Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K143326

    Validate with FDA (Live)

    Date Cleared
    2015-05-14

    (175 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The T.A.G. Dental Implant System is intended to replace single or multiple teeth in the fully or partially edentulous mandibular or maxillary alveolar process. The implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    Device Description

    The T.A.G.'s Dental Implant System is composed of three sub-families:
    (1) Massif - A self-tapping cylindrical screw type implant Available in lengths: 8 - 16 mm and diameters: 3.75 - 6.0 mm Note: length 6 mm not available for diameters below 4.2 mm.
    (2) Axis - A self-tapping conical implant. Available in lengths: 8 - 16 mm and diameters: 3.75 - 6.0 mm Note: length 8 mm not available for diameters below 3.75 mm, length 16 mm not available for diameters above 4.2
    (3) Crestone - A one piece implant Available in Lengths of 10 - 16 mm and Diameters 3.0 - 3.5 mm.
    The implants are provided sterilize for single patient use. Each implant is provided with cover screw inside the sterile package. All implants are made of titanium alloy grade 23 (Ti-6AI-4V-ELI).
    Provided are accessories which are used in dental implantation procedures. They are: Superstructures which are Healing Caps & Abutments made from Titanium Alloy TI 6AL 4V ELI, Stainless Steel, and/or PEEK. The superstructures are single patient use to be supplied non-sterile.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the T.A.G. Dental Implant System.

    It's important to note that the provided document is a 510(k) summary for a dental implant system, which is a medical device, not a software or AI-driven diagnostic device. Therefore, the concepts of "test set," "training set," "ground truth experts," "adjudication," and "MRMC studies" are typically not applicable in the context of hardware medical devices like dental implants. The "performance" here refers to mechanical and material performance, not diagnostic accuracy.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" and "reported device performance" in the typical sense of a diagnostic device with metrics like sensitivity, specificity, or AUC. Instead, it refers to compliance with established standards for mechanical performance.

    Acceptance Criteria (Implied)Reported Device Performance
    Compliance with FDA's Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental AbutmentsMet the requirements of the standards (ISO14801).
    Compliance with ISO 14801: Static and Cycling Loading Test of Dental ImplantsMet the requirements of the standards (ISO14801).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not specify the exact sample size for the mechanical testing. For ISO 14801, standard practice involves testing multiple samples per device configuration (e.g., different lengths, diameters) to establish statistical significance. The document states "testing, which was submitted."
    • Data Provenance: The testing was "performed at an accredited independent laboratory." The country of origin of the data is not explicitly stated, but the manufacturer (T.A.G. Medical Products) is located in Kibbutz Gaaton, Israel. The testing would, therefore, likely originate from a lab commissioned by the manufacturer, possibly in Israel or an international lab accredited to perform such tests. The study is prospective in the sense that the implants were manufactured and then tested.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This concept is not applicable to this type of device and study. "Ground truth" for mechanical testing is established by the physical testing results themselves and their compliance with engineering specifications, not by expert consensus or interpretation of images. An independent accredited laboratory performing the tests serves as the "expert" in terms of confirming the test methodology and results.

    4. Adjudication Method for the Test Set

    This concept is not applicable to this type of device and study. Adjudication typically refers to resolving discrepancies in human interpretation, which is not relevant for mechanical performance testing.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is specifically designed for diagnostic devices where human readers interpret medical images or data, and its goal is to assess the impact of a device (often AI-assisted) on human performance. Since the T.A.G. Dental Implant System is a physical implant, such a study is irrelevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This concept is not applicable. The device is a physical dental implant, not an algorithm. The performance testing described (mechanical loading) is inherently "standalone" in the sense that it assesses the device's physical properties directly, without human interpretation of results needing to be factored into the device's intrinsic function.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance testing of the dental implant system is the physical measurement and observation of the implant's behavior under controlled mechanical stress, as defined by the ISO 14801 standard. This standard specifies how dental implants should be subjected to static and dynamic (cycling) loading to evaluate their fatigue strength and resistance to fracture. The "truth" is whether the implant withstands the specified forces and cycles without failure, thus meeting the criteria of the standard.

    8. The Sample Size for the Training Set

    This concept is not applicable. Dental implants are hardware devices. There is no "training set" in the machine learning sense. The design and manufacturing process are informed by engineering principles, material science, and previous experience, but not through an iterative learning process with a "training set" of data like an AI algorithm.

    9. How the Ground Truth for the Training Set Was Established

    This concept is not applicable as there is no training set for a hardware medical device. The "ground truth" for the device's design and manufacturing is derived from established engineering principles, material specifications (e.g., Ti-6Al-4V-ELI), regulatory requirements (e.g., FDA guidance, ISO standards), and clinical experience with similar devices.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1