Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K110101

    Validate with FDA (Live)

    Date Cleared
    2011-04-11

    (88 days)

    Product Code
    Regulation Number
    888.3353
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Total hip components are indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative ioint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; nonunion; femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity. The subject devices are intended for single use only.

    Device Description

    Subject of this Traditional 510(k) Premarket Notification are the Smith & Nephew, Inc. 10/12 Taper Oxinium Femoral Heads. The subject devices are geometrically similar to the femoral head component in the Echelon Titanium Hip System cleared via submission K072817 and will include the addition of 32 mm and 36 mm diameter head sizes. This premarket notification seeks to add the 10/12 taper femoral head in oxidized zirconium material. Oxidized zirconium is the material of manufacture for 12/14 taper Femoral Heads cleared via submission K021673. The subject device will be offered in a size of 22 mm with offsets of -3, +0, and +3 mm; and is sizes ranging from 26 mm - 36mm, all with offsets of -5, +0, and +5 mm.

    AI/ML Overview

    The provided document describes a 510(k) premarket notification for the "10/12 Taper Oxinium Femoral Heads" and its substantial equivalence to previously cleared predicate devices. It focuses on the physical and mechanical performance of the device rather than AI/software performance. Therefore, many of the requested fields related to AI model evaluation are not applicable or cannot be extracted from this document.

    Here's the information that can be extracted from the provided text, along with an explanation for the inapplicable fields:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Performance Aspects Tested)Reported Device Performance
    Pre-Fatigue and Post-Fatigue assembly and disassemblyPerforms equal to or better than previously cleared devices (K021673, K072817)
    Environmental Fatigue and CorrosionPerforms equal to or better than previously cleared devices (K021673, K072817)
    Axial FatiguePerforms equal to or better than previously cleared devices (K021673, K072817)
    Range of MotionPerforms equal to or better than previously cleared devices (K021673, K072817)
    Device RigidityPerforms equal to or better than previously cleared devices (K021673, K072817)
    Wear PerformancePerforms equal to or better than previously cleared devices (K021673, K072817)

    2. Sample Size Used for the Test Set and Data Provenance

    This document describes physical and mechanical testing of a medical device (femoral head), not a software/AI device evaluated with a test set of data. Therefore, this information is not applicable. The "test set" here refers to physical components subjected to mechanical tests, not a data set.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable. Ground truth, in the context of expert review, typically relates to software predictions or image interpretations. For a physical device, performance is evaluated through engineering tests against established standards or predicate device performance, not expert consensus on data.

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication methods (like 2+1, 3+1) are used for resolving disagreements among human experts in evaluating data, which is not relevant to the mechanical testing of a femoral head.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. This study concerns the physical performance of a hip implant component, not an AI software or assistance system for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. No algorithm or AI is described or evaluated in this document.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For the performance data, the "ground truth" or reference for comparison is the performance of previously cleared predicate devices (Total Hip Femoral head - 12/14 Taper, K021673; and Echelon Titanium Hip System, K072817) in the specified mechanical and environmental tests. The underlying criteria for these tests would be derived from engineering standards and regulatory requirements for such devices.

    8. The Sample Size for the Training Set

    This information is not applicable. This document describes the testing of a physical medical device, not the training of an AI model.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable. There is no training set mentioned or implied as this is not an AI/software device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1