of Ground Truth Used

    Given that the device is a digitizer, the "ground truth" for its performance would typically involve comparing the digitized image to the original film's characteristics or a known standard. The document mentions "incorporates noise and linearity measurements" for the medical OD range, implying that instrumental measurements were used to verify the fidelity of the digitization process. There is no mention of "expert consensus, pathology, or outcomes data" as ground truth, as those would be relevant for diagnostic interpretation, not for a digitizing device.


    8. The Sample Size for the Training Set

    This information is not provided and is not applicable. The device is a film digitizer, which typically does not involve machine learning "training sets." Its operation is based on optical and electronic principles, not learned algorithms.


    9. How the Ground Truth for the Training Set Was Established

    This information is not provided and is not applicable. See point 8 above.