Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K233600

    Validate with FDA (Live)

    Date Cleared
    2024-02-05

    (88 days)

    Product Code
    Regulation Number
    892.1000
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Smart Fit Knee 3.0T coil is designed to be used in conjunction with a Philips 3.0T MR system to produce diagnostic images of the Knee that can be interpreted by a trained physician.

    Device Description

    The Smart Fit Knee 3.0T coil is designed to be used in conjunction with a Philips 3.0T MR system to produce diagnostic images of the Knee that can be interpreted by a trained physician.
    The Smart Fit Knee 3.0T is a 16-element coil designed for high-resolution imaging of the left or right knee. It is a phased-array, Rx volume coil providing an integrated solution with a base plate, an anterior and a posterior part. Positioning pads are also supplied to support comfortable positioning. The coil can be slightly rotated relative to its base plate to ease coil setup and enhance patient comfort. The coil is used independently and cannot be combined with any other coils. This coil is available for 3.0T MR systems and is compatible with Philips 3.0T MR Scanners.

    AI/ML Overview

    The provided text describes the Philips Smart Fit Knee 3.0T MRI coil, but it does not contain acceptance criteria for device performance or a detailed study proving the device meets specific performance criteria through clinical effectiveness measures.

    Instead, the document focuses on demonstrating substantial equivalence to a predicate device (HRK-127-8 KNEE ARRAY COIL K033567) for regulatory clearance. This is primarily done through technological comparisons and compliance with recognized consensus standards.

    Here's a breakdown of the information available and what is missing based on your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    • Acceptance Criteria: Not explicitly stated as quantifiable performance metrics for clinical effectiveness. The acceptance criteria in this document are implicitly related to compliance with recognized standards and showing similarity to the predicate device in design, intended use, and fundamental scientific technology.
    • Reported Device Performance:
      • "The performance test results demonstrate that the proposed Smart Fit Knee 3.0T meets the acceptance criteria and is adequate for its intended use." This is a general statement of compliance, not a report of specific numerical performance metrics.
      • "All clinical images on the proposed coils Smart Fit Knee 3.0T were evaluated by qualified radiologists. No issues with the clinical image quality were seen and images were considered have sufficient quality for diagnostic use." This broadly indicates acceptable image quality based on expert review, but no quantifiable performance metrics (e.g., sensitivity, specificity, SNR values) are provided.

    Table based on available information (focused on equivalence and general performance):

    Acceptance Criteria (Implicit)Reported Device Performance
    Compliance with relevant IEC, ISO, AAMI, NEMA standards.Meets ANSI AAMI ES60601-1, IEC60601-2-33, IEC60601-1-2, IEC60601-1-6, ISO 14971, IEC 62366-1, ANSI AAMI ISO10993-1, NEMA MS 1, NEMA MS 3, NEMA MS 9, NEMA MS 14 (list of standards provided in the Summary of Non-Clinical Performance Data).
    Image quality sufficient for diagnostic use.Evaluated by qualified radiologists; no issues with clinical image quality were seen; images considered to have sufficient quality for diagnostic use.
    Biocompatibility with intact human skin exposure < 24h.Biocompatibility testing performed against ISO10993-1; materials (PC and PU) safety proved in biocompatibility report.
    Similar fundamental scientific technology to predicate device.Phased-array Receive only coil, Head-Feet oriented (B0) magnetic field, 127.728MHz+/-0.75MHz frequency range, Overlap/pre-amp decouple/Active/Passive PIN diode decoupling, derived from MR scanner energy source.
    No new risks or sufficiently mitigated risks.Risk management activities show all risks are sufficiently mitigated, no new risks introduced, and overall residual risks are acceptable.

    2. Sample size used for the test set and the data provenance:

    • The document states that "All clinical images on the proposed coils Smart Fit Knee 3.0T were evaluated," but it does not specify the sample size (number of patients or knee images) for this clinical image quality evaluation test set.
    • Data Provenance: Not explicitly stated. Given that Philips Healthcare (Suzhou) Co., Ltd. is based in China, and the contacts appear to be from China and the Netherlands, the images could originate from various countries where Philips systems are used. The document does not specify if the data was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of experts: Not specified beyond "qualified radiologists."
    • Qualifications of experts: Only stated as "qualified radiologists." No specific details such as years of experience, subspecialty (e.g., musculoskeletal radiologist), or board certification are provided.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • The document mentions that clinical images were "evaluated by qualified radiologists." It does not specify an adjudication method (e.g., how disagreements were resolved if multiple radiologists reviewed the same images). It might imply a consensus reading or individual review without further adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC comparative effectiveness study was done. The document explicitly states: "The proposed Smart Fit Knee 3.0T did not require clinical study since substantial equivalence to the legally marketed predicate device was demonstrated in comparison."
    • This device is an MRI coil, a hardware component, not an AI-powered diagnostic algorithm. Therefore, "human readers improve with AI vs without AI assistance" is not applicable.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is an MRI coil, not an algorithm. Performance is assessed on its ability to produce diagnostic images, which are then interpreted by humans.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For image quality, the ground truth was expert opinion/review by qualified radiologists. Their determination that "No issues with the clinical image quality were seen and images were considered have sufficient quality for diagnostic use" served as the basis for performance.

    8. The sample size for the training set:

    • Not applicable. This device is a passive MRI coil (hardware). It does not involve machine learning algorithms that require a training set.

    9. How the ground truth for the training set was established:

    • Not applicable, as there is no training set for this hardware device.

    In summary, the provided document is a 510(k) summary focused on demonstrating substantial equivalence of a new MRI coil to a predicate device, primarily through non-clinical testing for compliance with standards and a general statement about image quality review by radiologists, rather than a clinical study with detailed performance metrics and ground truth establishment as would be expected for an AI/software device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1