Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K191895

    Validate with FDA (Live)

    Date Cleared
    2019-12-09

    (147 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Straumann® Mini Implants ø2.4 mm are suitable for oral endosteal implantation in the upper and lower jaw of fully or partially edentulous patients. The implants can be placed with immediate function when good primary stability is achieved. Strauman® Mini Implants are intended for the stabilization of removable dentures.

    Device Description

    The Straumann Mini Implants are tapered implants with an external diameter of 2.4 mm and lengths of 10, 12, and 14 mm. Standard manufacturing processes such machining, surface treatment, cleaning, packaging and sterilization were applied for the subject devices. There are no significant changes related to the implant material, surface treatment, fundamental operating principles, sterilization processes or procedures between the subject devices and the reference devices cleared under K083550. The main difference is the Optiloc® attachment element incorporated on the top of the implant. The implants are manufactured utilizing the Roxolid material and are finished with SLA® surface. The implant neck is machined and the Optiloc® attachment element of the implants is acting as a retention feature for dentures and it is coated using a Titanium Nitride (TiN) coating. Straumann® Mini Implants Ø2.4 mm are suitable for oral endosteal implantation in the upper and lower jaw of fully or partially edentulous patients. The implants can be placed with immediate function when good primary stability is achieved. The Straumann Mini Implants are intended for the stabilization of removable dentures. The removable dentures are connected to the Mini Implants through the incorporated Optiloc® attachment element.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the Straumann® Mini Implants. The information provided heavily focuses on the comparison to predicate devices to demonstrate substantial equivalence, rather than a clinical study evaluating the device's performance against specific quantitative acceptance criteria in a human population.

    Therefore, many of the requested details regarding acceptance criteria for device performance in a clinical setting, sample sizes for test sets, expert involvement for ground truth, and MRMC studies will not be present in this type of regulatory submission. The performance testing outlined here is primarily bench testing to demonstrate mechanical and material equivalence.

    Here's an analysis based on the provided document:

    Acceptance Criteria and Device Performance

    The document does not specify quantitative clinical performance acceptance criteria (e.g., success rates, complication rates) and corresponding reported device performance from a clinical trial. Instead, the "acceptance criteria" discussed are related to bench testing for demonstrating equivalence to predicate devices.

    Table 1: Acceptance Criteria (Bench Testing) and Reported Device Performance

    Test/CharacteristicAcceptance Criteria (Implicit from Equivalence Claim)Reported Device Performance (Summary from Bench Testing)
    Insertion TorqueReach a suitable implant insertion torque, comparable to predicate/reference devices.Average value of insertion torque measurements for each bone plate density and each implant length are "within the defined acceptance criteria."
    Wear Testing (Optiloc® attachment retention)Retention force loss of the Optiloc® blue ring within acceptable limits.The retention force loss of the Optiloc® blue ring on Straumann Mini Implants "passed the acceptance criteria."
    BiocompatibilityNo new issues raised compared to previously cleared materials.Roxolid material and TiN coating previously cleared (K083550, K190040).
    SterilizationAchieve a Sterility Assurance Level (SAL) of 10^-6.Validated in accordance with ISO 11137-1:2006 and ISO 11137-2:2013, achieving SAL of 10^-6.
    Shelf LifeMaintain integrity and sterility for 5 years.Determined through real-time and accelerated aging (ASTM F 1980) to be 5 years.
    PyrogenicityMeet pyrogen limit specifications.Testing limit of 20 EU/device met via LAL Endotoxin Analysis.
    Dynamic FatigueMechanical performance comparable to predicate devices under dynamic loading.Demonstrated equivalence to primary predicate devices in tests conducted according to ISO 14801 and FDA Guidance.

    Study Details (Based on Provided Document)

    Given that this is a 510(k) submission primarily relying on substantial equivalence through bench testing, a traditional clinical study with a "test set" in the context of AI/diagnostic device evaluation (as implied by the detailed questions) was not conducted or reported here. The performance testing discussed pertains to physical and mechanical characteristics of the dental implants.

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size: Not applicable in the context of a clinical test set for AI/diagnostic performance. The "test set" here refers to the samples of the physical device used for bench testing (e.g., number of implants for insertion torque, wear, and fatigue tests). Specific quantities are not provided, but these are typically small, controlled laboratory samples.
      • Data Provenance: Not applicable in terms of patient data. The data provenance is from laboratory bench testing.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. Ground truth for a physical device's mechanical properties is established through standardized engineering tests, not expert consensus on medical images or clinical outcomes.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. This concept applies to human reader studies or data labeling, not bench testing of a physical implant.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This is irrelevant to the evaluation of a dental implant's mechanical and material properties. The device is a physical implant, not an AI-assisted diagnostic tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • No. This also refers to AI/software performance, not a dental implant.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for this device is based on engineering standards and specifications (e.g., ISO 14801, ISO 11137), and the properties of the previously cleared predicate devices. For example, a successful fatigue test to a certain threshold is "ground truth" for mechanical durability.
    7. The sample size for the training set:

      • Not applicable. This device is not an AI/machine learning model, so there is no "training set."
    8. How the ground truth for the training set was established:

      • Not applicable. As noted above, there is no training set for this type of device.

    Summary of what the document primarily reports:

    The document, a 510(k) summary, demonstrates the substantial equivalence of the Straumann® Mini Implants to legally marketed predicate devices. This is achieved by:

    • Comparing Indications for Use: Showing the new device's indications are equivalent to a predicate (K031106), with one difference (not indicated for inter-radicular transitional applications) that does not raise new safety or effectiveness questions.
    • Comparing Technological Characteristics: Detailing similarities and differences in material, dimensions, coating, surface treatment, etc., with predicates and reference devices (Table 1).
    • Performance Testing (Bench Testing): Conducting various engineering tests (insertion torque, wear, biocompatibility, sterilization, shelf life, pyrogenicity, dynamic fatigue) and demonstrating that the device meets internal "acceptance criteria" for these tests, which are implicitly set to show equivalence to or non-inferiority against the predicate devices' known performance. The tests adhere to relevant FDA guidance and ISO standards.

    This type of submission is common for Class II medical devices where clinical trials are not explicitly required if substantial equivalence can be demonstrated through non-clinical means.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1