Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K221844

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-08-17

    (54 days)

    Product Code
    Regulation Number
    888.3080
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TDM Lumbar Interbody Fusion Cage System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1, who have had six months of non-operative treatment. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). This device is to be used with autogenous bone graft. TDM Lumbar Interbody Fusion Cage System is to be used with supplemental fixation.

    Device Description

    The TDM Lumbar Interbody Fusion Cage System (EVERGREEN PEEK PLIF Cage, LIME PEEK TLIF Cage, and PORTIA HYBRID TLIF cage) consists of intervertebral body fixation devices intended for use as an aid in spinal fixation. This system is fabricated and manufactured from PEEK as described by ASTM F2026. The Tantalum and Titanium marker used for this product is made to the voluntary standard of ASTM F560 and ASTM F136. The devices are available in a variety of different sizes and heights to match more closely the patient's anatomy. The ends of the implants have machined teeth which are designed to engage with the vertebral body end plates. The EVERGREEN PEEK PLIF Cages are to be implanted via posterior approach. Two PLIF PEEK cages are placed on each side of the interbody space (right and left). The LIME PEEK TLIF Cages and PORTIA HYBRID TLIF cages are dedicated to transforaminal approach. TLIF technique involves placing only one bone graft spacer in the middle of the interbody space, without retraction of the spinal nerves.

    AI/ML Overview

    The provided document is a 510(k) summary for the TDM Lumbar Interbody Fusion Cage System, a medical device used in spinal fusion procedures. It outlines the device's intended use, technological characteristics, and a comparison to predicate devices, focusing on non-clinical performance testing.

    Crucially, this document does not contain information about a study proving that an AI/software device meets acceptance criteria for diagnostic performance metrics (e.g., sensitivity, specificity, AUC). The performance data section explicitly states: "Non-clinical testing was performed to demonstrate that the subject TDM Lumbar Interbody Fusion Cage System is substantially equivalent to the predicate device. The following testing was performed in accordance with the ASTM F2077, and ASTM F2267: - Static Axial Compression - Dynamic Axial Compression - Static Torsion - Static Shear - Dynamic Shear - Subsidence."

    These are mechanical and biocompatibility tests for a physical implantable device, not performance studies for an AI/software as a medical device (SaMD) that would have diagnostic accuracy metrics.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and the study proving the device meets those criteria for an AI/software device based on the provided text. The document is for a physical orthopedic implant.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1