Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K190219

    Validate with FDA (Live)

    Date Cleared
    2019-05-13

    (98 days)

    Product Code
    Regulation Number
    866.3970
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DiaSorin Molecular Simplexa™ VZV Direct assay is intended for use on the LIAISON® MDX instrument for the qualitative detection of varicella-zoster virus (VZV) DNA in cerebrospinal fluid (CSF) from patients signs and/or symptoms of meningitis and/or encephalitis. This test is intended as an aid in the diagnosis of VZV infections of the central nervous system (CNS).

    Negative results do not preclude VZV infection and should not be used as the sole basis for treatment or other patient management decisions.

    The assay is not intended for use as a donor screening test. The assay is for professional use only.

    The Simplexa™ VZV Positive Control Pack is intended to be used as a control with the Simplexa™ VZV Direct kit.

    This control is not intended for use with other assays or systems.

    Device Description

    The Simplexa™ VZV Direct assay is a real-time polymerase chain reaction (PCR) system that enables the direct amplification and detection of VZV DNA from unprocessed cerebral spinal fluid (CSF) specimens without nucleic acid extraction. The system consists of the Simplexa™ VZV Direct assay, the LIAISON® MDX (with LIAISON® MDX Studio Software), the Direct Amplification Disc and associated accessories. In the Simplexa™ VZV Direct assay, fluorescent probes are used together with corresponding forward and reverse primers to amplify VZV and internal control targets. A well-conserved region of the VZV DNA polymerase gene is targeted to identify VZV DNA in the specimen. An internal control is used to detect PCR failure and/or inhibition.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance for Simplexa™ VZV Direct Assay

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly list "acceptance criteria" as a set of predefined thresholds. Instead, it presents performance metrics from various studies. Based on the provided clinical agreement study, the implicit acceptance criteria for clinical performance would likely be a high Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA). For analytical performance, the LoD is a key metric.

    Metric (Implicit Acceptance Criteria)Reported Device Performance (Simplexa™ VZV Direct)
    Clinical Performance
    Positive Percent Agreement (PPA)100.0% (12/12) (95% CI: 75.7% to 100.0%)
    Negative Percent Agreement (NPA)99.7% (623/625) (95% CI: 98.8% to 99.9%)
    PPA (Contrived Samples)100.0% (120/120) (95% CI: 96.9-100.0%)
    Analytical Performance
    Limit of Detection (LoD) VZV Strain 99392.03 TCID50/mL (1,614 copies/mL)
    Limit of Detection (LoD) VZV Strain Ellen0.001 TCID50/mL (1,505 copies/mL)
    Analytical Reactivity (VZV strains)100.0% agreement for 5 additional VZV strains
    Cross-Reactivity (159 microorganisms)0.0% detection (no cross-reactivity observed)
    Reproducibility (VZV) %CV0.5-4.6%
    InterferenceNo interference observed
    Inhibition by other microorganismsNo inhibition observed

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Clinical):
      • Prospective Samples: 637 clinical samples.
      • Contrived Samples: 240 contrived VZV positive samples.
    • Data Provenance:
      • Country of Origin: Not explicitly stated, but the submission is to the U.S. FDA by a company in Cypress, California, USA, and the testing was performed at "testing sites" and "DiaSorin Molecular, Cypress, CA", suggesting U.S.-based data.
      • Retrospective or Prospective: The study included both:
        • Prospective and prospectively banked collections from eight (8) collection sites (February 2018 to November 2018).
        • Contrived samples were used to supplement low prevalence true positive clinical samples.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not applicable. This device is an in vitro diagnostic (IVD) based on molecular detection. Ground truth for diagnostic assays like this is typically established by comparative methods rather than expert human interpretation of images or other subjective data. No human "experts" were used to establish the ground truth in the sense of clinical decision-making from images.

    4. Adjudication Method for the Test Set

    Not applicable in the sense of expert human review (e.g., 2+1, 3+1). The ground truth for the clinical samples was established using a composite reference method:

    • Two (2) validated real-time PCR assays.
    • Followed by confirmation of positive PCR amplification products with bi-directional sequencing.
    • Decision Rule: Samples were characterized as positive if one (1) or both PCR assays were positive AND confirmed by bi-directional sequencing. Samples were characterized as negative if both PCR assays were negative.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. This device is an in vitro diagnostic (IVD) designed for laboratory use to detect VZV DNA. Its performance is evaluated against reference methods, not against human readers (e.g., radiologists, pathologists). Therefore, improvement for human readers with AI assistance is not applicable in this context.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance)

    Yes, the primary performance studies presented focus on the standalone performance of the Simplexa™ VZV Direct assay. The "Clinical Agreement" section directly compares the Simplexa™ VZV Direct results to the composite reference method without explicit human intervention in the interpretation of the Simplexa™ VZV Direct results. The assay is an automated real-time PCR system.

    7. Type of Ground Truth Used

    The ground truth for the clinical agreement study was established using a composite reference method. This method involved:

    • Two (2) validated real-time PCR assays.
    • Confirmation of positive PCR amplification products with bi-directional sequencing.

    8. Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of machine learning or AI models. This device is an in vitro diagnostic (IVD) based on real-time PCR technology, which relies on molecular biology principles (primers and probes) rather than machine learning models that require distinct training and test sets. The assay design and optimization would involve internal development and validation, but not typically a "training set" in the AI sense.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as a "training set" for an AI model is not described or relevant for this type of molecular diagnostic device. The analytical characteristics and design of the PCR assay (e.g., primer and probe specificity) are established through various analytical studies (e.g., analytical sensitivity, specificity, cross-reactivity) rather than a ground-truthed training set for machine learning.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1