Found 2 results

510(k) Data Aggregation

    K Number
    K252383

    Validate with FDA (Live)

    Device Name
    Somfit D
    Date Cleared
    2025-08-28

    (28 days)

    Product Code
    Regulation Number
    868.2375
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Somfit D is a single-use, non-invasive prescription device for home use with patients suspected to have sleep-related breathing disorders. The Somfit D is a diagnostic aid for the detection of sleep-related breathing disorders, sleep staging (REM, N1, N2, N3, Wake), and snoring level. The Somfit D system acquires electrical data from three frontal electrodes, tri-axial accelerometer data, acoustical and plethysmographic data. The Somfit D calculates and reports to clinicians EEG/EOG channels, Sleep Stages, SpO2, Peripheral Arterial Tonometry signal, pulse rate, and snoring level. The Somfit D calculates and reports to clinicians derived parameters such as Peripheral Arterial Tonometry-derived Apnea Hypopnea Index, Obstructive Desaturation Index; and hypnogram-derived indices such as time in each sleep stage. Somfit D data is not intended to be used as the sole or primary basis for diagnosing any sleep-related breathing disorder, prescribing treatment, or determining whether additional diagnostic assessment is warranted. The Somfit D is not intended for use as life support equipment, for example vital signs monitoring in intensive care unit. The Somfit D is a prescription device indicated for adult patients aged 21 years and over.

    Device Description

    The Somfit D is a home-based sleep monitoring device which records signals from the patient's forehead. Somfit D is a wearable, low voltage, battery operated device which is attached to subject forehead via a self-adhesive and disposable skin electrode patch. The electrodes are placed on the anterior Prefrontal cortex (PFC) at the Fp1 and Fp2 positions according to the 10/20 EEG system. The device allows for recording of two frontal EEG signals, pulse rate, SPO2, PAT, PPG, motion, and snore. Somfit D uses a mobile phone application to acquire data wirelessly via Bluetooth BLE technology, then transfer into a secure cloud, for management, storage and post-processing. The software reports measured parameters in a format compatible with the American Academy of Sleep Medicine guidelines, including sleep time, ODI, pAHI and conventional graphical displays such as a hypnogram.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and Somfit D 510(k) Summary describe the acceptance criteria and the study that proves the device meets those criteria. However, it explicitly states that the Somfit D is substantially equivalent to the predicate device, Somfit (K231546), and therefore, the performance data for the Somfit D is derived from the studies conducted on the Somfit. The summary then refers to already submitted and approved studies for the Somfit device.

    Here's a breakdown of the requested information based on the provided document:

    Acceptance Criteria and Device Performance (Derived from Predicate Device, Somfit)

    Table 1: Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (Implicit from Predicate Studies)Reported Device Performance (Somfit D, by Equivalence to Somfit)
    Oximeter PerformanceISO 80601-2-61 complianceAchieved (Formal controlled desaturation study conducted as per standard, already submitted and approved for Somfit)
    PAT-derived Apnea-Hypopnea Index (pAHI)Meaningful validation as an HSAT device (implied by previous clearance of Predicate)Meaningful validation as an HSAT device (study conducted on Somfit, already submitted and approved)
    Oxygen Desaturation Index (ODI)Meaningful validation as an HSAT device (implied by previous clearance of Predicate)Meaningful validation as an HSAT device (study conducted on Somfit, already submitted and approved)
    Sleep Staging Concordance (REM, N1, N2, N3, Wake)Meaningful validation as an HSAT device (implied by previous clearance of Predicate)Meaningful validation as an HSAT device (study conducted on Somfit, already submitted and approved)
    Electrical SafetyIEC 60601-1:2005 (Third Edition) + COR1:2006 + COR2:2007 + A1:2012 complianceAchieved (Testing activities on Somfit, identical to Somfit D in this aspect)
    Electromagnetic Compatibility (EMC)IEC 60601-1-2:2014, EN 60601-1-2:2015 complianceAchieved (Testing activities on Somfit, identical to Somfit D in this aspect)
    Home Healthcare EnvironmentalIEC 60601-1-11:2015 complianceAchieved (Testing activities on Somfit)
    Electroencephalograph safety and performanceIEC 60601-2-26:2012 complianceAchieved (Testing activities on Somfit, identical to Somfit D in this aspect)
    Pulse oximeter safety and performanceISO 80601-2-61:2011 compliance (including functional simulator)Achieved (Testing activities on Somfit, identical to Somfit D in this aspect)
    Lithium Battery SafetyIEC 60086-4 (single use lithium batteries) complianceAchieved (Specific test report for Somfit D's CR2032 battery)
    Hardware Bench Testing / Electrical Parameters/Design SpecificationsVerification of electrical parameters and design specificationsAchieved (For Somfit)
    Software Functional Requirements / System IntegrationVerification of functional requirements and system integrationAchieved (For Somfit)
    BiocompatibilityOvernight use on intact skin (implied by materials and intended use)Met (For Somfit)

    Study Details (Pertaining to the Predicate Device, Somfit, as explicitly stated for Somfit D equivalence)

    1. Sample sizes used for the test set and the data provenance:

      • Oximeter Validation: Controlled desaturation study in accordance with ISO 80601-2-61. Specific sample size not specified in this document, but implied to be sufficient for standard compliance. Data provenance is implied to be from a "Hypoxia Lab." The document doesn't specify if it was retrospective or prospective, or country of origin, but clinical studies for FDA clearance are typically prospective to ensure controlled conditions.
      • Home Sleep Apnea Test Validation (pAHI, ODI, Sleep Staging): A "Multi-Site Clinical Study" was conducted. Specific sample sizes for each of these validations are not provided in this summary. The document states "clinical data for the purpose of the predicate device Somfit (K231546), already submitted and approved," indicating these details would be found in the original Somfit 510(k) submission. The provenance is from "Multi-Site Clinical Study," implying multiple locations, likely within the regulatory jurisdiction where the clearance was sought (e.g., US, Australia). It's implied to be prospective for validation purposes.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not explicitly stated in the provided 510(k) summary for any of the clinical validations (Oximeter, pAHI, ODI, Sleep Staging). It only mentions that the studies were "meaningful validations" and "concordance," implying comparison to a gold standard. For sleep staging, ground truth is typically established by certified polysomnography technologists and/or sleep physicians.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • This information is not explicitly stated in the provided 510(k) summary for any of the clinical validations.
    4. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC study is mentioned. The Somfit D (and its predicate Somfit) is described as a diagnostic aid that calculates and reports parameters to clinicians. It's not presented as an AI-assissted reading tool for human interpretation, but rather a device that quantifies specific physiological signals and derives standard indices. The human role is in interpretation of the reported data, not in an AI-assisted reading workflow.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, the device (Somfit D, through its predicate Somfit) performs standalone algorithmic analysis to calculate:
        • Sleep Stages (REM, N1, N2, N3, Wake)
        • SpO2
        • Peripheral Arterial Tonometry (PAT) signal
        • Pulse rate
        • Snoring level
        • PAT-derived Apnea Hypopnea Index (pAHI)
        • Obstructive Desaturation Index (ODI)
        • Hypnogram-derived indices (e.g., time in each sleep stage)
      • The "meaningful validation" studies for these parameters (pAHI, ODI, Sleep Staging) suggest a comparison of the device's algorithmic outputs against established ground truth.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Oximeter Validation: Performed in a "Hypoxia Lab," implying comparison to a highly accurate laboratory reference oximeter or blood gas analysis (gold standard for SpO2 measurements).
      • Home Sleep Apnea Test Validation (pAHI, ODI, Sleep Staging): While not explicitly stated, for HSAT devices, "ground truth" for pAHI and ODI is typically derived from comparison to full in-lab Polysomnography (PSG) data scored by a certified sleep technologist and/or interpreted by a board-certified sleep physician, which is considered the clinical gold standard. For sleep staging, the ground truth would be expert-scored PSG recordings.
    7. The sample size for the training set:

      • This information is not provided in the summary. The summary focuses on the validation studies, which imply the device's algorithms were already developed and trained.
    8. How the ground truth for the training set was established:

      • This information is not provided in the summary, as it pertains to the development phase rather than the validation phase described.
    Ask a Question

    Ask a specific question about this device

    K Number
    K231546

    Validate with FDA (Live)

    Device Name
    Somfit
    Date Cleared
    2023-11-30

    (184 days)

    Product Code
    Regulation Number
    868.2375
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Somfit system is a non-invasive prescription device for home use with patients suspected to have sleep-related breathing disorders. The Somfit is a diagnostic aid for the detection of sleep-related breathing disorders, sleep staging (REM, N1, N2, N3, Wake), and snoring level. The Somfit system acquires electrical data from three frontal electrodes, triaxial accelerometer data, acoustic and plethysmographic data. The Somfit calculates and reports to clinicians EEG/EOG channels, Sleep Stages, SpO2, Peripheral Arterial Tonometry (PAT) signal, pulse rate, and snoring level. The Somfit calculates and reports to clinicians derived parameters such as PAT-derived Apnea Hypopnea Index, Obstructive Desaturation Index; and hypnogram-derived indices such as time in each sleep stage. Somfit data is not intended to be used as the sole or primary basis for diagnosing any sleep-related breathing disorder, prescribing treatment, or determining whether additional diagnostic assessment is warranted. The Somfit is not intended for use as life support equipment, for example vital signs monitoring in intensive care unit. The Somfit is a prescription device indicated for adult patients aged 21 years and over.

    Device Description

    The Somfit is a home-based sleep monitoring device which records signals from the patient's forehead and surrounding environment. Somfit is a wearable, low voltage, battery operated device which is attached to subject forehead via a self-adhesive and disposable skin electrode patch. The electrodes are placed on the anterior Prefrontal cortex (PFC) at the Fp1 and Fp2 positions according to the 10/20 EEG system. The device allows for recording of two frontal EEG signals, pulse rate, SPO2, Peripheral Arterial Tonometry (PAT), PPG, motion, and snore. Somfit uses a mobile phone application to acquire data wirelessly via Bluetooth LE technology, then transfer into a secure cloud, for management, storage and postprocessing. The software reports measured parameters in a format compatible with AASM guidelines, including sleep time, pAHI and conventional graphical displays such as a hypnogram.

    AI/ML Overview

    The Somfit device underwent rigorous testing to establish its performance against predefined acceptance criteria, particularly for its ability to detect sleep stages and related parameters.

    Here's a breakdown of the acceptance criteria and the supporting study:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a formal "acceptance criteria" table with specific thresholds for each metric. However, the clinical performance section implicitly defines acceptance by demonstrating non-inferiority to the predicate device (Watch-PAT300) and superior or comparable performance against the gold standard (PSG) for key metrics.

    Based on the provided text, here's a table summarizing the performance metrics that act as implicit acceptance criteria:

    Implicit Acceptance Criteria & Reported Device Performance for Somfit

    MetricImplicit Acceptance Criteria (Comparative)Reported Somfit PerformanceReported Watch-PAT300 PerformanceGround Truth (PSG)
    PAT-derived AHI (pAHI)Non-inferiority to predicate (Watch-PAT300) compared to PSG.Mean difference with Watch-PAT300: 0.294 (95% CI: -2.661, 2.074). Mean difference with PSG: 0.658 (CI: -1.012, 2.326).Mean difference with PSG: 2.499 (CI: 0.732, 4.265).Polysomnography (PSG)
    Obstructive Desaturation Index (ODI)Superior or comparable accuracy to predicate.Mean difference with PSG: 0.658 (CI: -1.012, 2.326).Mean difference with PSG: 2.499 (CI: 0.732, 4.265).Polysomnography (PSG)
    N1/N2 Merged Sleep Stages AgreementHigh agreement with PSG; superior to predicate.80.92% agreement (95% CI: 79.96, 81.89)61.77% agreement (95% CI: 60.99, 62.55)Polysomnography (PSG) (consensus of three scoring sleep technologists)
    NREM/REM/Wake Stage Differential AgreementHigh agreement with PSG; superior to predicate.86.79% agreement (95% CI: 85.76, 87.81)74.42% agreement (95% CI: 73.51, 75.32)Polysomnography (PSG) (consensus of three scoring sleep technologists)
    Sleep/Wake Determination AgreementHigh agreement with PSG; superior to predicate.89.45% agreement (95% CI: 88.53, 90.38)82.09% agreement (95% CI: 81.37, 82.82)Polysomnography (PSG) (consensus of three scoring sleep technologists)
    SpO2 Accuracy (ARMS)Maintain or outperform predicate, satisfy ISO 80601-2-61.1.46%1.88% (Watch-PAT300)Gold standard SaO2 (invasive controlled desaturation study)
    Pulse Rate AccuracySatisfy ISO 80601-2-61.1.91 BPM (vs Masimo), 1.81 BPM (vs Nellcor)Not explicitly compared hereMasimo and Nellcor reference oximeters

    2. Sample Size and Data Provenance

    • Test Set Sample Size: 92 subjects for the comparative clinical study involving Somfit, Watch-PAT300, and PSG.
    • Data Provenance: The subjects were recruited from 3 site locations in Australia. The study was a prospective clinical study conducted to compare the devices' performance.

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: Three scoring sleep technologists.
    • Qualifications of Experts: The document states they were "scoring sleep technologists." Specific educational background or years of experience are not provided, but their role implies expertise in polysomnography scoring.

    4. Adjudication Method for the Test Set

    The adjudication method for sleep staging ground truth was expert consensus. The PSG epochs were classified based on the "consensus of three scoring sleep technologists." This implies a form of 3-way consensus. Further details on how discrepancies were resolved (e.g., majority rule, discussion to reach agreement) are not provided.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a traditional MRMC comparative effectiveness study was not explicitly described in the provided text in the context of human readers improving with AI vs. without AI assistance. The study evaluated the device's performance (Somfit and Watch-PAT300) against a human-scored gold standard (PSG), rather than measuring the improvement of human readers assisted by AI. The Somfit system is described as a diagnostic aid that calculates and reports parameters to clinicians, implying that clinicians review the output, but the study design was not focused on measuring changes in human reader performance.

    6. Standalone (Algorithm Only) Performance

    Yes, the study primarily evaluated the standalone performance of the Somfit algorithm (and Watch-PAT300) in deriving parameters like pAHI and sleep stages, comparing its output directly to the gold standard PSG and the predicate device. There was no human-in-the-loop component as part of the evaluated performance for regulatory submission.

    7. Type of Ground Truth Used

    The primary ground truth used for the clinical study was Polysomnography (PSG), which is considered the gold standard technology for diagnosing sleep disorders. Specifically, for sleep staging, it was expert consensus-scored PSG data. For SpO2, it was gold standard SaO2 data obtained from an invasively controlled desaturation study.

    8. Sample Size for the Training Set

    The document does not provide information on the sample size for the training set for the Somfit algorithms. It focuses solely on the clinical validation/test set.

    9. How the Ground Truth for the Training Set Was Established

    As the training set details are not provided, the method for establishing its ground truth is also not mentioned in the document.