510(k) Data Aggregation

    K Number
    K242834

    Validate with FDA (Live)

    Date Cleared
    2025-01-16

    (119 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stryker System 9 Sterile Battery Container is a reusable rigid sterilization container intended to allow for sterilization, by hospitals, surgery centers, and healthcare facilities, of the enclosed, compatible Stryker Sterile Battery Packs and maintain sterility of the enclosed device until used. The container is validated for use with compatible battery packs in low temperature hydrogen peroxide sterilization cycles.

    Device Description

    The Stryker System 9 Sterile Battery Container is a rigid, reusable device intended to enclose compatible Stryker sterile battery packs for hydrogen peroxide sterilization. The container consists of a stainless-steel perforated lid with a stainless-steel filter retention plate and a plastic (Radel®) base. The Sterile Battery container utilizes single use polypropylene filters and indicator cards. The container base is designed to exclusively fit onto the Stryker sterile battery charger to allow the sterile battery pack to charge while maintaining a sterile barrier. The Stryker System 9 Sterile Battery Container encompasses one distinct configuration in two sizes, small and large.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device (Stryker System 9 Sterile Battery Container). This type of submission relies on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving absolute safety and effectiveness through extensive clinical trials for novel devices.

    Therefore, the study described is a non-clinical performance testing suite designed to show that the device performs as intended and is as safe and effective as its predicate, particularly in the context of sterilization efficacy and maintenance of sterility. It is not a clinical study involving human patients or complex AI algorithms requiring MRMC studies, expert consensus on images, or detailed ground truth establishment in a diagnostic context.

    Here's an analysis based on the provided document, fulfilling the requested information where applicable. Many questions regarding AI/diagnostic device studies (like sample size for test sets, number of experts, adjudication methods, MRMC studies, training set details) are not applicable to this type of device and submission, as it's a physical sterilization container, not a diagnostic imaging AI.


    1. A table of acceptance criteria and the reported device performance

    TestingStandard in compliance withAcceptance CriteriaReported Device Performance
    Sterilization EfficacyISO 14937:2009, ISO 22441:2022Confirms sterile efficacyPass
    Toxicological Risk AssessmentISO 10993-17Confirms acceptable residualsPass
    Cleaning ValidationAAMI ST98:2022Cleaning instructions were validated and demonstrated that the subject devices could be visually and quantifiably cleanedPass
    Maintenance of SterilityAAMI ST77:2013/(R)2018Demonstrates that the subject device is capable of maintaining product sterility for the duration of the shelf lifePass
    Simulated Use Design ValidationISO 17664-1:2021, AAMI TIR 12:2020, ST77:2013/(R)2018Confirms that the device can undergo repeated processing to support that the device's performance is maintained following a minimum number of exposures to the proposed sterilization methodsPass
    Human Factors EngineeringIEC 62366-1 Ed 1.1 b 2020Confirms safety and effectiveness of the subject device for the intended users, uses, and use environmentsPass

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes (e.g., number of containers tested, number of sterilization cycles performed) for each non-clinical test. It generally states that "A suite of non-clinical testing was performed."

    • Data Provenance: The data comes from the manufacturer's (Stryker Instruments) internal testing as part of their premarket notification. The country of origin of the data is implied to be the USA, given the manufacturer's address in Michigan. These were prospective tests specifically conducted for this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable to the non-clinical performance testing of a sterilization container. "Ground truth" in the context of diagnostic devices (e.g., AI for medical imaging) involves expert interpretation or pathological confirmation. For this device, the "truth" is established by adherence to validated sterilization standards and measurable outcomes (e.g., sterility, absence of toxic residuals, successful cleaning). The expertise would lie with the engineers and microbiologists performing and validating these standard-driven tests, not a panel of clinical experts interpreting data for diagnostic purposes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical studies, particularly for establishing ground truth from multiple human readers for diagnostic AI, to resolve disagreements. Here, validation is against objective, measurable criteria defined by established standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. MRMC studies are specific to evaluating diagnostic AI systems, comparing human performance with and without AI assistance. This device is a physical sterile battery container, not an AI or diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This concept applies to AI algorithms. The "performance" of this device is its physical and chemical properties and its ability to maintain sterility, which are evaluated through direct testing, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance is adherence to established international and national standards for sterilization, cleaning, biocompatibility, and human factors. This includes:

    • Microbiological testing for sterility (e.g., confirming no microbial growth after sterilization cycles).
    • Chemical analysis for residuals.
    • Physical measurements and observations for cleaning effectiveness and structural integrity after simulated use.
    • Usability testing according to human factors standards.

    This is not a diagnostic device where pathology or expert consensus on images would define the ground truth.

    8. The sample size for the training set

    Not applicable. This device is a physical product, not a machine learning model, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.