Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K201504

    Validate with FDA (Live)

    Date Cleared
    2021-05-10

    (339 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Single-use medical poly (vinyl chloride) examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    The proposed device is Powder Free Vinyl Patient Examination Gloves which is made of poly vinyl chloride. The proposed device is clear, non-colored. The design of proposed device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D5250. The proposed device is non-sterile.

    AI/ML Overview

    This document is a 510(k) Pre-Market Notification for a medical device: "Single-use medical poly (vinyl chloride) examination glove." It's primarily a comparison to a predicate device to establish substantial equivalence, not a study proving device meeting acceptance criteria in the context of advanced AI/medical imaging devices. Therefore, many of the requested categories are not applicable.

    Here's an analysis based on the provided text, focusing on the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally based on meeting the specifications of various ASTM and ISO standards for medical examination gloves. The reported device performance indicates that the proposed device meets these criteria.

    Test ItemAcceptance Criteria (Pass Criteria)Reported Device Performance (Verdict)
    Dimension (mm)ASTM D5250Pass
    Overall LengthMin 230>230/ Pass
    WidthXS: 76±6; S: 85±5; M: 95±5; L: 105±5; XL: 115±5XS: 79-81/ Pass; S: 85-87 /Pass; M: 95-97/ Pass; L: 105-107/ Pass; XL: 115-117/ Pass
    Finger thickness0.050.09-0.12/ Pass
    Palm thickness0.080.08-0.09/ Pass
    Freedom from HolesMeet the requirements of ASTM D5151 AQL 1.50/125, 1/ 125, 4/315, 5/500, 3/200 leaks / Pass (Implies AQL 1.5 met)
    Powder ContentMeet the requirements of ASTM D6124 < 2.0mg0.3 – 0.42/ Pass
    Physical PropertiesASTM D412Pass
    Tensile Strength (MPa)Meet the requirements of ASTM D412 Min 1114-19/ Pass
    Ultimate Elongation (%)Meet the requirements of ASTM D412 Min 300%390 – 430/ Pass
    Biocompatibility
    IrritationNon-irritating (ISO 10993-10)Under the conditions of the study, not an irritant/ Pass
    SensitizationNon-sensitizing (ISO 10993-10)Under conditions of the study, not a sensitizer./ Pass
    CytotoxicityNon-cytotoxic (ISO 10993-5)Under conditions of the study, did not show potential toxicity to L-929 cells./ Pass

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document provides sample sizes for the "Freedom from Holes" test (e.g., 0/125, 1/125, 4/315, 5/500, 3/200 leaks). These numbers likely refer to the number of failed units out of the total tested for specific batches or sample groups to demonstrate compliance with AQL (Acceptable Quality Level) 1.5. For other tests like Dimensions, Powder Content, Physical Properties, and Biocompatibility, specific sample numbers are not explicitly stated, but are implied to be sufficient for compliance with the stated ASTM and ISO standards.
    • Data Provenance: The document does not specify the country of origin of the data. Given the applicant is Anhui Bytech Medical Supplies Co., Ltd. in China and the submission correspondent is Shanghai SUNGO Management Consulting Co., Ltd. also in China, it's highly probable the testing was conducted in China. The data is based on non-clinical performance testing, effectively a prospective quality control assessment of manufactured gloves against established standards.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    • This section is not applicable to this type of device submission. The "ground truth" for these tests are objective measurements against established technical standards (ASTM, ISO), not expert consensus on medical interpretation.

    4. Adjudication Method for the Test Set

    • This section is not applicable. The tests performed are objective measurements against defined criteria in standards, not subjective assessments requiring adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging devices where human readers interpret results with and without AI assistance. This document describes a physical medical device (examination glove).

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is not an AI algorithm. The device is a physical glove. Performance is measured directly from the physical characteristics of the glove.

    7. The Type of Ground Truth Used

    • The "ground truth" for the non-clinical performance and biocompatibility tests are the specifications and acceptable limits defined within the referenced international standards (ASTM D5250, ASTM D5151, ASTM D6124, ASTM D412, ISO 10993-10, ISO 10993-5). This is an objective, standardized basis for evaluation.

    8. The Sample Size for the Training Set

    • This concept is not applicable here. There is no "training set" in the context of an AI/machine learning model. The device's manufacturing process is designed to produce gloves meeting specifications, not "trained" on data.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable, as there is no training set for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K182043

    Validate with FDA (Live)

    Date Cleared
    2018-09-21

    (53 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Single-use medical poly (vinyl chloride) examination glove (Clear, Non-Colored) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    The proposed device is Powder Free Vinyl Patient Examination Gloves. The proposed device is clear, non-colored. The design of proposed device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D5250. The proposed device is non-sterile.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device: "Single-use medical poly (vinyl chloride) examination glove (Clear, Non-Colored)". This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving novel clinical benefit. Therefore, the information provided relates to performance against recognized standards and comparison to predicates, not a traditional clinical study with human readers or training sets as might be seen for AI algorithms.

    Here's an analysis based on the information provided:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document presents performance information in Tables 2, 3, and 4 (starting on page 6), comparing the proposed device to predicate devices and referencing ASTM and ISO standards for acceptance criteria.

    Acceptance Criterion (Standard Ref.)Proposed Device Performance (K182043)Predicate Device (K163168) PerformancePredicate Device (K180849) Performance
    Dimensions (Table 2)
    Length, mm (min)240230 (XS, S), 235 (M), 245 (L, XL)230 (S), 235 (M), 245 (L, XL)
    Width, mm ($\pm$5)80 (XS), 85 (S), 95 (M), 105 (L), 115 (XL)80 (XS), 85 (S), 95 (M), 105 (L), 115 (XL)85 (S), 95 (M), 105 (L), 115 (XL)
    Finger Thickness, mm (min)0.070.050.05
    Palm Thickness, mm (min)0.080.080.08
    Physical Properties (Table 3)
    Before Aging (ASTM D5250)
    Tensile Strength, MPa (min)151515
    Ultimate Elongation, % (min)350380380
    After Aging (ASTM D5250)
    Tensile Strength, MPa (min)151515
    Ultimate Elongation, % (min)350380380
    Freedom from Holes (ASTM D5151)Be free from holes, AQL=1.5Be free from holes, AQL=1.5Be free from holes, AQL=1.5
    Powder Content (ASTM D6124)0.33 mgMeet ASTM 5250 requirementsMeet ASTM 5250 requirements
    Biocompatibility (Table 4)
    Irritation (ISO 10993-10)Not an irritantNot an irritant / Comply with ISO 10993-10Comply with ISO 10993-10
    Sensitization (ISO 10993-10)Not a sensitizerNot a sensitizer / Comply with ISO 10993-10Comply with ISO 10993-10
    Cytotoxicity (ISO 10993-5)Did not show potential toxicity to L-929 cells/Comply with ISO 10993-5
    Label and LabelingMeet FDA's RequirementsMeet FDA's RequirementsMeet FDA's Requirements

    Differences noted and addressed:

    • Length (Analysis 1): The proposed device has a different length than the predicate, but test results according to ASTM 5250 showed it meets standard requirements.
    • Colorant (Analysis 2): The proposed device is clear/non-colored, while one predicate is white, blue, or yellow. The submission argues that color difference does not affect safety and effectiveness, supported by biocompatibility testing (ISO 10993-5 and ISO 10993-10).
    • Ultimate Elongation (Analysis 3): The proposed device has a different ultimate elongation value than the predicate, but test results according to ASTM 5250 showed it meets standard requirements.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document does not explicitly state the sample size for each specific test (e.g., number of gloves tested for tensile strength, freedom from holes, etc.). However, it refers to standards such as ISO 2859-1:1999 ("Sampling Procedures for Inspection by Attributes - Part I: Sampling Plans Indexed by Acceptable Quality Level (AQL) for Lot-by-Lot Inspection"), which would dictate the appropriate sample sizes for quality control testing based on lot size and AQL levels. For example, for "Freedom from Holes," an AQL of 1.5 is specified, which implies a specific sampling plan from ISO 2859-1.
    • Data Provenance: The tests were conducted by the "HEBEI TITANS HONGSEN MEDICAL TECHNOLOGY CO., LTD." in China, as indicated by the sponsor's identification. These are non-clinical (bench) tests supporting the device's conformance to established industry standards. The data is thus generated internally by the manufacturer for the purpose of this submission.
    • Retrospective or Prospective: These are laboratory/bench tests that are inherently "prospective" in the sense that they are performed on manufactured product samples to demonstrate compliance with specifications.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable to this type of device submission. The "ground truth" for medical gloves is established by predefined engineering specifications and performance criteria codified in international standards (e.g., ASTM D5250, ISO 10993 series). The tests are evaluated against these objective numerical or qualitative criteria, not by expert consensus in a clinical context.

    4. Adjudication method for the test set:

    Not applicable. This device is evaluated against objective technical standards (ASTM, ISO). There is no "adjudication" of expert opinions for a test set in this context.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This submission is for single-use medical examination gloves, which are not AI-enabled devices and do not involve human readers interpreting diagnostic images.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is not an AI algorithm. The device's performance is standalone in the sense that it meets physical and biocompatibility specifications on its own, without requiring human interpretation for its intended function.

    7. The type of ground truth used:

    The "ground truth" for this device's performance is defined by the objective, measurable parameters and criteria stipulated in the referenced international standards:

    • ASTM D6124 (Residual Powder)
    • ASTM D5151 (Detection of Holes)
    • ASTM D5250 (Poly(vinyl chloride) Gloves for Medical Application)
    • ISO 10993-10 (Irritation and Skin Sensitization)
    • ISO 10993-5 (In Vitro Cytotoxicity)
    • ISO 2859-1 (Sampling Procedures)

    These standards specify the methods for testing and the acceptable limits or qualities (e.g., minimum tensile strength, maximum powder content, AQL for freedom from holes, non-irritant/non-sensitizer for biocompatibility).

    8. The sample size for the training set:

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable. This is not a machine learning or AI device.