Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    DEN150011

    Validate with FDA (Live)

    Device Name
    Sonablate 450
    Date Cleared
    2015-10-09

    (200 days)

    Product Code
    Regulation Number
    876.4340
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sonablate® 450 is indicated for transrectal high intensity focused ultrasound (HIFU) ablation of prostatic tissue.

    Device Description

    The Sonablate® 450 is a computer-controlled device designed for transrectal delivery of high intensity focused ultrasound (HIFU) energy to the prostate to ablate regions of unwanted prostatic tissue via thermal ablation. The device makes use of integrated biplanar ultrasound imaging for real-time monitoring, planning, and pre- and post-ablation imaging of the prostate. The system consists of the following main components: Sonasource Console, Sonablate Probe 30/40, Sonachill, Sonablate Multi-Axis Stepper and Probe Arm, and Single Use Components (Probe Tip Kit and Water Path Kit).

    AI/ML Overview

    The provided text describes the Sonablate® 450, a High Intensity Focused Ultrasound (HIFU) system for prostate tissue ablation. The information covers device description, non-clinical studies (biocompatibility, shelf life, electrical safety, software), animal studies, and a clinical study. The acceptance criteria and the study proving the device meets these criteria are detailed, primarily within the "PERFORMANCE TESTING - BENCH" and "SUMMARY OF CLINICAL INFORMATION" sections, and implicitly derived from the "Special Controls" which outline the requirements for devices of this type.

    Here's a breakdown of the requested information:

    Acceptance Criteria and Device Performance

    The acceptance criteria for the Sonablate® 450 are primarily outlined in the "PERFORMANCE TESTING - BENCH" section and in the "SPECIAL CONTROLS" section. The reported device performance is presented in the "Results" columns of the tables in the "PERFORMANCE TESTING - BENCH" section and summarized in the "SUMMARY OF CLINICAL INFORMATION" and "Clinical Results Summary" sections.

    Table of Acceptance Criteria and Reported Device Performance

    Test/CharacteristicAcceptance CriteriaReported Device Performance
    Transducer Characterization
    Schlieren test (Beam profile & alignment)Image: b(4) - Specific numerical criteria are redacted but based on the overall context, it would relate to acceptable beam width and misregistration tolerance.All test samples (n=7) met the acceptance criteria.
    Electrostatic test (Focal length)Image: b(4) - Specific numerical criteria are redacted but would define the acceptable focal length range for each transducer (30mm and 40mm).All test samples (n=7) met the acceptance criteria. "These results support the targeting accuracy and reproducibility of the HIFU output."
    Force balance test (Total Acoustic Power- TAP)Image: b(4) - Specific output power range redacted.All test samples (n=7) met the acceptance criteria. (No acceptance criteria for focal intensity and acoustic pressure, as they were for modeling).
    Imaging Verification
    Phantom testing (Resolution, Penetration, SNR)Both transducers: Image: b(4) - Specific numerical criteria for resolution, penetration depth, and signal-to-noise ratio redacted.The test sample (n=1) met the acceptance criteria. "The resolution is sufficient for tissue differentiation of the prostate and surrounding anatomy."
    Imaging frame rate testImage: b(4) - Specific numerical criteria for linear and sector frame rate redacted.The test sample (n=1) met the acceptance criteria. "The frame rate is sufficient to provide feedback to the physician during probe positioning."
    Image display testUsing each transducer: the system performs transverse and longitudinal imaging using each probe, and displays them simultaneously.The test sample (n=1) met the acceptance criteria.
    HIFU Verification/Safety
    Ex vivo tissue phantom tests (Ablation volume/depth)b(4) - Specific numerical criteria for max ablation volume, depth, and lesion dimensions redacted. Criteria ensure volumes up to 40 cm³ at 40 mm depth, and creation of contiguous volume lesions of specified dimensions.The test sample (n=1) met the acceptance criteria. "The system is capable of targeting a tissue volume of up to 40 cm³, at a depth of up to 40 mm. The system is capable of creating single and volume lesions in x, y, and z dimensions that meet the acceptance criteria. These results support the targeting accuracy and reproducibility of the HIFU output."
    HIFU ON-time/OFF-time VerificationTo confirm the ON/OFF timing of the system's HIFU application cycle.The test sample (n=1) met the acceptance criteria. "The system meets ON time and OFF time requirements."
    Clinical Effectiveness (Primary Study Focus)Evidence of prostatic tissue ablation, as demonstrated by reductions in prostate volume and PSA level, and increase in the rate of negative biopsy, between baseline and 12 months post-HIFU (Implicitly, significant, statistically relevant changes are expected).Prostate Volume Reduction: Mean reduction of 11.8 cm³ (95% CI: 9.7, 14.0) in 73 patients, representing a mean reduction of 46% from baseline. PSA Reduction: 83% of 117 patients had a reduction in PSA post-ablation (95% CI: 74.6, 89.0). Negative Biopsy Rate: 61% of 117 patients had a negative post-ablation biopsy (95% CI: 51.2, 69.5). (Note: 30 non-biopsied patients were considered "positive").
    Clinical SafetyCharacterization of the adverse event profile, ensuring risks are outweighed by benefits.Adverse Events: 99% of patients experienced some application-related complication. Serious AEs included urinary retention (5%), hematuria (4%), rectal fistula (4%), UTI (4%), osteomyelitis (3%). Non-serious AEs included erectile dysfunction (71%), UTI (50%), urinary incontinence (47%), urinary retention (43%), hematuria (31%), urinary obstruction (20%), and urethral stricture (15%). Duration varied, some resolved, others required intervention. (Detailed table of AE percentages provided in the source text).

    2. Sample sizes used for the test set and the data provenance

    Test Set (Bench Testing):

    • Transducer Characterization: n=7 for Schlieren, Electrostatic, and Force balance tests.
    • Imaging Verification: n=1 for Phantom testing, Imaging frame rate test, and Image display test.
    • HIFU Verification/Safety: n=1 for Ex vivo tissue phantom tests and HIFU ON-time/OFF-time Verification Testing.

    Test Set (Animal Studies):

    • Study 1 ("Safety and Effectiveness of Tissue Ablation"): n=5 dogs.
    • Study 2 ("Efficacy Improvement Software Test Report Localization and Ablation"): n=4 dogs.
    • Study 3 ("Treatment Volume and Accuracy Study Report"): Preliminary dog study (n not specified, but likely small).
    • Study 4 ("Thermocouple Response to Treatment Test"): Preliminary dog study (n not specified, but likely small).
    • Study 5 ("Safety of HIFU Energy Application to Adjacent Tissues"): Preliminary dog study (n not specified, but likely small).

    Test Set (Clinical Study):

    • Clinical Dataset: Interim analysis of the initial 117 patients enrolled, with 116 undergoing HIFU application.
    • Data Provenance: Multi-center, prospective, single-arm study involving 20 sites across the U.S. and Canada. The study was structured as a clinical trial.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The provided text does not specify the number or qualifications of experts used to establish ground truth for the test set.

    • For bench and animal studies, "ground truth" is established through physical measurements, ex-vivo tissue analysis, and histological evaluation. These are typically performed by trained technical personnel and pathologists, but specific expert involvement (e.g., number and qualifications of pathologists for dog study histology) is not detailed.
    • For the clinical study, "ablation effectiveness" was assessed by analyzing changes in prostate volume, PSA level, and negative biopsy rate. PSA levels are objective measurements. Prostate volume changes and biopsy results would involve radiologists/urologists for imaging interpretation and pathologists for biopsy analysis. However, the exact number and qualifications of these healthcare professionals serving as ground truth experts are not given in the document.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The text does not describe any explicit adjudication method for observations or interpretations in the test set (bench, animal, or clinical).

    • For the clinical study, "investigator reported adverse events" were collected, implying individual site investigators reported. It does not mention a central adjudication committee for AEs or efficacy endpoints. The endpoint for negative biopsy rate notes that "30 patients who did not have a 12-month biopsy were considered 'positive'," which is a pre-defined rule for handling missing data, not an adjudication method for conflicting interpretations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader, multi-case (MRMC) comparative effectiveness study was not done. The Sonablate® 450 is a therapeutic device for tissue ablation, not an imaging diagnostic device assisted by AI. Therefore, the concept of human readers improving with AI assistance is not applicable in this context.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The Sonablate® 450 is a medical device for therapeutic ablation, not a diagnostic algorithm. Therefore, the concept of "standalone (algorithm only)" performance is not applicable in the typical sense of AI/ML software. The device's performance is inherently linked to its physical operation and the physician's control and monitoring. The "PERFORMANCE TESTING - BENCH" section details the device's functional performance in controlled (standalone power/accuracy) environments, but this is not an "algorithm only" evaluation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Bench Testing: Ground truth established via physical measurements (e.g., beam width, focal length, acoustic power, frame rate) and ex-vivo tissue analysis for lesion characterization.
    • Animal Studies: Ground truth established via histological evaluation of ablated tissue, temperature measurements using thermocouples, and comparison of planned vs. actual ablation zones.
    • Clinical Study: Ground truth for effectiveness was established using objective clinical outcomes data:
      • Prostate volume reduction (measured via imaging, presumably ultrasound, pre and post).
      • PSA level reduction (blood test, an objective biochemical marker).
      • Negative biopsy rate (histopathology, which is a gold standard for malignancy/tissue presence).
      • Ground truth for safety was established by investigator-reported adverse events and patient-completed questionnaires.

    8. The sample size for the training set

    The provided text does not include information about any training set. This is because the Sonablate® 450 is a hardware device for therapeutic ablation, not an AI/ML diagnostic or therapeutic algorithm that typically undergoes distinct training/test phases. While the system involves software that controls its functions, the documentation focuses on the verification and validation of this software, not its "training" on a dataset in the machine learning sense.

    9. How the ground truth for the training set was established

    As there is no explicit "training set" described for an AI/ML algorithm, the concept of establishing ground truth for it is not applicable in this context. The software's functionality is validated against design specifications and performance requirements through the various bench and animal tests, rather than being "trained" on a dataset with external ground truth.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1