Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K232039

    Validate with FDA (Live)

    Date Cleared
    2023-08-09

    (30 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition, these gloves were tested for use with Chemotherapy drugs and fentanyl citrate in accordance with ASTM D6978.

    Device Description

    The subject device is Nitrile Examination Gloves Sterile with Chemotherapy drugs and fentanyl citrate claims. The subject device is a patient examination glove made from nitrile compound, blue color, powder free and sterile. They are ambidextrous and come in different sizes - Extra Small, Medium, Large, Extra Large and XXL. The device meets all the specifications in ASTM D6319-19, Standard specification for Nitrile Examination Gloves. Additionally, the gloves have been tested for permeability to chemotherapy drugs and fentanyl citrate per ASTM D6978-05(2019).

    AI/ML Overview

    The document provided is a 510(k) Premarket Notification from the FDA for "Sterile Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs and Fentanyl Citrate." This document primarily focuses on demonstrating substantial equivalence to a predicate device, specifically by adding a claim for fentanyl citrate permeation. It performs non-clinical testing to support this claim, rather than a study involving AI or human readers for diagnostic purposes.

    Therefore, many of the requested details regarding AI device acceptance criteria, human reader studies, and AI-specific ground truth establishment are not applicable to this submission. This is a clearance for a physical medical device (gloves), not a diagnostic AI/ML device.

    However, I can extract the relevant information regarding the device's acceptance criteria and the non-clinical study conducted to prove it meets those criteria.


    Acceptance Criteria and Study for Sterile Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs and Fentanyl Citrate

    This medical device is a physical product (examination gloves), not an AI/ML diagnostic tool. The "acceptance criteria" here refer to the physical and chemical performance standards for the gloves, and the "study" refers to non-clinical bench testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the re-submitted device (K232039) are based on relevant ASTM and ISO standards, and for chemotherapy and opioid drug permeation, specific minimum breakthrough times (BDT). The reported device performance indicates that the gloves passed all these criteria.

    Referenced Standards / Test PerformedAcceptance CriteriaReported Device Performance
    Physical Dimensions (ASTM D6319-19)Passed
    LengthMinimum 220mm for size XS and S, 230mm for size M, L, XLPass (All size ≥290mm reported in 510(k) Summary, page 5)
    Palm WidthXS: 70±10mm; S: 80±10mm; M: 95±10mm; L: 110±10mm; XL: 120±10mm; XXL: 130±10mmPass (Reported ranges: XS: 76-78mm; S: 86-88 mm; M: 96-98mm; L: 106-108 mm; XL: 116-118 mm; XXL: 126-128 mm)
    Thickness (Finger & Palm)Finger: 0.05mm (min); Palm: 0.05mm (min)Pass (Reported ranges: Palm: 0.09-0.11mm; Finger: 0.13-0.14mm)
    Physical Properties (ASTM D6319-19, ASTM D412-16 (2021))Passed
    Tensile Strength (Before Aging)Min 14 MPaPass (Reported range: 15.8-20.8 MPa)
    Tensile Strength (After Accelerated Aging)Min 14 MPaPass (Reported range: 15.6-19.8 MPa)
    Ultimate Elongation (Before Aging)Min 500%Pass (Reported range: 500-560%)
    Ultimate Elongation (After Accelerated Aging)Min 400%Pass (Reported range: 400-500%)
    Freedom from holes (ASTM D6319-19, ASTM D5151-19)G-I, AQL 2.5Pass (Meet AQL2.5 requirements)
    Powder Residue (ASTM D6319-19, ASTM D6124-06 (2017))Max 2 mg/glovePass (≤ 2 mg, meet requirements)
    Chemical Permeation (ASTM D6978-05 (2019))Passed (Refer to specific drug breakdown for values)
    Chemotherapy DrugsSee "Indications for Use" table on pages 3 and 6, and "Chemotherapy Permeation Comparison Claim" table on pages 7-8 for specific minimum BDTs for 15 listed drugs.All reported BDTs met or exceeded the minimum requirements. For example, Carmustine (BCNU) had a BDT of 54.3 min (meeting criteria), and many others had >240 min.
    Opioid Drug (Fentanyl Citrate Injection)Minimum BDT >240 minutes>240 minutes
    Biocompatibility (ISO 10993-10:2010)No Skin sensitization and Skin irritationIs non-sensitization and Non-irritation
    Biocompatibility (ISO 10993-11:2017)Subject showed no adverse biological reactionno evidence of acute systemic toxicity.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the exact sample sizes used for each physical or chemical test. However, for tests like "Freedom from holes," it refers to AQL (Acceptable Quality Level) 2.5, which implies a sampling plan according to ASTM D5151-19. Similarly, other ASTM and ISO standards followed for physical properties and biocompatibility would specify sample sizes.

    The data for these tests would be generated prospectively as part of product development and quality control for the submission. The country of origin for the data generation is implied to be China, given the manufacturer's location.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This concept is not applicable here. Ground truth in the context of this submission refers to the established standards and methodologies of the ASTM and ISO tests (e.g., how to measure tensile strength, how to test for permeation). These are standardized laboratory tests, not subjective expert interpretations of images or clinical outcomes. The "experts" are the laboratory personnel who perform these tests according to the published standards.

    4. Adjudication Method for the Test Set

    Not applicable. This is not a study involving human readers or subjective interpretations requiring adjudication. The tests have predefined, objective measurement protocols.

    5. If a Multi Reader Multi Case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a physical medical device (gloves), not an AI-assisted diagnostic tool. No MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. There is no algorithm or AI component in this device. Performance is measured directly through bench testing.

    7. The type of ground truth used

    The "ground truth" for this device's performance is established by objective, standardized laboratory measurements against predefined performance specifications stated in ASTM (American Society for Testing and Materials) and ISO (International Organization for Standardization) standards. Specifically:

    • Physical Properties: Measured values (e.g., length, width, thickness, tensile strength, elongation) are compared directly against minimum/maximum thresholds defined in ASTM D6319-19 and ASTM D412-16.
    • Freedom from holes: Determined by water leak testing according to ASTM D5151-19, with acceptance based on AQL 2.5.
    • Powder Residue: Quantified according to ASTM D6124-06 (2017) against a maximum limit.
    • Chemical Permeation: Breakthrough Time (BDT) is measured for specific chemotherapy drugs and fentanyl citrate according to ASTM D6978-05 (2019) and compared against specified minimum times.
    • Biocompatibility: Assessed through in-vitro/in-vivo tests for irritation, sensitization (ISO 10993-10:2010), and acute systemic toxicity (ISO 10993-11:2017) to ensure no adverse biological reactions.

    8. The Sample Size for the Training Set

    Not applicable. As this is not an AI/ML device, there is no "training set."

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220250

    Validate with FDA (Live)

    Date Cleared
    2022-05-11

    (100 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sterile Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition these gloves were tested for use with Chemotherapy drugs in accordance with ASTM D6978 standards.

    Device Description

    The subject device is Nitrile Examination Gloves Sterile with Chemotherapy drugs claims. The subject device is a patient examination glove made from nitrile latex compound, blue color, powder free and sterile (Per 21 CFR 880.6250, class I). The device is direct contact and not to be worn for more than 24 hours. The device meets all the specifications in ASTM D6319-19, Standard specification for Nitrile Examination Gloves. Additionally, the gloves have been tested for biocompatibility per 10993-10, 10993-11 and permeability to chemotherapy drugs per ASTM D6978-05(2019).

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA regarding Sterile Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs.

    This document does not describe the acceptance criteria and study proving a software algorithm device meets these criteria, but rather the acceptance criteria and test results for a physical medical device (examination gloves).

    Therefore, many of the requested categories in your prompt (such as sample size for test/training sets, data provenance, number of experts for ground truth, adjudication method, MRMC study, standalone performance, etc.) are not applicable to this type of device and the information provided in the 510(k) summary.

    However, I can extract the relevant information about the acceptance criteria and the device's performance based on the provided document.


    Device: Sterile Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs

    Type of Device: Physical Medical Device (Examination Gloves), Class I, reserved

    Study Purpose: To demonstrate substantial equivalence to a predicate device (Riverstone Resources SDN BHD. Nitrile Examination Glove, Sterile Tested for use with Chemotherapy Drugs Claim, White (K190725)) by verifying that the subject device meets specific performance standards and has similar technological characteristics.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document presents several tables detailing the acceptance criteria (standards/specifications) and the results indicating that the device meets these criteria.

    Item/TestAcceptance Criteria (Standard/Specification)Reported Device Performance (Results)
    Dimensions
    LengthMinimum 230mmAll size ≥290mm
    Palm Width (XS)70 ± 10mm76-78mm
    Palm Width (S)80 ± 10mm86-88mm
    Palm Width (M)95 ± 10mm96-98mm
    Palm Width (L)110 ± 10mm106-108mm
    Palm Width (XL)120 ± 10mm116-118mm
    Palm Width (XXL)130 ± 10mm126-128mm
    Physical Properties (Thickness)
    Thickness (Palm)0.05mm (min)0.09-0.11mm
    Thickness (Finger)0.05mm (min)0.13-0.14mm
    Physical Properties (Tensile Strength)
    Tensile Strength, Before Aging14MPa, min15.8-20.8 MPa
    Tensile Strength, After Accelerated Aging14MPa, min15.6-19.8 MPa
    Physical Properties (Ultimate Elongation)
    Ultimate Elongation, Before Aging500%, min500-560%
    Ultimate Elongation, After Accelerated Aging400%, min400-500%
    Integrity
    Freedom from holesG-I, AQL 2.5 (per ASTM D5151-19, ASTM D6319-19)Meet and above AQL 2.5 requirements
    Powder Content≤ 2 mg per glove (per ASTM D6124-06(2017), ASTM D6319-19)≤ 2 mg, meet requirements
    Biocompatibility
    Primary Skin Irritation TestNot an irritant (per ISO 10993-10:2010)Not an irritant
    Dermal Sensitization AssayNot a sensitizer (per ISO 10993-10:2010)Not a sensitizer
    Acute Systemic Toxicity TestNo evidence of systemic toxicity (per ISO 10993-11:2017)No evidence of systemic toxicity
    Chemotherapy Drug Permeation (Minimum Breakthrough Time - BDT)
    Carmustine (BCNU) (3.3mg/ml)- (Predicate: 39 min)54.3 minutes
    Cyclophosphamide (20mg/ml)- (Predicate: >240 min)>240 minutes
    Doxorubicin Hydrochloride (2 mg/ml)- (Predicate: >240 min)>240 minutes
    Etoposide (20mg/ml)- (Predicate: >240 min)>240 minutes
    Fluorouracil (50mg/ml)- (Predicate: >240 min)>240 minutes
    Paclitaxel (6mg/ml)- (Predicate: >240 min)>240 minutes
    Thiotepa (10mg/ml)- (Predicate: 97.6 min)196.7 minutes
    Methotrexate (25mg/ml)- (Predicate: >240 min)>240 minutes
    Cisplatin (1mg/ml)- (Predicate: N/A)>240 minutes
    Vincristine Sulfate (1.0 mg/ml)- (Predicate: N/A)>240 minutes
    Cytarabine HCL (100mg/ml)- (Predicate: N/A)>240 minutes
    Mitoxantrone (2.0 mg/ml)- (Predicate: N/A)>240 minutes
    Mitomycin C (0.5 mg/ml)- (Predicate: N/A)>240 minutes
    Ifosfamide (50.0 mg/ml)- (Predicate: N/A)>240 minutes
    Dacarbazine (DTIC) (10.0 mg/ml)- (Predicate: N/A)>240 minutes

    (Note: For chemotherapy drug permeation, the "Acceptance Criteria" are implied by comparison to the predicate device and showing equivalent or better performance. The exact criteria are not explicitly stated for each drug beyond the measured BDT.)

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not explicitly stated as a number of "cases" or "samples" in the way a software algorithm test set would be. The testing is based on batches of gloves or representative samples according to the referenced ASTM/ISO standards. For example, "Freedom from holes" testing follows an AQL (Acceptance Quality Limit) which implies a sampling plan.
    • Data Provenance: The tests are non-clinical and conducted based on international standards (ASTM, ISO). The manufacturer is Grand Work Plastic Products Co., Ltd, located in Donggao Industrial Zone, Zanhuang, Hebei, China. The testing would have been performed there or at certified labs used by the manufacturer. The data is retrospective in the sense that it was collected as part of the device's development and qualification prior to the 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The "ground truth" for this physical device's performance (e.g., tensile strength, breakthrough time) is established by adherence to standardized physical and chemical testing methods (ASTM, ISO standards), not by human expert consensus or clinical judgment in the traditional sense of diagnostic AI. These standards define the test procedures and acceptance limits.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not a study requiring human reader adjudication. Performance is measured by laboratory equipment and standard protocols.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-based diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI-based diagnostic device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" is established through standardized physical, chemical, and biological testing methods as defined by:

    • ASTM D6319-19 (Standard Specification for Nitrile Examination Gloves)
    • ASTM D5151-19 (Standard Test Method for Detection of Holes in Medical Gloves)
    • ASTM D6124-06(2017) (Standard Test Method for Residual Powder on Medical Gloves)
    • ASTM D6978-05(2019) (Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs)
    • ISO 10993-10:2010 (Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization)
    • ISO 10993-11:2017 (Biological evaluation of medical devices - Part 11: Tests for systemic toxicity)

    8. The sample size for the training set

    Not applicable. This is not a machine learning device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is not a machine learning device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1