510(k) Data Aggregation

    K Number
    K143651

    Validate with FDA (Live)

    Date Cleared
    2015-03-18

    (85 days)

    Product Code
    Regulation Number
    866.2680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Focus Diagnostics Simplexa™ Group A Strep Direct assay is intended for use on the 3M Integrated Cycler for the in vitro qualitative detection of Group A Streptococcus (GAS) from throat swabs collected from human patients with signs and symptoms of pharyngitis, such as sore throat. This test is intended for use as an aid in the diagnosis of GAS infection. The assay is intended for use in hospital, reference, or state laboratory settings. The device is not intended for point-of-care use.

    Device Description

    The Simplexa™ Group A Strep Direct assay system is a real-time PCR system that enables the direct amplification and qualitative detection of Group A Strep bacterial DNA from throat swabs that have not undergone a nucleic acid extraction. The system consists of the Simplexa™ Group A Strep Direct assay, the 3M Integrated Cycler (with Integrated Cycler Studio Software), the Direct Amplification Disc (DAD) and associated accessories.

    In the Simplexa™ Group A Strep Direct assay, bi-functional fluorescent probe-primers are used together with corresponding reverse primers to amplify Group A Strep bacterial DNA and the Internal Control (DNA IC). The assay targets a conserved region of Group A Strep (pyroqenic exotoxin B gene) to identify this bacteria in the specimen. The DNA IC is used to detect PCR failure and/or inhibition.

    AI/ML Overview

    Here is a summary of the acceptance criteria and study information for the Simplexa™ Group A Strep Direct device, based on the provided text:


    Acceptance Criteria and Device Performance for Simplexa™ Group A Strep Direct

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria for the clinical study's sensitivity and specificity. However, the performance characteristics of the device and its predicate device are compared. For the purpose of this response, we will consider the reported "Performance Characteristics" of the predicate device as a benchmark or implicit acceptance criteria, though the device has slightly lower specificity than the predicate.

    MetricPredicate Device Performance / Benchmark (Simplexa™ Group A Strep Direct K133883)Simplexa™ Group A Strep Direct Performance (Clinical Prospective Study)
    Sensitivity96.5% (95% CI: 91.3% - 98.6%)97.4% (152/156) (95% CI: 93.6% to 99.0%)
    Specificity98.0% (95% CI: 97.0% - 98.6%)95.2% (1139/1196) (95% CI: 93.9% to 96.3%)
    Positive Predictive Value (PPV)Not explicitly stated for predicate in comparison table.72.7% (152/209) (95% CI: 66.3% to 78.3%)
    Negative Predictive Value (NPV)Not explicitly stated for predicate in comparison table.99.7% (1139/1143) (95% CI: 99.1% to 99.9%)

    Note: The device's sensitivity (97.4%) is numerically higher than the predicate's (96.5%), and its specificity (95.2%) is numerically lower than the predicate's (98.0%). The FDA's substantial equivalence determination implies these results were acceptable.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 1352 evaluable samples.
    • Data Provenance: Prospectively collected from four geographically diverse sites. The country of origin is not explicitly stated but can be inferred to be the USA given the FDA submission document.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not mention the use of "experts" in the traditional sense (e.g., radiologists) for establishing ground truth. Instead, for the clinical prospective study, the comparator culture method was used as the primary reference for ground truth, and discrepant results were further analyzed.

    4. Adjudication Method for the Test Set

    The primary comparison was against a "comparator culture method" performed at one central laboratory.
    For discrepant analysis, a "validated bidirectional sequencing assay" was performed.

    • 46 out of 57 samples where Simplexa™ was Positive and Culture was Not Detected were confirmed as Group A Strep Positive by sequencing.
    • 2 out of 4 samples where Simplexa™ was Not Detected and Culture was Detected were confirmed as Group A Strep Positive by sequencing.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This device is a molecular diagnostic assay (PCR-based system) for the qualitative detection of Group A Streptococcus, not an imaging device requiring human reader interpretation. Therefore, the concept of improving human readers with AI assistance is not applicable here.

    6. Standalone Performance Study

    Yes, a standalone performance study (algorithm only without human-in-the-loop performance) was performed. The entire clinical prospective study and analytical studies evaluate the device's performance independently. The device's results are compared directly against a culture method and a sequencing assay, not against human interpretation of the device's output.

    7. Type of Ground Truth Used

    • Primary Ground Truth: Comparator culture method.
    • Confirmatory Ground Truth for Discrepancies: Validated bidirectional sequencing assay.

    8. Sample Size for the Training Set

    The document describes the performance of the device and does not explicitly mention a "training set" in the context of machine learning. The studies described are validation studies (e.g., Limit of Detection, Analytical Reactivity, Cross Reactivity, Interference, Clinical Prospective Study) that assess the device's performance characteristics. If the device's internal algorithms underwent a "training" phase during its development, the details are not provided in this regulatory summary.

    9. How the Ground Truth for the Training Set Was Established

    As no specific "training set" is mentioned for an algorithm, the method for establishing its ground truth is not provided. The document focuses on the validation of the finalized device.