Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K083218

    Validate with FDA (Live)

    Device Name
    TC ENDO PORT
    Date Cleared
    2009-03-06

    (126 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TC Endo Port has applications in abdominal, thoracic, laparoscopic and gynecologic minimally invasive procedures to establish a path of entry for endoscopic instruments.

    Device Description

    TC Endoport trocars are sterile single patient use instruments consisting of a sleeve and obturator in sizes ranging from 5-12 mm in diameter. The TC Endoport obturator is made from plastic, and has a dilator tip which gently moves aside any internal viscera that may be adjacent to the abdominal or thoracic wall. The trocar sleeve contains two seals, an outer annular seal that accommodates instruments from 5-12 mm in diameter, and an internal seal. Together these seals minimize gas or fluid leakage when instruments are inserted or withdrawn through the trocar. A stopcock valve, or valved luer are provided with standard luer lock fittings, to provide for gas insufflation, desufflation, or infusion and aspiration of fluids.

    AI/ML Overview

    The provided document describes the acceptance criteria and performance of the "TC Endo Port" device, a surgical trocar. Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test NameAcceptance CriteriaReported Device Performance
    Sterilization ExposureAll samples must be capable of withstanding a 2X sterilization cycle.Implied to meet the criteria, as no failures are reported and the conclusion states "No new questions of safety or effectiveness were raised during the performance testing of the device." The device is also mentioned to be "EtO gas-permeable sterile."
    Simulated ImplantAll samples must be capable of being implanted in simulated conditions using the instructions for use. Insertion forces must be equal to or less than predicate device.Implied to meet the criteria, as no failures are reported and the conclusion states "No new questions of safety or effectiveness were raised during the performance testing of the device."
    Valve Tool Pass VerificationMust pass the range of recommended tools through the valve without damaging the valve or causing the loss of more than 10ml of fluid.Implied to meet the criteria, as no failures are reported and the conclusion states "No new questions of safety or effectiveness were raised during the performance testing of the device."
    Valve Cyclic TestMust pass the maximum sized tool through the valve ten times without damaging the valve or causing the loss of more than 10ml of fluid.Implied to meet the criteria, as no failures are reported and the conclusion states "No new questions of safety or effectiveness were raised during the performance testing of the device."
    Valve Pressure WithstandMust withstand a fluid pressure of 450mmHg without leaking.Implied to meet the criteria, as no failures are reported and the conclusion states "No new questions of safety or effectiveness were raised during the performance testing of the device."
    Balloon Diameter VerificationProfile of balloons must be within 15%-25% of expected values from design.Implied to meet the criteria, as no failures are reported and the conclusion states "No new questions of safety or effectiveness were raised during the performance testing of the device."
    Balloon Volume LossBalloons cannot lose more than 0.5ml over one hour.Implied to meet the criteria, as no failures are reported and the conclusion states "No new questions of safety or effectiveness were raised during the performance testing of the device."
    Balloon Inflation/Deflation TimeBalloons must inflate and deflate in under 15 seconds.Implied to meet the criteria, as no failures are reported and the conclusion states "No new questions of safety or effectiveness were raised during the performance testing of the device."
    Balloon Cyclic TestBalloons and inflation lumens must withstand 40 inflation/deflation cycles to 1 atm without failure.Implied to meet the criteria, as no failures are reported and the conclusion states "No new questions of safety or effectiveness were raised during the performance testing of the device."
    Balloon Minimum Burst StrengthBalloon should withstand 2X SF over 1 atm.Implied to meet the criteria, as no failures are reported and the conclusion states "No new questions of safety or effectiveness were raised during the performance testing of the device."
    Cannula to Hub StrengthMaterial interface should resist a tensile force of up to 3.4 lb (15 N).Implied to meet the criteria, as no failures are reported and the conclusion states "No new questions of safety or effectiveness were raised during the performance testing of the device."
    Inflation Lumen to Hub StrengthMaterial interface should resist a tensile force of up to 3.4 lb (15 N).Implied to meet the criteria, as no failures are reported and the conclusion states "No new questions of safety or effectiveness were raised during the performance testing of the device."
    Valve Body to Cannula StrengthMaterial interface should resist a tensile force of up to 3.4 lb (15 N).Implied to meet the criteria, as no failures are reported and the conclusion states "No new questions of safety or effectiveness were raised during the performance testing of the device."
    Dilator Insertion/Removal ForceMaterial interface should resist a tensile force of up to 3.4 lb (15 N).Implied to meet the criteria, as no failures are reported and the conclusion states "No new questions of safety or effectiveness were raised during the performance testing of the device."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size:
      • For "Sterilization Exposure" test: "All" samples.
      • For "Simulated Implant" test: 30 new devices + 1 Predicate device.
      • For all other listed tests: 30 samples each.
    • Data Provenance: The document does not specify the origin (e.g., country) of the data, nor does it explicitly state if the study was retrospective or prospective. Given that these are non-clinical, in-vitro performance tests for a new device, they are inherently prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The tests described are non-clinical performance evaluations of the device's physical properties and functionality, not diagnostic or clinical assessments requiring expert human interpretation for ground truth.

    4. Adjudication Method for the Test Set

    This information is not applicable/not provided. Adjudication methods (like 2+1, 3+1) are typically used in studies involving human interpretation of data where consensus is needed to establish ground truth. The tests performed for this device are objective measurements of physical and functional properties.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a Multi Reader Multi Case comparative effectiveness study was not done. The document explicitly states: "No clinical tests were conducted as part of this products development." This device is a surgical instrument, not an AI-assisted diagnostic tool.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. The device is an endoscope and accessories, a physical medical device, not a software algorithm.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The concept of "ground truth" as typically applied to diagnostic algorithms (e.g., pathology, expert consensus) does not directly apply here. For these non-clinical performance tests, "ground truth" is defined by the objective physical and functional parameters measured against pre-defined engineering specifications and direct comparison to a predicate device. For example, "withstanding a 2X sterilization cycle" or "not losing more than 10ml of fluid" are the objective criteria that act as the standard.

    8. The Sample Size for the Training Set

    This information is not provided and is not applicable. This is a physical medical device being evaluated through non-clinical performance testing against a predicate, not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable, as there is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1