Search Results
Found 3 results
510(k) Data Aggregation
(112 days)
Low Frequency Electric Stimulator: This Device uses microcurrent electrical to stimulate facial tissues for aesthetic purposes.
The Sunny Plus device is a Low Frequency Electric Stimulator that conveys current to the human body through a non-invasive electrode with a frequency of 10Hz. It is composed of the main body, handpiece with electrode tip, and foot switch. The device is operated through a graphic user interface on an LCD screen and is intended to be operated by medical professionals.
N/A
Ask a specific question about this device
(231 days)
The Sunny RF Handpiece (6.78MHz) is indicated for Dermatologic procedures for electrocoagulation and hemostasis.
The Pulsar handpiece (1MHz and 2MHz RF) is indicated to provide topical heating for the purpose of elevating tissue temperature for the treatment of selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.
The Sunny device is a high frequency electrosurgical unit that conveys current to the human body through non-invasive electrodes. The device includes three different high frequencies: 6.78MHz (RF), 1MHz (RF), and 2 MHz (RF). It is composed of the main body, Sunny handpiece, Pulsar handpiece, treatment tips, neutral electrode pad, refrigerant gas, and foot switch. The device is operated through a graphic user interface on an LCD screen and is intended to be operated by medical professionals.
Based on the provided text, the device in question is an electrosurgical unit (Sunny™) and not an AI-powered diagnostic device. Therefore, a significant portion of the requested information, such as multi-reader multi-case (MRMC) comparative effectiveness studies, ground truth establishment by experts, and sample sizes for training/test sets related to AI model validation, are not applicable.
The acceptance criteria and study that proves the device meets them are focused on hardware performance, safety, and substantial equivalence to a predicate device.
Here's an organized description of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Device: Sunny™ (Electrosurgical Cutting and Coagulation Device and Accessories)
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criterion (Implicit/Explicit from Text) | Reported Device Performance/Testing Result |
|---|---|
| Device Performance (Sunny RF Handpiece - 6.78MHz) | |
| Effectiveness for electrocoagulation and hemostasis (based on thermal effect) | Histopathology testing conducted to demonstrate the thermal effect of the device, showing it functions comparatively to the predicate (Thermage CPT) despite differences in max power and electrode size. |
| Energy Output Range (25 to 135 Joules) | Achieved. "Similar" to predicate's 20 to 137 Joules, with the small differences not impacting safety or effectiveness. |
| Device Performance (Pulsar Handpiece - 1MHz and 2MHz RF) | |
| Ability to provide topical heating for temporary relief of pain, muscle spasms, and increase in local circulation (based on temperature maintenance) | A temperature maintenance test was performed and confirmed that the device can maintain skin temperature of 40°C-45°C for 10 minutes on three different anatomical sites. The performance goal for the reference device (Nuera Tight RF) was 40-42°C +/- 2°C for 15 minutes, showing the Sunny device meets similar therapeutic temperature ranges, albeit for a slightly shorter duration. |
| Energy Output levels for 1MHz and 2MHz | 10 levels, specifically: Level 1 (1MHz: 15.2W, 2MHz: 7.6W), Level 10 (1MHz: 40.5W, 2MHz: 20.2W). |
| Power Density (60 J/cm² for Pulsar) | "Similar" to predicate's 45.6 J/cm². |
| Safety & Compliance | |
| Electrical Safety | Compliance with IEC 60601-1:2005 + A2:2020. |
| Electromagnetic Compatibility (EMC) | Compliance with IEC 60601-1-2:2014 + A1:2020 and IEC TR 60601-4-2 Edition 1.0 2016-05. |
| High-Frequency Surgical Equipment Specific Requirements | Compliance with IEC 60601-2-2:2017 + A1:2023. |
| Software Life Cycle Processes | Compliance with IEC 62304:2006+A1:2015. Software verification and validation testing conducted per FDA guidance. |
| Risk Management | Compliance with EN ISO 14971:2019. |
| Biocompatibility | Materials contacting intact patient skin (<24 hours) determined to be biocompatible through testing. Compliance with ISO 10993-1:2018. |
| Accuracy of RF energy | Bench testing performed to verify accuracy. |
| Predicate Equivalence (Sunny RF Handpiece) | Demonstrated equivalence to Thermage CPT in general design, frequency (6.78 MHz), and intended use for electrocoagulation and hemostasis, supported by thermal testing. |
| Predicate Equivalence (Pulsar Handpiece) | Demonstrated equivalence in intended use, mechanism of action, and general design to the predicate device, supported by tissue temperature maintenance testing despite differences in output power and frequency. Also heavily references Nuera Tight RF (K210867) as a reference device sharing frequencies and specific indications. |
2. Sample sizes used for the test set and the data provenance
- Test Set Sample Size: Not explicitly stated for bench tests (RF accuracy, temperature maintenance, histopathology). For biological evaluation (biocompatibility), it typically involves testing material samples, not patient data in the context of device performance.
- Data Provenance: Not specified regarding country of origin or whether it was retrospective/prospective. The testing appears to be primarily bench testing and materials testing, not human clinical data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- This question is not applicable as the provided document describes a hardware medical device (electrosurgical unit) and not an AI/software diagnostic tool. The "ground truth" for this device's performance is established through physical measurements, electrical output verification, thermal effect demonstration, and adherence to international safety and performance standards. Histopathology for thermal effect assessment would involve trained pathologists, but the number and qualifications are not specified nor is it a "ground truth" in the AI sense.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. This refers to adjudication of discrepancies in human expert labels for AI ground truth, which is not relevant here. The testing methods described are standard engineering and safety compliance tests.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. The device is an energy-based surgical tool, not an AI diagnostic aid. No MRMC study involving human readers assisting with AI was performed or is relevant to its intended use.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This pertains to AI algorithm performance. The device performs its intended function autonomously based on user input (setting power, mode), and its performance is evaluated by its physical effects (e.g., thermal effect, temperature maintenance), not as an algorithm generating diagnostic outputs.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- The "ground truth" here is established primarily by:
- Bench Testing Measurements: Direct measurement of electrical parameters (power, frequency, energy output).
- Histopathology: Microscopic examination of tissue after device application to demonstrate thermal effects (for electrocoagulation/hemostasis). This is a form of pathology.
- Temperature Measurement: Direct measurement of skin temperature during operation to confirm therapeutic heating.
- Compliance with Standards: Meeting the requirements and specifications outlined in various IEC and ISO standards (e.g., 60601-1 for basic safety, 14971 for risk management, 10993-1 for biocompatibility).
8. The sample size for the training set
- Not applicable. The device is a hardware product, not a machine learning model. There is no "training set" in the context of AI. Design and development follow engineering principles, not data-driven model training.
9. How the ground truth for the training set was established
- Not applicable. Refer to point 8.
Ask a specific question about this device
(266 days)
The Sunny Cup and Applicator is a receptacle placed in the vagina to collect blood and cellular debris that is exuded from the uterus via the cervix during menstruation.
The Sunny Cup and Applicator is an over the counter (OTC) medical device consisting of a reusable silicone menstrual cup with a lubricious coating and a reusable applicator with a nylon body and a thermoplastic polyurethane (TPU) tip with a lubricious coating. The Sunny Cup is a soft receptacle designed for placement within the vaginal canal to hold up to 25 mL of menstrual fluid and can remain within the body for up to 12 hours. The Sunny Cup is also designed with grip lines and a stem on the distal end to allow for easy location and manipulation of the Sunny Cup during removal.
The Applicator component enables the insertion process for the user and consists of a barrel, a pusher, and a pusher base, functioning similarly to a tampon applicator. For the insertion process, the Sunny Cup is folded and placed in the Applicator barrel. The pusher component is then used to deploy the Cup through the Applicator. The Sunny Cup then unfolds upon deployment to create a seal in the vaginal canal and collect menstrual fluid. After 12 hours, the Sunny Cup is removed manually by squeezing the grip lines to break the seal and pulling the Cup out by the grip lines along the base of the Cup. The Cup component should be cleaned via boiling at the start of each menstrual cycle, and the Cup and Applicator components should be cleaned with warm water and mild, fragrance-free soap between uses. The Sunny Cup and Applicator may be reused by a single patient for up to a year or until discoloration or damage is evident.
The provided text describes the "Sunny Cup and Applicator," a menstrual cup device, and its substantial equivalence to a predicate device, the "SckoonCup." However, the document does not contain information related to acceptance criteria, specific device performance metrics, or study details such as sample sizes for test/training sets, ground truth establishment, expert qualifications, or MRMC studies.
The document primarily focuses on demonstrating substantial equivalence to a predicate device, outlining comparisons in technological characteristics, materials, and intended use. The performance testing mentioned is "Bench performance testing" and "Biocompatibility Testing," which verified certain aspects but aren't presented with acceptance criteria or specific quantitative outcomes in the provided summary.
Therefore, most of the requested information cannot be extracted from this document, as it is a 510(k) summary, not a detailed study report.
Here's a summary of the available information and the missing details:
1. Table of Acceptance Criteria and Reported Device Performance:
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Not specified in the document. | - Bench Testing: Verified the Cup and Applicator can withstand anticipated applied forces during cup deployment. - Bench Testing: Verified the Cup met the 25 mL volume capacity. - Bench Testing: Verified the Cup met folding requirements throughout its use-life. - Biocompatibility Testing: Demonstrated non-cytotoxic, non-irritating, non-sensitizing, and non-systemically toxic for both cup and applicator. - Reprocessing and Cleaning Validation: Demonstrated acceptability of cleaning procedures and ability to withstand repeated reprocessing. |
Missing Details: Specific numerical acceptance criteria (e.g., minimum force withstood, specific ranges for chemical levels, cleanliness thresholds) are not provided. The performance is reported qualitatively as "met the prespecified requirements" or "demonstrated acceptability."
2. Sample size used for the test set and the data provenance:
- Sample Size for Test Set: Not specified.
- Data Provenance (country of origin, retrospective/prospective): Not specified.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Number of Experts: Not applicable, as no clinical study or expert-adjudicated test set is mentioned.
- Qualifications of Experts: Not applicable.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Adjudication Method: Not applicable, as no clinical study or expert-adjudicated test set is mentioned.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- MRMC Study: No, a MRMC study was not done. The device is a physical medical device (menstrual cup and applicator), not an AI-powered diagnostic or assistive tool.
- Effect size of human readers with/without AI: Not applicable, as no AI component is mentioned.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Standalone Performance: Not applicable. This is a physical medical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Type of Ground Truth: Not applicable, as the document describes bench testing and biocompatibility. The "ground truth" for these tests would be established by validated test methods and standards (e.g., ISO 10993 standards for biocompatibility), not clinical ground truth types like pathology or expert consensus.
8. The sample size for the training set:
- Sample Size for Training Set: Not applicable, as this is not an AI/machine learning device requiring a training set.
9. How the ground truth for the training set was established:
- How Ground Truth was Established: Not applicable.
Ask a specific question about this device