Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K221255

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-08-31

    (121 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SimpliShade Bulk Fill is indicated for:

    • Direct placement of anterior and posterior restorations.
    • Core build-ups.
    • Repair of defects in restorations.

    The SimpliShade Bulk Fill Flow is indicated for:

    • Base/liner under Class I and Class II direct restorations.
    • Pit and fissure sealant.
    • Restoration of small cavities.
    • Class III and V restorations.
    • Blocking out undercuts.
    • Small core build-ups.
    • Repair of resin and acrylic provisional restorations.
    Device Description

    SimpliShade™ Bulk Fill Composite is offered in two types of viscosities: the high viscosity formula will be branded as SimpliShade™ Bulk Fill; the low viscosity formula will be branded as SimpliShade™ Bulk Fill Flow.

    SimpliShade Bulk Fill is an intra-oral, light-cured, esthetic, resin-based nanohybrid dental restorative that is designed for the placement of direct restorations. SimpliShade Bulk Fill is designed to be used after placement of a dental adhesive, such as a product from the Kerr OptiBond™ family. SimpliShade Bulk Fill is available in Unidose™ compules and syringe delivery systems that are suitable for occlusal surfaces. The product offers allows a technique in which a cavity up to 5 mm in depth can be filled and cured in a single increment. SimpliShade Bulk Fill is offered in one universal shade to optically blend with a range of VITA™ shade guide equivalents. This composite can be polished with a wide variety of polishing discs, rubber cups, and brushes.

    SimpliShade Bulk Fill Flow is an intra-oral, light-cured, esthetic, resin-based nanohybrid dental restorative that is designed to be used as a base or liner in Class I and II restorations and must be overlayed with a composite resin to replace the occlusal surfaces. It is also suitable to be used on nonocclusal-contact surfaces, such as pit and fissures, core build-ups and Class III and V restorations. SimpliShade Bulk Fill Flow is available in Unidose™ compules and syringe delivery systems. The product allows a technique in which a cavity up to 4 mm in depth can be filled and cured in a single increment. SimpliShade Bulk Fill Flow is offered in one universal shade to optically blend with a range of VITAM shade guide equivalents. Once light cured, the composite can be finished and polished using conventional finishing and polishing instruments.

    AI/ML Overview

    The provided text is a 510(k) Summary for dental resin materials. It describes the acceptance criteria and study proving the device meets these criteria, but it focuses on non-clinical performance and does not include information about clinical studies with human participants or AI-assisted performance analysis. Therefore, I cannot generate information related to multi-reader multi-case studies, effect size of AI assistance, or standalone algorithm performance, as these are not present in the document.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are primarily defined by adherence to international standards and internal benchmarks. The reported device performance indicates that the subject devices (SimpliShade™ Bulk Fill and SimpliShade™ Bulk Fill Flow) met these criteria.

    Acceptance CritieriaReported Device Performance
    For both SimpliShade™ Bulk Fill & SimpliShade™ Bulk Fill Flow:
    Curing Intensity (mW/cm²)≥ 650 mW/cm² (Similar to Predicate Devices which were ≥ 550 mW/cm²)
    Wavelength (nm) for curing400 - 520 nm (Similar to Predicate Devices which were 400 - 500 nm)
    Curing Time (sec)20 sec (≥ 650 mW/cm²) (Similar to Predicate Devices which were ≥ 20 sec at ≥ 550 mW/cm²)
    Compressive Strength (MPa)Meet the criteria per internal standard (>260 Mpa)
    Rockwell Hardness (RHN)Meet the criteria per internal standard (>65 RHN)
    Flexural Strength (Mpa)Conform to ISO 4049 requirement
    Radio-opacity (mm of AI)Conform to ISO 4049 requirement
    Water Sorption (µg/mm³)Conform to ISO 4049 requirement
    Water Solubility (µg/mm³)Conform to ISO 4049 requirement
    Sensitivity to ambient lightConform to ISO 4049 requirement (Same as Predicate Devices)
    Color StabilityConform to ISO 4049 requirement (Same as Predicate Devices)
    Biocompatibility (ISO standards)Conform to ISO 10993-1, -3, -5, -6, -10, -11, -12, -23 and ISO 7405. (Similar to Predicate devices)
    Dentistry - Polymer-based restorative materials (ISO standard)Conform to ISO 4049:2019 (Similar to Predicate devices which were ISO 4049:2009)
    Specific to SimpliShade™ Bulk Fill:
    Depth of Cure (mm)Up to 5mm (Similar to Predicate Device)
    Recommended for load-bearing occlusal surfacesYes (Same as Predicate Device)
    Specific to SimpliShade™ Bulk Fill Flow:
    Depth of Cure (mm)Up to 4mm (Similar to Predicate Device)
    Recommended for load-bearing occlusal surfacesNo (Same as Predicate Device)

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes (number of specimens tested) for each non-clinical test. It states that "Performance bench tests according to ISO 4049...have been conducted" and "Additional tests...have been performed." The data provenance is non-clinical laboratory testing, conducted by the manufacturer, Kerr Corporation, to demonstrate substantial equivalence to legally marketed predicate devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a non-clinical study involving material properties, not diagnostic interpretation by human experts.

    4. Adjudication method for the test set

    Not applicable. Ground truth for material properties is established through standardized testing procedures, not by human adjudication of observations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This document describes the non-clinical performance of dental restorative materials. It does not involve AI, human readers, or clinical effectiveness studies in this context. The section "Clinical Performance Data" explicitly states: "Clinical data is not needed to characterize performance and establish substantial equivalence. The nonclinical test data characterize all performance aspects of the device based on well-established scientific and engineering principles. Clinical testing has not been conducted on the subject devices."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No. This document does not describe AI algorithm performance.

    7. The type of ground truth used

    The ground truth is established through physical and chemical property measurements according to established international standards (e.g., ISO 4049, ISO 10993 series) and internal company standards. This involves objective measurements rather than expert consensus, pathology, or outcomes data in a clinical sense.

    8. The sample size for the training set

    Not applicable. There is no mention of a training set as this is a non-clinical evaluation of material properties, not a machine learning model.

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set for a machine learning model, there is no ground truth established in that context.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1