Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K172850

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-11-17

    (58 days)

    Product Code
    Regulation Number
    880.6850
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aesculap Sterilcontainer System is a reusable sterilization container system (consisting of solid & perforated bottoms, a perforated lid w/ filter retention plates, and reusable polytetrafluoroethylene (PTFE) filters) intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device until used. This container system is compatible for use in pre-vacuum steam and IUSS (Immediate Use Steam Sterilization modalities. The SterilContainer System for includes accessories such as silicone mats, baskets, trays, and racks.

    Device Description

    The SterilContainer with PrimeLine Pro Lid is used in conjunction with Aesculap's Sterilcontainer Systems (K792558/K944864/K053389/K112671). The SterilContainer with PrimeLine Pro Lid is designed to be compatible for use with pre-vacuum steam and IUSS sterilization. The lid is manufactured from anodized aluminum and utilizes a reusable polytetrafluoroethylene (PTFE) filter. The SterilContainer with PrimeLine Pro Lid is offered in a various colors and range of sizes as the predicates including full, ¾, and half sizes lids.

    AI/ML Overview

    The Aesculap SterilContainer with PrimeLine Pro Lid is a reusable sterilization container system designed to sterilize and maintain the sterility of medical devices. The device was found to be substantially equivalent to its predicate devices based on non-clinical performance testing.

    Here's a breakdown of the acceptance criteria and study information:

    1. Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Sterilization EfficacyTesting demonstrated a 6 log reduction and a sterility assurance level (SAL) of $10^{-6}$ using the biological (BI) overkill method and half-cycle validation.
    Whole Package Microbial Aerosol ChallengeAfter exposure to a defined amount of aerosol microorganisms, the contents maintained sterility.
    Event Related Sterility MaintenanceThe test reports demonstrated that the SterilContainer with PrimeLine Pro Lid consistently maintained sterility of the contents after processing followed by 30, 180, and 365-day event-related storage under conditions which simulate hospital sterile package handling and storage conditions as long as the barrier is not compromised.
    Mechanical Cleaning (Protein Analysis)The study results validate that the mechanical cleaning instructions were effective in removing gross amounts of soil from the SterilContainer with PrimeLine Pro Lid to a protein level of less than $6.4 \mu g/cm^2$ per device.
    Mechanical Cleaning (Hemoglobin Analysis)The study results validate that the mechanical cleaning instructions were effective in removing gross amounts of soil from the SterilContainer with PrimeLine Pro Lid to a hemoglobin level of less than $2.2 \mu g/cm^2$ per device.
    Manual Cleaning (Protein Analysis)The study results validate that the manual cleaning instructions were effective in removing gross amounts of soil from the SterilContainer with PrimeLine Pro Lid to a protein level of less than $6.4 \mu g/cm^2$ per device.
    Manual Cleaning (Hemoglobin Analysis)The study results validate that the manual cleaning instructions were effective in removing gross amounts of soil from the SterilContainer with PrimeLine Pro Lid to a protein level of less than $2.2 \mu g/cm^2$ per device.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the specific sample sizes used for each test. It mentions that "Test were performed in best/mid/worst case sized regarding vent to volume ratio and container size." The data provenance is not specified, but the testing was conducted by a "qualified testing laboratory" in accordance with FDA guidance and AAMI standards. This indicates a prospective and controlled experimental design.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This device is not an AI/imaging device requiring expert interpretation for ground truth. The performance testing involves objective measures of sterilization efficacy, microbial challenge, sterility maintenance, and cleaning effectiveness, which do not typically rely on expert consensus for ground truth establishment in the same way an AI diagnostic tool would.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As noted above, this device does not involve human interpretation or subjective assessment that would require an adjudication method.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI device, so MRMC studies and the concept of human readers improving with AI assistance are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm-based device. The testing focuses on the physical and functional performance of the sterilization container system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the performance testing is based on established scientific and regulatory standards for sterilization, microbial challenge, sterility maintenance, and cleaning efficacy.

    • Sterilization Efficacy: Ground truth is a 6 log reduction and SAL of $10^{-6}$ as per biological indicator results and half-cycle validations.
    • Microbial Aerosol Challenge: Ground truth is the absence of microbial growth in the contents after exposure to a defined amount of aerosol microorganisms.
    • Event Related Sterility Maintenance: Ground truth is the absence of microbial growth in the contents after storage periods, indicating sterility was maintained.
    • Cleaning Effectiveness: Ground truth is a protein level of less than $6.4 \mu g/cm^2$ and a hemoglobin level of less than $2.2 \mu g/cm^2$ per device, as determined by laboratory analysis.

    These are objective, measurable outcomes based on established scientific principles and regulatory benchmarks.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device, so there is no concept of a "training set."

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set for this device, a method for establishing its ground truth is not relevant.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1