Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K160641

    Validate with FDA (Live)

    Date Cleared
    2017-03-08

    (366 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Solitaire™ Platinum Revascularization Device is intended to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

    Device Description

    The subject 6-40-10 Solitaire™ Platinum Revascularization Device is designed to restore blood flow in patients experiencing ischemic stroke due to large intracranial vessel occlusion. The Solitaire™ Platinum Revascularization Device is designed for use in the neurovasculature such as the Internal Carotid Artery (ICA), M1 and M2 segments of the middle cerebral artery, basilar, and the vertebral arteries. The distal nitinol portion of the Solitaire™ Platinum Revascularization Device facilitates clot retrieval. The Solitaire™ Platinum Revascularization Device has Platinum Iridium radiopaque markers on the working cell length, proximal and distal ends. The subject Solitaire™ Platinum Revascularization Device is a portfolio expansion to the predicate Solitaire™ Platinum Revascularization Device (K153071).

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Device: Solitaire™ Platinum Revascularization Device (6x40 mm)


    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document describes both biocompatibility testing and non-clinical bench testing. There is also clinical data leveraged from a registry.

    Biocompatibility Acceptance Criteria & Performance:

    Test CategoryTest DescriptionMethodAcceptance CriteriaReported Device Performance (Conclusion)
    CytotoxicityL929 MTT CytotoxicityISO 10993-5Viability is $\geq$ 70%.Acceptance criteria met
    SensitizationGuinea Pig Maximization SensitizationISO 10993-10Test article does not elicit a sensitization response.Acceptance criteria met
    IrritationIntracutaneous Irritation TestISO 10993-10Differences in the mean test and control scores of the extract dermal observations are $<$ 1.0.Acceptance criteria met
    Acute Systemic ToxicityAcute Systemic Injection TestISO 10993-11No abnormal clinical signs and weight loss in excess of 10%. Temperature rise $\geq$ 0.5°CAcceptance criteria met
    Materials Mediated Rabbit Pyrogen(Not specified)(Not specified)Acceptance criteria met
    Hemo-compatibilityHemolysisISO 10993-4No significant differences between the test article extract and negative control article results. The test article is considered non-hemolyticAcceptance criteria met
    Partial Thromboplastin Time(Not specified)Clotting times are similar to the negative control and the reference material (HDPE) indicating the device materials are not an activator of the intrinsic coagulation pathway.Acceptance criteria met
    Platelet and Leukocyte Count(Not specified)Test article does not adversely affect the platelet and leukocyte components of the blood compared to the reference material.Acceptance criteria met
    Complement Activation C3a and SC5b-9 Assay(Not specified)Levels of C3a and SC5b-9 are comparable and less than the positive control.Acceptance criteria met
    Thrombosis(Not specified)Thrombo-resistance properties are acceptable in clinical use.Acceptance criteria met
    GenotoxicityBacterial Mutagenicity TestISO 10993-3Test article is considered non-mutagenicAcceptance criteria met
    In-vitro Mouse Lymphoma Assay(Not specified)Test article is considered non-mutagenicAcceptance criteria met
    In-vivo Mouse Micronucleus Assay(Not specified)Test article is considered non-mutagenicAcceptance criteria met

    Non-Clinical Bench Testing Acceptance Criteria & Performance:

    Test DescriptionMethodAcceptance CriteriaReported Device Performance (Conclusion)
    Total System LengthTotal system length was measured from the distal tip of the distal marker to the proximal tip of the delivery system.The stent length should be adequate for its intended use for delivery via standard microcatheters.Acceptance criteria met
    Marker-to-Marker LengthMarker-to-marker length is representative of the total stent length. Total stent length was measured 100% in process.The stent length should be adequate for its intended use for delivery via standard microcatheters.Acceptance criteria met
    Delivery ForceDelivery force was measured through a representative tortuous anatomical model.The stent must be below delivery force specification.Acceptance criteria met
    Resheathing ForceResheathing force was measured through a representative tortuous anatomical model.The stent must be below re-sheathing force specification.Acceptance criteria met
    Multiple Resheathing DurabilityMultiple resheathing durability was evaluated on the ability to withstand delivery and withdrawal forces in a representative tortuous model beyond the recommended number of passes and re-sheathings allowed per the Instructions for Use.The stent must be able to be deployed and resheathed up to four times.Acceptance criteria met
    Body Finger Marker Coil TensileBody finger marker coil tensile was performed to verify the strength of the laser weld of the Platinum/Iridium marker coil to the Nitinol body finger.The body finger marker coil tensile should be greater than or equal to existing tensile strength specification.Acceptance criteria met
    Radial ForceRadial force was measured 100% in process.The stent must be within existing radial force specification.Acceptance criteria met
    RadiopacityRadiopacity was verified by both a material and dimensional analyses.The radiopaque body markers must be visible using standard catheter laboratory equipment.Acceptance criteria met
    Ease of Use (Fluorosafe Marker Location)Ease of use (fluorosafe marker location) was measured from the distal tip of the distal marker to the distal end of the fluorosafe marker.Less required fluoroscopy is better than more required fluoroscopy.Acceptance criteria met
    ParticulateParticulate was evaluated for generation under simulated use in a representative tortuous anatomical model.The stent must not generate clinically unacceptable particulate.Acceptance criteria met

    Clinical Performance (Leveraged from STRATIS Registry 4x40 Subgroup):

    Confirmatory Clinical DataAcceptance Criteria (Implied)Reported Device Performance
    Safety Data:
    Symptomatic ICHAs low as reasonably achievable, comparable to predicate2.8% (7/252)
    Mortality at 90 daysAs low as reasonably achievable, comparable to predicate16.2% (48/296)
    Study Device-Related SAEsAs low as reasonably achievable, comparable to predicate0.3% (1/296)
    Performance Data:
    Successful revascularization (TICI 2b-3)High success rate, comparable to predicate90.4% (227/251)
    Functional Independence (mRS 0-2) at 90 daysHigh rate, comparable to predicate60.3% (167/277)

    2. Sample Size Used for the Test Set and Data Provenance

    • Non-Clinical Animal Testing: The study used an "in-vivo Yorkshire cross swine model." The number of animals is not explicitly stated, but it evaluated 4 out of 8 proposed target sites (Right Vertebral, Left Vertebral, Right Internal Thoracic, Left Internal Thoracic, Right Costocervical, Left Costocervical, Renal, Ascending Pharyngeal).
      • Provenance: Prospective, animal model.
    • Confirmatory Clinical Data (Leveraged from STRATIS Registry):
      • Sample Size: A total of 984 patients were enrolled in the STRATIS Registry. From this, the analysis for the 4x40 Solitaire devices (predicate and reference) included 296 patients who were treated only with these devices.
      • Provenance: Prospective, multi-center, observational registry conducted in the United States (55 sites).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Animal Testing: The document does not specify the number of experts or their qualifications for evaluating the animal testing results (e.g., Luminal Thrombus, Endothelial Cell Coverage, etc.).
    • Clinical Data (STRATIS Registry):
      • Clinical Events Committee: An independent Clinical Events Committee was used for adverse event adjudication. The number of experts or their specific qualifications (e.g., type of physician, years of experience) are not detailed.
      • Core Lab: An independent Core Lab was used for procedural image assessment. The number of experts or their specific qualifications are not detailed.

    4. Adjudication Method for the Test Set

    • Clinical Data (STRATIS Registry): The document states that an "independent Clinical Events Committee for adverse event adjudication and an independent Core Lab for procedural image assessment" were used. This indicates an independent adjudication process, but the specific method (e.g., 2+1, 3+1 consensus) is not detailed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, an MRMC comparative effectiveness study was not done as described for AI assistance. The clinical data leveraged was from an observational registry evaluating clinical outcomes of the device in humans. This study was not comparing human readers with and without AI assistance; rather, it was assessing the device's performance in a real-world clinical setting.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • Not applicable. This document describes a medical device (revascularization device), not an algorithm or AI software. Therefore, there is no "standalone" algorithm-only performance to report. The performance discussed for the clinical data is the device's performance in situ within a human patient, with clinicians performing the procedure.

    7. The Type of Ground Truth Used

    • Biocompatibility: In vitro and in vivo laboratory test results (e.g., cell viability, immune responses, clotting times, mutagenicity).
    • Bench Testing: Engineering measurements and evaluations against design specifications.
    • Animal Testing: Histopathological and physiological observations in an animal model (e.g., luminal thrombus, endothelial cell coverage, inflammation).
    • Clinical Data (STRATIS Registry):
      • Adverse Events: Adjudicated by an independent Clinical Events Committee, likely based on medical records, clinical observations, and definitions of adverse events.
      • Procedural Image Assessment: Adjudicated by an independent Core Lab, likely measuring revascularization success (TICI scores) from imaging data.
      • Functional Independence (mRS 0-2): Patient outcomes measured by modified Rankin Scale (mRS) at 90 days.

    8. The Sample Size for the Training Set

    • Not applicable. For this type of medical device (physical revascularization device), there is no "training set" in the context of an AI/ML algorithm. The "training set" concept is relevant for data-driven models that learn from data.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As there is no training set for an AI/ML algorithm, this question is not relevant to the provided document.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1