Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K153537

    Validate with FDA (Live)

    Date Cleared
    2016-06-17

    (190 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sol-M TB Syringe is intended to be used for medical purposes to inject fluids into or withdraw fluids from the body.
    The Sol-M Insulin Syringe with needle, with the calibration unit of insulin for U-100, is a device intended for medical purposes for the manual aspiration of insulin, and for the injection of insulin into parts of the body below the surface skin.

    Device Description

    The Sol-M Syringe is a sterile, single use, standard 3 piece piston syringe with a fixed needle. The Sol-M 0.3, 0.5 and 1 ml Insulin Syringe will be labeled with a Units Scale calibrated for U-100 Insulin. The 0.5 and 1 ml TB Syringe will be labeled with a mL Scale. The piston type syringe is a plastic disposable syringe made of the following components: 1. Barrel – The barrel has a scale showing the capacity of the syringe as well as the company brand name with the Do Not Reuse symbol. 2. Plunger - The plunger is used to aspirate fluid into the barrel or push fluid out of the barrel. 3. Stopper - The Stopper maintains the fluid in the barrel between the base of the barrel and Plunger. 4. Cap The Cap is used to cover and protect the tip of the barrel from being damaged.

    AI/ML Overview

    This is a 510(k) premarket notification for a medical device (syringes), not an AI/ML device. Therefore, the requested information about acceptance criteria, study details, sample sizes, ground truth establishment, expert involvement, MRMC studies, and standalone performance is generally not applicable in the context of AI/ML device evaluations. Traditional medical device clearances like this one focus on demonstrating substantial equivalence to a legally marketed predicate device through non-clinical bench testing.

    Based on the provided document, here's what can be extracted related to performance and testing, framed in the closest possible way to your request, but acknowledging the difference in context:

    1. A table of acceptance criteria and the reported device performance

    For the Sol-M Insulin Syringes:

    Acceptance Criteria CategoryStandard/RequirementReported Device Performance (Compliance)
    Delivery AccuracyISO 8537:2007 Sterile single use syringes, with and without needles, for insulinComplies with ISO 8537:2007
    Stainless Steel Needle TubingISO 9626:1991 Stainless steel needle tubing for the manufacture of medical devices/Amendment:2001Complies with ISO 9626:1991

    For the Sol-M TB Syringes:

    Acceptance Criteria CategoryStandard/RequirementReported Device Performance (Compliance)
    Manual Use SyringesISO 7886-1:1993 Sterile Hypodermic syringes for single use Part 1: Syringes for manual useComplies with ISO 7886-1:1993
    Hub/Needle Bond StrengthPer ISO 7886 (likely referring to the relevant section within 7886-1 or similar)Complies with ISO 7886
    Stainless Steel Needle TubingISO 9626:1991 Stainless steel needle tubing for the manufacture of medical devices/Amendment:2001Complies with ISO 9626:1991

    For both devices (general characteristics):

    Acceptance Criteria CategoryStandard/RequirementReported Device Performance (Compliance)
    Lubricant amount/cm2Predicate: 1 ml – 0.078 Mg/cm2Subject device: 1ml-0.078mg/cm2 (Matches predicate)
    BiocompatibilityPer ISO 10993-1Complies with ISO 10993-1
    LabelingPer 21 CFR 801Complies with 21 CFR 801
    MaterialsBarrel, Plunger, Cap: Polypropylene; Gasket: SantopreneMatches predicate
    Gradations LegibilityBold markingsBold markings (Matches predicate)
    Barrel TransparencyClearClear (Matches predicate)

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document mentions "Bench tests were conducted to verify that the proposed devices met all design specifications." However, it does not specify the sample sizes for these bench tests. The data provenance is not explicitly mentioned as country of origin, but these are non-clinical tests performed to regulatory standards. The testing is prospective in the sense that new samples of the device were tested against established standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable as this is a traditional medical device (syringes) and not an AI/ML device relying on expert-annotated ground truth for performance evaluation in the way an AI would. Performance is measured against engineering and material standards (e.g., ISO standards).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable in the context of traditional device bench testing against ISO standards. Test results are typically objective measurements.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable, as this is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable, as this is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance is established by international and national standards for medical syringes (e.g., ISO 8537, ISO 7886-1, ISO 9626, ISO 10993-1, 21 CFR 801). The device's physical and functional characteristics are objectively measured against these established parameters.

    8. The sample size for the training set

    Not applicable, as this is not an AI/ML device and does not involve training data.

    9. How the ground truth for the training set was established

    Not applicable, as this is not an AI/ML device and does not involve a training set or its ground truth.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1