Found 1 results

510(k) Data Aggregation

    K Number
    K142096

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-10-17

    (77 days)

    Product Code
    Regulation Number
    868.5925
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The T-Piece Resuscitator is a gas powered emergency resuscitator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway. It is intended for use with patients weighing greater than 10kg (>22lb).

    Device Description

    The Mercury T-Piece Resuscitator is manually operated, gas powered resuscitator for use with patients greater than 10 kg (>22 lb).

    It is a simple T-piece, with a manometer and the ability to adjust Peak Inspiratory Pressure (PIP) and Positive End-Expiratory Pressure (PEEP). It incorporates a pressure relief valve for excessive pressure.

    The T-Piece Resuscitator can be connected to the patient via a face mask, Supraglottic airway or endotracheal tube.

    AI/ML Overview

    The provided document describes the Mercury T-Piece Resuscitator and its substantial equivalence to predicate devices, focusing on non-clinical testing for performance. This is a 510(k) premarket notification summary, which means the device is being compared to an already legally marketed device (predicate device) and does not involve AI or complex analytical studies in the way modern AI/ML medical devices would.

    Therefore, many of the requested categories in the prompt are not applicable to this document. I will focus on the information that can be extracted directly from the provided text.

    Here's the breakdown based on the provided document:

    Acceptance Criteria and Device Performance

    The acceptance criteria are generally aligned with the performance requirements outlined in ISO 10651-5 - Particular Requirements for Basic Safety and Essential Performance for Gas Powered Resuscitators. The "Reported Device Performance" column reflects the specifications of the proposed Mercury Medical T-Piece Resuscitator as presented in comparison tables with its predicate devices.

    Acceptance Criteria (Derived from Predicate & ISO 10651-5)Reported Device Performance (Mercury Medical T-Piece Resuscitator)
    Indications for Use: Provide emergency respiratory support by face mask or tube; patients > 10kg (>22lbs).The T-Piece Resuscitator is a gas-powered emergency resuscitator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway. It is intended for use with patients weighing greater than 10kg (>22lb).
    Environment of Use: Hospital, sub-acute facilities, and pre-hospital (EMS).Hospital, sub-acute facilities, and pre-hospital (EMS).
    Patient Population: Patients greater than 10 Kg (>22 lbs.).Patients greater than 10 Kg (>22 lbs.)
    Ventilation Frequency: Manually delivered by the user up to 60 BPM.Manually delivered by the user up to 60 BPM.
    Maximum Pressure Relief: Factory set at 40 cm H2O (predicate range 5-80 cm H2O).60 cm H2O (Factory set at 40 cm H2O).
    Delivered Pressure: Up to 60 cm H2O (predicate range 5-80 cm H2O).Up to 60 cm H2O.
    Delivered Volume: (Predicate range 190 - 675 ml with flow rates between 4 - 36 lpm).60 - 700 ml with flow rates between 5 - 35 lpm.
    Inspiratory Resistance: (Predicate: Less than - 5 cm H2O).2 cm H2O at minimum PEEP setting @ 60 lpm.
    Expiratory Resistance: (Predicate: Less than + 5 cm H2O).2.4 cm H2O at minimum PEEP setting @ 60 lpm.
    Oxygen Concentration with optional blender: (Predicate range: 21-100%).21 - 98% based upon blender setting.
    Manometer Range: (Predicate range: -20 to 80 cm H2O).Up to 60 cmH2O (Cleared under K954486).
    Manometer Accuracy: (Predicate: +/- 2 cm H2O).+/- 3 cm H2O up to 15 cm H2O; +/- 5 cm H2O > 15 cm H2O.
    Peak Inspiratory Pressure (PIP): (Predicate range: 5-80 cm H2O).0-60 cm H2O.
    Positive End-Expiratory Pressure (PEEP): (Predicate uses CPAP instead; reference NeoTee has PEEP).0 to 60 cm H2O.
    Operational Gas Flow Rate: (Predicate range: 0-60 lpm).0 - 35 lpm.
    Operational time with 400 L cylinder: (Predicate @ 36 lpm - 10 minutes).@ 35 lpm - 11 minutes.
    Dead space of circuit: (Predicate: < 15 ml).< 7.5 ml.
    Operating Temperature Range: (Predicate range: 10 to 40°C).-18 to 60°C.
    Storage Temperature: (Predicate range: -20 to 60°C).-40 to 50°C.
    Immersion Resistance: Temporary submersion of circuit does not affect functionality.Temporary submersion of circuit does not affect functionality.
    Vomitus Resistance: Meets ISO 10651-5.Tested per ISO 10651-5.
    Water Immersion: Meets ISO 10651-5.Tested per ISO 10651-5.
    Oxygen Concentration: Meets ISO 10651-5.Tested per ISO 10651-5.
    Pressure Limitation: Meets ISO 10651-5.Tested per ISO 10651-5.
    Storage/Operating Conditions: Meets ISO 10651-5.Tested per ISO 10651-5.
    Drop: Meets ISO 10651-5.Tested per ISO 10651-5.

    The study that proves the device meets the acceptance criteria is a series of "Bench testing" performed according to ISO 10651-5 for gas powered resuscitators. The document explicitly states: "We have demonstrated that the T-Piece Resuscitator meets the performance and design specifications requirements as outlined in requirements and specifications as outlined in ISO 10651-5 - Particular Requirements for Basic Safety and Essential Performance for Gas Powered Resuscitators."


    Additional Information on the Study:

    1. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

      • The document describes "bench testing" which implies tests on physical devices rather than patient data. Therefore, details like "country of origin of the data" or "retrospective/prospective" are not applicable. The sample size for the bench tests is not specified in the provided summary.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • This is not applicable as the study involves bench testing of a physical medical device, not a diagnostic or AI algorithm relying on expert interpretation of data. The ground truth is established by the performance standards outlined in ISO 10651-5.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • This is not applicable to a physical device performance test assessed against engineering standards.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • This is not applicable. This is a physical medical device (resuscitator), not an AI/ML diagnostic tool.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • This is not applicable. This is a physical medical device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for the device's performance is based on engineering and safety standards, specifically ISO 10651-5 - Particular Requirements for Basic Safety and Essential Performance for Gas Powered Resuscitators.
    7. The sample size for the training set:

      • This is not applicable. The device is a physical resuscitator, not an AI/ML algorithm that requires a "training set."
    8. How the ground truth for the training set was established:

      • This is not applicable, as there is no training set for a physical device.