510(k) Data Aggregation

    K Number
    K120891

    Validate with FDA (Live)

    Date Cleared
    2012-10-01

    (192 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The tangoRS™ is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar and sacral spine (T1-S2). When used as a posterior spine thoracic/lumbar system, the tangoRS is intended for the following indications: degenerative disc disease (as defined by back pain of discogenic with degeneration of the disc confirmed by patient history and radiographic studies), spinal stenosis, spondylolisthesis, spinal deformities (i.e., degenerative scoliosis, kyphosis, and/or lordosis), fracture, spinal tumor, pseudarthrosis and failed previous fusion.

    Device Description

    The tangoRS™ consists of rods, polyaxial pedicle screws and crosslinks with locking set screws. The components are available in various sizes to accommodate differing patient anatomy. Rods are available in one diameter and a variety of lengths. Polyaxial screws are available in a variety of diameter-length combinations. Crosslinks are offered in four lengths. The purpose of this submission is to add components (a screw size and crosslink options) and two indications for use.

    AI/ML Overview

    The provided 510(k) summary for the tangoRS™ multifunctional posterior system describes its performance through mechanical testing. Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criterion (Implicit)Reported Device Performance
    Biocompatibility (Material safety)Components manufactured from titanium alloy (Ti-6Al-4V ELI) as described by ASTM F136.
    Mechanical Strength and Durability (Static Compression Bending)tangoRS™ performs as well as or better than predicate devices according to ASTM F1717 testing.
    Mechanical Strength and Durability (Dynamic Compression Bending)tangoRS™ performs as well as or better than predicate devices according to ASTM F1717 testing.
    Mechanical Strength and Durability (Static Torsion)tangoRS™ performs as well as or better than predicate devices according to ASTM F1717 testing.
    Substantial Equivalence to Predicate Devices (Technological Characteristics & Intended Use)Same intended use, basic design (rod-based, screw anchors), material (titanium alloy), and comparable sizes to predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify a "sample size" in terms of individual patients or clinical cases for a test set. The performance evaluation is based on mechanical testing of "worst case tangoRS™ constructs." The exact number of constructs tested is not provided, but it would typically involve multiple samples for each test condition to ensure reproducibility.
    • Data Provenance: The data is generated from in-vitro (benchtop) mechanical testing, not from patient-derived data. Therefore, there's no country of origin for clinical data, and it is not retrospective or prospective in the clinical sense.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable to this type of submission. Mechanical testing does not involve human experts establishing "ground truth" on patient data. The "ground truth" for mechanical testing is established by the specified ASTM F1717 standard and the performance of predicate devices.

    4. Adjudication Method for the Test Set

    This is not applicable as the study is mechanical testing, not a clinical study involving human judgment that would require adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. This submission relies on mechanical bench testing, not clinical studies involving human readers.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    This is not applicable. The device is a physical medical implant (pedicle screw system), not an algorithm or software. Therefore, there's no concept of "standalone performance" in the context of an algorithm. The performance described is inherent to the device's physical and mechanical properties.

    7. Type of Ground Truth Used

    The "ground truth" used for this type of device and submission is based on:

    • ASTM F1717 standards: These provide the accepted methodology and criteria for testing spinal implant constructs.
    • Performance of predicate devices: The tangoRS™ system is compared against the mechanical performance of previously cleared predicate devices (flamenco™ {K102853}, CD HORIZON® {K031655/K041460}, Expedium {K041119/K062196}, tangoRS™ {K052385}). The implicit "ground truth" is that if the new device performs "as well as or better than" these established devices under standardized testing, it is considered safe and effective.

    8. Sample Size for the Training Set

    This is not applicable. There is no "training set" in the context of mechanical testing for a physical implant. The design of the tangoRS™ system is based on established engineering principles and knowledge gained from predicate devices, but there isn't a "training set" of data in the machine learning sense.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable as there is no "training set" for this type of device and study. The physical design and materials are chosen based on known biomechanical requirements and historical performance of similar devices.