Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K252057

    Validate with FDA (Live)

    Date Cleared
    2025-08-28

    (58 days)

    Product Code
    Regulation Number
    870.5150
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Symphony Thrombectomy System is intended for:

    • The non-surgical removal of fresh, soft emboli and thrombi from blood vessels.
    • Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.

    The Symphony Thrombectomy System is intended for use in the peripheral vasculature and for the treatment of pulmonary embolism.

    The Symphony 16F 82cm Thrombectomy System is intended for:

    • The non-surgical removal of fresh, soft emboli and thrombi from blood vessels.
    • Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.

    The Symphony 16F 82cm Thrombectomy System is intended for use in the peripheral vasculature. It is not for use in the pulmonary vasculature.

    Device Description

    The Symphony™ Thrombectomy System is comprised of the following devices:

    • 24F Symphony Catheter
    • 24F Symphony Standard Dilator
    • 24F Symphony Advance™ Long Dilator
    • 24F Symphony ProHelix™
    • 16F Symphony Catheter
    • 16F Symphony Dilator
    • 16F Symphony ProHelix™
    • Symphony Clot Container
    • TRUVIC Generator
    • TRUVIC Canister
    • TRUVIC Tubeset

    The Symphony™ 16F 82cm Thrombectomy System is comprised of the following devices:

    • 16F 82cm Symphony Catheter
    • 16F 82cm Symphony Length Matched Dilator
    • Symphony Clot Container
    • TRUVIC Generator
    • TRUVIC Canister
    • TRUVIC Tubeset

    Both Systems are designed to remove thrombus/embolus (also referred to as 'clot') from the peripheral vasculature using controlled aspiration. The Symphony Catheters and the Symphony 16F 82cm Catheter target aspiration from the TRUVIC Generator directly to the thrombus. The Symphony ProHelix may be used to facilitate aspiration and removal of the thrombus through the 16F Symphony Catheter or through the 24F Symphony Catheter. The Symphony ProHelix is not used with the 16F 82cm Symphony Catheter.

    The Symphony Catheters and Symphony Dilators are introduced through a vascular access sheath into the peripheral vasculature and guided over a guidewire to the site of the thrombus in the peripheral vasculature or pulmonary anatomy. The 16F 82cm Catheter and compatible Dilator are introduced through a vascular access sheath into the peripheral vasculature and guided over a guidewire to the site of thrombus in the peripheral vasculature. The Symphony 16F 82cm Catheter and compatible Dilator are not intended for use in the pulmonary anatomy. Both the Symphony Catheters and the Symphony 16F 82cm Catheter are used with the TRUVIC Generator, connected using the TRUVIC Tubeset and the TRUVIC Canister, to aspirate thrombus.

    As needed, the Symphony ProHelix may be introduced through a Symphony 16F Catheter or a Symphony 24F Catheter to assist with thrombus removal. The Symphony ProHelix is not used with the 16F 82cm Symphony Catheter. The Symphony ProHelix is manually advanced through the Symphony Catheter over a guidewire, remaining inside the Symphony Catheter during the procedure. During aspiration, the handle on the proximal end of the Symphony ProHelix is manually rotated, which rotates the tip of the Symphony ProHelix to facilitate thrombus removal through the Symphony Catheter. The tips of the devices are visible under fluoroscopy.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the Imperative Care, Inc. Symphony™ Thrombectomy System, specifically for the expanded pulmonary embolism indication, based on the provided 510(k) clearance letter:


    1. Table of Acceptance Criteria and Reported Device Performance

    MetricAcceptance Criteria (Performance Goal)Reported Device Performance
    Primary Efficacy Endpoint: Mean Reduction in RV/LV Between Baseline and 48-Hours (assessed by CT Angiography)Lower one-sided 97.5% confidence interval (CI) bound > 0.200.44 ± 0.42 (Mean ± SD) Lower one-sided 97.5% CI for Mean: 0.36 Met: Yes (0.36 > 0.20)
    Primary Safety Endpoint: Composite 48-Hour Major Adverse Events (MAE) Rate (all-cause major bleeding, device-related mortality, and device-related serious adverse events including clinical deterioration, pulmonary vascular injury, or cardiac injury)Upper one-sided 97.5% CI bound < 15.0%0.9% (1/109) Upper one-sided 97.5% CI: 5.7% Met: Yes (5.7% < 15.0%)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Clinical Study (SYMPHONY-PE):

      • ITT Cohort (Safety Analysis): 109 subjects
      • mITT Cohort (Efficacy Analysis - device only used, no thrombolytics/other devices within 48h): 106 subjects
      • Primary Efficacy Analysis Population (mITT with paired RV/LV measurements): 102 subjects
    • Data Provenance: The SYMPHONY-PE clinical study was a prospective, multicenter, single-arm, open label study conducted across 19 sites in the U.S.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Ground Truth for Clinical Endpoints:
      • Primary Efficacy Endpoints (RV/LV measurements): Adjudicated by an independent imaging core lab. The specific number and qualifications of experts within this core lab are not detailed in the provided document.
      • Primary Safety Endpoints (Major Adverse Events): Adjudicated by an Independent Safety Board (ISB). The specific number and qualifications of experts on this board are not detailed in the provided document.
      • Clinical Events Committee (CEC): Also reviewed safety endpoints data. The number and qualifications are not detailed.

    4. Adjudication Method for the Test Set

    • The document states that the primary efficacy endpoint (mean reduction in RV/LV) was "independent core-lab adjudicated."
    • The primary safety endpoint (MAE rate) was "adjudicated by the Independent Safety Board (ISB)."
    • No specific method like "2+1" or "3+1" concensus is described for the adjudication processes. It implies that the core lab and ISB performed their respective adjudications, likely following their own established protocols.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No, an MRMC comparative effectiveness study was not done. The SYMPHONY-PE study was a single-arm study evaluating the safety and effectiveness of the device itself against pre-specified performance goals, not comparing its performance or human reader performance with and without AI assistance, or against another device.
    • Therefore, no effect size for human reader improvement with AI vs. without AI assistance is reported, as AI is not explicitly mentioned as part of the system or as an aid to human readers in this context.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • No, a standalone (algorithm-only) performance study was not done. The Symphony™ Thrombectomy System is a medical device for mechanical thrombectomy, not an AI or software algorithm that provides a diagnostic output in a standalone manner. The clinical study evaluated the device in vivo during actual procedures.

    7. The Type of Ground Truth Used

    • Clinical Outcomes Data: For the SYMPHONY-PE clinical study, the ground truth was based on clinical outcomes measured in human subjects.
      • Efficacy: Reduction in RV/LV ratio, assessed by CT Angiography (CTA) and adjudicated by an independent imaging core lab. This is a quantitative, image-based clinical measurement.
      • Safety: Rates of Major Adverse Events (MAE) including bleeding, mortality, and device-related SAEs, adjudicated by an Independent Safety Board and CEC. This is based on clinical observations and event reporting.

    8. The Sample Size for the Training Set

    • Not applicable / Not provided. The Symphony™ Thrombectomy System is a mechanical device, not a machine learning or AI model. Therefore, there is no "training set" in the context of algorithm development.
    • The document does describe extensive in-vitro bench testing and in-vivo GLP pre-clinical animal testing to validate the device's performance prior to human clinical trials.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable / Not provided. As mentioned above, there is no "training set" for this mechanical device. The ground truth for its development and verification was established through various engineering tests and animal studies, which are designed to validate physical performance, material properties, and biocompatibility against established standards and safety profiles.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1