Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K090479

    Validate with FDA (Live)

    Date Cleared
    2009-07-07

    (133 days)

    Product Code
    Regulation Number
    870.2770
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TANITA Body Composition Analyzer is indicated for use in the measurement of weight and impedance, and the estimation of body mass index (BMI), total body fat percent, total body water percent and weight, muscle mass (skeletal and smooth), physique rating, bone mass, visceral fat rating with healthy range, basal metabolic rate (BMR), daily caloric intake (DCI), metabolic age, and target body fat percent with predicted weight and fat mass, using BIA (Bioelectrical Impedance Analysis).

    The device is indicated for use for healthy children 5-17 years old and healthy adults with active, moderately active, to inactive lifestyles.

    Device Description

    The TANITA Body Composition Analyzer SC-331 is a composition analyzer that utilizes BIA (bioelectrical impedance analysis) to determine body mass index (BMI), total body fat percent, total body water percent and weight, muscle (skeletal and smooth), physique rating, bone mass, visceral fat rating with healthy range, basal metabolic rate (BMR), daily caloric intake (DCI), metabolic age, and target body fat percent with predicted weight and fat mass for use by adults and children.

    AI/ML Overview

    This document focuses on the Tanita Body Composition Analyzer Model SC-331, a device that uses Bioelectrical Impedance Analysis (BIA) to measure and estimate various body composition parameters. The provided text is a 510(k) summary, which is a premarket notification to the FDA to demonstrate that a device is substantially equivalent to a legally marketed predicate device.

    Here's an analysis of the acceptance criteria and study information, based on the provided text:


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Tanita Body Composition Analyzer SC-331 are implicitly established by demonstrating substantial equivalence to its predicate devices (TANTA Body Fat Analyzer Professional and Consumer Models K014009 and K040778, and the TANITA Body Composition Analyzer: Model BC-53X).

    The table directly from the 510(k) summary compares the specifications and indicated outputs of the SC-331 with its predicate devices. The "reported device performance" is the statement that the SC-331 "performs equivalently" to these predicate devices based on "the results of [its] whole body BIA."

    Feature/ParameterPredicate Device (K014009) Performance/SpecificationPredicate Device (K040778) Performance/SpecificationPredicate Device (BC-53X) Performance/SpecificationTANITA Body Composition Analyzer SC-331 Performance/Specification (Reported)Acceptance Criteria (Implied by Substantial Equivalence and Comparison)
    Analytical Method / MeasurementPatented "Foot-to-Foot" BIA, In house BIA and DEXA referencePatented "Foot-to-Foot" BIA, In house BIA and DEXA referencePatented "Foot-to-Foot" BIA, In house BIA and DEXA referencePatented "Foot-to-Foot" BIA, In house BIA and DEXA referenceMust use "Foot-to-Foot" BIA, utilize in-house BIA and DEXA references.
    Measurement Frequency50kHz50kHz50kHz50kHzMust operate at 50kHz.
    Number of Electrodes4444Must use 4 electrodes.
    Weight CapacityNew / 270 kg / 600 lb270 kg / 600 lb150 kg / 330 lb600 lb / 270 kg or 450 kg / 1,000 lbMust meet or exceed predicate weight capacities. (SC-331 exceeds some.)
    Weight IncrementsNew / 100 g / 0.2 lb100 g / 0.2 lb100 g / 0.2 lb0.2 lb / 100 g or 0.1 lb / 50 gMust offer increments comparable to or better than predicate devices.
    Body Fat % Increments0.1%0.1%0.1%0.1%Must have 0.1% body fat increments.
    Input Age5 - 99 (5-17: Child, 18-99: Adult)7 - 99 (7-17: Child, 18-99: Adult)7 - 99 (7-17: Child, 18-99: Adult)5 - 99 (5-17: Child, 18-99: Adult)Must support similar or wider age ranges. (SC-331 offers wider child range).
    Input Height3' - 7' 11.5" / 90 - 249.9cm3' - 7' 11.5" / 90 - 249.9cm3' 4.0" - 7' 3.0" / 100-220cm3' - 7' 11.5" / 90 - 249.9cmMust cover similar or wider height ranges. (SC-331 has wider range).
    Input Body TypeStandard / AthleteStandard / AthleteStandard / AthleteStandard / AthleteMust accommodate "Standard" and "Athlete" body types.
    Indicated Measurements/EstimatesSimilar to predicates (Weight, Impedance, FAT %, FAT M, FFM, Muscle M, Physique Rating, TBW, TBW %, BMR/DCI, Metabolic Age, Visceral Fat Rating, Bone Mass, BMI)Yes (matching detailed table rows)Yes (matching detailed table rows)Yes (matching detailed table rows)Must have the same or a superset of the measurements/estimates as the predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not explicitly state the sample size used for the test set. It mentions "the results of [TANITA's] whole body BIA" as the basis for performance claims.
    The data provenance is not specified. It does not mention the country of origin of the data, nor whether it was retrospective or prospective. The term "in house BIA and DEXA reference" suggests internal validation data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The statement "in house BIA and DEXA reference" for analytical method/measurement implies that DEXA (Dual-energy X-ray absorptiometry) was used as a reference standard, which is considered a gold standard for body composition. However, it does not specify how many experts, if any, were involved in generating or interpreting this ground truth data, nor their qualifications.

    4. Adjudication Method for the Test Set

    The document does not describe any adjudication method for the test set.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study involving human readers and AI assistance was not done nor is it applicable to this device. This is a body composition analyzer, not an imaging analysis AI intended to assist human readers. The assessment is purely on the device's accuracy in measuring body composition parameters.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance assessment was implicitly performed. The device itself, the TANITA Body Composition Analyzer SC-331, is an algorithm-only device in the sense that it automatically processes BIA measurements to output body composition parameters. The study involves comparing its outputs ("whole body BIA") to "in house BIA and DEXA reference" measurements, meaning its performance is evaluated directly against a reference standard without human interpretation as part of the primary measurement.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The type of ground truth used is DEXA (Dual-energy X-ray absorptiometry) reference. The document states "in house BIA and DEXA reference" as part of its analytical method. DEXA is widely considered a gold standard for body composition analysis, providing highly accurate measurements of bone mineral density, lean soft tissue mass, and fat mass.

    8. The Sample Size for the Training Set

    The document does not specify the sample size used for the training set. While it mentions "in house BIA and DEXA reference" for the analytical method, it doesn't differentiate between data used for model development/training and data used for testing/validation.

    9. How the Ground Truth for the Training Set was Established

    The document implies that the ground truth for establishing the algorithms (which would be analogous to a training set for model development) was established using "in house BIA and DEXA reference." This means that the device's impedance measurements were correlated with and validated against highly accurate DEXA measurements to develop and refine the equations used to estimate various body composition parameters. This process would typically involve a dataset where subjects undergo both BIA measurements with the device and DEXA scans, with the DEXA results serving as the ground truth.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1